Search Results
Found 4 results
510(k) Data Aggregation
(303 days)
Fiagon Navigation System
The Fiagon Navigation System is intended as an aid for locating anatomical structures in either open or percutaneous neurosurgical procedures. The Fiagon Navigation System is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of cranial surgery can be identified relative to a CT or MR based model of the anatomy.
Example procedures include, but are not limited to:
Cranial Procedures:
- Craniotomies/Craniectomies (e.g., Tumor Resection)
- Skull Base Procedures
- Cranial Biopsies
- General Catheter Shunt Placement
- Pediatric Catheter Shunt Placement
The user should consult the "Bench Accuracy" section of the User Manual to assess whether the accuracy of the system is suitable for their needs.
The Fiagon Navigation System displays the position of the instruments in preoperative scans (e.g., CT, MRI, fluoroscopy) utilizing electromagnetic tracking technology. The instrument with integrated sensor and the patient equipped with localized within an electromagnetic field generated by a field generator. The principle of navigation is based on electromagnetic spatial measuring of localizer element in a generated electromagnetic field.
The display of navigation information requires an image-to-patient registration procedure. During registration procedure, the navigation system determines the coordinate transformation between the intraoperative position of the patient and the position of the preoperative scan by fiducial marker, anatomical landmark or surface matching. Thereafter, the spatial position of the instrument is displayed superimposed to the image data. The navigation is updated with a rate of 15 to 45 Hz.
The components of the navigation system are
- Navigation unit with Navigation software. It has interfaces for screen, mouse and the components 2 – 4 below.
-
- Navigation sensor (Headrest with field generator)
-
- Navigation instrument (ShuntPointer, BiopsyPointer190 and BiopsyPointer 250)
- Patient reference localizer (Localizer Set Bone Anchor and Localizer Adhesive Pad)
The navigation unit is connected to a medical monitor. The unit runs the navigation software. Preoperative radiological images of the patient (DICOM CT, CBCT, MR) are imported to the system by means of CD-ROM, USB storage media or LAN network and displayed in appropriate way defined by the software.
The navigation unit includes the spatial measuring device electronics as well. This has connections to the field generating device (navigation sensor), the patient localizer and the navigation instrument.
The patient reference localizer and navigation instrument are tracked within the generated field by localizer elements integrated in the devices.
The patient reference localizer is fixed to the patient's anatomy and references it, while the instrument is tracked in relation to the patient localizer and thus to the patient's anatomy.
The provided text describes the Fiagon Navigation System, a stereotaxic instrument used as an aid for locating anatomical structures in neurosurgical procedures. The document focuses on demonstrating the substantial equivalence of this device to existing predicate devices, rather than a standalone AI-driven diagnostic system. As such, many of the requested elements pertaining to AI model validation, such as training set details, multi-reader multi-case studies, and expert consensus for ground truth, are not applicable to this type of medical device submission.
The document primarily details bench testing to assess the accuracy of the navigation system.
Here's an breakdown of the available information based on your request:
Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device appear to be implicitly defined by demonstrating accuracy similar to predicate devices. The key performance metrics are Target Registration Error (TRE) and Angular Registration Error (ARE).
Performance Metric | Acceptance Criteria (Implied: similar to predicate) | Reported Device Performance (Localizer Set Bone Anchor) | Reported Device Performance (Localizer Adhesive Pads) |
---|---|---|---|
Target Registration Error (TRE) |
Ask a specific question about this device
(120 days)
Fiagon Navigation System
The Fiagon Navigation System is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. The Fiagon Navigation System is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy.
Example procedures include, but are not limited to:
ENT Procedures: Transphenoidal access procedures. Intranasal procedures. Sinus procedures, such as Maxillary antrostomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies. ENT related anterior skull base procedures.
The Fiagon Navigation System displays position instruments in preoperative scans (e.g., CT, MRI, fluoroscopy) utilizing electromagnetic tracking technology. The position of the instrument with integrated sensor and the patient equipped with localizers are localized within an electromagnetic field generated by a field generator. The principle of navigation is based on electromagnetic spatial measuring of localizer element in a generated electromagnetic field.
The display of navigation information requires an image-to-patient registration procedure. During registration procedure, the navigation system determines the coordinate transformation between the intraoperative position of the patient and the position of the preoperative scan by fiducial marker, anatomical landmark or surface matching. Thereafter the spatial position of the instrument is displayed superimposed to the image data. The navigation information is updated with a rate of 15 to 45 Hz.
The provided document is a 510(k) Premarket Notification from the FDA for a medical device called the Fiagon Navigation System. This type of submission is for demonstrating substantial equivalence to a predicate device, not for establishing novel performance or clinical efficacy through the types of detailed studies typically associated with AI/ML model validation. Therefore, the document does not contain the information required to answer all parts of your request, particularly those related to AI/ML specific evaluations (like training set details, expert consensus for ground truth, MRMC studies, etc.).
However, I can extract the relevant information regarding acceptance criteria and performance data for this device as presented in the submission.
Here's a breakdown based on the provided text:
Acceptance Criteria and Reported Device Performance
The submission focuses on bench testing to demonstrate that the modified device (with WiFi connectivity and iPad remote control with automatic registration) performs as intended and is substantially equivalent to its predicate. The "acceptance criteria" are implied by the types of tests conducted and their successful completion.
1. Table of Acceptance Criteria and the Reported Device Performance:
Feature/Test | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Positioning Accuracy | Must perform comparably to the predicate device in locating anatomical structures precisely for navigation. | "Positioning accuracy test for target registration" was conducted. The device "functioned as intended and similar to the predicate." |
Wireless Coexistence | Must not interfere with other wireless devices and must maintain functionality when coexisting with them. | "Wireless coexisting testing" was conducted. The device "functioned as intended and similar to the predicate." |
Software Functionality | Software must operate without defects and perform all intended functions (e.g., displaying navigation information, updating at specified rates, handling registration). | "Software testing" was conducted. The device "functioned as intended and similar to the predicate." |
Overall Performance & Equivalence | The modified system must perform as intended and in a similar manner compared to the predicate, and not raise new safety or effectiveness questions. | "In all instances, the device functioned as intended and similar to the predicate, supporting the substantial equivalence to the predicate device." "The modified device does not present any new issues of safety or effectiveness." |
Study Details (as inferable from the document)
2. Sample size used for the test set and the data provenance:
- The document mentions "bench testing" but does not specify sample sizes (e.g., number of tests, number of targets, number of wireless interferences) or data provenance (e.g., country of origin, retrospective/prospective). This level of detail is typically not required for a 510(k) submission focused on substantial equivalence of a modified hardware/software system.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not provided. The testing appears to be quantitative bench testing of accuracy and functionality, not a study requiring expert interpretation of medical images or clinical outcomes.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- This is not applicable to the type of bench testing described.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC study was not done. This device is a navigation system that displays instrument position relative to pre-operative scans, not an AI/ML diagnostic or assistive tool that would involve "human readers" interpreting images.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- The tests described ("Positioning accuracy test for target registration", "Wireless coexisting testing", "Software testing") are essentially standalone performance evaluations of the device's capabilities. However, this is not an AI algorithm in the contemporary sense. The "algorithm" here refers to the underlying calculations for electromagnetic tracking and image-to-patient registration.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For positioning accuracy, the ground truth would likely be established by a precisely measured physical reference (e.g., a known target position in a phantom) rather than clinical expert consensus or pathology. For wireless coexistence and software testing, the ground truth is whether the system performs according to specifications.
8. The sample size for the training set:
- This is not applicable and not provided. This device is not described as an AI/ML system requiring a distinct "training set" in the context of deep learning.
9. How the ground truth for the training set was established:
- This is not applicable, as there's no mention of a "training set" for an AI/ML model for this device.
Ask a specific question about this device
(336 days)
Fiagon Navigation System
The Fiagon Navigation System is intended as an aid for precisely locating anatomical structures in either open or percutaneous neurosurgical procedures. The Fiagon Navigation System is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of cranial surgery can be identified relative to a CT or MR based model of the anatomy.
Example procedures include, but are not limited to:
Cranial Procedures:
- Craniotomies/Craniectomies (e.g., Tumor Resection)
- Skull Base Procedures
- Cranial Biopsies
- General Catheter Shunt Placement
The Fiagon Navigation System displays the position instruments in preoperative scans (e.g., CT, MRI, fluoroscopy) utilizing electromagnetic tracking technology. For cranial procedures, the use of this device is restricted to rigid fixation with a patient reference localizer attached directly to the skull clamp. The position of the instrument with integrated sensor and the patient referencing localizer (attached to the skull clamp) are localized within an electromagnetic field generated by a field generator. The principle of navigation is based on electromagnetic spatial measuring of localizer element in a generated electromagnetic field.
The display of navigation information requires an image-to-patient registration procedure. During registration procedure, the navigation system determines the coordinate transformation between the intraoperative position of the patient and the position of the preoperative scan by fiducial marker, anatomical landmark or surface matching.
Thereafter, the spatial position of the instrument is displayed superimposed to the image data. The navigation information is updated with a rate of 15 to 45 Hz.
The device Fiagon Navigation System utilizing similar technology than the proposed device has been previously cleared for a different intended use. This device is listed as a reference device.
The components of the navigation system are
-
- Navigation unit with Navigation software. It has interfaces for screen, mouse and the components 2 - 4.
-
- Navigation sensor (Headrest with field generator)
-
- Navigation instrument
- Patient reference localizer (with fixation of the localizer on the skull clamp using the adhesive pad)
The navigation unit is connected to a medical monitor. The unit runs the navigation software. Preoperative radiological images of the patient (DICOM CT, CBCT, MR) is imported to the system by means of CD-ROM. USB storage media or LAN network and displayed in appropriate way (defined by the software).
The navigation unit compromises the spatial measuring device electronics as well. This has connections to the field generating device (navigation sensor), the patient reference localizer and the navigation instrument.
Patient reference localizer and navigation instrument are tracked within the generated field by localizer elements integrated in the devices.
The patient reference localizer is fixed to the skull clamp and references the patient's anatomy, while the instrument is tracked in relation to the patient reference localizer and thus to the patient's anatomy.
Here's a breakdown of the acceptance criteria and study information for the "Fiagon Navigation System" based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Bench Accuracy) | Reported Device Performance (Bench Accuracy) |
---|---|
Position Mean: |
Ask a specific question about this device
(161 days)
FIAGON NAVIGATION SYSTEM
The Fiagon Navigation System is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. The Fiagon Navigation System is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy.
Example procedures include, but are not limited to:
ENT Procedures; Transphenoidal access procedures. Intranasal procedures. Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies. ENT related anterior skull base procedures
The Fiagon Navigation System displays the position instruments in preoperative scans (e.g., CT, MRI, fluoroscopy) utilizing electromagnetic tracking technology. The position of the instrument with integrated sensor and the patient equipped with localizers are localized within an electromagnetic field generated by a field generator. The principle of navigation is based on electromagnetic spatial measuring of localizer element in a generated electromagnetic field.
The display of navigation information requires an image-to-patient registration procedure. During registration procedure, the navigation system determines the coordinate transformation between the intraoperative position of the patient and the position of the preoperative scan by fiducial marker, anatomical landmark or surface matching,
Thereafter the spatial position of the instrument is displayed superimposed to the image data. The navigation information is updated with a rate of 15 to 45 Hz.
The components of the navigation system are
- Navigation unit with Navigation software. It has interfaces for screen, mouse and the components 2 - 4.
- Navigation sensor (Headrest with field generator)
- Navigation instrument,
- Patient reference localizer (with fixation material) :
The navigation unit, is connected to a medical monitor. The unit runs the navigation software. Preoperative radiological images of the patient (DICOM CT, CBCT, MR) is imported to the system by means of CD-ROM, USB storage media or LAN network and displayed in appropriate way (defined by the software)
The navigation unit compromises as well the spatial measuring device electronics. This has connections to the field generating device (navigation sensor), the patient localizer and the navigation instrument.
Patient reference localizer and navigation instrument are tracked within the generated field by localizer elements integrated in the devices.
The patient reference localizer is fixed to the patients anatomy and references it, while the instrument is tracked in relation to the patient localizer and thus to the patient's anatomy.
Here's a summary of the acceptance criteria and study details for the Fiagon Navigation System, based on the provided text:
Acceptance Criteria and Device Performance
Acceptance Criteria (Predicate Bench/Clinical Accuracy) | Reported Device Performance (Fiagon Navigation System) |
---|---|
Bench Accuracy: 0.8 mm to 1.0 mm | Bench Accuracy: 0.9 mm (SD 0.34 mm) |
Field Distortion Detection: 1.0 mm | Field Distortion Detection: |
Ask a specific question about this device
Page 1 of 1