Search Results
Found 4 results
510(k) Data Aggregation
(112 days)
FUJIFILM Ultrasonic Endoscope EG-740UT is intended to provide ultrasonic images of submucosal and peripheral organs of the upper gastrointestinal tract for observation, diagnosis, and endoscopic treatment. This product is intended to be used with a FUJIFILM ultrasonic processor.
This product is not intended for use on children and infants.
FUJIFILM Ultrasonic Endoscope EG-740UT is comprised of three general sections: a control portion, an insertion portion and an umbilicus. The control portion controls the angulation of the endoscope. This portion also controls the flexibility of the distal end in the endoscope. The insertion contains glass fiber bundles, several channels and a charge-coupled device (CCD) image sensor in its distal end. The channels in the insertion portion assist in delivering air/suction as well as endoscope accessories, such as forceps. The glass fiber bundles allow light to travel through the endoscope and emit light from the tip of the insertion portion to illuminate the body cavity. This provides enough light to the CCD image sensor to capture an image and display it on the monitor. The umbilicus consists of electronic components needed to operate the endoscope when plugged into the video processor, light source, and ultrasonic processor. The endoscopes are used in combination with FUJIFILM's video processors, light sources, ultrasonic processors, and peripheral devices such as monitor, printer, foot switch, and cart.
Here's an analysis of the acceptance criteria and study information for the FUJIFILM Ultrasonic Endoscope EG-740UT, based on the provided FDA 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Metric | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Resolution | At 5mm of working distance: 0.08mm of line pair on the square wave chart is readable. | Met (implied by "The proposed device met performance specifications in the following additional testing: ... Resolution") |
| At 100mm of working distance: 1.4mm of line pair on the square wave chart is readable. | Met (implied by "The proposed device met performance specifications in the following additional testing: ... Resolution") | |
| Field of View | (Not explicitly stated as an acceptance criterion in the provided text, but was tested) | Met (implied by "The proposed device met performance specifications in the following additional testing: Field of view") |
| Bending Capability | (Not explicitly stated as an acceptance criterion in the provided text, but was tested) | Met (implied by "The proposed device met performance specifications in the following additional testing: Bending capability") |
| Rate of Air Supply | (Not explicitly stated as an acceptance criterion in the provided text, but was tested) | Met (implied by "The proposed device met performance specifications in the following additional testing: ... Rate of air supply") |
| Rate of Water Supply | (Not explicitly stated as an acceptance criterion in the provided text, but was tested) | Met (implied by "The proposed device met performance specifications in the following additional testing: ... Rate of water supply") |
| Rate of Balloon Water Supply | (Not explicitly stated as an acceptance criterion in the provided text, but was tested) | Met (implied by "The proposed device met performance specifications in the following additional testing: ... Rate of balloon water supply") |
| Rate of Suction | (Not explicitly stated as an acceptance criterion in the provided text, but was tested) | Met (implied by "The proposed device met performance specifications in the following additional testing: ... Rate of suction") |
| Rate of Balloon Suction | (Not explicitly stated as an acceptance criterion in the provided text, but was tested) | Met (implied by "The proposed device met performance specifications in the following additional testing: ... Rate of balloon suction") |
| Diameter of Forceps Channel | (Not explicitly stated as an acceptance criterion in the provided text, but was tested) | Met (implied by "The proposed device met performance specifications in the following additional testing: ... Diameter of forceps channel") |
| Viewing Direction | (Not explicitly stated as an acceptance criterion in the provided text, but was tested) | Met (implied by "The proposed device met performance specifications in the following additional testing: ... Viewing direction") |
| LG Output | (Not explicitly stated as an acceptance criterion in the provided text, but was tested) | Met (implied by "The proposed device met performance specifications in the following additional testing: ... LG output") |
| Axial Resolution | (Not explicitly stated as an acceptance criterion in the provided text, but was tested) | Met (implied by "The proposed device met performance specifications in the following additional testing: ... Axial resolution") |
| Lateral Resolution | (Not explicitly stated as an acceptance criterion in the provided text, but was tested) | Met (implied by "The proposed device met performance specifications in the following additional testing: ... Lateral resolution") |
| Penetration Depth | (Not explicitly stated as an acceptance criterion in the provided text, but was tested) | Met (implied by "The proposed device met performance specifications in the following additional testing: ... Penetration depth") |
| Biocompatibility | (Implied by ISO 10993 standards and FDA guidance) | Evaluated using ISO 10993-1, ISO 10993-5, ISO 10993-10, and FDA guidance. |
| Electrical Safety & EMC | (Implied by IEC 60601 standards) | Evaluated using ANSI/AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-18, and IEC 60601-2-37. |
| Laser Safety & Photobiological Safety | (Implied by IEC 60825-1 and IEC 62471 standards) | Evaluated using IEC 60825-1 and IEC 62471. |
| Endoscope Specific Testing | (Implied by ISO 8600-1) | Conducted according to ISO 8600-1:2015. |
| Usability | (Implied by IEC 62366-1 and FDA guidance) | Conducted according to IEC 62366-1:2015 and FDA guidance. |
| Cleaning, Disinfection, and Sterilization | Validation in accordance with FDA guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" | Validated according to FDA guidance. |
2. Sample Size Used for the Test Set and Data Provenance
The provided 510(k) summary does not specify a "test set" in the context of clinical data for performance evaluation. The performance data presented are primarily from bench testing and compliance with established standards for various aspects like resolution, biocompatibility, electrical safety, etc. It does not appear to involve human patient data or a clinical study test set in the traditional sense.
Therefore, information on sample size for a test set and data provenance (country, retrospective/prospective) is not applicable or not provided in this document for evaluating device performance against clinical outcomes. The studies cited are for engineering, safety, and reprocessing validations.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
As there's no clinical "test set" with ground truth established by experts for diagnostic performance, this information is not applicable or not provided in this document. The evaluations are against technical specifications and regulatory standards.
4. Adjudication Method for the Test Set
Since there is no clinical "test set" requiring expert judgment for ground truth, an adjudication method is not applicable or not provided.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC study is mentioned in the provided text. The document focuses on the technical performance and safety of the device itself, not on differences in human reader performance with or without AI assistance. Therefore, information on effect size for human readers is not applicable or not provided.
6. Standalone (Algorithm Only) Performance Study
No standalone algorithm performance study is mentioned. The device is an endoscope, a medical instrument, not an AI algorithm. Therefore, this is not applicable or not provided.
7. Type of Ground Truth Used
The "ground truth" for the performance evaluation in this document is established by:
- Technical specifications/measurements: For metrics like resolution, field of view, bending capability, flow rates, etc. These are objective measurements against predefined engineering targets.
- Compliance with consensus standards: For biocompatibility (ISO 10993), electrical safety (IEC 60601), laser safety (IEC 60825, IEC 62471), endoscope specifics (ISO 8600-1), usability (IEC 62366-1), and reprocessing (FDA guidance). These standards themselves define the acceptable "truth" for device characteristics and safety.
8. Sample Size for the Training Set
As this is a medical device (endoscope) undergoing 510(k) clearance, and not an AI/ML algorithm, there is no concept of a "training set" in the context of learning data. The device's design and manufacturing are based on established engineering principles and prior device models, not on a machine learning training process.
9. How the Ground Truth for the Training Set Was Established
Since there is no "training set," this information is not applicable or not provided.
Ask a specific question about this device
(268 days)
FUJIFILM Ultrasonic Endoscopes EG-580UT and EG-580UR are intended to provide ultrasonic images of submucosal and peripheral organs of the upper gastrointestinal tract for observation, diagnosis, and endoscopic treatment. These products are intended to be used with a FUJIFILM ultrasonic processor. These products are not intended for use on children and infants.
FUJIFILM Ultrasonic Endoscopes EG-580UT and EG-580UR are comprised of three general sections: a control portion, an insertion and an umbilicus. The control portion controls the angulation of the endoscope. This portion also controls the flexibility of the distal end in the endoscope. The insertion contains glass fiber bundles, several channels and a chargecoupled device (CCD) image sensor in its distal end. The insertion portion assist in delivering air/suction as well as endoscope accessories, such as forceps. The glass fiber bundles allow light to travel through the endoscope and emit light from the tip of the insertion portion to illuminate the body cavity. This provides enough light to the CCD image sensor to capture an image and display it on the monitor. The umbilicus consists of electronic components needed to operate the endoscope when plugged into the video processor, light source, and ultrasonic processor. The endoscopes are used in combination with FUJIFILM's video processors, light sources, ultrasonic processors and peripheral devices such as monitor, printer, foot switch, and cart. All of these combinations were previously cleared in K181763.
The provided text is a 510(k) summary for the FUJIFILM Ultrasonic Endoscopes EG-580UT and EG-580UR. It describes the device, its intended use, and a comparison to predicate devices, focusing on the substantial equivalence argument.
However, the document does NOT contain information about an AI-powered device or an AI performance study. Instead, it describes a medical device undergoing a 510(k) submission, which primarily focuses on demonstrating substantial equivalence to a legally marketed predicate device through engineering performance, electrical safety, biocompatibility, and reprocessing validation.
Therefore, for the specific questions related to AI device acceptance criteria and performance study details (sample size for test set, data provenance, number of experts, adjudication, MRMC studies, standalone performance, ground truth, training set size), the provided text does not contain the information required to answer them.
The document primarily focuses on the physical and functional aspects of the endoscopes, such as control portion, insertion, umbilicus, and their compatibility with other FUJIFILM equipment. The performance data mentioned are related to:
- Electrical safety: evaluated using standards like ANSI/AAMI ES 60601-1:2012, IEC 60601-1-2:2007, etc.
- Biocompatibility: evaluated using standards like ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-10:2010.
- Endoscope specific testing: evaluated using standards like ISO 8600-1:2015, ISO 8600-3:1997, and ISO 8600-4:2014.
- Cleaning, high-level disinfection, and sterilization: evaluated according to AAMI TIR12:2010, AAMI TIR30:2011.
- Additional functional testing: including field of view, resolution, bending capability, working length, diameter of forceps channel, rates of suction, air supply, water supply, balloon water supply, balloon suction, viewing direction, LG output, axial resolution, lateral resolution, and penetration depth.
These are all standard tests for a physical medical device and do not involve AI or machine learning algorithms.
Ask a specific question about this device
(141 days)
EB-530US is intended for the observation, diagnosis, and endoscopic treatment of the trachea, bronchial tree and surrounding organs using ultrasonic images. It is used with a FUJIFILM ultrasonic processor, video processor, light source, other peripheral equipment and endoscopic accessories. It is not intended for use on children and infants.
FUJIFILM Ultrasonic Endoscope EB-530US is an ultrasonic bronchoscope that emits ultrasound waves and scans the reflected signals to provide ultrasonic images when used in combination with an ultrasonic processor.
Here's an analysis of the acceptance criteria and study information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Specific Test/Parameter | Reported Device Performance | Study Type |
|---|---|---|---|
| Electrical Safety | ANSI/AAMI ES 60601-1:2012 | Met pre-defined criteria | Non-clinical |
| IEC 60601-1-2:2007 | Met pre-defined criteria | Non-clinical | |
| IEC 60601-1-6:2013 | Met pre-defined criteria | Non-clinical | |
| IEC 60601-2-18:2009 | Met pre-defined criteria | Non-clinical | |
| IEC 60601-2-37:2015 | Met pre-defined criteria | Non-clinical | |
| Biocompatibility | ISO 10993-1:2009 | Met pre-defined criteria | Non-clinical |
| ISO 10993-5:2009 | Met pre-defined criteria | Non-clinical | |
| ISO 10993-10:2010 | Met pre-defined criteria | Non-clinical | |
| Storage & Transportation | Expanded temperature range (-20°C to 60°C) | Validated | Non-clinical |
| Expanded humidity range (10 to 85% RH) | Validated | Non-clinical | |
| Performance Testing | Field of view | Met pre-defined criteria | Non-clinical |
| Forceps channel diameter | Met pre-defined criteria | Non-clinical | |
| Axial resolution | Met pre-defined criteria | Non-clinical | |
| Bending capability | Met pre-defined criteria | Non-clinical | |
| Viewing direction | Met pre-defined criteria | Non-clinical | |
| Lateral resolution | Met pre-defined criteria | Non-clinical | |
| Suction rate | Met pre-defined criteria | Non-clinical | |
| Resolution | Met pre-defined criteria | Non-clinical | |
| Penetration depth | Met pre-defined criteria | Non-clinical | |
| Working length | Met pre-defined criteria | Non-clinical | |
| LG output | Met pre-defined criteria | Non-clinical |
2. Sample size used for the test set and the data provenance
The document does not specify human clinical trials or a "test set" in the context of patient data. The performance testing appears to be entirely non-clinical (engineering and laboratory tests), likely using a sample of the manufactured device itself or components. Therefore, information about human data provenance is not applicable.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. The reported performance testing is non-clinical and does not involve expert evaluation for ground truth in a clinical context.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as no clinical test set requiring adjudication by experts is described.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study was done, and the device is an ultrasonic endoscope, not an AI-assisted diagnostic tool. Therefore, the effect size of human readers improving with AI assistance is not applicable.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This is not an AI algorithm. The device is a medical instrument. Therefore, this question is not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the non-clinical performance tests, the "ground truth" would be established by objective measurements against engineering specifications, calibrated instruments, and established safety and performance standards (e.g., IEC standards for electrical safety, ISO standards for biocompatibility).
8. The sample size for the training set
Not applicable, as this is not an AI/machine learning device requiring a training set.
9. How the ground truth for the training set was established
Not applicable, as this is not an AI/machine learning device.
Ask a specific question about this device
(24 days)
EG-530UT2 and EG-530UR2 are intended to provide ultrasonic images of submucosal and peripheral organs of the upper gastrointestinal tract for observation, diagnosis, and endoscopic treatment. The product is intended to be used with a Fujifilm ultrasonic processor. This product is not intended for use on children and infants.
FUJIFILM Ultrasonic Endoscopes EG-530UT2 and EG-530UR2 are upper gastrointestinal endoscopes that emit ultrasound waves and scan the reflected signals to provide ultrasonic images when used in combination with an ultrasonic processor.
The provided text describes a 510(k) premarket notification for FUJIFILM Ultrasonic Endoscopes EG-530UT2 and EG-530UR2. This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving that a new AI/software device meets specific performance criteria through a clinical study.
Therefore, the requested information regarding acceptance criteria for an AI/software device and a study to prove it meets them cannot be fully extracted from this document. The document lists performance tests for the physical endoscopic devices (e.g., field of view, bending capability, resolution, air/water supply rates), but these are for hardware performance, not AI/software performance.
However, I can extract the information related to the device performance and acceptance criteria for the physical endoscopy device as described in the document.
Here's a breakdown of what can be extracted and what cannot be, based on the provided text:
What Can Be Extracted (for the physical endoscope device):
-
A table of acceptance criteria and the reported device performance:
The document states: "Fujifilm conducted the following performance testing on the proposed devices EG-530UT2 and EG-530UR2 to ensure that the modified devices perform equivalently to the predicate devices:- Field of view
- Viewing direction
- Bending capability
- Resolution
- Air supply rate
- Water supply rate
- Axial resolution
- LG output
- Suction rate
- Working length
- Lateral resolution
- Penetration depth
- Forceps channel diameter
In all cases, the devices met the pre-defined acceptance criteria for the test."
Therefore, the table would look like this (specific values are not provided in the document, only the claim of meeting criteria):
Acceptance Criteria (Measured Parameter) Reported Device Performance Field of view Met pre-defined criteria Viewing direction Met pre-defined criteria Bending capability Met pre-defined criteria Resolution Met pre-defined criteria Air supply rate Met pre-defined criteria Water supply rate Met pre-defined criteria Axial resolution Met pre-defined criteria LG output Met pre-defined criteria Suction rate Met pre-defined criteria Working length Met pre-defined criteria Lateral resolution Met pre-defined criteria Penetration depth Met pre-defined criteria Forceps channel diameter Met pre-defined criteria -
Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Sample Size: Not explicitly stated for each performance test. The document refers to "the proposed devices EG-530UT2 and EG-530UR2" performing these tests, suggesting testing on the physical devices themselves rather than a separate dataset of patient cases.
- Data Provenance: Not specified. This is a pre-market submission, and the tests are likely internal lab/bench testing of the device hardware.
What Cannot Be Extracted (as it's not an AI/software performance study):
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for physical device performance typically comes from engineering specifications and measurements, not expert human interpretation of medical images.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document does not describe an AI-assisted diagnostic device or an MRMC study.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable, as this is not an AI algorithm.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable for an AI/software context. Ground truth for physical device specs (like resolution, field of view) is based on engineering measurements against design specifications.
- The sample size for the training set: Not applicable. This document is not about AI training.
- How the ground truth for the training set was established: Not applicable.
In summary: The provided FDA 510(k) clearance document for FUJIFILM Ultrasonic Endoscopes EG-530UT2 and EG-530UR2 pertains to demonstrating substantial equivalence of a physical medical device (endoscope) to a predicate, primarily through engineering performance testing (e.g., field of view, bending capability, resolution, fluid rates), electrical safety, and biocompatibility. It is not a document describing the validation of an AI/software device, and therefore the specific criteria related to AI performance studies cannot be found within it.
Ask a specific question about this device
Page 1 of 1