Search Results
Found 1 results
510(k) Data Aggregation
(235 days)
FUJIFILM Double Balloon Endoscope EI580BT
This device is intended for the visualization of the upper and lower digestive tracts, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, duodenum, small intestine, and rectum.
FUJIFILM Double Balloon Endoscope EI-580BT is comprised of three general sections: a control portion, an insertion portion and an umbilicus. The control the angulation of the endoscope. This portion also controls the flexibility of the distal end in the endoscope. The insertion portion contains glass fiber bundles, several channels and a complementary Charge-Coupled Device (CCD) image sensor in its distal end. The channels in the insertion assist in delivering air/suction as well as endoscope accessories, such as forceps and balloons. The glass fiber bundles allow light to travel through the endoscope and emit light from the tip of the insertion to illuminate the body cavity. This provides enough light to the CCD image sensor to capture an image and display it on the monitor. The umbilicus consists of electronic components needed to operate the endoscope when plugged in to the video processor and the light source. The endoscope is used in combination with FUJIFILM's video processors, light sources and peripheral devices such as monitor, printer, foot switch, and cart. All of these combinations were previously cleared in K143556 and K172916.
The provided text describes the regulatory clearance of the FUJIFILM Double Balloon Endoscope EI-580BT and outlines the performance data submitted to demonstrate substantial equivalence to a predicate device. However, it does not contain information related to a study proving the device meets specific acceptance criteria based on human-in-the-loop performance, AI assistance, or an algorithm's diagnostic accuracy.
The performance data mentioned in the document focuses on:
- Electrical safety: Compliance with standards like ANSI/AAMI ES 60601-1:2012, IEC 60601-1-2:2007, etc.
- Biocompatibility: Compliance with ISO 10993 standards.
- Endoscope specific testing: Compliance with ISO 8600 standards.
- Cleaning, high-level disinfection, and sterilization: Compliance with AAMI TIR standards and FDA guidance.
- Usability testing: Compliance with IEC 62366-1:2015.
- Bench testing: Validation of features like "Advanced Force Transmission" and "Adaptive Bending," and performance specifications for field of view, bending capability, rates of air/water/suction, working length, channel diameter, viewing direction, resolution, and LG output.
Therefore, I cannot fulfill the request for information regarding acceptance criteria and a study proving the device meets these criteria in the context of diagnostic accuracy, human-in-the-loop performance, AI assistance, or algorithm-only performance, as this information is not present in the provided text.
The document pertains to the clearance of an endoscope, which is a medical instrument for direct visualization, not an AI or image-analysis diagnostic device. The "performance data" discussed are engineering and safety specifications typical for such a device, not metrics like sensitivity, specificity, AUC, or reader performance improvement often associated with AI/diagnostic software.
Ask a specific question about this device
Page 1 of 1