Search Results
Found 2 results
510(k) Data Aggregation
(114 days)
FLXfit
The EOI FLXfit™ 15 and FLXfit™ Intervertebral body fusion device is indicated for interbody fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). These patients should be skeletally mature and have completed six months of non-operative treatment. The EOI FLXfit™ 15 and FLXfit™ devices are intended to be used with supplemental spinal fixation system and with autogenous bone graft.
Important: FLXfit™ 15 must be applied in combination with posterior fusion system such as the EOI Spinal System.
EOI FLXfit15 Intervertebral Body Fusion Device is made from titanium (Ti6Al4V-ELI per ASTM F136). This is an articulated-expanded device with height ranges from 8mm through 13mm. The device also expands in height to provide lordosis angle correction of up to 15°.
This document describes a 510(k) premarket notification for the FLXfit™ 15 intervertebral body fusion device. While it details the device and its intended use, it does not contain information about a study that assesses its performance against specific acceptance criteria in a clinical or AI-related context. Instead, it focuses on the device's mechanical properties and its substantial equivalence to a predicate device.
Therefore, I cannot fulfill your request for:
- A table of acceptance criteria and reported device performance (in a clinical/AI sense).
- Sample size used for the test set and data provenance.
- Number of experts used to establish ground truth and their qualifications.
- Adjudication method for the test set.
- MRMC comparative effectiveness study results.
- Standalone (algorithm only) performance.
- Type of ground truth used.
- Sample size for the training set.
- How ground truth for the training set was established.
The document primarily discusses performance testing summary which refers to mechanical testing of the device, not clinical performance or an AI study.
Here's what the document does provide regarding performance testing:
1. Acceptance Criteria and Reported Device Performance (Mechanical Testing):
Acceptance Criteria (Test) | Reported Device Performance (Implied Meeting of Standards) |
---|---|
Dynamic and static compression per ASTM F2077 | The document states that testing was conducted to "demonstrate that the modifications... provide adequate and substantially equivalent mechanical strength for their intended use." Implies the device met the standards of ASTM F2077. |
Dynamic and static shear compression per ASTM F2077 | The document states that testing was conducted to "demonstrate that the modifications... provide adequate and substantially equivalent mechanical strength for their intended use." Implies the device met the standards of ASTM F2077. |
Subsidence per ASTM F2267 | The document states that testing was conducted to "demonstrate that the modifications... provide adequate and substantially equivalent mechanical strength for their intended use." Implies the device met the standards of ASTM F2267. |
2. Sample Size Used for the Test Set and Data Provenance:
* Test Set Sample Size: Not specified for mechanical testing.
* Data Provenance: Not applicable in the context of clinical or AI data. The tests are physical mechanical tests conducted on device prototypes.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
* Not applicable as this is mechanical testing to ASTM standards, not clinical ground truth established by experts.
4. Adjudication Method for the Test Set:
* Not applicable.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
* No, this document does not describe an MRMC study or any AI component.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
* No, this document does not describe any standalone algorithm performance.
7. The type of ground truth used:
* For mechanical testing, the "ground truth" is defined by the specific performance parameters and thresholds outlined in the referenced ASTM standards (F2077, F2267).
8. The sample size for the training set:
* Not applicable. There is no AI model or training set discussed.
9. How the ground truth for the training set was established:
* Not applicable.
Ask a specific question about this device
(211 days)
FLXFIT INTERVERTEBRAL BODY FUSION DEVICE
The EOI FLXfit Intervertebral body fusion device is indicated for interbody fusion in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). These patients should be skeletally mature and have completed six months of non-operative treatment. The EOI FLXfit device is intended to be used with supplemental spinal fixation system and with autogenous bone graft.
EOI FLXfit Intervertebral Body Fusion Device is made From Titanium (Ti6A14V-ELI per ASTM norm F136). This is an articulated-expanded device with height ranges from 8mm through 14mm. The device also expands in height to provide lordosis angle correction of up to 10°. The FLXfit components are manufactured from titanium alloy (ASTM F136). It is implanted via a Transforaminal approach and supplemented by posterior fixation. The device is made from two titanium segments that are serrated on the superior and inferior surfaces. The titanium segments are attached (linked) with titanium wedge pins that provide the means to form the implant into its final articulated shape.
The provided document is a 510(k) summary for the FLXfit Intervertebral Body Fusion Device, which focuses on demonstrating substantial equivalence to predicate devices through mechanical testing. It is not a study that evaluates the performance of a device using clinical or reader studies, especially for AI/ML devices. Therefore, a significant portion of the requested information, such as sample sizes for test sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, and ground truth for clinical performance, is not applicable or available in this document.
However, I can extract the acceptance criteria and reported device performance from the provided mechanical testing data.
1. Table of Acceptance Criteria and Reported Device Performance
The document describes performance data based on mechanical testing standards, rather than clinical efficacy. The acceptance criterion for this type of submission is typically "substantial equivalence" to predicate devices, meaning the device performs at least as well as or better than the predicate devices under the specified tests. The reported performance is that the device met these criteria.
Acceptance Criteria | Reported Device Performance |
---|---|
Static Axial Compression (ASTM F2077) | The FLXfit's strength was found sufficient for its intended use and substantially equivalent to legally marketed predicate devices. |
Static Compression-Shear (ASTM F2077) | The FLXfit's strength was found sufficient for its intended use and substantially equivalent to legally marketed predicate devices. |
Dynamic Axial Compression (ASTM F2077) | The FLXfit's strength was found sufficient for its intended use and substantially equivalent to legally marketed predicate devices. |
Dynamic Compression-Shear (ASTM F2077) | The FLXfit's strength was found sufficient for its intended use and substantially equivalent to legally marketed predicate devices. |
Subsidence (ASTM F2267 and ASTM F2077) | No specific quantitative reported performance, but results show the strength is sufficient and substantially equivalent. |
Usability cadaveric study (clinical application, expansion mechanism, intra-operative subsidence) | No specific quantitative reported performance, but the study was conducted to evaluate clinical application, and the overall conclusion is substantial equivalence. |
2. Sample size used for the test set and the data provenance
- Test Set Sample Size: Not applicable. The "test set" here refers to physical devices undergoing mechanical testing, not a dataset of patient cases for an AI/ML device. The document does not specify the number of devices tested for each mechanical test.
- Data Provenance: The data comes from "non-clinical testing" and a "usability cadaveric study." No country of origin is specified, but it can be inferred to be from the manufacturer's testing facilities or contract labs. The nature is experimental, not retrospective or prospective clinical data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. For mechanical testing, "ground truth" is established by the specifications of the ASTM standards themselves, and the measurement of physical properties, not by expert consensus on clinical cases. The usability study involved cadavers, likely evaluated by surgical or biomechanical experts, but no details are provided on their number or qualifications.
4. Adjudication method for the test set
Not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies to resolve discrepancies among human readers for establishing ground truth from patient data. This document describes mechanical testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a medical device (intervertebral body fusion device), not an AI/ML diagnostic or assistive tool. Therefore, no MRMC study or AI-related effectiveness analysis was performed or described.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. As noted, this is a physical medical device, not an algorithm.
7. The type of ground truth used
- For Mechanical Testing: The ground truth is defined by the requirements of the ASTM standards (e.g., F2077, F2267) which specify acceptable load-bearing capacities, fatigue life, and other mechanical properties for intervertebral body fusion devices. Substantial equivalence to predicate devices also serves as a comparative "ground truth" for performance.
- For Usability Cadaveric Study: The ground truth would be the direct observation and evaluation of the device's clinical application, expansion mechanism, and intra-operative subsidence in a cadaveric model by qualified personnel (likely surgeons or biomechanical engineers).
8. The sample size for the training set
Not applicable. There is no "training set" in the context of mechanical testing or regulatory submission for a physical medical device. This term applies to machine learning models.
9. How the ground truth for the training set was established
Not applicable. There is no training set mentioned in this document.
Ask a specific question about this device
Page 1 of 1