510(k) Data Aggregation

    K Number
    K070390
    Date Cleared
    2007-04-05

    (55 days)

    Product Code
    Regulation Number
    892.1650
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    FLUOROPRO RF DIGITAL IMAGING SYSTEM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for obtaining fluoroscopic radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts.

    Device Description

    The FluoroPro RF Digital Imaging System is a high resolution, digital imaging system designed for digital videography. It is intended to replace conventional film techniques in multipurpose or dedicated applications where general fluoroscopy, interventional fluoroscopy or angiography or cardiac imaging procedures are performed. The FluoroPro RF Digital Imaging System allows the operator to view and enhance 1000 line fluoroscopy. High resolution digital spot images (1024) may be acquired at single or rapid acquisition rates, up to 30 fps. Images may be viewed and enhanced enabling the operator to bring out diagnostic details difficult or impossible to see using conventional imaging techniques. Images can be stored locally for medium term storage. The FluoroPro RF Digital Imaging System enables the operator to hardcopy image with a laser printer or send images over a network for longer term storage. The major system components include: a fluoroscopic TV camera, monitors, and an image processor.

    AI/ML Overview

    The provided documentation describes the FluoroPro RF Digital Imaging System, a fluoroscopic X-Ray system, and its substantial equivalence to a predicate device. The core of the submission focuses on demonstrating similarity to a legally marketed device rather than proving performance against specific acceptance criteria in a comparative clinical study.

    Here's an analysis based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The submission does not present explicit "acceptance criteria" in the traditional sense of a clinical trial (e.g., target sensitivity, specificity). Instead, it demonstrates performance by showing equivalence to a predicate device based on technical characteristics. The table below summarizes these characteristics and how the new device compares to the predicate.

    CharacteristicPredicate Device (Infimed Orion Digital Imaging System, K012490)FluoroPro RF Digital Imaging System (New Device)Performance Against "Acceptance Criteria" (Equivalence)
    Intended UseAcquisition, display, storage, and transmission of fluoroscopic imagesSAMEMeets (Identical)
    Power source120 VAC 50/60 HZ 7 amps120 VAC 50/60 HZ 2.5 ampsDifferent, but considered acceptable (likely an improvement in efficiency)
    Image acquisitionUp to 15 FPS (spot), up to 30 fps (fluoro)SAMEMeets (Identical)
    File compatibilityDICOMSAMEMeets (Identical)
    CCDInnovision CCDThales CCDDifferent, but not deemed a substantial difference affecting safety/effectiveness
    Digital Resolution1024 x 1024 12 bitSAMEMeets (Identical)
    Performance StandardUS FDASAMEMeets (Identical)
    Electrical safetySAMESAMEMeets (Identical)

    Study that Proves the Device Meets the Acceptance Criteria:

    The documentation states: "The results of bench and user testing indicates that the new device is as safe and effective as the predicate devices." and "After analyzing both bench and user testing data, it is the conclusion of Virtual Imaging that the FluoroPro RF Digital Imaging System is as safe and effective as the predicate device, has few technological differences, and has no new indications for use, thus rendering it substantially equivalent to the predicate device."

    This indicates that the "study" was primarily bench testing (evaluating technical specifications, performance parameters, and compliance with standards) and user testing (likely validation by trained personnel in a simulated or controlled environment to confirm functionality and usability). The goal of these tests was to demonstrate substantial equivalence to the predicate device, not to meet specific quantitative clinical performance metrics like sensitivity or specificity.

    2. Sample sized used for the test set and the data provenance

    The document does not provide details on the sample size used for the "bench and user testing." It also does not specify the provenance of any data (e.g., country of origin, retrospective/prospective), as it relies on technical and operational comparisons rather than clinical data from human subjects.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. Given that the submission focuses on substantial equivalence based on technical specifications and functional testing, it's unlikely that "ground truth" for diagnostic accuracy was established by medical experts in the way it would be for an AI-powered diagnostic device. The "user testing" likely involved qualified technicians or doctors evaluating the system's operational aspects, image quality for diagnostic interpretation (compared to the predicate), and ease of use.

    4. Adjudication method for the test set

    This information is not provided. As "ground truth" from experts for diagnostic accuracy in a clinical context isn't explicitly mentioned, an adjudication method for such a ground truth would not be detailed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted or reported in this submission. This device is a digital imaging system (hardware and software for image acquisition and display), not an AI-powered diagnostic tool, so the concept of human readers improving with AI assistance is not applicable here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a standalone algorithm performance study was not done. The device itself is an imaging system; its "performance" is tied to its ability to acquire and display images comparable to a predicate device, and it is intended for use by a "qualified/trained doctor or technician." It is not an autonomous diagnostic algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For this type of device (a digital fluoroscopic x-ray system), the "ground truth" used for validation would primarily relate to its technical performance and image quality, not disease diagnosis. This would involve:

    • Technical Specifications: Confirming that the system meets its stated parameters (resolution, frame rate, power consumption, etc.) through engineering tests.
    • Image Quality Assessment: Evaluating the acquired images against established standards or by comparison to the predicate device for clarity, contrast, and diagnostic interpretability by a qualified user. This often involves phantoms and possibly human users assessing image quality visually.
    • Safety Standards: Adherence to electrical safety and other relevant performance standards (e.g., US FDA performance standards mentioned).

    There is no mention of ground truth established via expert consensus, pathology, or outcomes data related to diagnostic accuracy from patient images.

    8. The sample size for the training set

    This information is not applicable and therefore not provided. The FluoroPro RF Digital Imaging System is a digital X-ray system, not a machine learning or AI algorithm that requires a "training set" in the computational sense.

    9. How the ground truth for the training set was established

    This information is not applicable and therefore not provided, as there is no "training set" for this device.