510(k) Data Aggregation
(39 days)
FLUOBEAM® LX Imaging System (FB-LX); FLUOBEAM**®** LX Red Imaging System (FB-LXR)
FLUOBEAM® LX and FLUOBEAM® LX Red are intended to provide real-time near infrared (NIR) fluorescence imaging of tissue during surgical procedures. Upon intravenous administration and use of an ICG consistent with its approved labeling, the FLUOBEAM® LX and FLUOBEAM® LX Red are indicated for use in capturing and viewing fluorescent images for the visualization of vessels, blood flow and tissue perfusion before, during and after organ transplant, plastic, micro- and reconstructive surgeries.
The FLUOBEAM® LX and FLUOBEAM® LX Red can also be used to assist in the imaging of parathyroid glands and can be used as an adjunctive method to assist in the location of parathyroid glands due to the auto-fluorescence of this tissue.
Use of the FLUOBEAM® LX and FLUOBEAM® LX Red devices are intended to assist, not replace, experienced visual assessment, and biopsy with conventional histopathological confirmation per standard of care. The system is not to be used to confirm the absence of parathyroid tissue or glands and is only to be used to assist in locating visually identified gland/tissues.
FLUOBEAM® LX and FLUOBEAM® LX Red are imaging systems intended to provide realtime near infrared (NIR) fluorescence imaging of tissue during surgical procedures. The FLUOBEAM® LX and FLUOBEAM® LX Red are indicated for use in capturing and viewing fluorescent images for the visualization of vessels, blood flow and tissue perfusion before, during and after organ transplant, plastic, micro- and reconstructive surgeries.
FLUOBEAM® LX and FLUOBEAM® LX Red can also be used to assist in the imaging of parathyroid glands and can be used as an adjunctive method to assist in the location of parathyroid glands due to the auto-fluorescence of this tissue. Use of the FLUOBEAM® LX and FLUOBEAM® LX Red devices are intended to assist, not replace, experienced visual assessment, and biopsy with conventional histopathological confirmation per standard of care. The system is not to be used to confirm the absence of parathyroid tissue or glands and is only to be used to assist in locating visually identified gland/tissues.
FLUOBEAM® LX and FLUOBEAM® LX Red enable surgeons to observe fluorescent images of parathyroid glands, blood vessels and related tissue perfusion. Fluorescence can be observed as a result of natural fluorescence of parathyroid glands or as a result of a fluorescent product, indocyanine green (ICG), injected intravenously into patients before the surgery allowing the perfusion assessment.
Class 1 infrared laser light is used to excite the fluorescent tissues of parathyroid glands or the ICG and illuminate the regions of a patient's body to be observed. A camera inside the optical head captures the fluorescent image that is used to visualize the parathyroid glands or assess the blood vessels and related tissue perfusion. FLUOBEAM® LX and FLUOBEAM® LX Red consist of the following components: a hardware part with a camera unit (optical head) linked by a specific cable to a control box and a software part with FLUOSOFT™ LX or FLUOSOFT™ LX Red imaging software. The optical head contains a video camera and light sources (laser and LEDs) and is used by hand. The control box receives the video signal of the fluorescent image from the optical head, it digitizes it and sends it to a computer that outputs it on a display screen and/or records it. Adjustments of the fluorescent image are possible either by the optical head or via the FLUOSOFT™ LX imaging software and the FLUOSOFT™ LX Red imaging software on the computer.
This special 510(k) premarket notification is intended to re-frame the indications for use statement to be consistent with specific indications of legally marketed ICG products, eliminating the need for the co-packaging with ICG.
The provided document is a 510(k) Premarket Notification from the FDA, specifically concerning a "Special 510(k)" for the FLUOBEAM® LX and FLUOBEAM® LX Red Imaging Systems.
Crucially, this document states: "No performance data are needed to support the modified indications for use. As noted above, there are no technological changes associated with the proposed labeling changes. Additionally, no new surgical procedures or tissue types are being referenced in the modified indications for use."
This means the submission is not presenting new performance studies or data to demonstrate the device meets acceptance criteria. Instead, it's leveraging the substantial equivalence to previously cleared devices (K190891 and K230898) and a reference device (K223020) because the changes are limited to refining the "Indications for Use" statement to align with existing ICG product labeling, thereby eliminating the need for co-packaging with ICG.
Therefore, for the information requested in your prompt, based solely on the provided text, we cannot fill in most of the table or answer many of the questions as no new performance study data is presented.
Here's what can be extracted from the document:
1. Table of acceptance criteria and the reported device performance:
Acceptance Criteria (Not explicitly stated as such for new performance, but implied by substantial equivalence to predicates) | Reported Device Performance (Implied by substantial equivalence to predicates) |
---|---|
Ability to provide real-time near infrared (NIR) fluorescence imaging of tissue during surgical procedures. | Device provides real-time NIR fluorescence imaging. |
Visualization of vessels, blood flow and tissue perfusion with ICG. | Device enables visualization of vessels, blood flow, and tissue perfusion with ICG. |
Assistance in imaging parathyroid glands and location due to auto-fluorescence. | Device assists in imaging and locating parathyroid glands through auto-fluorescence. |
Consistent with the performance of predicate devices FLUOBEAM® LX (K190891) and FLUOBEAM LX Red (K230898). | Device performance is substantially equivalent to predicate devices. |
2. Sample size used for the test set and the data provenance:
- Sample size: Not applicable, as no new performance study data is presented. The submission relies on substantial equivalence.
- Data provenance: Not applicable.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable, as no new performance study data is presented.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable, as no new performance study data is presented.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is an imaging system, not an AI/CADe/CADx device that assists human readers in interpretation or diagnosis. It aids in visualization during surgery. No MRMC study is mentioned.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is not an algorithm-only device; it's an imaging system where a human observes the generated images.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not explicitly stated for new data, as none was required. For the original clearance of the predicate devices, ground truth would likely have involved direct visual confirmation during surgery, correlation with anatomical knowledge, and potentially histopathological confirmation where applicable (e.g., for parathyroid tissue). The document reiterates that the device "is not to be used to confirm the absence of parathyroid tissue or glands and is only to be used to assist in locating visually identified gland/tissues," implying that standard clinical and pathological evaluations remain the definitive ground truth for such aspects.
8. The sample size for the training set:
- Not applicable, as no new performance study data for a machine learning model is presented.
9. How the ground truth for the training set was established:
- Not applicable, as no new performance study data for a machine learning model is presented.
In summary: This 510(k) submission primarily focuses on a labeling change for an existing device, asserting that no new performance data is needed because "there are no technological changes associated with the proposed labeling changes. Additionally, no new surgical procedures or tissue types are being referenced in the modified indications for use." Therefore, the detailed performance study information requested is not present in this specific FDA clearance document.
Ask a specific question about this device
(119 days)
FLUOBEAM**®** LX Red
FLUOBEAM® LX Red is intended to provide real-time near infrared (NIR) fluorescence imaging of tissue during surgical procedures. The FLUOBEAM® LX Red is indicated for use in capturing fluorescent images for the visual assessment of blood flow in adults as an adjunctive method for the evaluation of tissue perfusion, perfused organs, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive and organ transplant surgeries.
The FLUOBEAM® LX Red can also be used to assist in the imaging of parathyroid glands and can be used as an adjunctive method to assist in the location of parathyroid glands due to the auto-fluorescence of this tissue. Use of the FLUOBEAM® LX Red device is intended to assist, not replace, experienced visual assessment, and biopsy with conventional histopathological confirmation per standard of care. The system is not to confirm the absence of parathyroid tissue or glands and is only to be used to assist in locating visually identified gland/tissues.
FLUOBEAM® LX Red is an imaging system intended to provide real-time near infrared (NIR) fluorescence imaging of tissue during surgical procedures. The FLUOBEAM® LX Red is indicated for use in capturing and viewing fluorescent images for the visual assessment of blood flow in adults as an adjunctive method for the evaluation of tissue perfusion, perfused organs, and related tissue-transfer circulation in tissue and free flaps used in plastic, micro- and reconstructive and organ transplant surgeries.
FLUOBEAM® LX Red can also be used to assist in the imaging of parathyroid glands and can be used as an adjunctive method to assist in the location of parathyroid glands due to the autofluorescence of this tissue. Use of the FLUOBEAM® LX Red device is intended to assist, not replace, experienced visual assessment, and biopsy with conventional histopathological confirmation per standard of care. The system is not to be used to confirm the absence of parathyroid tissue or glands and is only to be used to assist in locating visually identified gland/tissues.
FLUOBEAM® LX Red enables surgeons to observe fluorescent images of parathyroid glands, blood vessels and related tissue perfusion. Fluorescence can be observed thanks to natural fluorescence of parathyroid glands or thanks to a fluorescent product, indocyanine green (ICG), injected intravenously into patients before the surgery allowing the perfusion assessment.
Class 1 infrared laser light is used to excite the fluorescent tissues of parathyroid glands or the ICG and illuminate the regions of a patient's body to be observed. A camera inside the optical head captures the fluorescent image that is used to visualize the parathyroid glands or assess the blood vessels and related tissue perfusion. FLUOBEAM® LX Red consists of the following components: a hardware part with a camera unit (optical head) linked by a specific cable to a control box and a software part with FLUOSOFT™ LX Red imaging software. The optical head contains a video camera and light sources (laser and LEDs) and is used by hand. The control box receives the video signal of the fluorescent image from the optical head, it digitizes it and sends it to a computer that outputs it on a display screen and/or records it. Adjustments of the fluorescent image are possible either by the optical head or via the FLUOSOFT™ LX Red imaging software on the computer.
The modified device FLUOBEAM® LX Red has therefore exactly the same principle of operation of the predicate device. The modified device FLUOBEAM® LX Red is a device modification of the FLUOBEAM® LX device. Compared to the predicate device FLUOBEAM® LX, change of wavelength and detection range is being implemented for modified device FLUOBEAM® LX Red to improve performance in terms of sensitivity for the location of parathyroid glands due to the auto-fluorescence of this tissue while maintaining sensitivity to indocyanine green (ICG).
The FLUOBEAM® LX Red is an imaging system designed to provide real-time near-infrared (NIR) fluorescence imaging of tissue during surgical procedures. It is indicated for visual assessment of blood flow using indocyanine green (ICG) as an adjunctive method, and for assisting in the location of parathyroid glands due to their autofluorescence.
Here’s a breakdown of the acceptance criteria and the study information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
Acceptance Criteria Category | Specific Criteria | Reported Device Performance |
---|---|---|
Safety | Laser Safety | Meets IEC 60825-1 (Class 1 laser product) |
Performance - Imaging | Homogeneity of excitation illumination pattern | Demonstrated equivalence through bench testing |
Performance - Imaging | Live image quality (spatial resolution) | Demonstrated equivalence through bench testing |
Performance - Imaging | Live image quality (acquisition frame rate) | Demonstrated equivalence through bench testing |
Performance - Imaging | Fluorescence sensitivity | Demonstrated equivalence through bench testing |
Clinical Performance (ICG Fluorescence) | ICG fluorescence imaging of blood flow in comparison to predicate device | Performance supported by clinical tests on 5 patients |
Clinical Performance (Autofluorescence) | Sensitivity for location of parathyroid glands due to autofluorescence | Improved performance compared to predicate device (implied by change in wavelength/detection range) |
Clinical Performance (ICG Sensitivity) | Maintaining sensitivity to indocyanine green (ICG) | Maintained sensitivity compared to predicate device (implied by change in wavelength/detection range) |
Overall Performance | Function as intended and perform comparably to predicate device | Tests demonstrated that FLUOBEAM® LX Red met acceptance criteria, functioned as intended, and performed comparably to the predicate device FLUOBEAM® LX. |
2. Sample size used for the test set and the data provenance:
- Test Set Sample Size: 5 patients for the clinical tests on ICG fluorescence imaging.
- Data Provenance: Not explicitly stated, but given it's for a European company (FLUOPTICS SAS, France), the data is likely from a European country. The text does not specify if it was retrospective or prospective, but clinical tests are generally prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not provided in the document. The general indication statement mentions that the device is intended to "assist, not replace, experienced visual assessment, and biopsy with conventional histopathological confirmation per standard of care," suggesting that expert assessment and histology are the ground truth, but doesn't specify the number or qualifications of experts involved in the study itself.
4. Adjudication method for the test set:
- This information is not provided in the document.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a multi-reader multi-case (MRMC) comparative effectiveness study involving human readers and AI assistance was not mentioned. This device is an imaging system, not an AI-driven algorithm with a human-in-the-loop component.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- This device is an imaging system, not solely an algorithm. Its performance is intrinsically linked to the physical hardware (camera, light source) and software for image acquisition and display. Therefore, a "standalone algorithm only" performance study, as typically understood for AI, is not applicable in this context. The document describes bench tests for image quality and sensitivity, and clinical tests for its imaging capabilities.
7. The type of ground truth used:
- For ICG fluorescence imaging for blood flow assessment: The ground truth is implied by comparison to the predicate device in clinical settings, and by the "visual assessment of blood flow." Ultimately, the standard of care for confirming tissue perfusion and viability would involve clinical observation, surgical assessment, and potentially other diagnostic tools.
- For parathyroid gland location: The ground truth is implied to be "visually identified gland/tissues" by experienced visual assessment and "biopsy with conventional histopathological confirmation per standard of care."
8. The sample size for the training set:
- This information is not provided as the document does not mention any machine learning or AI models with distinct training sets. The device's performance is established through engineering design, bench testing, and limited clinical validation.
9. How the ground truth for the training set was established:
- This information is not applicable as the document does not mention a training set in the context of machine learning or AI.
Ask a specific question about this device