Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K203393
    Device Name
    FAST-FIX FLEX
    Date Cleared
    2021-02-01

    (75 days)

    Product Code
    Regulation Number
    878.5000
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FAST-FIX FLEX Meniscal Repair System is indicated for use in meniscal repairs, allograft transplant procedures, and anchoring the allograft to the meniscal rim during allograft transplant procedures.

    Device Description

    FAST-FIX® FLEX is comprised of an implant-suture construct incorporating #2-0 nonabsorbable Ultra-high Molecular Weight Polyethylene (UHMWPE) suture having a pre-tied one-way sliding knot between two non-absorbable polyetheretheketone (PEEK) polymer implants. The suture-implant construct is preloaded into a needle delivery inserter. The needle delivery inserter is offered in two device configurations of curved and reverse curved, whose needle can be modified by the end user with a supplied Bend Tool. The Bend Tool is a sterile, single use optional non-measuring accessory intended to control the bend radius and limit the bend angle to ensure reliable deployment of implants after modification. A sterile, single use slotted cannula is also supplied with the system.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the "FAST-FIX FLEX Meniscal Repair System." This submission is for a surgical device, not an AI/software device. Therefore, many of the requested criteria (e.g., test set, ground truth for AI, MRMC studies) are not applicable.

    However, I can extract the information relevant to the performance testing of this surgical device.

    Here's the breakdown based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a quantitative table of specific acceptance criteria values with corresponding measured device performance numbers. Instead, it states that the device "met performance specifications" and "met all acceptance criteria."

    Test CategoryAcceptance Criteria (General Description)Reported Device Performance (General Description)
    BiocompatibilityPer ISO 10993-1:2018 requirementsAll acceptance criteria were met; deemed biologically safe
    SterilizationSuitability for terminal sterilization under same conditions as predicateCompared to predicate, demonstrated suitability.
    Implant Repair TestingMet performance specificationsMet performance specifications
    Implant Bridge StrengthMet performance specificationsMet performance specifications
    Tensile Strength through Cyclic LoadingMet performance specificationsMet performance specifications
    Implant Failure TestingMet performance specificationsMet performance specifications

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes (e.g., number of devices, number of tests) used for the performance testing. It mentions "nonclinical performance data" but does not detail the size of the test sets for each specific test.

    Data Provenance: Not explicitly stated regarding geographical origin or retrospective/prospective nature. For non-clinical device testing, this typically refers to lab-based experimental data.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. This is a physical surgical device, and its performance is evaluated through material and mechanical testing, not by human experts establishing "ground truth" on images or clinical cases in the way AI models are evaluated.

    4. Adjudication Method for the Test Set

    Not applicable. See point 3. Testing involves objective measurements against predefined specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    Not applicable. See point 3. MRMC studies are relevant for diagnostic AI/imaging devices, not surgical implants. The comparison is made against a predicate device based on technological characteristics and non-clinical performance data, not human reader performance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Not applicable. See point 3.

    7. The Type of Ground Truth Used

    For this surgical device, the "ground truth" or "reference standard" is the predefined performance specification for each test (e.g., minimum tensile strength, specific sterilization parameters, biocompatibility standards). These are typically derived from engineering principles, relevant ISO standards (like ISO 10993-1:2018 for biocompatibility), and comparison to the predicate device's known performance.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device; therefore, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. See point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1