510(k) Data Aggregation

    K Number
    K163705
    Device Name
    EzRay Air W
    Manufacturer
    Date Cleared
    2017-01-24

    (26 days)

    Product Code
    Regulation Number
    872.1800
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    EzRay Air W

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EzRay Air W (Model: VEX-S300W) is an intra-oral dental X-ray system (extra-oral X-ray source system) intended for use by a trained and qualified dentist or dental technician for both adult and pediatric subjects for producing diagnostic dental radiographs for treatment of diseases of the teeth, jaw, and other oral structures using intra-oral image receptors.

    Device Description

    The EzRay Air W (Model: VEX-S300W) is an intra-oral dental X-ray system intended for intra-oral imaging, It consists of X-ray generator, X-ray controller, beam limiting device, operation panel and mechanical arm. The Xray controller allows for accurate exposure control, and the adjustable mechanical arm allows for easy positioning. The system can be used with an imaging system.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the EzRay Air W (Model: VEX-S300W) intra-oral dental X-ray system. This document is focused on demonstrating substantial equivalence to a legally marketed predicate device rather than undergoing extensive clinical trials or performance studies that would typically involve acceptance criteria for an AI-powered device.

    Therefore, the information requested, particularly regarding acceptance criteria for an AI device, sample sizes for test and training sets, expert ground truth establishment, MRMC studies, and standalone performance, is not applicable to this submission. This is a submission for an X-ray imaging device itself, not an AI or software device that processes images.

    However, I can extract the available performance information related to the device's technical specifications and how its performance was compared to the predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance (Focus on X-ray system performance, not AI software):

    Acceptance Criteria (related to X-ray performance)Reported Device Performance (EzRay Air W)Predicate Device (VEX-S100W) PerformanceReference Device (EzRay Air, VEX-P300) Similarities
    Minimum Source to skin distance: >= 200 mm200 mm200 mm (default), 300 mm (option)200 mm
    X-ray field Size (default)60 mm round60 mm round60 mm round
    Focal spot0.4 mm0.4 mm0.4 mm
    Minimum half-value layer1.5 mm Al1.5 mm Al1.5 mm Al
    Electric Power VoltageAC 100-240 VAC 100-120 V / 200 -240 VRechargeable 22.2 V DC Li-ion polymer battery pack
    Rated Current10 A (at AC 250 V)16 A (at AC 250 V)N/A
    Exposure time0.05 - 0.5 seconds in 0.01 increments0.04 - 2.0 seconds in 0.01 increments0.05 - 0.5 seconds in 0.01 increments
    Tube current2.5 or 3.0 mA fixed4 - 7 mA2.5 mA fixed
    Tube voltage65 kVp fixed50 - 70 kVp60 or 65 kVp fixed
    Operation modeManual Mode, Auto ModeManual ModeManual Mode, Auto Mode
    Compliance with Safety StandardsIEC 60601-1, IEC 60601-1-3, IEC 60601-2-65, IEC 60601-1-2, 21 CFR 1020.30, 1020.31IEC 60601-1, IEC 60601-1-3, IEC 60601-2-65, IEC 60601-1-2, 21 CFR 1020.30, 1020.31IEC 60601-1, IEC 60601-1-3, IEC 60601-2-65, IEC 60601-1-2, 21 CFR 1020.30, 1020.31
    Accuracy of loading factorsMet essential performance requirements (e.g.,