Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K242389
    Date Cleared
    2024-10-10

    (59 days)

    Product Code
    Regulation Number
    886.4300
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    EyeGility™ Inserter for Preloaded enVista IOLs

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EyeGility Inserter for enVista preloaded is indicated for folding and inserting of enVista IOL models approved for use with this IOL insertion device as indicated in the IOL approved labeling.

    Device Description

    The EyeGility Inserter is a sterile, single-use device used to fold and insert an intraocular lens through surgical procedure into a human eye. The Inserter provides a tubular pathway through an incision over the iris, allowing delivery of an IOL into the capsular bag.

    AI/ML Overview

    The provided text describes the EyeGility™ Inserter for Preloaded enVista® IOLs and its non-clinical testing to demonstrate substantial equivalence.

    Here's the information broken down as requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    Testing CategoryTestStandardsAcceptance Criteria (Implied by "Pass")Reported Device Performance
    Delivery Verification Testing and Stability TestingFold and Recovery TestISO 11979-3Met established requirementsPass
    Dioptric Power and Image Quality Post DeliveryISO 11979-2Met established requirementsPass
    Lens Dimensions Post DeliveryISO 11979-3Met established requirementsPass
    Surface and Bulk HomogeneityISO 11979-3Met established requirementsPass
    Cosmetic InspectionN/AFree from specified cosmetic defectsPass
    BiocompatibilityBiocompatibility AssessmentISO 10993-1BiocompatiblePass
    Particulate StudyISO 10993-1Within acceptable particulate levelsPass
    Coating Transfer StudyISO 10993-1No unacceptable coating transferPass
    Leachable ExtractionsISO 10993-18Within acceptable leachable limitsPass
    EO SterilizationValidation Product Adoption AnalysisISO 11135Sterilization validatedPass
    EO ResidualsETO/ECH Transfer Engineering StudyISO 10993-7Within acceptable residual limitsPass
    Bacterial EndotoxinBacterial Endotoxin TestANSI/AAMI ST72Within acceptable endotoxin limitsPass
    PackagingTransport Stability TestingISO 11607-1, ISO 11979-6, ASTM F2096Maintained integrity and sterility after transportPass

    2. Sample Size Used for the Test Set and Data Provenance
    The document does not explicitly state the sample sizes for each specific test. It broadly mentions "functional testing" and "nonclinical testing." The data provenance is not mentioned, but given it's a premarket notification for the US market (FDA), such tests are typically conducted by the manufacturer in a controlled laboratory setting. The study is prospective in the sense that the device was designed and then tested against established standards.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
    This information is not applicable. The device is an inserter for Intraocular Lenses (IOLs) and the testing performed is non-clinical, focusing on the physical performance, biocompatibility, and sterility of the device against established standards (e.g., ISO standards). Ground truth, in the context of expert review, is typically relevant for interpretative devices (e.g., AI for medical imaging diagnosis), which is not the case here.

    4. Adjudication Method for the Test Set
    Not applicable, as this is a non-clinical device that is tested against objective, measurable performance standards rather than expert adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    No. The provided text explicitly states: "No clinical testing was completed in support of the subject device." This implies no human readers or AI assistance was involved in a clinical comparative effectiveness study for this device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
    No. This is a physical medical device (an inserter) and does not involve an algorithm or AI.

    7. The type of ground truth used
    The "ground truth" for the non-clinical testing of this device is defined by the acceptance criteria established by international standards (e.g., ISO, ANSI/AAMI, ASTM) for medical devices and IOLs. These standards dictate acceptable performance for parameters like folding and recovery, optical quality, dimensions, biocompatibility, sterilization, and packaging integrity.

    8. The sample size for the training set
    Not applicable. This is a physical medical device and does not involve machine learning or a "training set."

    9. How the ground truth for the training set was established
    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1