Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K242578
    Date Cleared
    2025-01-03

    (127 days)

    Product Code
    Regulation Number
    868.5165
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    EVOLVE Nitric Oxide Delivery System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Evolve DS (delivery system) delivers INOMAX® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed Injector Module, which enables tracking of the gas delivery system waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with ventilators and respiratory care devices for which the Evolve DS has been validated. The Evolve DS provides continuous integrated monitoring of inspired NO2 and NO with a comprehensive alarm system. The Evolve DS also provides monitoring and alarms for the drug delivery system.

    The Evolve DS incorporates a battery that provides up to 4 hours of uninterrupted NO delivery in the absence of external power. The Evolve DS also incorporates an integrated electronic blender that functions as a backup delivery device to provide an adjustable INOmax dose with user supplied oxygen to a manual resuscitator or gas delivery system. The electronic blender incorporates a separate control and delivery pathway that serves as a redundant mechanism for nitric oxide delivery in the event of a main system fault.

    The Evolve DS must only be used in accordance with the indications, warnings and precautions described in the nitric oxide drug packaging inserts and labeling and is indicated for use in term and near-term (>34weeks gestation) neonates with hypoxic respiratory failure associated with clinical or echocardiographic evidence of pulmonary hypertension. The Evolve DS is indicated for a maximum of 14 days use. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU), and secondary targeted clinical setting is the transport of neonates.

    Device Description

    The EVOLVE Nitric Oxide (NO) Delivery System (DS) is used for administration and monitoring of INOMAX (nitric oxide for inhalation). It is comprised of a main delivery device (EVOLVE), an electronic blender (eINOblender) for backup use, and a Cart which holds the EVOLVE as well as INOMAX drug cylinders, an oxygen cylinder and miscellaneous parts.

    The EVOLVE Nitric Oxide Delivery System utilizes Technology component technology to deliver Nitric Oxide gas to the patient. The components consist of a main delivery device (EVOLVE), an electronic blender (eINOblender) for backup use, and a Cart which holds the EVOLVE as well as INOMAX drug cylinders, an oxygen cylinder and miscellaneous parts. In this revision of the EVOLVE Nitric Oxide Delivery System, the changes to the device includes the labeling for compatibility with respiratory care devices.

    AI/ML Overview

    The provided FDA document describes a 510(k) premarket notification for the EVOLVE Nitric Oxide Delivery System. This submission is for an updated version of a previously cleared device (K240410 and K222930), primarily focusing on expanded labeling for compatibility with additional ventilator and breathing devices.

    Therefore, the document does not contain information about acceptance criteria and a study that proves the device meets those criteria in the context of device performance, as it pertains to a labeling update for compatibility. Clinical studies were explicitly stated as "not applicable to this submission."

    Here's a breakdown of what can be extracted based on your request, and what cannot:

    1. A table of acceptance criteria and the reported device performance:

    This information is not provided in the document. The filing is for a change in labeling for compatibility with more ventilator devices, not for demonstrating new performance criteria for the core device. It implicitly relies on the original clearance for the device's performance specifications.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample size for test set: Not explicitly stated as a numerical value for patients or cases. The "test set" in this context refers to the additional ventilator and breathing devices with which compatibility was tested.
    • Data Provenance: The testing was "conducted across all platforms to demonstrate that the EVOLVE Nitric Oxide Delivery System performs within published specifications." This suggests laboratory or bench testing rather than clinical data from a specific geographic region or a prospective/retrospective study on human subjects.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This information is not provided. The testing described is focused on the device's functional performance and compatibility with other medical equipment, not on clinical interpretations that would require expert consensus for ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not provided, as the nature of the "test set" does not involve clinical judgment requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable/provided. The device described is a nitric oxide delivery system, not an AI-assisted diagnostic or therapeutic device that would involve human readers or image interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This information is not applicable/provided for the same reasons as above. The device doesn't involve an algorithm in the sense of AI for diagnostic or interpretive tasks.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    The "ground truth" for the compatibility testing appears to be the published specifications of the EVOLVE Nitric Oxide Delivery System and the functional operation of the connected ventilator devices. The testing aimed to demonstrate that the EVOLVE system "performs within published specifications" when used with the additional ventilators.

    8. The sample size for the training set:

    This is not applicable/provided. The submission is not for a machine learning or AI-based device that would typically have a training set.

    9. How the ground truth for the training set was established:

    This is not applicable/provided.

    Summary of what is available from the document:

    • Device Name: EVOLVE Nitric Oxide Delivery System
    • Purpose of Submission: Updated labeling for compatibility with additional ventilator and breathing devices.
    • Testing Conducted: "Ventilator / Gas Delivery System Validation Test Protocol" was used to validate hazard mitigation and demonstrate performance within published specifications across various ventilator platforms. This was the same protocol used in previous clearances (K222930 and K240410).
    • Clinical Studies: Explicitly stated as "not applicable to this submission."
    • Predicate Device: EVOLVE Nitric Oxide Delivery System (K240410). The current device is considered substantially equivalent.

    In conclusion, the supplied document does not describe the types of performance studies (e.g., diagnostic accuracy, clinical effectiveness) and associated acceptance criteria that would typically involve many of the points in your request. Instead, it focuses on technical compatibility validation for a medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240410
    Date Cleared
    2024-05-28

    (106 days)

    Product Code
    Regulation Number
    868.5165
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    EVOLVE Nitric Oxide Delivery System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Evolve DS must only be used in accordance with the indications, contraindications, warnings and precautions described in the nitric oxide drug packaging inserts and labeling and is indicated for use in term and near-term (>34weeks gestation) neonates with hypoxic respiratory failure associated with clinical or echocardiographic evidence of pulmonary hypertension. The Evolve DS is indicated for a maximum of 14 days use. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU), and secondary targeted clinical setting is the transport of neonates.

    Device Description

    The EVOLVE Nitric Oxide (NO) Delivery System (DS) is used for administration and monitoring of INOMAX (nitric oxide for inhalation). It is comprised of a main delivery device (EVOLVE), an electronic blender (eINOblender) for backup use, and a Cart which holds the EVOLVE as well as INOMAX drug cylinders, an oxygen cylinder and miscellaneous parts.

    AI/ML Overview

    This submission is for an updated labeling for compatibility with additional ventilator and breathing devices for the EVOLVE Nitric Oxide Delivery System, not a new device or a software-driven diagnostic. Therefore, the detailed information about acceptance criteria, device performance, sample sizes, expert involvement, and ground truth establishment typically required for AI/ML device studies is not available in this document.

    The document primarily focuses on demonstrating substantial equivalence to a previously cleared device (K222930) by evaluating the impact of new ventilator compatibility labeling.

    Here's an analysis based on the provided text, addressing the requested points where information is available or applicable:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide a specific table of acceptance criteria and reported device performance in the context of an algorithmic diagnostic. Instead, it refers to the performance of the overall EVOLVE Nitric Oxide Delivery System, which is a hardware-based medical device.

    The "Summary of Non-Clinical Tests" section mentions:
    "The Mallinckrodt Ventilator / Gas Delivery System Validation Test Protocol was used to validate the hazards identified from risk input were mitigated. Testing was conducted across all platforms to demonstrate that the EVOLVE DS performs within published specifications. Ultimately, the requirements necessary for the operation of the EVOLVE Nitric Oxide Delivery System passed."

    This indicates that the acceptance criteria were likely related to the device working within its published specifications when interfaced with the new ventilators, and the reported device performance was that it "passed" these requirements. However, specific quantitative metrics are not provided.

    Acceptance Criteria (Inferred)Reported Device Performance
    Device performs within published specifications when used with additional ventilator devices."Ultimately, the requirements necessary for the operation of the EVOLVE Nitric Oxide Delivery System passed."
    Hazards identified from risk input are mitigated.Validation testing confirmed mitigation.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable and therefore not provided in the document. The study was a non-clinical validation test of device compatibility, not a clinical study on patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable. The "ground truth" here would be the functional performance of the device and its compatibility with ventilators, established through engineering and technical validation rather than expert clinical consensus on diagnostic outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies with human readers to resolve discrepancies in diagnoses or interpretations. This was a non-clinical validation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. The submission is not for an AI device or a diagnostic requiring human reader interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a standalone algorithm performance study was not done. This device is a hardware delivery system, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" used for this non-clinical validation was the functional performance of the device and its compatibility with the listed ventilators according to its published specifications and safety requirements. This would be established through engineering tests and measurements, rather than clinical ground truth types like pathology or expert consensus.

    8. The sample size for the training set

    This information is not applicable. There is no training set as it's not an AI/ML device.

    9. How the ground truth for the training set was established

    This information is not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222930
    Date Cleared
    2023-12-06

    (436 days)

    Product Code
    Regulation Number
    868.5165
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    Evolve Nitric Oxide Delivery System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1