Search Results
Found 3 results
510(k) Data Aggregation
(86 days)
Esophageal Cooling Device
The Esophageal Cooling Device is a thermal regulating device, intended to:
- connect to a Gaymar Medi-Therm III Conductive Hyper/Hypothermia System or Stryker Altrix Precision Temperature Management System to control patient temperature, and
- provide gastric decompression and suctioning.
The Esophageal Cooling Device is a multi-lumen silicone tube that is placed in the esophagus in a similar manner to a standard orogastric tube to modulate a patient's temperature, while simultaneously maintaining access to the stomach to allow gastric decompression and drainage and maintain the functionality of the standard orogastric tube. Modulation and control of patient temperature is achieved by connecting the Esophageal Cooling Device to an external heat exchanger. Two lumens connect to the exchanger. A third central lumen connects to wall suction and is used for standard gastric decompression. The Esophageal Cooling Device is made of standard medical-grade silicone. It is a single-use, disposable, non-implantable device with an intended duration of use of 36 hours or less.
This document is a 510(k) premarket notification for the Esophageal Cooling Device (ECD), seeking to expand its indications for use to include compatibility with an additional external heat exchanger, the Stryker Altrix Precision Temperature Management System.
Here's a breakdown of the requested information based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state formal acceptance criteria with numerical targets. Instead, the non-clinical performance testing focuses on demonstrating an aspect suitable for maintaining thermal transfer capabilities.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Suitable for maintaining thermal transfer capabilities with the Stryker Altrix Precision Temperature Management System | Flow rate testing demonstrated the Esophageal Cooling Device is suitable for maintaining thermal transfer capabilities. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not specified for the "Flow rate" test. The entire document refers to a single device (Esophageal Cooling Device model ECD01-A).
- Data Provenance: The testing was non-clinical performance testing. The provenance (country of origin, retrospective/prospective) is not specified, but typically non-clinical tests are conducted in a laboratory setting by the manufacturer.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. This was a non-clinical performance test, likely involving engineering or technical personnel, not medical experts establishing ground truth in a clinical context.
4. Adjudication Method for the Test Set
Not applicable. This was a non-clinical performance test.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This document describes a medical device, not an AI or imaging diagnostic tool that would typically involve human readers.
6. If a Standalone (i.e., algorithm only without human-in-the loop performance) was Done
Not applicable. This document describes a medical device, not an algorithm. The testing focused on the device's physical performance (flow rate) in conjunction with an external system.
7. The Type of Ground Truth Used
The ground truth for the non-clinical performance testing was based on the physical properties and engineering specifications of the device and the associated external heat exchanger, specifically the objective measurement of "flow rate." It's an objective measurement rather than expert consensus, pathology, or outcomes data.
8. The Sample Size for the Training Set
Not applicable. This device does not involve a "training set" in the context of an algorithm or machine learning.
9. How the Ground Truth for the Training Set Was Established
Not applicable as there is no training set for this type of device.
Ask a specific question about this device
(140 days)
Esophageal Cooling Device
Model #: ECD01-A:
The Esophageal Cooling Device is a thermal regulating device, intended to:
- connect to a Gaymar Medi-Therm III Conductive Hyper/Hypothermia System to control patient temperature, and
- provide gastric decompression and suctioning.
Model #: ECD02-A:
The Esophageal Cooling Device is a thermal regulating device, intended to:
- connect to a Cincinnati Sub-Zero Blanketrol II or Blanketrol III Hyper-Hypothermia System to control patient
temperature, and - provide gastric decompression and suctioning.
The Esophageal Cooling Device (ECD) is a multi-lumen silicone tube that is placed in the esophagus in a similar manner to a standard orogastric tube to modulate a patient's temperature, while simultaneously maintaining access to the stomach to allow gastric decompression and drainage and maintain the functionality of the standard orogastric tube. Modulation and control of patient temperature is achieved by connecting the ECD to an external heat exchanger. Two lumens connect to the external heat exchanger. A third central lumen connects to wall suction and is used for standard gastric decompression. The ECD is made of standard medical-grade silicone. It is a single-use, disposable, nonimplantable device.
The provided text is a 510(k) summary for the Esophageal Cooling Device (ECD). It details the device's characteristics, indications for use, and a comparison to a predicate device, along with a list of non-clinical performance tests. However, this document does not describe acceptance criteria for a device's performance in terms of metrics like sensitivity, specificity, accuracy, or any performance targets against a ground truth.
Instead, the "acceptance criteria" can be inferred as the successful completion of various engineering and safety tests, and the "study" is the non-clinical performance testing conducted by the manufacturer. These tests are primarily to demonstrate that the device is safe, functions as intended, and is substantially equivalent to a previously cleared device.
Therefore, many of the requested fields cannot be directly answered from the given input, as the document focuses on regulatory clearance for a physical medical device, not a diagnostic or AI-driven system that would typically undergo studies with performance metrics against a ground truth.
Here's an attempt to answer the questions based on the provided text, while acknowledging its limitations for certain categories:
1. A table of acceptance criteria and the reported device performance
Since this document describes a physical medical device (Esophageal Cooling Device) and the "studies" are non-clinical engineering and safety tests, the acceptance criteria are generalized to "suitability for intended use" rather than specific numerical performance metrics typically seen for diagnostic devices or AI systems.
Acceptance Criteria (Inferred from Test Objective) | Reported Device Performance |
---|---|
Biocompatibility: Materials suitable for intended use | "All materials used to manufacture the ECD are suitable for the intended use." |
Tensile Force: Suitable for insertion/removal | "ECD is suitable for insertion into and removal from the esophagus." |
Burst Strength: Suitable for normal operating pressures | "ECD is suitable for normal operating pressures." |
Ultimate Material Strength: Suitable for specified operating conditions | "ECD is suitable for normal operating pressures under worst-case manufacturing conditions." |
Leakage: Prevention of stomach filling with coolant | "ECD is suitable for preventing the patient's stomach from filling with coolant." |
Resistance to Vacuum: Suitable for normal operating pressures | "ECD is suitable for normal operating pressures." |
Flow Rate: Maintenance of thermal transfer capabilities | "ECD is suitable for maintaining thermal transfer capabilities." |
Packaging Integrity: Maintenance of cleanliness | "ECD packaging is suitable for maintaining the cleanliness of the ECD." |
Overall: No new questions of safety or effectiveness (compared to predicate) | "The ECD01-A and ECD02-A have technological characteristics that do not raise new questions of safety or effectiveness when compared to the predicate device." |
Shelf Life: Increased from 1 year to 3 years | "Non-clinical performance testing was also conducted to show the technological characteristics of the subject devices are suitable to increase the shelf life from 1 year to 3 years." |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify sample sizes (e.g., number of devices tested) for the non-clinical performance tests. These are typically engineering tests, likely involving a small number of physical samples from manufacturing lots. Data provenance in terms of country of origin or retrospective/prospective is not applicable as these are laboratory/benchtop tests, not human studies.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. This device is not an imaging or diagnostic device that requires expert ground truth establishment for a test set. The "ground truth" here is based on engineering specifications and existing regulatory standards for medical device safety and performance.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This concept applies to studies where expert opinions (e.g., medical image interpretations) are being adjudicated. For engineering tests of a physical device, results are typically determined by measurement against established standards, not expert consensus or adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI-assisted diagnostic device, nor is it a device that involves "human readers."
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical medical device, not an algorithm.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
For biocompatibility, the ground truth is established by ISO 10993 standards (e.g., cell viability for cytotoxicity, irritation scores for irritation). For mechanical tests (tensile, burst, strength, leakage, vacuum, flow rate), the ground truth is established by industry-accepted engineering standards and internal specifications derived from the device's intended use and comparison to the predicate device. The ultimate "ground truth" for regulatory clearance is "substantial equivalence" to the predicate device in terms of safety and effectiveness.
8. The sample size for the training set
Not applicable. This device does not use a "training set" in the context of machine learning or algorithms. Its design and manufacturing are based on established engineering principles and prior device models.
9. How the ground truth for the training set was established
Not applicable, as there is no "training set." The design and testing are based on engineering principles and regulatory guidances relevant to medical devices.
Ask a specific question about this device
(410 days)
ESOPHAGEAL COOLING DEVICE
The Esophageal Cooling Device is a thermal regulating device, intended to:
- · connect to a Gaymar Medi-Therm III Conductive Hyper/Hypothermia System to control patient temperature, and
- · provide gastric decompression and suctioning.
The Esophageal Cooling Device (ECD) is a silicone tube with three lumens that is placed in the esophagus (Figure 1). The intended function of the device is to control a patient's temperature. while simultaneously maintaining access to the stomach to allow gastric decompression and drainage. Modulation and control of patient temperature is intended to be achieved by connecting the ECD to an external heat exchanger and circulating temperaturecontrolled fluid (distilled water). Two lumens (the cooling lumens) connect to the external heat exchanger and are in contact with the esophageal tissues. A third central lumen connects to wall suction and is used for standard gastric decompression. A web supports the inner gastric lumen and separates the cooling lumens.
The ECD is made of standard medical-grade silicone. It is a single-use, disposable, nonimplantable device with a stated intended duration of use of 36 hours or less.
The ECD is intended to be used in conjunction with the Gaymar Medi-Therm III Conductive Hyper/Hypothermia System (cleared under K100585) with the following operating specifications:
- · Dead Head Pressure: Maximum 9 psi (62 kPa)
- Flow: Minimum 16 gallons/h (60.6 L/hr)
- · Water Temperature Control Range: 4°C 42°C
The Gaymar System supplies temperature-controlled water through a connector hose to the ECD. An accessory probe interfaces between the Gaymar System and the patient to sense patient temperature, which is displayed on the Gavmar System's control panel. The Gaymar device controls water temperature by mixing hot and cold water using hot and cold solenoid valves under microprocessor control and includes a circulating pump, heater, and refrigeration system.
Here's a summary of the acceptance criteria and the study information for the Esophageal Cooling Device, structured as requested:
Acceptance Criteria and Device Performance
Acceptance Criteria Category | Specific Test / Characteristic | Acceptance Criteria | Reported Device Performance / Results |
---|---|---|---|
Biocompatibility | Cytotoxicity Evaluation | Non-cytotoxic | Non-cytotoxic |
Kligman Maximization Test (Sensitization) | Sensitization rate = 0% | Sensitization rate = 0%, Sensitization grade = “Weak” | |
Intracutaneous Injection Test (Irritation) | No difference between mean test article score and mean control score | No difference between mean test article score and mean control score | |
Mechanical Integrity | Tensile force | ≥15 N | Passed; ≥15 N |
Burst strength | ≥15 psi | Passed; ≥15 psi | |
Ultimate material strength | ≥5 MPa | Passed; ≥5 MPa | |
Leakage | No visible evidence of leakage | Passed; no visible evidence of leakage | |
Resistance to vacuum | No visible evidence of occlusion | Passed; no visible evidence of occlusion | |
End cap axial deflection testing (puncture/irritation) | ≤ mean force required to cause end cap axial deflection for a currently marketed enteral feeding connector | Passed; ≤ mean force required to cause end cap axial deflection for a currently marketed enteral feeding connector | |
Thermal Performance | Flow rate (thermal transfer capacity) | ≥31 L/h | Passed; ≥31 L/h |
Hardware verification (temperature probe accuracy) | Temperature probe accurate to within ± 1.0°C | Passed; temperature probe accurate to within ± 1.0°C | |
Hardware verification (coolant temperature adjustment) | Coolant temperature at lowest setting reached 4.5°C within 30 minutes | Passed; coolant temperature at lowest setting reached 4.5°C within 30 minutes | |
Hardware verification (safety features: alarms, indicators) | All alarms and indicators sounded appropriately (high/low temperature alarms, coolant flow occlusion alarm) | Passed; all alarms and indicators sounded appropriately | |
Shelf-Life/Packaging | Shelf life | Device maintains performance after 1 year accelerated aging (by repeating bench performance testing) | Passed (bench performance testing repeated after one year of accelerated aging) |
Transportation simulation | Packaging capable of withstanding shipping without product damage (in accordance with ISTA 2A 2011) | Passed in accordance with ISTA 2A 2011 | |
Packaging integrity | Packaging capable of preventing contamination of the device from outside sources (in accordance with ASTM F2096-11) | Passed in accordance with ASTM F2096-11 | |
Animal Study Outcomes | Esophageal mucosa tissue integrity | No adverse effects from the Esophageal Cooling Device identified in gross or histological analyses at necropsy. | No adverse effects from the Esophageal Cooling Device were identified in gross or histological analyses, demonstrating that the device did not cause harm to the esophageal lining following use (except for one death due to endotracheal tube issues, not device related). |
Successful cooling and rewarming | Swine successfully cooled and rewarmed; average deviation from goal temperature within acceptable limits during maintenance. | All five swine were cooled successfully (average rate of temperature decrease of 1.3 °C/hour). Warming rates averaged 0.4℃/hour. Average deviation from goal temperature was 0.2℃ during the maintenance phase. | |
Gastric decompression and suctioning | Gastric fluids successfully collected. | Between 300 and 825 mL of gastric fluids were collected from each animal. | |
Clinical Observations | Adequate device performance in achieving and maintaining goal temperatures (based on clinical summaries) | Adequate device performance at achieving and maintaining goal temperatures. | Each of the ten temperature charts showed adequate device performance at achieving and maintaining goal temperatures (target temperatures 33℃ - 36.5℃). |
Absence of adverse events or device malfunctions (based on clinical summaries) | No adverse events or device malfunctions reported. | No adverse events or device malfunctions were reported. | |
Gastric suctioning worked as intended without need for additional tube insertion (based on clinical summaries) | Gastric suctioning worked as intended without the need for additional tube insertion. | The gastric suctioning was reported by the centers to have worked as intended without the need for additional tube insertion. |
Study Details for Esophageal Cooling Device
1. Sample Size and Data Provenance
- Test Set (Bench/Animal):
- Bench Testing: The document does not specify a precise sample size (N) for each individual bench test, but refers to "final manufactured product" exposed to worst-case conditions.
- Animal Study: 5 female domestic swine (61-70 kg).
- Data Provenance: The animal study was published in "Resuscitation 2013; 8; p. 1619-1624" by Kulstad et al. The country of origin for the animal study is not explicitly stated in the provided text, but it's a published research study. The study was prospective in nature, involving experimental temperature modification in the animals.
- Test Set (Clinical Data Summaries):
- Sample Size: 16 patient summaries.
- Data Provenance: Obtained from centers outside the United States where the device is commercially available. These were summaries of therapeutic cooling, maintenance, and rewarming, making them retrospective clinical observations rather than a formal prospective clinical trial.
2. Number of Experts and Qualifications for Ground Truth (Clinical Data)
- Number of Experts: Not applicable/not specified for establishing ground truth in a formal sense. The clinical data provided consists of summaries from various centers where the device was in use. The "truth" in these summaries relates to observed temperature control and absence of reported adverse events, presumably documented by the healthcare professionals at those institutions.
- Qualifications of Experts: Not specified. The summaries were from "healthcare professionals with training in the use of orogastric tubes and the use of the Gaymar Medi-Therm III Conductive Hyper/Hypothermia System," as noted in the device limitations for general use.
3. Adjudication Method for the Test Set (Clinical Data)
- Adjudication Method: None explicitly stated. The document refers to "clinical data summaries," "reports," and "feedback from centers," implying a collection of independently generated clinical observations rather than a centralized, adjudicated test set.
4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Was it done? No. This type of study is typically associated with diagnostic AI tools where human readers interpret cases with and without AI assistance to measure improvement. The Esophageal Cooling Device is a hardware device for therapeutic temperature management, not an interpretive diagnostic tool.
- Effect Size of Human Readers' Improvement: Not applicable, as no such study was performed or is relevant for this device.
5. Standalone Performance Study (Algorithm only)
- Was it done? Yes, a form of standalone performance was done for the device itself through the extensive bench testing and the animal study.
- Bench Testing: Demonstrated the device's functional performance (mechanical integrity, thermal transfer, compatibility with the external system, material strength, leakage, vacuum resistance, etc.) in a controlled environment without direct human intervention beyond setting up the tests.
- Animal Study: Evaluated the device's ability to induce, maintain, and reverse therapeutic hypothermia, as well as its safety profile (esophageal tissue integrity, gastric suctioning) in a physiological system. While humans operated the external heat exchanger, the performance measured was of the device itself and its interaction with the biological system.
6. Type of Ground Truth Used
- Bench Testing: Engineering specifications, physical measurements, and industry standards (e.g., ISO, ASTM, pre-defined thresholds for force, pressure, temperature, flow rates).
- Animal Study:
- Tissue integrity: Pathology (gross pathological and histological analysis by a veterinary pathologist at necropsy).
- Temperature control: Direct measurements of swine body temperature and external heat exchanger settings.
- Gastric suctioning: Quantifiable volume of gastric fluids collected.
- Clinical Data Summaries: Physician observations, patient temperature charts, and reported adverse events/device malfunctions from clinical practice, which can be considered a form of outcomes data or reported clinical experience.
7. Sample Size for the Training Set
- Training Set: Not applicable. The document describes pre-market testing and clinical observations for a new medical device, not a machine learning algorithm that requires a "training set." The device itself is not an AI/ML algorithm. The "training" for the device would be its iterative design, engineering, and manufacturing process based on established principles, not data.
8. How the Ground Truth for the Training Set Was Established
- Ground Truth for Training Set: Not applicable, as there is no algorithms with a training set for this device.
Ask a specific question about this device
Page 1 of 1