Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K182649
    Date Cleared
    2018-11-09

    (46 days)

    Product Code
    Regulation Number
    876.5980
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Enteral Feeding Syringes with Enfit Connectors

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Enteral Feeding Syringe with ENFit Connector is indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syringes are intended to be used in clinical or home care settings by users ranging from laypersons (under the supervision of a clinician) to clinicians in all age groups.

    Device Description

    The proposed device, Enteral Feeding Syringes with ENfit Connectors, is provided sterile, single use. It has 3 components: (1) barrel, (2) plunger, and (3) piston. The Enteral Feeding Syringes with ENfit Connectors is available in various size, including 1ml, 2.5ml, 5ml, 10ml, 20ml, 60ml and 100ml. In addition, it contains low dose enteral specific connectors. The Enteral Feeding Syringes with ENfit Connectors is sterilized by Ethylene Oxide Gas to achieve a SAL of 10-6 and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of 5 years.

    AI/ML Overview

    The provided text describes a 510(k) summary for Enteral Feeding Syringes with ENFit Connectors, which are medical devices, not AI/ML-driven software. Therefore, the questions regarding acceptance criteria, study details, expert involvement, and ground truth for an AI device are not applicable to this document. The document details the substantial equivalence of the proposed device to a predicate device based on non-clinical tests and conformance to established standards.

    Here's an analysis of the provided information, focusing on the device's conformity to standards rather than AI-specific criteria.

    Device: Enteral Feeding Syringes with ENFit Connectors

    1. Table of Acceptance Criteria and the Reported Device Performance

    The acceptance criteria for this medical device are based on compliance with several ISO standards and specific performance tests. The device's reported performance demonstrates compliance with these standards.

    Acceptance Criteria (Standard / Test)Reported Device Performance
    ISO 7886-1:2017 Sterile hypodermic syringes for single use - Part 1: Syringes for manual useComplies with the standard
    ISO 80369-3:2016 Small-bore connectors for liquids and gases in healthcare applications — Part 3: Connectors for enteral applicationsComplies with the standard
    ISO 80369-20:2015 Small-bore connectors for liquids and gases in healthcare applications — Part 20: Common test methodsComplies with the standard
    ISO 10993-5:2009 Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicityNo Cytotoxicity
    ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitizationNo Irritation to Skin, No significant evidence of sensitization
    Dimension analysisComplies with requirements
    Dosing accuracy testComplies with requirements
    Connector incompatibility testComplies with requirements
    Sterilization Assurance Level (SAL)10⁻⁶ (EO Sterilized)
    Endotoxin Limit20 EU per device

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes for each non-clinical test conducted. It states that "Non clinical tests were conducted to verify that the proposed device met all design specifications" and that the "test results demonstrated that the proposed device complies with the following standards." The provenance of the data is from in-house non-clinical testing performed by Jiangsu Kangbao Medical Equipment Co., Ltd.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable as the device is not an AI/ML system requiring expert-established ground truth for a test set. The validation relies on technical specifications and international standards, which are evaluated by engineers and quality control personnel.

    4. Adjudication method for the test set

    This question is not applicable as there is no mention of a human-based adjudication process for the non-clinical tests described. Compliance is determined by adherence to established measurement and testing protocols defined by the ISO standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable as the device is not an AI/ML system, and no human readership studies are referenced.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable as the device is a physical medical instrument, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For this physical device, the "ground truth" for non-clinical performance and safety is defined by the established international standards (ISO standards) and engineering specifications. For biocompatibility, this is determined by laboratory tests for cytotoxicity, irritation, and sensitization as per ISO 10993. For sterilization, it's about achieving a specific SAL.

    8. The sample size for the training set

    This question is not applicable as the device is not an AI/ML system and does not involve a training set.

    9. How the ground truth for the training set was established

    This question is not applicable as the device is not an AI/ML system and does not involve a training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1