510(k) Data Aggregation

    K Number
    K142539
    Manufacturer
    Date Cleared
    2015-01-08

    (121 days)

    Product Code
    Regulation Number
    876.5980
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    EnteraLite Infinity Enteral Pump Delivery Set, 1200 mL Enteral Feeding Delivery Set

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The devices in this product family are used to dispense liquid nutrients (feeding solution) at a preprogrammed pump or user controlled rate. These enteral feeding sets interface with the patient's feeding tube and may use gravity or an enteral feeding pump to dispense feeding solution. The devices may include a bag to contain the feeding solution and/or a spike to connect to a prefilled container.

    Device Description

    These enteral feeding sets are designed to be used with or without an enteral feeding pump, (pump is not in scope for this 510k) model specific. Refer to diagrams below for explanation of components. The enteral sets terminate in a bonded conical connector, the ENFit connector; that is designed to be incompatible with ISO 594-1/2 and other non-enteral feeding connections. The ENFit connector is a new connector and it is anticipated that not all patient feeding tubes will be compatible with it, so the administration sets include an adapter that connects to the ENFit connector on one end while providing the traditional "Christmas Tree" shaped transitional stepped connector on the other.

    AI/ML Overview

    This document appears to be a 510(k) summary for an enteral feeding set, which is a medical device. Based on the provided text, the device is not an AI/ML device. Therefore, the specific criteria for AI/ML device studies (such as sample sizes for test/training sets, number of experts for ground truth, adjudication methods, MRMC studies, or standalone performance) are not applicable or detailed in this document.

    The document primarily focuses on demonstrating substantial equivalence to a predicate device, as required for 510(k) submissions. This is achieved through non-clinical performance data.

    Here's an analysis of the acceptance criteria and supporting study information, focusing on the available details from the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document lists various non-clinical tests performed to support substantial equivalence. While specific quantitative acceptance criteria are not explicitly stated in a table format in the provided text, the performance is reported as "Passed" or "Acceptable" for each test.

    Test CategoryReported Device Performance
    Mechanical/Physical Integrity
    Bond StrengthPassed
    Leak TestingPassed
    Volumetric accuracy testingPassed
    Dimensional analysisPassed
    Accelerated AgingPassed
    Biocompatibility
    BiocompatibilityPassed
    Connection Safety
    Misconnection assessmentPassed
    Enteral Connector Risk Management ReportAcceptable
    Human Factors
    Human Factors TestingAcceptable
    Failure Modes and Effects Analysis (FMEA)Acceptable
    ISO 80369-3 Compliant Testing
    Falling drop positive pressure liquid leakagePassed
    Stress crackingPassed
    Resistance to separation from axial loadPassed
    Resistance to unscrewingPassed
    Resistance to overridingPassed
    Disconnection by unscrewingPassed
    Falling drop positive pressure liquid leakage after 20 cycles of connection and separationPassed

    2. Sample size used for the test set and the data provenance

    Not applicable for an AI/ML device. For this physical medical device, specific sample sizes for tests like bond strength or leak testing are not provided in the summary. The provenance would be the manufacturing facility where the tests were performed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a physical device, and its performance is evaluated against engineering specifications and industry standards, not through expert-labeled ground truth in the context of an AI/ML model.

    4. Adjudication method for the test set

    Not applicable. This is a physical device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This is not an AI/ML device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. This is not an AI/ML device.

    7. The type of ground truth used

    The "ground truth" for this device is defined by:

    • Engineering specifications: The device's design inputs and required performance characteristics.
    • Industry standards: Notably ISO 80369-3 for small-bore connectors.
    • Predicate device performance: The new device is compared to a legally marketed predicate device (K012147) to demonstrate similar safety and effectiveness.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML device.


    Summary for the provided document:

    The document describes an enteral feeding set, which is a physical medical device, not an AI/ML device. Therefore, most of the requested information pertaining to AI/ML device studies (such as test/training set sample sizes, expert ground truth, adjudication methods, MRMC studies, or standalone algorithm performance) is not relevant or available in this submission.

    The device's conformity is established through non-clinical performance data, showing that it "Passed" or was "Acceptable" for a series of tests related to its physical integrity, biocompatibility, connection safety (especially ISO 80369-3 compliance), human factors, and risk management. The "ground truth" here refers to the device meeting its design specifications and applicable industry standards. The submission asserts substantial equivalence to a predicate device based on these non-clinical tests, without requiring clinical testing.