510(k) Data Aggregation

    K Number
    K230076
    Date Cleared
    2023-03-08

    (57 days)

    Product Code
    Regulation Number
    878.4630
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Enhanced AURORA™ Medical Diode System, and related accessories

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Enhanced AURORA™ Medical Diode system and related accessories is indicated for use in targeted PUVA photochemistry and UVB phototherapy for the treatment of skin conditions including psoriasis, vitiligo, atopic dermatitis, eczema, and seborrheic dermatitis. In addition, the system UVB channel is indicated for the treatment of leukoderma.

    Device Description

    The Enhanced AURORA™ Medical Diode system and related accessories is designed to allow a clinician to easily apply UVA or UVB light to the patient's affected skin. The clinician selects the desired treatment type (UVA or UVB light) and applies the light through the appropriate handpiece by pointing the exit aperture towards the affected skin area.

    The device is comprised of four main components:

    • A base station
    • A touchscreen where the user can enter system commands
    • Two handpieces connected to the base by power cables. One handpiece contains diode emitters that, when activated, emit UVA light. The other handpiece emits UVB light. Either handpiece will only operate if the appropriate wavelength is selected, so a user cannot accidentally activate the incorrect handpiece. Activation of energy occurs on depression of a trigger in the handpiece.
    • A printer accessory that can print out verification of treatment parameters for hard copy records.
    AI/ML Overview

    The document describes the acceptance criteria and study for the Enhanced AURORA™ Medical Diode system.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    UVA energy dosage output within 10% of user-selected power settingsConfirmed within 10% of user-selected power settings
    UVB energy dosage output within 10% of user-selected power settingsConfirmed within 10% of user-selected power settings

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state a 'test set' in the context of clinical data or patient samples. The performance data is based on bench testing only. Therefore, there is no sample size for a test set of patient data, nor is there information about data provenance (country of origin, retrospective/prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The performance data is derived from bench testing of the device's output, not from expert interpretation of medical images or patient data requiring ground truth establishment by experts.

    4. Adjudication method for the test set

    Not applicable, as a test set requiring adjudication by experts (e.g., for diagnostic accuracy) was not conducted. The assessment was based on physical measurements during bench testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a phototherapy system, not an AI-powered diagnostic tool, and therefore, an MRMC comparative effectiveness study involving human readers and AI assistance was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is a medical phototherapy system, not an algorithm. Its performance is assessed by its physical output characteristics during bench testing, not by an algorithm's standalone performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance was the user-selected power settings, against which the actual UVA and UVB energy dosage output was measured. This is a technical performance measurement, not a clinical ground truth.

    8. The sample size for the training set

    Not applicable. This document describes a medical device for phototherapy, not a machine learning model requiring a training set.

    9. How the ground truth for the training set was established

    Not applicable, for the same reason as above.