Search Results
Found 9 results
510(k) Data Aggregation
(115 days)
EnSite X EP System, AutoMark Index Software License (ENSITE-AMI-01); EnSite X EP System, AutoMark Index
Ask a specific question about this device
(99 days)
EnSite X EP System
The EnSite X EP System is a suggested diagnostic tool in patients for whom electrophysiology studies have been indicated.
The EnSite X EP System provides information about the electrical activity of the heart and displays catheter location during conventional electrophysiological procedures.
The EnSite™ X EP System is a catheter navigation and mapping system. A catheter navigation and mapping system is capable of displaying the 3-dimensional (3-D) position of conventional and Sensor Enabled™ (SE) electrophysiology catheters, as well as displaying cardiac electrical activity as waveform traces and as three-dimensional (3D) isopotential and isochronal maps of the cardiac chamber.
The contoured surfaces of the 3D maps are based on the anatomy of the patient's own cardiac chamber. The system creates a model by collecting and labeling the anatomic locations within the chamber. A surface is created by moving a selected catheter to locations within a cardiac structure. As the catheter moves, points are collected at and between all electrodes on the catheter. A surface is wrapped around the outermost points.
The provided FDA 510(k) clearance letter for the EnSite™ X EP System (K251234) details the device's regulatory pathway and general testing conducted. However, it does not contain the specific information required to populate a table of acceptance criteria and reported device performance. It focuses on the regulatory aspects, substantial equivalence to a predicate device, and the general types of testing performed (e.g., software verification, amplifier design verification, system design validation) to demonstrate that the device meets user requirements and its intended use.
The document states: "Design verification activities were performed and met their respective acceptance criteria to ensure that the devices in scope of this submission are substantially equivalent to the predicate device." However, the specific acceptance criteria (e.g., a numerical threshold for accuracy or precision) and the reported device performance values against those criteria are not presented in this public clearance letter.
Similarly, the letter does not provide details regarding:
- Sample sizes used for test sets (beyond stating "design verification" and "system design validation" were performed).
- Data provenance (country of origin, retrospective/prospective).
- Number of experts, their qualifications, or adjudication methods for establishing ground truth for any test set.
- Whether a multi-reader multi-case (MRMC) comparative effectiveness study was done, or any effect size for human readers.
- Whether standalone (algorithm-only) performance was assessed.
- The type of ground truth used (expert consensus, pathology, outcomes data).
- The sample size for the training set.
- How ground truth for the training set was established.
This type of detailed performance data is typically found within the confidential 510(k) submission itself, not routinely published in the public clearance letter.
Therefore,Based on the provided FDA 510(k) clearance letter for the EnSite™ X EP System, the following information can be extracted regarding the device's acceptance criteria and the study that proves it meets those criteria:
Key Takeaway: The provided FDA 510(k) clearance letter asserts that acceptance criteria were met through various design verification and validation activities, demonstrating substantial equivalence to a predicate device. However, it does not disclose the specific numerical acceptance criteria or the quantitative results of the device's performance against those criteria. The details below are based on what is stated or can be inferred from the document.
1. Table of Acceptance Criteria and Reported Device Performance
As per the provided document, specific numerical acceptance criteria and reported device performance data are not explicitly stated or detailed. The document generally states:
"Design verification activities were performed and met their respective acceptance criteria to ensure that the devices in scope of this submission are substantially equivalent to the predicate device."
And
"System Design Validation to confirm the system could meet user requirements and its intended use after modifications"
Without specific numerical cut-offs or performance metrics (e.g., accuracy, precision, error rates), a table cannot be populated as requested. The clearance indicates that internal testing demonstrated the device met pre-defined acceptance criteria, but those criteria and the actual performance results are not publicly available in this document.
Acceptance Criteria Category (Presumed) | Specific Acceptance Criteria (Not specified in document) | Reported Device Performance (Not specified in document) | Met? (Inferred from clearance) |
---|---|---|---|
System Functionality | (e.g., Catheter position display accuracy, Cardiac electrical activity waveform fidelity, 3D map creation accuracy) | (Specific quantitative results, e.g., X mm accuracy) | Yes (Implied by clearance) |
Safety & Effectiveness | (e.g., Conformity to electromagnetic compatibility, software robustness, risk mitigation effectiveness) | (e.g., Passes all EMC tests, no critical software bugs identified) | Yes (Implied by clearance) |
User Requirements | (e.g., System usability, interface responsiveness) | (e.g., Demonstrates ability to meet intended use) | Yes (Implied by clearance) |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size Used for Test Set: Not specified in the provided document. The document mentions "Design verification activities" and "System Design Validation" but does not give the number of cases, patients, or data points used for these tests.
- Data Provenance (e.g., country of origin of the data, retrospective or prospective): Not specified in the provided document.
3. Number of Experts Used to Establish Ground Truth and Qualifications
- Number of Experts: Not specified.
- Qualifications of Experts: Not specified.
- (It's common for electrophysiology systems that ground truth would be established by electrophysiologists, but this document does not confirm that.)
4. Adjudication Method for the Test Set
- Adjudication Method: Not specified. (e.g., 2+1, 3+1, none)
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- MRMC Study: No indication that an MRMC comparative effectiveness study was performed or required for this 510(k) clearance. The focus of this submission is on substantial equivalence to a predicate device, which often relies on non-clinical testing for software updates or minor changes, rather than clinical efficacy studies comparing human readers with and without AI assistance.
- Effect Size of Human Readers Improvement with AI vs. Without AI Assistance: Not applicable/Not provided, as an MRMC study is not mentioned.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Standalone Performance: The document describes "Software Verification at unit, software and system level" and "Amplifier Design Verification," which are types of standalone-like algorithmic or component-level testing. However, the exact metrics and results for pure "algorithm-only" performance (e.g., for automated mapping or analysis features if present) are not detailed. The system is described as a "diagnostic tool" that "provides information" and "displays catheter location," implying human interaction is integral.
7. The Type of Ground Truth Used
- Type of Ground Truth: Not explicitly stated. Given the nature of an EP system, ground truth would likely involve a combination of:
- Validated phantom models: For physical accuracy of catheter tracking and mapping.
- Clinical expert consensus: For validating the interpretation of electrical activity and the accuracy of generated 3D maps or anatomical models.
- Reference measurements: From other validated systems or direct measurements during testing.
- The document implies ground truth was used for "Design verification" and "System Design Validation," which "confirm the system could meet user requirements."
8. The Sample Size for the Training Set
- Training Set Sample Size: Not applicable/Not specified. This 510(k) is for a software update (v5.0) to an existing system (EnSite™ X EP System, predicate K242016). The document describes changes related to compatibility with new catheters and ultrasound systems, rather than the development of entirely new AI/ML algorithms requiring a "training set" in the conventional sense of deep learning. While software is involved, the primary testing discussed is verification and validation, not model training.
9. How the Ground Truth for the Training Set Was Established
- Ground Truth for Training Set Establishment: Not applicable/Not specified, as the document does not indicate the use of a "training set" in the context of machine learning model development. The 'ground truth' concept would apply more to the test and validation steps, as discussed in point 7.
Ask a specific question about this device
(30 days)
EnSite X EP System
The EnSite ™ X EP System is a suggested diagnostic tool in patients for whom electrophysiology studies have been indicated.
The EnSite ™ X EP System provides information about the electrical activity of the heart and displays catheter location during conventional electrophysiological (EP) procedures.
When used with the TactiSys™ Quartz Equipment, the EnSite™ X EP System Contact Force Module is intended to provide visualization of force information from compatible catheters.
The EnSite ™ X EP Surface Electrode Kit is indicated for use with the EnSite ™ X EP System in accordance with the EnSite™ X EP System indications for use.
The EnSite™ X EP System is a catheter navigation and mapping system. A catheter navigation and mapping system is capable of displaying the 3-dimensional (3-D) position of conventional and Sensor Enabled™ (SE) electrophysiology catheters, as well as displaying cardiac electrical activity as waveform traces and as three-dimensional (3D) isopotential and isochronal maps of the cardiac chamber.
The contoured surfaces of the 3D maps are based on the anatomy of the patient's own cardiac chamber. The system creates a model by collecting and labeling the anatomic locations within the chamber. A surface is created by moving a selected catheter to locations within a cardiac structure. As the catheter moves, points are collected at and between all electrodes on the catheter. A surface is wrapped around the outermost points.
Here's an analysis of the provided text regarding the acceptance criteria and study proving device performance.
Important Note: The provided document is an FDA 510(k) clearance letter and a 510(k) Summary. It does not contain detailed information about a comprehensive clinical study to establish performance, especially not in the depth typically seen for AI/ML device clearances (e.g., MRMC studies, specific performance metrics against ground truth for diagnostic accuracy). This document primarily focuses on demonstrating substantial equivalence to a predicate device, specifically for software updates to an existing system. Therefore, many of the requested details, particularly those related to a new AI/ML model's performance validation, are not present in this type of submission.
Based on the provided document, here's what can be inferred and what information is missing:
The device in question is the EnSite™ X EP System, and the 510(k) (K242016) is a "Special 510(k)" for software updates (Version 3.1) to an already cleared predicate device (EnSite™ X EP System, K231415). Special 510(k)s are typically used for changes that do not alter the fundamental technology or indications for use.
The "study" referenced in this context is primarily software verification and validation testing to ensure the updated software maintains the safety and effectiveness of the previously cleared predicate device. It's not a clinical performance study in the sense of evaluating a new diagnostic algorithm's accuracy against a clinical ground truth with human agreement.
1. Table of Acceptance Criteria & Reported Device Performance
Acceptance Criteria (Inferred) | Reported Device Performance |
---|---|
Functional Performance: | |
Software functions as intended | "Software Verification at unit, software and system level" was performed. |
Compatibility with specific hardware (Advisor™ HD Grid X, Sensor Enabled™ Farawave catheters) | "Compatibility with Advisor™ HD Grid X, Sensor Enabled™ Farawave catheter visualization" was a documented update. |
Safety & Effectiveness: | |
No new questions of safety or effectiveness are raised. | "All risks were mitigated to acceptable levels. No new questions of safety or effectiveness were raised." |
Device remains substantially equivalent to predicate. | "The testing completed and submitted in this Special 510(k) provides objective evidence the subject device is substantially equivalent to the predicate device." |
Compliance with relevant standards (IEC 62304, IEC 60601-1-2, ISO 14971) | Device "developed and tested in accordance with the following industry guidance documents and standards: [...] IEC 62304: 2015-06 Edition 1.1, Medical Device Software - Software Life Cycle Processes; IEC 60601-1-2:2020-09 Edition 4.1, Medical Electrical Equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic disturbances – Requirements and tests; ISO 14971:2019 Medical Devices - Application of Risk Management to Medical Devices." |
Performance of updated feature functionality maintained | "Performance Testing of updated feature functionality" was performed. |
Bug fixes are effective and do not introduce new issues. | "Minor software updates (bug fixes) and enhancements" were implemented. |
Design verification activities met acceptance criteria. | "Design verification activities were performed and met their respective acceptance criteria to ensure that the devices in scope of this submission are substantially equivalent to the predicate device." |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not specified in terms of clinical cases or data points. The "test set" here refers more to the scope of software testing rather than a clinical dataset for diagnostic performance.
- Data Provenance: Not specified. Given it's a software update to an existing system, the "data" would likely be the system's internal functions, compatibility with hardware, and simulated/recorded electrophysiology data used for testing. It's not a retrospective or prospective clinical study on patient data in the typical sense of evaluating a new AI/ML diagnostic model.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Number of Experts: Not specified. Given the nature of a software update, ground truth for software functionality would be established by software engineers, quality assurance personnel, and potentially subject matter experts (e.g., electrophysiologists) for clinical workflow and output validity, but not in the context of diagnostic "ground truth" derived from patient-specific interpretations.
- Qualifications of Experts: Not specified.
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable or not specified in this type of submission. The "ground truth" is based on system specifications and expected software behavior, not human interpretation requiring adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- MRMC Study: No, an MRMC comparative effectiveness study was not performed as described for AI assistance. This submission is for software updates to a diagnostic system, not a new AI-powered diagnostic algorithm designed to assist human readers directly in interpreting images or data.
- Effect Size of Improvement: Not applicable, as no such study was conducted.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- Standalone Study: Not directly applicable in the typical sense of evaluating an AI algorithm's diagnostic performance. The device is a "Programmable Diagnostic Computer" used in electrophysiology procedures. Its performance is evaluated by ensuring its internal computations, mapping, and display functions are accurate and consistent with its design specifications and the predicate device. It's not a standalone "algorithm" providing a diagnosis independent of human interaction.
7. Type of Ground Truth Used
- Type of Ground Truth: For software verification, the ground truth is primarily based on design specifications, established functional requirements, and the expected behavior of the system as defined by its engineers and clinical subject matter experts. This is not clinical "ground truth" like pathology, expert consensus on imaging, or outcomes data, as this is a functional system, not an AI diagnostic algorithm.
8. Sample Size for the Training Set
- Sample Size for Training Set: Not applicable. This document does not describe the development or training of a new AI/ML model. The updates are to existing software components (bug fixes, compatibility, minor enhancements).
9. How the Ground Truth for the Training Set Was Established
- Establishing Ground Truth for Training Set: Not applicable, as no new AI/ML training data or process is described.
Ask a specific question about this device
(86 days)
EnSite X EP System
The EnSite™ X EP System is a suggested diagnostic tool in patients for whom electrophysiology studies have been indicated.
The EnSite™ X EP System provides information about the electrical activity of the heart and displays catheter location during conventional electrophysiological (EP) procedures.
EnSite™ X EP System Contact Force Software License:
When used with the TactiSys™ Quartz Equipment, the EnSite™ X EP System Contact Force Module is intended to provide visualization of force information from compatible catheters.
EnSite™ X EP System Surface Electrode Kit:
The EnSite™ X EP Surface Electrode Kit is indicated for use with the EnSite™ X EP System in accordance with the EnSite™ X EP System indications for use.
EnSite™ X EP System, TactiFlex™ Ablation Catheter, Sensor Enabled™, Software Upgrade and EnSite™ X EP System, TactiFlexTM Ablation Catheter, Sensor EnabledTM, Software License:
The EnSite™ X EP System TactiFlex™ Ablation Catheter, Sensor Enabled™ Software Module is indicated for use with the EnSite™ X EP System in accordance with the EnSite™ X EP System indications for use.
The EnSite™ X EP System is a catheter navigation and mapping system. A catheter navigation and mapping system is capable of displaying the 3-dimensional (3-D) position of conventional and Sensor Enabled™ (SE) electrophysiology catheters, as well as displaying cardiac electrical activity as waveform traces and as three-dimensional (3D) isopotential and isochronal maps of the cardiac chamber.
The contoured surfaces of the 3D maps are based on the anatomy of the patient's own cardiac chamber. The system creates a model by collecting and labeling the anatomic locations within the chamber. A surface is created by moving a selected catheter to locations within a cardiac structure. As the catheter moves, points are collected at and between all electrodes on the catheter. A surface is wrapped around the outermost points.
The provided document, a 510(k) premarket notification letter from the FDA regarding the EnSite™ X EP System, does not contain the detailed information typically required to answer your specific questions about device acceptance criteria and a study proving those criteria were met for an AI/ML medical device.
The document describes the device as a "Programmable Diagnostic Computer" for electrophysiology (EP) studies and notes software revisions for the subject device (EnSite™ X EP System v3.0) compared to its predicate (EnSite™ X EP System v2.0.1). However, the software updates described are primarily workflow enhancements, minor display updates, and data import/export expansions, not features that would typically involve a large-scale AI/ML performance study with specific acceptance criteria like sensitivity, specificity, or AUC as you described.
The "Non-Clinical Testing Summary" mentions "Software Verification at unit, software and system level," "Performance testing of updated feature functionality," "Preclinical Validation Testing to confirm the system could meet user requirements," and "Human Factors Evaluations." While these are tests, they are framed in the context of ensuring the updated software functions as intended and meets user requirements, rather than demonstrating a performance benefit against a specific clinical outcome with an AI algorithm.
Therefore, based solely on the provided text, I cannot extract the information required for your questions.
Here's why each point cannot be answered:
- A table of acceptance criteria and the reported device performance: The document does not provide specific quantitative acceptance criteria (e.g., sensitivity, specificity, accuracy) for an AI algorithm, nor reported device performance against such criteria. The "acceptance criteria" mentioned are more general for software verification and validation.
- Sample sizes used for the test set and the data provenance: There is no mention of a specific test set, its size, or the provenance of any data used for performance evaluation in the context of an AI/ML algorithm.
- Number of experts used to establish the ground truth for the test set and qualifications: This information is absent as no detailed performance study with a test set requiring ground truth is described.
- Adjudication method for the test set: Not applicable, as no such test set or study is described.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance: The document does not describe any MRMC study or any study evaluating human reader improvement with AI assistance. The device is a "Programmable Diagnostic Computer" for EP studies, not explicitly an AI-assisted diagnostic tool in the sense of image interpretation where MRMC studies are common.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not explicitly mentioned in the context of a standalone AI algorithm performance. The device itself is a system that assists in EP studies.
- The type of ground truth used: No ground truth for an AI algorithm's performance is described.
- The sample size for the training set: Not applicable, as there's no mention of an AI model's training set.
- How the ground truth for the training set was established: Not applicable, as there's no mention of an AI model's training set.
In summary, the provided document is a regulatory approval letter and a summary of the device's technical specifications and non-clinical testing. It does not elaborate on an AI/ML component meeting specific performance criteria through a clinical validation study as implied by your questions. The software changes are presented as enhancements to an existing system, rather than the introduction of a new AI-driven diagnostic capability requiring a comprehensive AI performance study.
Ask a specific question about this device
(88 days)
EnSite X EP System TactiFlex Ablation Catheter, Sensor Enabled, Software Upgrade; EnSite X EP System
The EnSite™ X EP System is a suggested diagnostic tool in patients for whom electrophysiology studies have been indicated.
The EnSite™ X EP System provides information about the electrical activity of the heart and displays catheter location during conventional electrophysiological (EP) procedures.
EnSite™ X EP System Contact Force Software License:
When used with the TactiSys™ Quartz Equipment, the EnSite™ X EP System Contact Force Module is intended to provide visualization of force information from compatible catheters.
EnSite™ X EP System Surface Electrode Kit:
The EnSite™ X EP Surface Electrode Kit is indicated for use with the EnSite™ X EP System in accordance with the EnSite™ X EP System indications for use.The EnSite™ X EP System TactiFlex™ Ablation Catheter Sensor Enabled™ Software Module is indicated for use with the EnSite™ X EP System in accordance with the EnSite™ X EP System indications for use.
EnSite™ X EP System, TactiFlex™ Ablation Catheter, Sensor Enabled™, Software Upgrade and EnSite™ X EP System, TactiFlexTM Ablation Catheter, Sensor EnabledTM, Software License:
The EnSite™ X EP System TactiFlex™ Ablation Catheter, Sensor Enabled™ Software Module is indicated for use with the EnSite™ X EP System in accordance with the EnSite™ X EP System indications for use.
The EnSite™ X EP System is a catheter navigation and mapping system. A catheter navigation and mapping system is capable of displaying the 3-dimensional (3-D) position of conventional and Sensor Enabled™ (SE) electrophysiology catheters, as well as displaying cardiac electrical activity as waveform traces and as three-dimensional (3D) isopotential and isochronal maps of the cardiac chamber.
The contoured surfaces of the 3D maps are based on the anatomy of the patient's own cardiac chamber. The system creates a model by collecting and labeling the anatomic locations within the chamber. A surface is created by moving a selected catheter to locations within a cardiac structure. As the catheter moves, points are collected at and between all electrodes on the catheter. A surface is wrapped around the outermost points.
The provided text is a 510(k) summary for the EnSite™ X EP System, which is a catheter navigation and mapping system. The submission describes a software upgrade that enables compatibility with a new ablation catheter (TactiFlex™ Ablation Catheter, Sensor Enabled™). However, it does not provide specific acceptance criteria or a detailed study proving that the device meets those acceptance criteria.
Instead of specific acceptance criteria and detailed study results, the document generally states that "Design verification activities were performed and met their respective acceptance criteria to ensure that the devices in scope of this submission are safe and effective." It also lists various types of testing performed and the standards followed.
Therefore, many of the requested details cannot be extracted from this document.
Here's what can be inferred or explicitly stated based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not provide a table of acceptance criteria with corresponding performance metrics. It indicates that "Design verification activities were performed and met their respective acceptance criteria to ensure that the devices in scope of this submission are safe and effective."
2. Sample size used for the test set and the data provenance
The document does not specify the sample size for any test set or the data provenance (e.g., country of origin, retrospective/prospective). It mentions "Preclinical Validation Testing to confirm the system could meet user requirements and its intended use after modifications" but does not give details.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not mentioned in the document.
4. Adjudication method for the test set
Not mentioned in the document.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
A multi-reader multi-case (MRMC) comparative effectiveness study is not mentioned. The device is a "Programmable Diagnostic Computer" for electrophysiology, assisting in catheter navigation and mapping, not an AI for image interpretation or diagnosis that would typically involve human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The document states, "Software Verification at a software and system level to test the new software features added for the display of the TactiFlex™ Ablation Catheter, Sensor Enabled™ Force Number Refresh Rate Force Direction Indicator". This indicates software-level testing, which could be considered standalone performance for the specific software features implemented. However, detailed performance metrics are not given.
7. The type of ground truth used
The document does not explicitly state the "ground truth" used for testing, but given the nature of the device (catheter navigation and mapping, force display), it would likely involve:
- Engineering validation/simulations: For verifying force number refresh rates and direction indicator accuracy (implied by "Software Verification").
- User requirements and intended use: For "Preclinical Validation Testing to confirm the system could meet user requirements and its intended use."
- Safety standards: For "Electrical Safety Testing."
8. The sample size for the training set
The document does not mention a training set, as the changes described are software adaptations for a new catheter, rather than the development of a machine learning model that would require a training set.
9. How the ground truth for the training set was established
Not applicable, as no training set is mentioned in the context of the described software upgrade.
Ask a specific question about this device
(113 days)
EnSite X EP System
The EnSite™ X EP System is a suggested diagnostic tool in patients for whom electrophysiology studies have been indicated.
The EnSite™ X EP System provides information about the electrical activity of the heart and displays catheter location during conventional electrophysiological (EP) procedures.
EnSite™ X EP System Contact Force Software License:
When used with the TactiSys™ Quartz Equipment, the EnSite™ X EP System Contact Force Module is intended to provide visualization of force information from compatible catheters.
EnSite™ X EP System Surface Electrode Kit:
The EnSite™ X EP Surface Electrode Kit is indicated for use with the EnSite™ X EP System in accordance with the EnSite™ X EP System indications for use.
The EnSite™ X EP System is a catheter navigation and mapping system. A catheter navigation and mapping system is capable of displaying the 3-dimensional (3-D) position of conventional and Sensor Enabled™ (SE) electrophysiology catheters, as well as displaying cardiac electrical activity as waveform traces and as three-dimensional (3D) isopotential and isochronal maps of the cardiac chamber.
The contoured surfaces of the 3D maps are based on the anatomy of the patient's own cardiac chamber. The system creates a model by collecting and labeling the anatomic locations within the chamber. A surface is created by moving a selected catheter to locations within a cardiac structure. As the catheter moves, points are collected at and between all electrodes on the catheter. A surface is wrapped around the outermost points.
The provided text does not contain details about acceptance criteria or a study proving the device meets those criteria for the EnSite™ X EP System's AI software.
The document is a 510(k) summary for the EnSite™ X EP System, specifically for version 2.0. It primarily focuses on demonstrating substantial equivalence to a predicate device (EnSite™ X v1.1.1). While it mentions "software was revised to include the following updates" which sound like new features (e.g., "Displays activation time based on the Near Field Detection Method," "Displays peak frequency maps," "Automatically assesses activation times for outliers using the Outlier Filter"), it does not provide any specific quantitative acceptance criteria or detailed study results related to the performance of these new software features or any embedded AI/algorithmic components.
The "Non-Clinical Testing Summary" broadly states that "Design verification activities were performed and met their respective acceptance criteria to ensure that the devices in scope of this submission are safe and effective." It lists types of testing performed such as "Software Verification at unit, software and system level," "Performance Testing of updated feature functionality," and "Preclinical studies to evaluate substantial equivalence," and "Preclinical Validation Testing to confirm the system could meet user requirements and its intended use after modifications." However, it does not elaborate on the specific acceptance criteria for these tests, nor does it provide any data (e.g., accuracy, sensitivity, specificity, or inter-reader agreement statistics) to demonstrate how these criteria were met.
Therefore, I cannot fulfill the request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria, as the necessary information is not present in the provided document. The document primarily focuses on regulatory approval based on "substantial equivalence" rather than a detailed performance study of specific AI algorithm metrics.
Ask a specific question about this device
(30 days)
EnSite X EP System
The EnSite™ X EP System is a suggested diagnostic tool in patients for whom electrophysiology studies have been indicated.
The EnSite™ X EP System provides information about the electrical activity of the heart and displays catheter location during conventional electrophysiological (EP) procedures.
The EnSite™ X EP System is a catheter navigation and mapping system. A catheter navigation and mapping system is capable of displaying the 3-dimensional (3-D) position of conventional and Sensor Enabled™ (SE) electrophysiology catheters, as well as displaying cardiac electrical activity as waveform traces and as three-dimensional (3D) isopotential and isochronal maps of the cardiac chamber.
The contoured surfaces of the 3D maps are based on the anatomy of the patient's own cardiac chamber. The system creates a model by collecting and labeling the anatomic locations within the chamber. A surface is created by moving a selected catheter to locations within a cardiac structure. As the catheter moves, points are collected at and between all electrodes on the catheter. A surface is wrapped around the outermost points.
I apologize, but the provided text does not contain the detailed information necessary to answer your request regarding the acceptance criteria and the study that proves the device meets those criteria.
The document is an FDA 510(k) clearance letter and an accompanying 510(k) summary for the EnSite™ X EP System. While it states that "Design verification and validation activities were performed to ensure the EnSite X v1.1.1 software release is safe and effective" and lists some "Types of Testing Performed" (Software Verification, Preclinical Validation Testing, Installation Validation), it does not provide:
- A table of specific acceptance criteria and detailed device performance metrics. The document primarily focuses on demonstrating substantial equivalence to a predicate device, rather than presenting a performance study with quantitative results against specific criteria.
- Sample sizes used for testing.
- Data provenance (country of origin, retrospective/prospective).
- Information on experts used for ground truth (number, qualifications).
- Adjudication methods.
- Details on Multi-Reader Multi-Case (MRMC) comparative effectiveness studies or effect sizes.
- Standalone performance data (algorithm only).
- Specific type of ground truth used (e.g., pathology, outcomes data).
- Sample size for the training set (as this is a software update for an existing system, deep learning training data might not be relevant in the same way as for a novel AI device).
- How ground truth for the training set was established.
The document states that the software updates include "Patch input signals switched from driven to undriven signals," "Composite signal respiration waveform," "Bio Impedance Scaling algorithm updates," "Respiration Gating algorithm updates," and "Metal distortion threshold changes," along with "Additional bug fixes." The testing described is general software verification and preclinical validation to ensure user requirements are met and risks are mitigated, rather than a clinical performance study with specific quantitative acceptance criteria.
Ask a specific question about this device
(113 days)
EnSite X EP System
The EnSite™ X EP System is a suggested diagnostic tool in patients for whom electrophysiology studies have been indicated.
The EnSite™ X EP System provides information about the electrical activity of the heart and displays catheter location during conventional electrophysiological (EP) procedures.
EnSite™ X EP System Contact Force Software License:
When used with the TactiSys™ Quartz Equipment, the EnSite™ X EP System Contact Force Module is intended to provide visualization of force information from compatible catheters.
EnSite™ X EP System Surface Electrode Kit:
The EnSite™ X EP Surface Electrode Kit is indicated for use with the EnSite™ X EP System in accordance with the EnSite™ X EP System indications for use.
The EnSite™ X EP System is a catheter navigation and mapping system. A catheter navigation and mapping system is capable of displaying the 3-dimensional (3-D) position of conventional and Sensor Enabled™ (SE) electrophysiology catheters, as well as displaying cardiac electrical activity as waveform traces and as three-dimensional (3D) isopotential and isochronal maps of the cardiac chamber. The contoured surfaces of the 3D maps are based on the anatomy of the patient's own cardiac chamber. The system creates a model by collecting and labeling the anatomic locations within the chamber. A surface is created by moving a selected catheter to locations within a cardiac structure. As the catheter moves, points are collected at and between all electrodes on the catheter. A surface is wrapped around the outermost points.
The provided document is a 510(k) summary for the EnSite™ X EP System, outlining its substantial equivalence to a predicate device. This type of submission focuses on demonstrating safety and effectiveness compared to an already legally marketed device, not necessarily on novel AI algorithm performance studies as might be found in a De Novo submission.
Therefore, the specific information requested about acceptance criteria and study details often associated with AI/ML device performance (like sample size for test sets, data provenance, expert ground truth adjudication, MRMC studies, standalone performance, and training set details) is not directly available in this document.
The document discusses "Design verification activities" and "Performance Testing of updated feature functionality" for the software updates, but these are general engineering and software validation tests rather than clinical performance studies demonstrating diagnostic accuracy with human readers or standalone AI performance.
Here's how to address the requested points based on the available information:
1. A table of acceptance criteria and the reported device performance
The document states that "Design verification activities were performed and met their respective acceptance criteria to ensure that the devices in scope of this submission are safe and effective." However, it does not provide a table specifying the explicit acceptance criteria for each software update or the detailed reported performance metrics against those criteria. It lists the types of testing, implying that the outcomes of these tests met their internal acceptance criteria.
2. Sample sized used for the test set and the data provenance
Not explicitly stated for the "performance testing of updated feature functionality." The document mentions "Bench studies to evaluate substantial equivalence" and "Preclinical Validation Testing to confirm the system could meet user requirements." These usually involve in-vitro or simulated data, rather than large clinical test sets with specified patient populations. The provenance of such data (e.g., country of origin, retrospective/prospective) is not mentioned.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable or not specified. This level of detail for expert ground truth is typically provided for diagnostic accuracy studies involving human interpretation of clinical data, which is not the primary focus of this 510(k) for software updates to an existing electrophysiology system.
4. Adjudication method for the test set
Not applicable or not specified.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study is mentioned. This submission is about software updates to an existing system, not the introduction of an AI algorithm requiring a comparative effectiveness claims with human readers. The new features (e.g., calculated waveforms, activation direction arrows, wave speed maps, deflection direction indicators, real-time map points) are presented as direct display enhancements or functional improvements rather than AI-driven diagnostic assistance to human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This refers to an "EnSite™ X EP System" which is a catheter navigation and mapping system, intended to be used by trained medical professionals. It's an interactive diagnostic tool, not a standalone AI algorithm producing a diagnosis without human interaction. Therefore, a standalone (algorithm only) performance study as typically understood for AI/ML devices is not applicable or described here. Its function is to provide information for a human clinician.
7. The type of ground truth used
For the software updates, the "ground truth" would likely be defined by internal engineering specifications, established scientific principles of electrophysiology, and potentially comparisons against existing validated methods or simulations, rather than clinical 'ground truth' such as pathology or long-term outcomes data, which is more relevant for diagnostic accuracy claims. The validation would ensure the calculated waveforms, map displays, etc., are accurate representations of the underlying electrophysiological data according to accepted standards.
8. The sample size for the training set
Not applicable. The document describes software updates for an electrophysiology system, not a machine learning algorithm that undergoes a training phase with a specific dataset.
9. How the ground truth for the training set was established
Not applicable, as there is no mention of a machine learning model with a training set.
Ask a specific question about this device
(121 days)
EnSite X EP System, Advisor VL Circular Mapping Catheter, Sensor Enabled, Advisor FL Circular Mapping
EnSite™ X EP System: The EnSite™ X EP System is a suggested diagnostic tool in patients for whom electrophysiology studies have been indicated. The EnSite™ X EP System provides information about the electrical activity of the heart and displays catheter location during conventional electrophysiological (EP) procedures.
EnSite™ X EP System Contact Force Software License: When used with the TactiSys™ Quartz Equipment, the EnSite™ X EP System Contact Force Module is intended to provide visualization of force information from compatible catheters.
EnSite™ X EP System Surface Electrode Kit: The EnSite™ X EP Surface Electrode Kit is indicated for use with the EnSite™ in accordance with the EnSite™ X EP System indications for use.
Advisor™ VL Circular Mapping Catheter, Sensor Enabled™: Advisor™ VL Circular Mapping Catheter, Sensor Enabled™ is a steerable electrophysiology catheter with integrated sensors. The catheter is used for recording intracardiac stimulation during diagnostic electrophysiology studies. The cather can be used to map the atrial regions of the heart.
Advisor™ FL Circular Mapping Catheter, Sensor Enabled™: The Advisor™ FL Circular Mapping Catheter, Sensor Enabled™ is steerable electrophysiology catheter with integrated sensors. The catheter is used for recording intracardiac stimulation during diagnostic electrophysiology studies. The catheter can be used to map the atrial regions of the heart.
Advisor™ HD High Density Mapping Catheter, Sensor Enabled™: The Advisor™ HD Grid Mapping Catheter, Sensor Enabled™, is indicated for multiple electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electricular regions of the heart.
The EnSite™ X EP System is a catheter navigation and mapping system. A catheter navigation and mapping system is capable of displaying the 3-dimensional (3-D) position of conventional and Sensor Enabled™ (SE) electrophysiology catheters, as well as displaying cardiac electrical activity as waveform traces and as three-dimensional (3D) isopotential and isochronal maps of the cardiac chamber.
The contoured surfaces of the 3D maps are based on the anatomy of the patient's own cardiac chamber. The system creates a model by collecting and labeling the anatomic locations within the chamber. A surface is created by moving a selected catheter to locations within a cardiac structure. As the catheter moves, points are collected at and between all electrodes on the catheter. A surface is wrapped around the outermost points.
Advisor™ VL Circular Mapping Catheter, Sensor Enabled™ (Advisor VL) is a variable radius, circular mapping catheter. It has an adjustable 4 French (F) distal loop size with a diameter ranging from 15mm - 25mm with models containing both ten (10) equidistant or twenty (20) paired platinum-iridium electrodes. The catheter has integrated sensors with two impedance-based navigational electrodes and two magnetic sensors located at the distal end of the shaft. The catheter is intended to be used with the EnSite Precision™ Cardiac Mapping System, or the EnSite™ X EP System.
Advisor™ FL Circular Mapping Catheter, Sensor Enabled™ (Advisor FL, SE) is a circular mapping catheter for performing electrophysiology mapping procedures and providing pacing signals to the heart during electrophysiology procedures. The catheter handle and shaft design allows for improved maneuverability. A magnetic sensor in the distal shaft pocket provides compatibility with visualization and navigation systems. The catheter is compatible with Abbott's EnSite Precision™ Cardiac Mapping System, MediGuide™ System, or EnSite™ X EP System.
The Advisor™ HD Grid Mapping Catheter, Sensor Enabled™, is a sterile, single use, irrigated, high-density mapping catheter with a 7.5F shaft and an 8F distal shaft deflectable section. It is available in a D-F bi-directional curve model that is deflected using the actuator located on the catheter handle. The catheter working length is 110 cm. The device consists of a paddle-shaped distal tip with 16 electrodes, two distal shaft ring electrodes, two magnetic sensors, polymer braided shaft, handle, fluid lumen extension with a luer, and an electrical connector. The catheter also has an introducer tool intended to compress and guide the distal paddle into, and withdraw from, the hemostasis valve of an introducer sheath. The catheter is compatible with the EnSite™ Velocity, EnSite Precision™, and EnSite™ X EP Cardiac Mapping Systems and other accessories, including the connecting cable and commercially available irrigation pumps.
The provided FDA 510(k) summary (K202066) focuses on the substantial equivalence of the EnSite X EP System and associated catheters to previously cleared predicate devices. It largely relies on the similar intended use, indications for use, fundamental scientific technology, and performance of the subject devices compared to the predicate devices. The primary change described is compatibility with the EnSite™ X EP System and updated labeling.
However, the document does not contain specific acceptance criteria, reported device performance metrics, or details of a study structured to prove the device meets pre-defined acceptance criteria in the manner that would be expected for a novel AI/ML diagnostic device with a specific performance claim (e.g., sensitivity, specificity for a disease).
Instead, the non-clinical testing summary focuses on design verification activities, compliance with industry standards, and in vivo preclinical studies to evaluate substantial equivalence, and human factors evaluations. This suggests that the "acceptance criteria" here are more about demonstrating that the new system and its components function as intended and do not raise new questions of safety or effectiveness compared to the predicate(s).
Given this context, I will extract the information available and note where specific details regarding AI/ML performance studies or traditional clinical performance metrics are not present, as they do not appear to be the primary focus of this 510(k).
Acceptance Criteria and Study for K202066
1. Table of Acceptance Criteria and Reported Device Performance
As mentioned, this 510(k) does not present performance in terms of specific diagnostic metrics (e.g., sensitivity, specificity, accuracy) with numerical acceptance criteria. Instead, the "acceptance criteria" are implied by successful completion of design verification activities, adherence to standards, and performance in preclinical studies demonstrating substantial equivalence to predicate devices. The reported device performance generally refers to its ability to function as intended and similar to predicate devices.
Acceptance Criterion (Implied) | Reported Device Performance |
---|---|
Safe and effective operation of the EnSite™ X EP System. | Demonstrated through design verification, software verification, performance testing, and in vivo preclinical studies. |
Compliance with relevant industry guidance and standards (e.g., IEC 62304, ISO 14971). | Confirmed through adherence to listed standards and guidance documents. |
Functional performance of each device in the EnSite™ X EP System (e.g., functional, shipment). | Met through specific performance testing. |
Accuracy of catheter position and orientation (similar to predicate). | Demonstrated through bench and in-vivo preclinical data comparing subject, predicate, and reference devices. |
Proper functioning of cardiac mapping and model creation (similar to predicate). | Demonstrated through bench and in-vivo preclinical data comparing subject, predicate, and reference devices. |
Compatibility of Sensor Enabled™ Catheters with the EnSite™ X EP System. | Confirmed via labeling updates and EEPROM functional tests. |
User interface usability as intended by defined user groups. | Verified through Human Factors Evaluations. |
No new questions of safety or effectiveness are raised compared to predicate devices. | Concluded based on predicate comparison and non-clinical testing results. |
2. Sample Size Used for the Test Set and Data Provenance
The document mentions "In vivo Preclinical Studies" for the EnSite™ X EP System and Sensor Enabled™ Catheters. However, it does not specify the sample size (e.g., number of patients, cases, or animal subjects) used for these studies, nor does it explicitly state the country of origin or whether the data was retrospective or prospective. The term "preclinical" typically refers to animal studies rather than human clinical data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
The document does not specify the use of experts to establish ground truth for a test set in the context of diagnostic interpretation. The studies described are more focused on the functional performance and substantial equivalence of the hardware and software components rather than a diagnostic accuracy study requiring expert human interpretation as ground truth.
4. Adjudication Method for the Test Set
Since the document does not describe a diagnostic study requiring expert interpretation or ground truth establishment in a clinical sense, there is no mention of an adjudication method (e.g., 2+1, 3+1).
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
The document does not indicate that an MRMC comparative effectiveness study was done. The focus is on demonstrating that the device itself performs comparably to predicate devices, which may include functional performance in in-vivo settings, but not a study of human readers' performance with and without AI assistance for improving diagnostic outcomes.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
While the EnSite™ X EP System performs functions involving data processing ("programmable diagnostic computer"), the K202066 submission does not present it as a standalone AI/ML diagnostic algorithm with specific performance metrics (e.g., sensitivity, specificity) proven in a standalone study. The system provides "information about the electrical activity of the heart and displays catheter location" and "visualization of force information." The testing described confirms the system's functional integrity and similarity to predicate devices, rather than a standalone diagnostic performance claim often associated with AI/ML algorithms.
7. Type of Ground Truth Used
For the "in vivo Preclinical Studies" and bench testing, the ground truth would likely be based on:
- Physical measurements and established physiological parameters: For evaluating catheter position, orientation, and electrical activity recording accuracy.
- Comparison to predicate device performance: Demonstrating similar outputs and behavior.
- Engineering specifications and design requirements: For functional tests of hardware and software components.
There is no mention of "pathology" or "outcomes data" being used as ground truth for the evaluations described.
8. Sample Size for the Training Set
The document does not provide any information regarding a training set size. This indicates that the regulatory submission is likely not for a device that relies on continuously learning or adaptive AI/ML algorithms that require specific training data sets in the typical sense. The "software verification" and "system level" testing suggest traditional software development and testing, rather than a machine learning model's training and validation.
9. How the Ground Truth for the Training Set Was Established
Since there is no mention of a training set or a machine learning component requiring one, there is no information provided on how ground truth for a training set was established.
Ask a specific question about this device
Page 1 of 1