Search Results
Found 1 results
510(k) Data Aggregation
(184 days)
The EmboTrap® II Revascularization Device is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
The EmboTrap® II Revascularization Device is composed of a retrievable, self-expanding, Nitinol shaped section at the distal end of a tapered Nitinol shaft. It is designed to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke. The EmboTrap® II Revascularization Device is supplied sterile, and is intended for single-use only by physicians trained in neuro-interventional catheterization and the treatment of ischemic stroke. It is indicated for patients who are ineligible for intravenous tissue plasogen activator (IV t-PA) or who fail IV t-PA therapy.
Here's a breakdown of the acceptance criteria and the study proving the device meets those criteria, based on the provided text:
Acceptance Criteria and Device Performance
Criteria Category | Acceptance Criteria | Reported Device Performance and Conclusion |
---|---|---|
Biocompatibility | All biocompatibility tests (MTT Cytotoxicity, ISO Guinea Pig Maximization Sensitization, ISO Intracutaneous, ISO Systemic Toxicity, USP Rabbit Pyrogen, ASTM Hemolysis, Complement Activation Assay) must meet pre-assigned acceptance criteria. | All biocompatibility tests completed met the pre-assigned acceptance criteria. The device was found to be non-cytotoxic, not a sensitizer, not an irritant, non-toxic, non-pyrogenic, non-hemolytic, and had acceptable levels of C3a and SC5b-9. |
Sterilization | Device must be EO-sterilized to achieve a minimum SAL of 10-6. | The sterilization process was successfully validated and process monitoring controls assure a minimum SAL of 10-6. |
Shelf Life | Shelf life studies must establish that the product and packaging remain functional and sterile for a period of 3 years. | Shelf life studies established that the product and packaging remain functional and sterile for a 3-year shelf life. |
In Vitro (Bench) Testing | Device must fulfill all pre-determined product performance specification requirements for: Dimensional Testing (Radial Force, Outer Cage Recovery, Device Durability/Integrity, Joint Strength), Flexibility & Kink Resistance Testing (Torque Durability, Corrosion Resistance, Radiopacity, Tip Flexibility, Re-sheathing Force), Simulated Use/Performance Testing (Coating Integrity, Particulate Evaluation). Performance must be comparable to predicate devices. | The device fulfilled all pre-determined product performance specifications. Side-by-side bench testing showed the EmboTrap® II to be comparable or superior to its predicate devices in terms of performance and functionality for all tested characteristics (System Dimensions, Tip Flexibility, Re-sheathing force, Delivery and re-sheathing force during simulated use, Durability Testing, System Kink Resistance, Kink Resistance - Deployed Shaped Section, Distal Coil Tensile Testing, System Tensile Testing, Torque Testing in a Microcatheter, Torque Strength Testing, Radial Force Testing, Radiopacity, Clot retrieval and performance, and Physician usability testing). |
In Vivo (Animal) Studies | Usability, effectiveness, and safety must be assessed in a swine model compared to predicate devices. Local and end organ tissue response must be comparable to predicate devices. | Acute performance on Day 0 showed equivalent usability and performance to the predicate device. Histological evaluation after 3 and 28 days demonstrated comparable local and end organ tissue response between the EmboTrap® II and predicate devices. |
Clinical Efficacy (Primary Endpoint) | Revascularization rate (mTICI 2b or greater in the target vessel following 3 or less passes of the EmboTrap® device without rescue therapy) must be statistically significantly greater than a performance goal (NL) of 0.56 (56%). Null Hypothesis (H0: PEmboTrap ≤ 0.56) should be rejected in favor of the Alternative Hypothesis (H1: PEmboTrap > 0.56) at a one-sided Exact test for a binomial proportion at the 5% significance level. | The successful revascularization rate was 72% (137/191) with a 95% confidence interval of [65%, 78%]. The p-value was |
Ask a specific question about this device
Page 1 of 1