Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K243443
    Manufacturer
    Date Cleared
    2025-03-19

    (133 days)

    Product Code
    Regulation Number
    892.1680
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Digital X-ray detector, EXPD-N Series, is designed for use in digital imaging solutions for general radiographic diagnosis of human anatomy. This device is intended for use in all general diagnostic procedures, replacing film or screen-based radiographic systems for both adult and paediatric patients. It is not intended for use in mammography.

    Device Description

    In comparison to existing devices, the new detectors incorporate a Flexible a-Si in the TFT material within the panel. The primary difference from the conventional glass a-Si panel is that the electronic circuits, such as silicon, are deposited on a plastic substrate instead of a glass substrate during the manufacturing of the TFT panel. Since only the material of the substrate on which the silicon is deposited changes, the overall image performance remains unaffected. Another difference is the pixel pitch. While existing products feature only a pixel pitch of 140μm, the new models include an option with a pixel pitch of 100um. The resolution of an X-ray detector has a significant impact on MTF (Modulation Transfer Function) and sensitivity.

    AI/ML Overview

    This medical device submission is for an X-ray detector, not an AI/ML device. Therefore, the typical acceptance criteria and study requirements for AI/ML devices, such as those related to multi-reader multi-case studies, standalone performance, and ground truth establishment with expert consensus or pathology, are not applicable here.

    The submission focuses on establishing substantial equivalence to a predicate device based on technical characteristics and physical performance, confirming it is suitable for general radiographic diagnosis.

    Here's a breakdown of the provided information, tailored to the context of a non-AI X-ray detector:

    1. Table of Acceptance Criteria and the Reported Device Performance

    The acceptance criteria are implicitly defined by demonstrating substantial equivalence to the predicate device in terms of technical characteristics and performance metrics relevant to X-ray image quality. The table below compares the subject device's performance to the predicate device, highlighting where performance is similar or improved.

    ItemAcceptance Criteria (Implied by Predicate Device K193017 Performance)Subject Device (EXPD-N Series) Reported Performance
    Intended UseGeneral radiographic diagnosis, replaces film/screen-based systems, adult & pediatric, Not for mammography.General radiographic diagnosis, replaces film/screen-based systems, adult & pediatric, Not for mammography. (Same)
    Anatomical SitesGeneral RadiographyGeneral Radiography (Same)
    Dimensions (mm)EVS 3643W/WG/WP: 460(W) x 386(L) x 15(H)EVS 4343W/WG/WP: 460(W) x 460(L) x 15(H)EXPD 3643N/NP/NU/N1/U1: 460(W) x 386(L) x 15.5(H)EXPD 4343N/NP/NU/N1/U1: 460(W) x 460(L) x 15.5(H) (Slight difference in thickness, otherwise similar)
    Pixel Pitch140 μm140 μm (for N/NP/NU models)100 μm (for N1/U1 models) (Improved resolution option added)
    Image Size (pixels)EVS 3643W/WG/WP: 2,560 x 3,072EVS 4343W/WG/WP: 3,072 x 3,072EXPD 4343N/NP/NU: 3,072 × 3,072 (Same)EXPD 3643N/NP/NU: 2,560 × 3,072 (Same)EXPD 4343N1/U1: 4,302 × 4,302 (Improved with 100μm pixel pitch)EXPD 3643N1/U1: 3,534 × 4,302 (Improved with 100μm pixel pitch)
    Active Area (mm)EVS 3643W/WG/WP: 430 x 358EVS 4343W/WG/WP: 430 x 430EXPD 4343N/NP/NU: 430.2mm × 430.2mm (Similar)EXPD 3643N/NP/NU: 353.4mm × 430.2mm (Similar)EXPD 4343N1/U1: 430.08mm × 430.08mm (Similar, adapted for 100μm pixel pitch)EXPD 3643N1/U1: 358.4mm × 430.08mm (Similar, adapted for 100μm pixel pitch)
    TFT Materiala-Si, IGZOa-Si, Flexible a-Si, IGZO (New Flexible a-Si material introduced, otherwise similar)
    Cycle Time< 8.5 sec.< 8.5 sec. (Same)
    Power Supply100240V, 50/60 Hz100240V, 50/60 Hz (Same)
    Dynamic Range16 bits16 bits (Same)
    MTFEVS 4343W: 50.0% at 2.0 lp/mmEVS 3643W: 42.5% at 2.0 lp/mmetc.EXPD 4343N/NP/NU/3643N/NP/NU: 35% at 2.0 lp/mm (Comparable, some variations)EXPD 4343N1/U1/3643N1/U1: 40% at 2.0 lp/mm (Improved with 100μm pixel pitch)
    DQEEVS 4343W: 52.8% at 1.0 lp/mmEVS 3643W: 53.3% at 1.0 lp/mmetc.EXPD 4343N/NP/NU/3643N/NP/NU: 60% at 0.5 lp/mm (Comparable, different spatial frequency reported)EXPD 4343N1/U1/3643N1/U1: 60% at 0.5 lp/mm (Comparable, different spatial frequency reported)
    Resolution3.5 lp/mmEXPD 4343N/NP/NU/3643N/NP/NU: 3.5 lp/mm (Same)EXPD 4343N1/U1/3643N1/U1: 5 lp/mm (Improved with 100μm pixel pitch)
    Communication MethodWire, Wireless IEEE 802.11a/g/n (2.4 GHz / 5 GHz)Wire, Wireless IEEE 802.11n/ac (2.4 GHz / 5 GHz) (Updated wireless standard)
    Software (Image Processing program)EConsole 1 (K152172)EConsole 2 (K240243) (Newer software version)

    Study Proving Acceptance Criteria:

    The study proving the device meets the acceptance criteria is a non-clinical performance testing and expert review study.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Test Set: Not explicitly stated as a number of images. The document refers to "clinical images acquired from various anatomical sites."
    • Data Provenance: The document does not explicitly state the country of origin for the clinical images. It indicates the data was retrospective, as it refers to "clinical images acquired."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

    • Number of Experts: Not explicitly stated, but referred to as "qualified expert clinicians."
    • Qualifications of Experts: "qualified expert clinicians." (Specific specialties or years of experience are not provided.)

    4. Adjudication Method for the Test Set:

    • Adjudication Method: "clinical images acquired from various anatomical sites were reviewed by qualified expert clinicians to assess imaging performance." No specific adjudication method like 2+1 or 3+1 is mentioned. It implies a qualitative assessment by experts.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This device is an X-ray detector, not an AI-powered diagnostic tool, so a study on AI assistance for human readers is not relevant to this submission.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No, a standalone (algorithm only) performance study was not done. This device is an X-ray detector; its performance is evaluated as a physical imaging device, not a standalone algorithm. The "performance" refers to image quality metrics like MTF, DQE, and resolution.

    7. The Type of Ground Truth Used:

    • Type of Ground Truth: The ground truth for image quality assessment was established through expert qualitative assessment of clinical images. The stated goal was to demonstrate that "the device produces accurate X-ray images suitable for diagnostic purposes."

    8. The Sample Size for the Training Set:

    • Sample Size for Training Set: Not applicable. This is a hardware device (X-ray detector) and its associated image processing software (EConsole 2). While the software itself might have been developed using some data, the submission does not detail a "training set" in the context of an AI/ML model for diagnostic interpretation. The submission indicates physical performance testing and expert review of acquired images.

    9. How the Ground Truth for the Training Set Was Established:

    • How Ground Truth for Training Set Was Established: Not applicable, as this is not an AI/ML device with a defined "training set" requiring ground truth for model development. The focus is on the physical performance characteristics of the detector and the suitability of its output images for diagnostic purposes.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1