Search Results
Found 1 results
510(k) Data Aggregation
K Number
K023184Manufacturer
Date Cleared
2002-11-18
(55 days)
Product Code
Regulation Number
862.3040Type
TraditionalPanel
Clinical ChemistryReference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
The Bayer ADVIA IMS Ethanol (ETOH) assay is an in vitro diagnostic device intended to quantitatively measure ethanol concentration in human serum or plasma (Lithium Heparin). Measurement of ethanol is used in the diagnosis of ethanol (alcohol) toxicity and overdose.
Device Description
Not Found
AI/ML Overview
The Bayer ADVIA® IMS™ Ethanol assay is an in vitro diagnostic device intended to quantitatively measure ethanol concentration in human serum or plasma (Lithium Heparin). This assay is used in the diagnosis of ethanol (alcohol) toxicity and overdose.
1. Table of Acceptance Criteria and Reported Device Performance
Performance Metric | Acceptance Criteria (Predicate Device TDx) | Reported Device Performance (ADVIA® IMS™ Ethanol) |
---|---|---|
Imprecision (Total CV%) | ||
Level ~40 mg/dL | 5.9% | 5.3% (at 22.0 mg/dL) |
Level ~100 mg/dL | 3.2% | 2.2% (at 88.9 mg/dL) |
Level ~250 mg/dL | 3.2% | 1.8% (at 289.5 mg/dL) |
Correlation (vs. Predicate TDx - Serum) | ||
Regression Equation | Not explicitly stated, but high R-value expected | Y = 0.964x - 0.04 |
Syx (mg/dL) | Not explicitly stated | 4.8 |
R | Not explicitly stated, but close to 1 expected | 0.999 |
Sample Range (mg/dL) | Not explicitly stated | 22.9 to 433.0 |
Correlation (Plasma vs. Serum - ADVIA IMS) | ||
Regression Equation | Not explicitly stated | Y = 1.005x - 1.75 |
Syx (mg/dL) | Not explicitly stated | 2.99 |
R | Not explicitly stated, but close to 1 expected | 1.000 |
Sample Range (mg/dL) | Not explicitly stated | 10.0 to 516.2 |
Interfering Substances | Minimal interference ( |
Ask a Question
Ask a specific question about this device
Page 1 of 1