Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K092090
    Manufacturer
    Date Cleared
    2009-11-30

    (139 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    ERBELIFT HAND PUMP AND FLEXIBLE PROBE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ERBELIFT Hand Pump in conjunction with an ERBELIFT Flexible Probe is indicated for the induction of sterile normal saline into the submucosa to lift polyps or other mucosal lesions using direct visualization through a flexible endoscope.

    Device Description

    The ERBELIFT™ Hand Pump and Flexible Probe are used for the delivery of pressurized sterile normal saline to the submucosa to lift polyps or other mucosal lesions. A control syringe is filled with sterile normal saline. The filled syringe is attached to the Flexible Probe via a hub with a standard luer lock and loaded into the Hand Pump. The Flexible Probe is primed with the saline and then the distal end of the Probe is placed down the working channel of the endoscope. Under direct visualization, the Probe's tip is placed against the target tissue without puncturing the mucosa. Then, the trigger of the Hand Pump is squeezed to produce the pressurized saline from the control syringe and through the Probe into the bed of the polyp or lesion to achieve a desired tissue lift.

    The Hand Pump is made of nylon and stainless steel. It is 8-3/4" (22 cm) long, 6" (15 cm) wide, and having a thickness of 2-1/2" (6 cm). The Pump per squeeze of the handle delivers approximately 0.75 ml of normal saline. The Flexible Probe is made of Polyetheretherketone (PEEK) and nylon (Note: Medical grade inks and reducer are used to imprint logos on the device.). It has a 1.3 mm Outer Diameter and length of 260 cm. The distal tip is tapered to approximately 0.005" and delivers a jet stream of 120 um. The Hand Pump with the Flexible Probe can generate pressures in the approximate range of 145 to 650 psi which correlates to flow rates of 17 to 38 ml/minute.

    The Hand Pump is reusable with the Flexible Probe being provided as sterile and single use.

    AI/ML Overview

    The provided 510(k) summary for the Byrne Medical Incorporated ERBELIFT™ Hand Pump and Flexible Probe does not contain the detailed information required to fill out a table of acceptance criteria and reported device performance with quantitative metrics. The document focuses on demonstrating substantial equivalence to predicate devices through similarities in intended use, materials, and form, and briefly mentions functional bench and animal lab testing. However, it does not provide specific acceptance criteria or performance metrics from these tests.

    Therefore, the following sections will indicate where the information is not available in the provided text.


    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The provided document states that "Functional bench and animal lab testing has demonstrated that the ERBELIFT Hand Pump and Flexible Probe lifts the submucosa safely and effectively," but it does not provide specific acceptance criteria or quantitative performance results for these tests. Therefore, this table cannot be fully completed from the given text.

    Performance MetricAcceptance CriteriaReported Device Performance
    Mechanical Performance
    Pressure Range (Hand Pump)Not explicitly stated as acceptance criteria, but described as a characteristicApproximately 145 to 650 psi
    Flow Rate (Hand Pump)Not explicitly stated as acceptance criteria, but described as a characteristic17 to 38 ml/minute
    Volume per squeeze (Hand Pump)Not explicitly stated as acceptance criteria, but described as a characteristicApproximately 0.75 ml
    Delivery method to submucosaSafe and effective lifting of submucosaDemonstrated in functional bench and animal lab testing
    Tissue traumaDemonstrate less penetration tissue damage compared to predicatesSmaller diameter and fluid stream with no mechanical needle injection causes less penetration tissue damage. Easier to find tissue plane and create a lift.
    Biocompatibility
    Probe materials biocompatibilityDemonstrate biocompatibility of specific materialsDemonstrated for the specific materials of the Flexible Probe (due to slight differences from predicates)
    Sterilization
    Sterility (Flexible Probe)SterileProvided as sterile (Ethylene Oxide)
    Material/Design Characteristics
    Outer Diameter (Flexible Probe)Not explicitly stated as acceptance criteria, but described as a characteristic1.3 mm
    Length (Flexible Probe)Not explicitly stated as acceptance criteria, but described as a characteristic260 cm
    Distal tip taper (Flexible Probe)Not explicitly stated as acceptance criteria, but described as a characteristicApproximately 0.005"
    Jet stream size (Flexible Probe)Not explicitly stated as acceptance criteria, but described as a characteristic120 um

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not specified in the provided document. The document mentions "Functional bench and animal lab testing" but does not give details on the number of samples, animals, or tests conducted.
    • Data Provenance: Not specified. The document does not mention the country of origin of the data or whether the studies were retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • This information is not available in the provided 510(k) summary. The document mentions "Functional bench and animal lab testing" but does not detail how ground truth or performance assessment was established, nor does it refer to human expert involvement in establishing ground truth for the device's functional performance.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This information is not available in the provided 510(k) summary. Given that the testing mentioned is "functional bench and animal lab testing" for a medical device's physical performance, an adjudication method for human interpretation is unlikely to be applicable or described here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The ERBELIFT™ Hand Pump and Flexible Probe is a physical medical device for injecting saline, not an AI-powered diagnostic or interpretive tool. Therefore, a study comparing human readers with and without AI assistance is not relevant to this device and is not mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • No, standalone algorithm performance testing is not applicable as this device is a physical medical instrument, not a software algorithm. The device's operation involves a human user (endoscopist) directly.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The term "ground truth" as typically used in AI/diagnostic studies is not directly applicable here. For this physical device, the "truth" would be objective measurements and observations from the "Functional bench and animal lab testing." This would likely involve:
      • Bench Testing: Direct physical measurements (e.g., pressure, flow rate, volume delivered, jet stream characteristics) and visual inspection of fluid delivery.
      • Animal Lab Testing: Direct visualization of submucosal lift, potentially histological analysis of tissue for damage, and observation of overall safety and effectiveness in an in-vivo model.
    • The document implies that the "truth" for safety and effectiveness was derived from these direct experimental observations rather than expert consensus on interpretation of data.

    8. The sample size for the training set

    • Not applicable / Not available. The ERBELIFT™ Hand Pump and Flexible Probe is a physical medical device, not a machine learning model, so there is no "training set" in the context of data used for algorithm development. The development process would have involved design, prototyping, and testing, but not training data for an algorithm.

    9. How the ground truth for the training set was established

    • Not applicable / Not available. As explained in item 8, there is no "training set" for this physical medical device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1