Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K202879
    Date Cleared
    2021-03-03

    (156 days)

    Product Code
    Regulation Number
    880.2800
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    EOGas 4 Endo-SteriTest RRBI

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EOGas 4 Endo-SteriTest Rapid Readout Biological Indicator consists of a self-contained biological indicator inoculated with viable Bacillus atrophaeus bacterial spores that is placed in a dedicated biological indicator receptacle mounted on the purge probe in the sterilizer. It monitors the efficacy of the 6-hour gas exposure at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer.

    Device Description

    The EOGas 4 Endo-SteriTest Rapid Readout Biological Indicator (RRBI) consists of a single-use self-contained biological indicator (SCBI) placed in a reusable biological indicator (BI) receptacle. It is designed for monitoring the efficacy of the 6-hour gas exposure at 50°C in an EOCas 4 Ethylene Oxide Gas Sterilizer.

    The Bionova BT110 Rapid Readout Biological Indicator (K191021) consists of a plastic vial that serves as the culture tube and a cap including a filter material port to allow ethylene oxide to enter the vial. The plastic vial contains Bacillus atrophaeus spores inoculated onto a paper carrier, and a glass ampoule containing culture medium and a pH indicator. There is a chemical indicator printed on the unit label of the SCBI to indicate EO exposure by changing color from brown/red to green.

    Following manufacturer's instructions, the operator inserts the Bionova BT110 Rapid Readout Biological Indicator into the reusable BI receptacle on the dedicated purge probe of the EOGas 4 Ethylene Oxide Gas Sterilizer, and initiates a 6-hour gas exposure at 50°C. After cycle completion, the SCBI is retrieved and activated by crushing the glass ampoule. The chemical indicator on the SCBI changes from brown/red to a green color after ethylene oxide exposure.

    The activated SCBI and an unprocessed control are incubated in a Terragene Bionova IC10/20FR, IC10/20FRLCD or MiniBio Auto-Reader Incubator for 4 hours to detect fluorescent activity or 48 hours to detect color change. Evidence of microbial growth by presence of fluorescent activity or color change from blue to yellow must be interpreted as a failure to meet the conditions necessary for sterilization (cycle failed); no fluorescence or no color change indicates conditions for sterilization were achieved (cycle passed).

    AI/ML Overview

    The provided document describes the EOGas 4 Endo-SteriTest RRBI, a rapid readout biological indicator for monitoring ethylene oxide sterilization cycles. The submission seeks to demonstrate substantial equivalence to a predicate device.

    Here's a breakdown of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Functionality1) Critical parameters include time, temperature, gas concentration, and relative humidity 35-90% are met; 2) Device is appropriate for monitoring the efficacy of the sterilization process claimed.Pass
    Shelf LifeMaintains performance specifications (resistance characteristics and correctly indicate pass/fail in cycles) throughout the stated shelf life of 2 years; Stability demonstrates reasonable assurance for effectiveness.Pass
    Viable Spore Population$\ge 1.0 \times 10^6$ Bacillus atrophaeus (ATCC 9372) spores.$\ge 1.0 \times 10^6$
    Resistance Characteristics(Implicitly) Must be greater than the same biological indicator placed in the worst-case location of endoscope validation loads.Met (Implied by "Pass" of full functionality)
    Growth InhibitionNo inhibition of viable spore growth by carrier or primary packaging materials.Met (Implied by "Pass" of full functionality)
    Holding Time Assessment(Details not explicitly stated, but expected for BI performance)Met (Implied by "Pass" of individual studies)
    Reduced Incubation Time Validation(Details not explicitly stated, but expected for RRBI performance)Met (Implied by "Pass" of individual studies)
    Recovery Protocols for Recovery Medium(Details not explicitly stated, but expected for BI performance)Met (Implied by "Pass" of individual studies)
    Visual Readout Stability(Details not explicitly stated, but expected for RRBI performance)Met (Implied by "Pass" of individual studies)
    In-Field Evaluation(Details not explicitly stated, but expected for BI performance)Met (Implied by "Pass" of individual studies)

    Notes on Acceptance Criteria:

    • The document explicitly lists "Functionality" and "Shelf Life" tests with their acceptance criteria and "Pass" results.
    • For the individual studies performed for the Bionova BT110 Rapid Readout Biological Indicators (which is a component of the EOGas 4 Endo-SteriTest RRBI), the document states that "The results of all studies met the established acceptance criteria," without enumerating the specific criteria for each.

    2. Sample Sizes Used for the Test Set and Data Provenance

    The document does not provide specific sample sizes (e.g., number of BIs or sterilization cycles) for the "test set" studies. It mentions that "The EOGas 4 Endo-SteriTest RRBI has been validated using applicable tests in FDA 2007 'Guidance for Industry and FDA Staff: Biological Indicator (BI) Premarket Notification [510(k)] Submissions', and ANSI/AAMI/ISO 11138-1:2017 'Sterilization of health care products -Biological indicators - Part 1: General requirements'." These standards typically define the requirements for testing, including sample sizes, but the specific numbers used in this submission are not stated.

    Data provenance is not explicitly mentioned but is assumed to be prospective bench testing conducted by the manufacturer as part of the validation process for regulatory submission. There is no mention of country of origin for data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. This device is a biological indicator designed to directly measure sterilization efficacy, not to be interpreted by human experts like medical images. The "ground truth" for a biological indicator is the direct observation of microbial growth (or lack thereof) after exposure to a sterilization process, relative to positive and negative controls.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, the device's output (microbial growth/no growth, or fluorescent activity/no fluorescent activity) is a direct, objective measurement, not subject to subjective interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. This is not an AI-assisted diagnostic or imaging interpretation device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is not an AI algorithm. The device itself (the biological indicator) is a standalone measurement tool.

    7. The Type of Ground Truth Used

    The ground truth used for biological indicators is the viability of the Bacillus atrophaeus bacterial spores.

    • Failed Sterilization: Indicated by observable fluorescent activity or a color change from blue to yellow in the culture medium, signifying spore survival and growth.
    • Successful Sterilization: Indicated by no fluorescence and no color change, signifying spore inactivation.
    • Positive controls (unprocessed BIs) are expected to show growth, and negative controls (processed BIs from a successful cycle) are expected to show no growth, confirming the validity of the test.

    8. The Sample Size for the Training Set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192980
    Date Cleared
    2020-11-12

    (385 days)

    Product Code
    Regulation Number
    880.2800
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    EOGas 4 Endo-SteriTest

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EOGas 4 Endo-SteriTest consists of a self-contained biological indicator inoculated with viable Bacillus atrophaeus bacterial spores that is placed in a dedicated biological indicator receptacle mounted on a gold-colored purge probe in the sterilizer. It monitors the efficacy of the 6-hour gas exposure at 50°C in the EOGas 4 Ethylene Oxide Gas Sterilizer.

    Device Description

    The EOGas 4 Endo-SteriTest consists of a single-use self-contained biological indicator (SCBI) placed in a reusable biological indicator (BI) receptacle. It is designed for monitoring the efficacy of the 6-hour gas exposure at 50°C in an EOGas 4 Ethylene Oxide Gas Sterilizer.

    The SCBI, the EZTest-Gas Biological Indicator, consists of a plastic vial that serves as the culture tube, and a cap including a filter material port to allow ethylene oxide to enter the vial. The plastic vial contains Bacillus atrophaeus spores inoculated onto a paper carrier, and a glass ampoule containing modified soybean casein digest broth and phenol red acting as a pH indicator. There is a chemical indicator printed on the unit label of the SCBI to indicate EO exposure.

    Following manufacturer's instructions, the operator inserts the EZTest-Gas Biological Indicator into the reusable BI receptacle on the dedicated purge probe of the EOGas 4 Ethylene Oxide Gas Sterilizer, and initiates a 6-hour gas exposure at 50℃. After cycle completion, the EZTest-Gas Biological Indicator is retrieved and activated by crushing the glass ampoule. The chemical indicator on the SCBI changes from blue to a green/brown color depending on the duration of ethylene oxide exposure. The activated SCBI and an unprocessed control are incubated at 30- 35°C for 48 hours and monitored for any color change and/or turbidity. Evidence of microbial growth by color change from red-orange to yellow and/or turbidity must be interpreted as a failure to meet the conditions necessary for sterilization (cycle failed); no color change or turbidity indicates conditions for sterilization were achieved (cycle passed).

    AI/ML Overview

    The provided text describes the acceptance criteria and the results of studies conducted for the EOGas 4 Endo-SteriTest, a biological indicator for ethylene oxide sterilization.

    Here's a breakdown of the requested information based on the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    EZTest-Gas Biological Indicator Performance Characteristics
    Viable Spore Population Assays≥ 10^6 microorganisms per unit; Stable for the stated 2-year shelf lifeMeets the specifications for spore population according to 2007 FDA BI guidance and ANSI/AAMI/ISO 11138-1
    Resistance Characteristics Study (D-value)D-value ≥ 3 min; Survival/Kill Time: 15 min; D-values remain ±20% of the stated value for the claimed 2-year shelf lifeMeets the specifications for resistance characteristics according to 2007 FDA BI guidance and ANSI/AAMI/ISO 11138-1
    Growth InhibitionAll exposed carriers and packaging components are not bacteriostatic in a worst-case situationCompliance with 2007 FDA BI guidance and ANSI/AAMI/ISO 11138-1
    Holding Time AssessmentThe resistance characteristics are not altered significantly over a 7-day hold time prior to incubation.Compliance with 2007 FDA BI guidance
    Recovery Protocols, Medium and Incubation TimeGrowth of an inoculum of 10-100 spores of Bacillus atrophaeus after being subjected to the sterilization process.Compliance with 2007 FDA BI guidance and ANSI/AAMI/ISO 11138-1; Data generated supports an incubation time of 48 hours for EZTest-Gas biological indicators.
    D-Value Comparison for EZTest-Gas Biological Indicator
    D-value (in vacuum BIER vs. A-BIER)D-values are within 20% of the original stated values for all lots of biological indicators tested.There is no significant difference in the D-values measured in a vacuum BIER vs. an A-BIER.
    Performance Testing for the EOGas 4 Endo-SteriTest
    Half Dose ValidationDevice is appropriate for monitoring the efficacy of the sterilization process claimedPartial kill of biological indicators in PCD in half dose cycles, and complete kill in full dose cycles; Biological indicators at the worst-case locations in the endoscope loads are completely inactivated in both half and full dose cycles; EO concentration in the full dose cycles are twice the concentration in the half dose cycles for each endoscope load.
    Pass/Fail Results from the EOGas 4 SterilizerCorrectly indicate pass/fail in cyclesData demonstrate the EOGas 4 Endo-SteriTest correctly indicates passed and failed 6-hour gas exposures in the EOGas 4 throughout the stated shelf life.
    Endpoint Color Stability of the Chemical IndicatorStability to provide reasonable assurance of effectivenessThe endpoint green/brown color was stable for at least 28 days when EZTest-Gas BIs were stored at 20-25°C or at 35-39°C after the EOGas 4 sterilization cycles.
    Shelf LifeStability of the D-Value and spore population; Correctly indicate pass/fail in cycles throughout the stated shelf lifeMaintains performance specifications throughout the stated shelf life of 2 years

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the specific sample sizes (number of biological indicators, number of runs, etc.) for each individual test. It mentions that "lots of biological indicators" were tested for D-value comparison, and refers to "half dose cycles" and "full dose cycles" for the half-dose validation. It does not provide information on the country of origin or whether the data was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is a biological indicator for sterilization. Its performance is evaluated through microbial growth/inactivation and chemical indicator changes, not by human expert interpretation of images or other data typically requiring expert consensus.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. The ground truth for this device is based on objective biological and chemical responses, not subjective expert judgment that would require adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a biological indicator for monitoring sterilization efficacy, not an AI-assisted diagnostic device that involves human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is a standalone device in the context of it generating its own result (color change, microbial growth/no growth). Its performance is evaluated directly through its biological and chemical responses to sterilization conditions, independent of human intervention for its core function.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth used for this biological indicator is primarily the inactivation or survival of Bacillus atrophaeus bacterial spores when subjected to defined sterilization conditions, and the physical/chemical change of an indicator. This is an objective, measurable biological and chemical response, rather than expert consensus, pathology, or outcomes data.

    8. The sample size for the training set

    Not applicable. This device is a biological indicator, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As this is not a machine learning model, there is no training set or associated ground truth establishment process in that context. The "ground truth" for evaluating the BI's performance is established by microbiological standards for spore inactivation.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1