Search Results
Found 1 results
510(k) Data Aggregation
(160 days)
ENDOEYE HD II (WA50040A, WA50042A, WA50050A, WA50052A)
This instrument has been designed to be used with a video system center, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs.
The ENDOEYE HD II - High Definition Digital Video Laparoscope is a rigid video telescope used for endoscopic diagnosis, treatment, video observation and surgery within the thoracic and abdominal cavities including female reproductive organs. For laparoscopic applications, the video telescope is inserted via a trocar into the patient. The ENDOEYE HD II is used with a video system center, light source and monitor to achieve its intended function. In addition, the ENDOEYE HD II can be inserted in compatible instrument trays for reprocessing. The ENDOEYE HD II can provide an image with either white light or narrow band imaging and WA50040A and WA50042A include a heater function at the distal tip to reduce fogging of the lens.
This document describes the premarket notification (510(k)) for the Olympus ENDOEYE HD II, a rigid video telescope used for endoscopic diagnosis, treatment, video observation, and surgery. This is primarily a regulatory submission for a medical device and not a study proving an AI/ML device meets acceptance criteria.
Therefore, many of the requested details, such as acceptance criteria for AI/ML performance metrics, sample sizes for AI/ML test/training sets, expert adjudication methods for AI/ML ground truth, and MRMC studies, are not applicable to this document. This document focuses on demonstrating substantial equivalence to a predicate device through engineering performance data rather than sophisticated clinical or AI/ML performance studies.
However, based on the provided text, I can infer and state the following regarding "acceptance criteria" and "study" in the context of this device:
1. A table of acceptance criteria and the reported device performance:
The document doesn't provide a table of specific numerical acceptance criteria for each performance test, nor does it present detailed quantitative performance results in a table. Instead, it states that "All samples tested met their predefined acceptance criteria." The acceptance criteria would be linked to the standards mentioned for each test.
Acceptance Criteria Category (Inferred from tests) | Reported Device Performance (Summary Statement) |
---|---|
Electrical Safety (IEC 60601-1 Ed 3.1, IEC 60601-2-18 Ed 3.0) | All samples tested met their predefined acceptance criteria. |
Electromagnetic Compatibility (EMC) (IEC 60601-1-2 Ed 4.0) | All samples tested met their predefined acceptance criteria. |
Thermal Safety (IEC 60601-1 Ed 3.1, IEC 60601-2-18 Ed 3.0) | All samples tested met their predefined acceptance criteria. |
Mechanical Performance (ISO 8600-3 Ed 1, ISO 8600-4 Ed 2, ISO 8600-5 Ed 1) | All samples tested met their predefined acceptance criteria. |
Transport and Shipping (ASTM D4169-16) | All samples tested met their predefined acceptance criteria. |
Expected Service Life (IEC 60601-1 Ed 3.1, IEC 60601-2-18 Ed 3.0) | All samples tested met their predefined acceptance criteria. |
Illumination System (IEC 60601-1 Ed 3.1, IEC 60601-2-18 Ed 3.0) | All samples tested met their predefined acceptance criteria. |
Software Verification (IEC 62304 Ed 1.1) | All samples tested met their predefined acceptance criteria. |
Usability Validation for Instructions for Use (FDA Guidance) | All samples tested met their predefined acceptance criteria. |
Risk Analysis (ISO 14971:2007) | Risk analysis carried out in accordance with established internal acceptance criteria. |
2. Sample size used for the test set and the data provenance:
- Sample size: The document does not specify the exact sample sizes (number of units) used for each individual test. It generally refers to "All samples tested."
- Data provenance: Not explicitly stated regarding country of origin or specific patient data. The tests are lab-based design verification and validation tests rather than clinical data tests. It's a regulatory submission from a German manufacturer (Olympus Winter & Ibe GmbH) to the US FDA. The tests are "Performance Data" and are likely performed in a controlled, prospective manner as part of the design and manufacturing process.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. This device is a hardware endoscopic camera. The "ground truth" for its performance is established by meeting engineering specifications and recognized international standards (e.g., IEC, ISO, ASTM) through objective physical and software testing, not by expert human interpretation of medical images or conditions.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable. As the "ground truth" is established by adherence to engineering standards and objective measurements, there is no need for expert adjudication of medical findings.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This is not an AI-enabled device. Therefore, no MRMC study looking at human reader improvement with or without AI assistance was conducted or is relevant for this regulatory submission.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This device is a camera system, not an algorithm that performs a standalone diagnostic or analytical function.
7. The type of ground truth used:
- The ground truth is based on engineering specifications, recognized international standards (e.g., IEC, ISO, ASTM), and internal company acceptance criteria for hardware and software performance. This includes electrical safety, EMC, thermal safety, mechanical performance, transport resistance, service life, illumination, and software functionality.
8. The sample size for the training set:
- Not applicable. This is not an AI/ML device that requires a "training set."
9. How the ground truth for the training set was established:
- Not applicable. See point 8.
Ask a specific question about this device
Page 1 of 1