510(k) Data Aggregation

    K Number
    K040668
    Device Name
    E-Z SCAN AB5500+
    Manufacturer
    Date Cleared
    2004-04-13

    (29 days)

    Product Code
    Regulation Number
    892.1560
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    E-Z SCAN AB5500+

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The E-Z scan AB 5500+ ultrasound system is a multi-purpose computer-based ultrasonic diagnostic system ophthalmic applications, intended to both visualize the interior of the eye by means of ultrasound and to mak measurements inside the eye, including the measurement of axial length for determination of IOL power.

    The Intended Use of the A_Mode probe for the E-Z scan AB 5500+ ultrasound system is to make measurements inside the eye, including the measurement of axial length for determination of IOL power, and to visualize the eye by means of A-scans.

    The intended use of the B-Mode for the E-Z scan AB 5500 is to visualize the interior of the eye by means of ultrasound and to make measurements inside the eye by B-scan ultrasound.

    Device Description

    The E-Z Scan AB5500* combines a contact B-scanner used for the visualization by ultrasound of the eye and orbit and an A-scan used for intraocular measurements. The intended use of this system includes the localization and visualization of ophthalmic disorders and measurement of the eye and orbit.

    The E-Z Scan AB5500+ is a conventional ophthalmic B-scan system using a motordriven transducer and angle sensor for scanning and a conventional contact A-scan system. The transducer frequency is 10 MHz. It uses a motor-driven 10 MHz transducer with an attached angle encoder. The display is on a video touch screen, also used for controlling the system. The entire device is computer-controlled by an internal microprocessor. The A-scan uses a separate 10 MIIz transducer and its own pulserreceiver.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Sonomed E-Z Scan AB5500+, an ophthalmic A and B scan system. Here's an analysis of the acceptance criteria and the study information based on the given document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriterionReported Device PerformanceComments
    SafetyDetermined equivalent to legally marketed predicate device.Non-clinical tests included ultrasonic emissions tests. Clinical tests were "Not required."
    EfficacyDetermined equivalent to legally marketed predicate device.Non-clinical tests included accuracy and validation tests. Clinical tests were "Not required."
    Intended UseThe device is intended for localization and visualization of ophthalmic disorders and measurement of ocular distances, including axial length for IOL power determination.This matches the intended use of the predicate devices.
    Technological CharacteristicsThe E-Z Scan AB5500+ is a conventional ophthalmic B-scan system using a motor-driven transducer and angle sensor (10 MHz) and a conventional contact A-scan system with a separate 10 MHz transducer. Displays on a video touch screen and is computer-controlled.This information describes the device's technical specifications and is considered acceptable for demonstrating substantial equivalence to the predicate.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document explicitly states "Clinical tests: Not required" for both safety and efficacy. This indicates that no clinical test set was used for this 510(k) submission. The substantial equivalence was based on non-clinical tests (ultrasonic emissions, accuracy, and validation) and comparison to existing predicate devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Since no clinical test set was used, there is no information on experts establishing ground truth for a test set. Ground truth was not established via expert review of clinical data for this submission.

    4. Adjudication Method for the Test Set

    As no clinical test set was used, there was no adjudication method for a test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No MRMC comparative effectiveness study was mentioned or performed, as clinical tests were "Not required."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The device is an ultrasound system directly operated by a human, not an algorithm. Therefore, a standalone algorithm performance study is not applicable and was not done.

    7. The Type of Ground Truth Used

    For the safety and efficacy assessment, the ground truth was based on the performance and established safety/efficacy of the legally marketed predicate devices (Sonomed Ophthalmic B-scan B-3000 and Sonomed A-scan A-2000), as well as non-clinical test results (ultrasonic emissions, accuracy, and validation).

    8. The Sample Size for the Training Set

    There is no mention of a training set in the document. This is common for predicate-based 510(k) submissions for conventional medical devices that do not involve AI/ML algorithms.

    9. How the Ground Truth for the Training Set Was Established

    As no training set was used, ground truth for a training set was not established.