Dual Lumen 5Fr HydroPICC Catheter (PICC-251CM) (PICC-251CM); Dual Lumen 5Fr HydroPICC Catheter (PICC
Dual Lumen 5Fr HydroPICC Catheter (PICC-251CM): Indicated for short-or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media. Rated for maximum power injection flow rate of 4.0ml/s.
Dual Lumen 5Fr HydroPICC Catheter (PICC-252CM): Indicated for short-or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to the administration of fluids, medications, and nutrients; the sampling of blood; central venous pressure monitoring; and power injection of contrast media. Rated for maximum power injection flow rate of 5.0ml/s.
The HydroPICC catheters are a family of peripherally inserted central catheters (PICC) made of radiopaque hydrophilic material with a suture wing, Luer lock hub, and extension tubing made from materials commonly used in catheter manufacturing. The catheters are provided in kit configurations with the necessary accessories for placement in clinical environments. The maximum power injection flow rate for the lumen is indicated on the extension tube clamp.
HydroPICC has been shown to be effective in reducing thrombus accumulation and thrombotic occlusions. These reductions were evaluated using in vitro and in vivo models. Pre-clinical evaluations do not necessarily predict clinical performance with respect to thrombus formation. HydroPICC 5F Dual Lumen Catheter Components include: HydroPICC 5Fr Dual Lumen Catheter Assembly, Dispensing Tube, 2 Channel Clip, and Guidewire Introducer and Straightener.
The provided FDA 510(k) clearance letter pertains to a medical device (PICC catheter), not an AI/Software as a Medical Device (SaMD) product. Therefore, the content of the letter does not contain the information requested in your prompt regarding acceptance criteria and studies for AI/SaMD, such as:
- A table of acceptance criteria and reported device performance (for AI metrics like sensitivity, specificity, etc.)
- Sample sizes used for test sets and data provenance (for image/data-driven AI)
- Number of experts and their qualifications for ground truth establishment
- Adjudication method for ground truth
- MRMC comparative effectiveness study results or effect sizes
- Standalone algorithm performance
- Type of ground truth used (e.g., pathology, outcomes data)
- Sample size for training set
- How ground truth for the training set was established
The 510(k) summary for the HydroPICC Catheter focuses on physical device performance characteristics. It mentions:
- Acceptance Criteria/Performance (Implied through testing): The device was tested to demonstrate acceptable performance for the modified power injection ratings and device length.
- Study Type: Non-clinical tests were performed, specifically "Power injection Cyclical and Static Burst testing as well as dimensional verification testing."
- Conclusion: These tests demonstrated that "the modifications to the power injection rating and the length do not adversely impact product performance."
In summary, this document does not contain the specific information you requested relating to the acceptance criteria and study proving performance for an AI/SaMD device. The information provided is for a physical medical device clearance based on engineering and performance bench testing.