Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K192258
    Date Cleared
    2019-12-30

    (132 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Disposable Valve Sets

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Disposable Valve Sets includes either one air/water and suction valve or one air/water, suction and biopsy valve.

    Disposable Air/Water Valve: This device intended to be fitted to an endoscope air/water channel to enable the operator to control inflow of medical gases and water, whilst preventing backflow during a GI Endoscopic procedure.

    Disposable Suction Valve: The device intended to be fitted to an endoscope suction channel to enable the operator to control suction whilst preventing inflow of air during a GI Endoscopic procedure.

    Disposable Biopsy Valve: This device is intended to be fitted to an endoscope biopsy port to enable access for/exchange of endoscopic devices while maintaining insufflation and minimizing leakage of bio material during an endoscopic procedure.

    Device Description

    Disposable Valve Sets includes either one air/water and suction valve or one air/water, suction and biopsy valve. So we divided into two models for these devices. One suction valve, air/water valve and biopsy valve means model B, one suction valve and air/water valve means model A. These devices are supplied sterile. These valves easily incorporate into infection prevention policies as a single use item, and these valves are suitable for Olympus 140/160/190/240/260 series endoscopes.

    AI/ML Overview

    The document describes the acceptance criteria and study to demonstrate substantial equivalence for the Disposable Valve Sets (intended for use with endoscopes) by Zhejiang Chuangxiang Medical Technology Co., LTD. against the predicate device Endorate™ Valve Sets (K181509) by Smardata Suzhou Co., Ltd.

    Here's the detailed breakdown as requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" alongside "reported device performance" in a quantitative manner (e.g., specific thresholds and measured values). Instead, it states that "In-vitro Testing has been performed and all components, subassemblies, and/or full devices met the required specifications for the completed tests." and "The testing successfully demonstrated essential performance is achieved before and after the shelf life test."

    The following tests were performed and implicitly form the basis of the acceptance criteria by ensuring the device meets required specifications:

    Test CategorySpecific Tests PerformedReported Device Performance
    Disposable Suction Valve1. Endoscope CompatibilityMet required specifications (implicit)
    2. Vacuum Leak TestMet required specifications (implicit)
    3. Suction Flow TestMet required specifications (implicit)
    4. Depression Force TestMet required specifications (implicit)
    Disposable Air/Water Valve1. Endoscope CompatibilityMet required specifications (implicit)
    2. Water Flow TestMet required specifications (implicit)
    3. Air Flow TestMet required specifications (implicit)
    4. Leak TestMet required specifications (implicit)
    5. Depression Force TestMet required specifications (implicit)
    Disposable Biopsy Valve1. Endoscope CompatibilityMet required specifications (implicit)
    2. Leak TestMet required specifications (implicit)
    3. Squeegee Leak TestMet required specifications (implicit)
    SterilizationEO residual measurement (ISO 10993-7)Met criteria defined in ISO 10993-7
    Sterility Assurance Level (SAL)10⁻⁶
    Shelf-LifeAccelerated testing (ASTM F1980-16), Packaging (ISO 11607-1, ISO 11607-2)Essential performance achieved before and after shelf life
    BiocompatibilityCytotoxicity, Sensitization, Intracutaneous Irritation (ISO 10993-1, FDA Guidance)Device demonstrated biocompatibility
    Overall PerformanceIn-vitro tests for performance and biocompatibilityPerforms as well as the predicate in all testing performed

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the exact sample sizes used for each of the in-vitro performance tests. It generally states that "In-vitro Testing has been performed" on "components, subassemblies, and/or full devices."

    The data provenance is not explicitly stated in terms of country of origin or retrospective/prospective nature. However, given that "Zhejiang Chuangxiang Medical Technology Co., LTD." is located in Hangzhou, Zhejiang, China, it is highly probable that the testing was conducted in China. These would be considered prospective laboratory tests using newly manufactured device samples.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This device is a physical medical instrument (valve sets), not an AI algorithm or diagnostic tool that relies on expert interpretation of data or images to establish a "ground truth." Therefore, the concept of "number of experts used to establish the ground truth" is not applicable in this context. The "ground truth" for this device's performance is established by objective engineering and safety standards (e.g., flow rates, leak resistance, mechanical force, biocompatibility standards).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in studies involving human interpretation (e.g., radiology reads), especially for AI clinical trials. The performance of these physical valves is assessed against predefined engineering and material standards through objective laboratory measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device (endoscope valve sets), not an artificial intelligence-powered diagnostic or assistive tool. MRMC studies are used to evaluate AI's impact on human reader performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device. The concept of "standalone algorithm performance" does not apply. The performance tests evaluate the device itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance tests of the Disposable Valve Sets is based on objective engineering and safety specifications and recognized international standards. This includes:

    • Physical performance criteria: Such as vacuum leak rates, flow rates (water and air), depression force, and leak resistance (squeegee leak test). These are quantifiable measurements.
    • Sterilization standards: Meeting criteria defined in ISO 10993-7 for EO residual and achieving a SAL of 10⁻⁶.
    • Shelf-life standards: Compliance with ASTM F1980-16 for accelerated aging and ISO 11607-1/2 for packaging.
    • Biocompatibility standards: Compliance with FDA Guidance and ISO 10993-1 for cytotoxicity, sensitization, and intracutaneous irritation.

    There is no "expert consensus" or "pathology" involved in determining if a valve leaks or if it's biocompatible beyond interpreting the results of the objective tests against established thresholds.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not an AI algorithm. Therefore, there is no "training set" in the context of machine learning. The device design and manufacturing processes are validated through engineering principles and testing.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1