510(k) Data Aggregation

    K Number
    K161691
    Date Cleared
    2017-03-20

    (273 days)

    Product Code
    Regulation Number
    862.1475
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Direct LDL Cholesterol (LDL)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For the quantitative in vitro determination of LDL-cholesterol concentration in human plasma and serum. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders mellitus), atherosclerosis and various liver and renal diseases.

    This in vitro diagnostic device is intended for prescription use only.

    Device Description

    The LDL Cholesterol kit assay consists of ready to use reagent solutions. CATALOGUE NUMBER: CH8312

    R1. Enzyme Reagent 1 4 x 20 mL R2. Enzyme Reagent 2 4 x 9 mL

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    Device Name: Direct LDL Cholesterol (LDL)

    Acceptance Criteria CategorySpecific MetricAcceptance Criteria (Implied/Direct)Reported Device Performance
    PrecisionTotal CV % (within run, among run, among day)Generally expected to be low for clinical assays, with specific targets often dependent on concentration levels.QC 1 (92.0 mg/dl): 5.9% total CV
    QC 2 (135.9 mg/dl): 4.6% total CV
    QC 3 (186.7 mg/dl): 4.4% total CV
    Serum pool 1 (65.0 mg/dl): 5.9% total CV
    Serum pool 2 (154.0 mg/dl): 5.0% total CV
    Serum pool 3 (200.1 mg/dl): 5.0% total CV
    Serum pool 4 (343.7 mg/dl): 5.3% total CV
    Linearity/Reportable RangeLinear Regression Correlation Coefficient (r)Close to 1.0 (indicating a strong linear relationship)r = 0.997
    Reportable RangeDefined range where results are linear.21 - 740 mg/dl
    Detection LimitLimit of Blank (LoB)Very low, ideally close to zero, to ensure no signal from blank.1.94 mg/dl
    Limit of Detection (LoD)Low enough to reliably detect the analyte.3.19 mg/dl
    Limit of Quantitation (LoQ)Low enough for precise and accurate quantification at low concentrations (typically ≤20% imprecision).16.1 mg/dl (with ≤20% imprecision)
    Analytical SpecificityInterference (% of Control)% of Control ± 10% for potential interferents at specified concentrations.Hemoglobin: No significant interference up to 1000mg/dl.
    Total Bilirubin: No significant interference up to 60mg/dl.
    Conjugate Bilirubin: No significant interference up to 60mg/dl.
    Triglycerides: No significant interference up to 500mg/dl.
    Intralipid®: No significant interference up to 500mg/dl.
    Ascorbic Acid: No significant interference up to 6mg/dl.
    Method ComparisonLinear Regression Correlation Coefficient (r)Close to 1.0 when compared to a predicate device, indicating substantial equivalence.r = 0.998 (compared to predicate device)
    Matrix ComparisonLinear Regression Correlation Coefficient (r)Close to 1.0 when comparing serum and lithium heparin plasma, indicating equivalent performance across matrices.r = 0.998 (serum vs. lithium heparin plasma)
    TraceabilityConformance to reference materials/standardsTraceable to an internal master reference material. Not certified by CRMLN (stated as a disclaimer in labeling).Traceable to an internal master reference. Labeling states "device has not been certified by the CRMLN."

    Study Details

    1. Sample size used for the test set and the data provenance:
      • Precision (Analytical Performance): 80 determinations for each of 7 pools/QC levels (total of 560 determinations). The samples included control material and "unaltered human serum samples that were spiked with LDL cholesterol concentrations or diluted to achieve concentrations based on established ranges" (e.g.,