Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K241642
    Date Cleared
    2024-08-02

    (56 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Diode Laser Hair Removal System (MBT-Diode Laser)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode Laser Hair Removal System (Model: MBT-Diode Laser) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term. stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    Diode Laser Hair Removal System (Model: MBT-Diode Laser) is a surgical device, which is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI). The Dermatological Diode Laser Systems utilize a semiconductor diode with invisible infrared radiation as a laser source to emit 808nm wavelength laser which is absorbed by melanin. The laser power is delivered to the treatment area via laser handle. The emission laser is activated by footswitch or handle button.

    The Diode Laser Hair Removal System consists of the following major components

    • The main console unit that incorporates the main CPU board, power supply modules, laser power supply, laser device, cooling system and switching module
    • Handpiece
    • Footswitch
    AI/ML Overview

    The provided text is a 510(k) Summary for a Diode Laser Hair Removal System (MBT-Diode Laser). It describes the device, its intended use, and a comparison to a predicate device to demonstrate substantial equivalence.

    Here's an analysis of the acceptance criteria and study information, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not outline specific, quantified acceptance criteria for the new device as would be seen in a clinical study. Instead, it demonstrates substantial equivalence to a predicate device by comparing technical specifications and adherence to recognized standards.

    The table below summarizes the technical comparison provided, where "acceptance criteria" can be interpreted as demonstrating equivalence or justification for differences compared to the predicate/reference devices:

    FeatureAcceptance Criteria (Comparison to Predicate/Reference)Reported Device Performance (Proposed Device)
    General
    Device NameNA (for comparison)Diode Laser Hair Removal System
    Classification RegulationSAME as predicate (21 CFR 878.4810)21 CFR 878.4810
    Classification PanelSAME as predicate (General & Plastic Surgery)General & Plastic Surgery
    ClassSAME as predicate (II)II
    Product CodeSAME as predicate (GEX)GEX
    Indications for UseSAME as predicate: Hair removal, permanent hair reduction on all skin types (Fitzpatrick I-VI), including tanned skin. Permanent hair reduction defined as long-term, stable reduction measured at 6, 9, and 12 months post-treatment.Same as criterion
    Performance (Technical)
    Laser TypeSAME as predicate (Diode Laser)Diode Laser
    Laser ClassificationSAME as predicate (Class IV)Class IV
    Laser WavelengthSAME as predicate (808nm)808nm
    Spot SizeSIMILAR to predicate (15mm x 15mm vs. 2.25cm², 1.44cm²)15mm x 15mm
    FluenceDifferent from predicate (0-120 J/cm²), but justified as lower (55.96 J/cm²) and similar to reference device (48 J/cm²) for effectiveness. Safety ensured by compliance with standards.1-55.96 J/cm²
    FrequencySAME as predicate (1-10Hz)1-10Hz
    Pulse DurationDifferent from predicate (10-400ms), but within predicate range (2-240ms) and considered a minor difference not affecting effectiveness/safety. Safety ensured by compliance with standards.2-240ms
    Power SupplyDifferent from predicate (220/110 VAC/50Hz-60Hz), but electrical safety and EMC tests conducted ensure normal operation.110 VAC/50Hz-60Hz
    DimensionDifferent from predicate (112 cm x 42 cm x 60 cm), but mechanical performance accepted by IEC 60601-1.650mm x 580mm x 1210mm
    WeightDifferent from predicate (63Kg), but mechanical performance accepted by IEC 60601-1.68Kg
    Safety
    Electrical SafetyComply with IEC 60601-1, IEC 60601-2-22 (SAME as predicate)Complies
    EMCComply with IEC 60601-1-2 (SAME as predicate)Complies
    Laser SafetyComply with IEC 60601-2-22, IEC 60825 (SAME as predicate)Complies
    CytotoxicityNo Cytotoxicity (SAME as predicate; aligns with ISO 10993-5)No Cytotoxicity
    SensitizationNo evidence of sensitization (SAME as predicate; aligns with ISO 10993-10)No evidence of sensitization
    IrritationNo evidence of irritation (SAME as predicate; aligns with ISO 10993-23)No evidence of irritation

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "No clinical study is included in this submission."
    Therefore, there is no "test set" in the context of clinical performance data. The evaluation relies on non-clinical (bench) testing and comparison to predicate devices, not human subject data.

    Since no clinical study was conducted, there is no data provenance (country of origin, retrospective/prospective) for a test set.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. As no clinical study was conducted and no clinical test set was used, there was no ground truth for a test set established by experts.

    4. Adjudication Method for the Test Set

    Not applicable. As no clinical study was conducted and no clinical test set was used, there was no adjudication method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. The document explicitly states: "No clinical study is included in this submission." Therefore, no MRMC study was conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The device is a hardware product (Diode Laser Hair Removal System), not an algorithm or software-only device. Therefore, the concept of "standalone algorithm performance" as typically applied to AI/software as a medical device (SaMD) is not applicable here. The performance is evaluated based on its physical/electrical characteristics and safety standards.

    7. The Type of Ground Truth Used

    For the non-clinical tests regarding electrical safety, EMC, laser safety, and biocompatibility, the "ground truth" is defined by compliance with internationally recognized standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 60601-2-22, ISO 10993 series). The device's performance is measured against the requirements of these standards.

    8. The Sample Size for the Training Set

    Not applicable. This is a hardware device submission, not an AI/ML algorithm requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. No training set was used.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1