Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K182225
    Date Cleared
    2019-03-04

    (200 days)

    Product Code
    Regulation Number
    862.1215
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Dimension Vista® High-Sensitivity Troponin I (TNIH) assay is for in vitro diagnostic use in the quantitative measurement of cardiac troponin I in human plasma using the Dimension Vista system. The assay can be used to aid in the diagnosis of acute myocardial infarction (AMI).

    Device Description

    The Dimension Vista® TNIH assay is a homogeneous, sandwich chemiluminescent immunoassay based on LOCI® technology. The LOCI reagents include two synthetic bead reagents and two biotinylated anti-cardiac troponin I monoclonal antibody fragments. The first bead reagent (Sensibeads) is coated with streptavidin and contains photosensitizer dye. The second bead reagent (Chemibeads) is coated with a third anticardiac troponin I monoclonal antibody and contains chemiluminescent dye. Sample is incubated with Chemibeads and biotinylated antibodies to form bead-cardiac troponin Ibiotinylated antibody sandwiches. Sensibeads are added and bind to the biotin to form bead-pair immunocomplexes. Illumination of the complex at 680 nm generates singlet oxygen from Sensibeads which diffuses into the Chemibeads, triggering a chemiluminescent reaction. The resulting signal is measured at 612 nm and is a direct function of the cardiac troponin I concentration in the sample. Lithium heparin plasma specimens may be used. The reagent is stored unopened at 2 – 8 °C, is stable sealed on system for 30 days and opened on the system for 7 days. Calibration is performed every 30 days for a reagent lot.

    AI/ML Overview

    The provided document describes the Siemens Healthcare Diagnostics Dimension Vista High-Sensitivity Troponin I (TNIH) Assay. The following information regarding its acceptance criteria and a study proving its performance is extracted:

    Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance CriteriaReported Device Performance
    Detection Limit
    - Limit of Blank (LoB)95th percentile of all values, measured using a non-parametric approach.1.0 pg/mL
    - Limit of Detection (LoD)Non-parametric approach following CLSI EP17-A2.2.0 pg/mL (consistent with range of 0.7 - 1.9 pg/mL)
    Limit of Quantitation (LoQ)Analyte level with a within-lab CV of ≤ 20.0%.3.0 pg/mL (Lot-1: 1.2 pg/mL, Lot-2: 2.4 pg/mL, Lot-3: 2.2 pg/mL)
    10% CV LimitAnalyte level with a within-lab CV of ≤ 10.0%.Consistent with 10.0 pg/mL (Lot-1: 2.7 pg/mL, Lot-2: 5.7 pg/mL, Lot-3: 5.4 pg/mL)
    PrecisionAll precision goals were met. (Specific goals for SD and %CV not explicitly detailed but implied by "All precision goals were met" statement).Repeatability: Plasma 1 (48.9 pg/mL): 1.12 SD, 2.3% CV; Plasma 2 (157.7 pg/mL): 1.55 SD, 1.0% CV; QC (8088.5 pg/mL): 99.54 SD, 1.2% CV.
    Within-Lab: Plasma 1: 3.05 SD, 6.2% CV; Plasma 2: 2.60 SD, 1.6% CV; QC: 200.36 SD, 2.5% CV.
    LinearityThe p values of nonlinear terms in the quadratic and cubic fit equations are non-significant (p ≤ 0.05). If p-value > 0.05, then allowable bias is ≤ 10% or 3 pg/mL, whichever is greater.Confirmed linearity from 3.0 - 25,000.0 pg/mL.
    InterferencesBias exceeding 10% is considered interference.Exogenous Compounds: No interference detected from Hemoglobin (400 mg/dL), Bilirubin (conjugated 30 mg/dL, unconjugated 40 mg/dL), Lipemia (3000 mg/dL) at approximately 40 pg/mL and 1350 pg/mL cTnI.
    Numerous therapeutic/toxic substances: No interference detected at specified low/therapeutic and high/toxic concentrations (details in document for each substance).
    High Dose Hook EffectNot explicitly stated as a numerical criterion, but implies absence of hook effect.No hook effect found at 1,000,000 pg/mL troponin.
    Dilution RecoverySupported use of diluent for over-range samples. (Specific numerical criteria not explicitly stated).Testing supported use of the diluent for over-range samples.
    Calibration StabilityP-value of regression slope ≥ 0.05 OR drift ≤ LoQ OR drift ≤ 10% for values up to 20,000 pg/mL OR drift ≤ 13% for values > 20,000 pg/mL.Calibration interval measured to be 30 days.
    Open Well StabilityP-value of regression slope ≥ 0.05 OR drift ≤ LoQ OR drift ≤ 10% for values up to 20,000 pg/mL OR drift ≤ 13% for values > 20,000 pg/mL.Stability of reagents opened onboard the instrument was 7 days per well set.
    Sample StabilityLower bound of the one-sided 95% confidence interval of the regression line must be ≤ -10% AND all individual data points must have a bias of ≤ -20% when compared to time zero.Separated samples stable for 8 hours at room temperature, 24 hours at 2-8 °C, up to 40 days at or below -20 °C (non-frost free), and up to 1 year at or below -70 °C.
    Clinical Performance (Sensitivity)Not explicitly stated as acceptance criteria, but clinical performance is assessed for diagnostic accuracy.Pooled Gender (58.9 pg/mL 99th percentile): ranged from 79.0% (0-24h).
    Male (78.5 pg/mL 99th percentile): ranged from 74.0% (0-24h) to 92.5% (0-24h) to 94.4% (0-
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1