Search Results
Found 5 results
510(k) Data Aggregation
(121 days)
Digital Video Processor
The Digital Video Processor is designed to be used with Vathin medical endoscopes, and other ancillary equipment for endoscopic diagnosis, treatment and video observation.
The Digital Video Processor is used in conjunction with Vathin endoscopes, and other accessories for endoscopic image processing during endoscopic diagnostic and therapeutic procedures. When used, it receives and processes image signals from a digital endoscope and transfers these signals to a connected monitor.
The provided text is a 510(k) Premarket Notification submission for a medical device called "Digital Video Processor" (K243117). It describes the device, its intended use, and compares it to a predicate device (Digital Video Monitor, K231135).
However, the document does not contain the specific information required to complete a table of acceptance criteria and reported device performance related to a study proving the device meets these criteria in the context of an AI/algorithm-based device as requested in the prompt.
The document primarily focuses on demonstrating substantial equivalence to a predicate device through:
- Comparison of technological characteristics: This involves comparing features like trade name, classification, product code, regulation number, models, intended use, application field, intended user, max resolution, display type, USB connection, video output, and image/video capture.
- Summary of Non-clinical tests: This section lists general categories of testing performed, such as Electrical safety and electromagnetic compatibility (EMC), Software Verification and Validation Testing, and Performance testing (e.g., Field of view, Depth of field, Geometric distortion, Image intensity uniformity, Color performance).
There is no mention of:
- Specific acceptance criteria with quantitative thresholds.
- Reported device performance data against those thresholds.
- Details of a study involving AI/algorithm performance (e.g., sensitivity, specificity, AUC).
- Sample sizes for test sets or training sets.
- Data provenance (country of origin, retrospective/prospective).
- Number of experts or their qualifications for ground truth.
- Adjudication methods.
- MRMC comparative effectiveness studies or effect sizes.
- Standalone algorithm performance.
- Type of ground truth used (expert consensus, pathology, outcomes data).
- How ground truth was established for training sets.
The "Digital Video Processor" appears to be a hardware component for processing endoscopic images, not an AI/algorithm-driven device itself that would require clinical performance studies for diagnostic accuracy. The performance tests listed are related to image quality and system functionality, not diagnostic performance.
Therefore, given the provided text, I cannot fulfill the request for a table of acceptance criteria and the study details as it pertains to an AI/algorithm-based device. The document clears a "Digital Video Processor" as an endoscope accessory, focusing on its functional equivalence, safety, and EMC, not a diagnostic AI tool.
Ask a specific question about this device
EPX-4440FN Digital Video Processor with Light Source
The EPX-4440FN Digital Video Processor with Light Source is used for endoscopic or laparoscopic observation, diagnosis, treatment, and image recording. It is intended to process electronic signals transmitted from a video endoscope or laparoscope (a video camera in an endoscope or laparoscope). This product may be used on all patients requiring endoscopic examination or laparoscopic surgical procedures and when using a FUJIFILM medical endoscope or laparoscope, light source, monitor, recorder and various peripheral devise. FICE (Flexible spectral-Imaging Color Enhancement) is an adjunctive tool for gastrointestinal endoscopic examination which can be used to supplement FUJIFILM white light endoscopy. FICE is not intended to replace histopathologic sampling as a means of diagnosis.
The EPX-4440FN Digital Video Processor with Light Source is comprised of:
- o VP-4440FN Video Processor with DK-4440E Data Keyboard
- o XL-4450FN Light Source
The EPX-4440FN Digital Video Processor with Light Source consists of the VP-4440FN Video Processor and the XL-4450FN Light Source. The VP-4440FN Video Processor relays the image from a laparoscope or endoscope to a video monitor. The processor also controls the light projected to the body cavity. The XL-4450FN Light Source employs a Xenon lamp with an emergency back-up Halogen lamp. Brightness control is performed by the user.
The provided text is a 510(k) premarket notification for a medical device called the EPX-4440FN Digital Video Processor with Light Source. It does not include acceptance criteria for device performance in the form of specific metrics (e.g., sensitivity, specificity, accuracy for a diagnostic task), nor does it describe a study proving the device meets particular clinical or diagnostic performance criteria.
Instead, this document focuses on demonstrating substantial equivalence to predicate devices based on intended use, technological characteristics, and compliance with recognized consensus standards for safety and essential performance. The "Performance Data" section describes engineering-level bench testing to ensure the device's basic functionality, safety, and electromagnetic compatibility.
Therefore, I cannot fulfill the request as the necessary information (acceptance criteria for clinical performance and a study proving those criteria are met) is not present in the provided text.
The "Performance Data" section states:
"Additionally, performance testing showed that the EPX-4440FN Digital Video Processor with Light Source met all the acceptance criteria."
However, it does not list what those acceptance criteria were in terms of specific performance metrics relevant to an AI/diagnostic device (e.g., accuracy, sensitivity, specificity, positive predictive value, negative predictive value). It also doesn't describe any studies that would typically be associated with evaluating such criteria.
The studies mentioned are related to:
- Compliance with recognized consensus standards:
- ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012 (Medical electrical equipment - General requirements for basic safety and essential performance)
- IEC 60601-1-2:2007 (Electromagnetic compatibility)
- IEC 60601-2-18 Edition 3.0 2009-08 (Particular requirements for endoscopic equipment)
- ISO 14971:2007 (Medical Device Risk Management)
- IEC 62471: 2006 (Photobiological safety of lamps and lamp systems)
- AAMI ANSI IEC 62304:2006 (Medical device software - Software lifecycle processes)
These are primarily engineering and safety standards, not studies designed to assess clinical diagnostic performance or how an AI algorithm (if present as a core diagnostic component) performs against a clinical ground truth.
Regarding your numbered points, based on the provided text:
- A table of acceptance criteria and the reported device performance: Not present. The document states "met all the acceptance criteria" but does not define these criteria or report specific performance values.
- Sample size used for the test set and the data provenance: Not applicable, as no clinical performance study involving a test set is described. The "performance testing" referenced is likely bench testing for engineering compliance.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no clinical performance study involving ground truth establishment is described.
- Adjudication method for the test set: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done: No. The device is a "Digital Video Processor with Light Source" and its FICE function is described as an "adjunctive tool... not intended to replace histopathologic sampling as a means of diagnosis." This indicates it's not a primary AI-based diagnostic tool requiring MRMC studies for comparative effectiveness of human readers with/without AI assistance.
- If a standalone (i.e. algorithm only without human-in-the loop performance) was done: No, this device is not presented as a standalone diagnostic algorithm.
- The type of ground truth used: Not applicable.
- The sample size for the training set: Not applicable, as this device's description does not suggest a machine learning model that would require a "training set" in the sense of AI/diagnostic algorithm development.
- How the ground truth for the training set was established: Not applicable.
Ask a specific question about this device
(212 days)
EPX-4440HD DIGITAL VIDEO PROCESSOR
The VP-4440HD unit is used for endoscopic observation, diagnosis, treatment, and image recording. It is intended to process electronic signals transmitted from a video endoscope). This product may be used on all patients requiring endoscopic examination and when using a Fujinon/Fujifilm medical endoscope, light source, monitor, recorder and various peripheral devices. FICE is an adjunctive tool for gastromtestination which can be used to supplement FUJIFILM white light endoscopy. FICE is not intended to replace histopathological sampling as a means of diagnosis.
This system is intended to be used in coniunction with Fuiifilm endoscopes to provide illumination, visual display and data storage during endoscopic procedures.
The EPX-4440HD Digital Video Processor with FICE and light source consists of three components used in conjunction with one another:
The VP-4440HD Video Processor: The Processor relays the image from the endoscope to a video monitor. Projection can be either analog or digital at the user's preference. The Processor incorporates internal or external digital storage capacity. The Processor also controls the light projected to the body cavity. The Processor provides for optional structural enhancement at the user's option. Spectral and structural enhancements are achieved through proprietary software. The device is AC operated at a power setting of 120V/60MzJ0.8A. The Processor is housed in a steel-polycarbonate case measuring 390x105x460mm.
The XL-4450 Light Source: The Fujinon/Fujifilm endoscope employs fiber bundles to transmit light from the light source and subsequently to the body cavity. The Light Source employs a 300W Xenon lamp with a 75W emergency back-up Halogen lamp. Brightness control is performed by the user. The device is AC operated at a power setting of 120V/60MzJ3.3A. The Light Source is housed in a steel-polycarbonate case measuring 390x155x450mm.
The DK-4440E Keyboard: The Keyboard is used to enter pertinent procedural information (patient, physician, date, etc.) for display on the video monitor and digital/analog storage systems. The Keyboard is also used to control operational features of the VP-4440HD Processor. The Keyboard resembles a standard computer keyboard in size and shape.
The VP-4440HD Digital Video Processor with FICE and light source is the same as the cleared EPX-4440HD Digital Video Processor and Light source (K102466) device, with the exception of the addition of the imaging algorithm termed "FICE" (Flexible spectral-Imaging Color Enhancement). As with the predicate device, the subject processor receives an image from the endoscope via a connector cable. The subject processor is capable of bi-directional communication that allows importation and exportation of various patient and procedural data between EMR (Electronic Medical Record) systems.
FICE is an image processing feature that a licensed medical practitioner (user) can utilize, as an adiunctive tool, to supplement the white light endoscopic examination. FICE extracts spectral images of specific wavelength components from the original full spectral (white light) image through an image processing algorithm that, assigns a spectral wavelength component in the Red (R), Green (G) and Blue (B) spectrum and displays the enhanced color image. FICE may enhance color contrast to improve visibility of the selected structures, borders of areas of interests.
1. Table of Acceptance Criteria and Reported Device Performance
Criterion | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Non-clinical Performance | ||
Software Validation | In accordance with IEC 62304 | "All testing criteria were met, and in all instances the device functioned as intended." |
Electrical Safety | In accordance with IEC 60601 | "All testing criteria were met, and in all instances the device functioned as intended." |
Functional Testing (FICE) | Contrast enhancement, dynamic range, resolution, noise, artifact creation, and color reproduction should meet specified levels/standards. | "All testing criteria were met, and in all instances the device functioned as intended." |
Clinical Performance (Image Quality of FICE presets) | Image quality of each FICE preset should be at least comparable to optical filtering of the white light source (Olympus EVIS EXERA II 180 System's Narrow Band Imaging - NBI). | "Overall means between all readers demonstrated at least comparable image quality of each FICE setting in one or more locations within the gastrointestinal tract." |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set (Clinical Study): Each subject underwent two (2) gastrointestinal endoscopic examinations during a single clinical procedure. The document doesn't explicitly state the total number of subjects or individual images in the test set.
- Data Provenance: Prospective clinical trial. The country of origin of the data is not specified.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Number of Experts: Multiple blinded investigators. The exact number is not specified.
- Qualifications of Experts: Not specified.
4. Adjudication Method for the Test Set
- Adjudication Method: Images were analyzed independently by "multiple blinded investigators using a 4-point Likert scale." The document does not specify a formal adjudication method (e.g., 2+1, 3+1 consensus). It appears to rely on the "overall means between all readers."
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
- MRMC Study: Yes, a reader concurrence study was performed, which is a type of MRMC study comparing FICE (an AI/image processing feature) to optical filtering (NBI).
- Effect Size: The study aimed to establish "substantial equivalence of diagnostic visualization (e.g. image quality) to the predicate optical filtering method." The reported outcome is that "Overall means between all readers demonstrated at least comparable image quality of each FICE setting in one or more locations within the gastrointestinal tract." This indicates FICE performed similarly to or as well as NBI for image quality; a specific quantitative effect size of human reader improvement with AI assistance vs. without is not provided, as the study focused on comparability rather than improvement. The FICE feature itself is described as an "adjunctive tool," not necessarily a comprehensive AI diagnosis system.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Based on the provided text, the FICE feature is described as an "adjunctive tool for gastrointestinal endoscopic examination which can be used to supplement FUJIFILM white light endoscopy." The clinical study focused on readers evaluating image quality generated by FICE. There is no indication of a standalone performance study of the FICE algorithm without human interpretation.
7. The Type of Ground Truth Used
- Type of Ground Truth: The clinical study compared the "overall image quality" observed by multiple blinded investigators using a Likert scale. This suggests a form of expert consensus/reader perception for image quality as the ground truth. It's important to note that the FICE feature is "not intended to replace histopathological sampling as a means of diagnosis," indicating that the "ground truth" for diagnosis in a clinical sense would still be pathology. However, for the purpose of this study measuring the effectiveness of the FICE image processing, reader evaluation of image quality served as the primary measure.
8. The Sample Size for the Training Set
- The document describes performance testing for an image processing feature (FICE) which is integrated into a video processor. It does not explicitly mention a "training set" in the context of machine learning model development. This suggests the FICE algorithm might be based on pre-defined spectral filters or rule-based image processing rather than a deep learning model that requires a dedicated training set. If there was a training phase for the algorithm, its sample size is not disclosed.
9. How the Ground Truth for the Training Set Was Established
- As a dedicated "training set" is not mentioned and the FICE algorithm's nature isn't fully detailed as a deep learning model, the method for establishing ground truth for a training set is not applicable or not provided in this document.
Ask a specific question about this device
(268 days)
FUJINON EPX-4440HD DIGITAL VIDEO PROCESSOR
The VP-4440HD unit is used for endoscopic observation, diagnosis, treatment, and image recording. It is intended to process electronic signals transmitted from a video endoscope (a video camera in an endoscope). This product may be used on all patients requiring endoscopic examination and when using Fujinon medical Endoscope, light source, monitor, recorder and various peripheral devices.
The XL-4450 light source is used for endoscopic observation, diagnosis, treatment, and image recording. It is intended to provide illumination to an endoscope. The light source also functions as a pump to supply air through the endoscope while inside the body to assist in obtaining clear visualization to facilitate diagnostic examination. This product may be used on all patients requiring endoscopic examination and when using Fujinon medical Endoscope, video processor, monitor, recorder and various peripheral devices.
This system is intended to be used in conjunction with Fujinon endoscopes to provide illumination, visual display and data storage during endoscopic procedures.
The EPX-4440HD Digital Video Processor consists of three components used in conjunction with one another:
The VP-440HD Video Processor: The Processor relays the image from the endoscope to a video monitor. Projection can be either analog or digital at the user's preference. The Processor also incorporates internal or external digital storage capacity. The Processor also controls the light projected to the body cavity. The Processor also provides for optional structural enhancement at the user's option. Spectral and structural enhancements are achieved through proprietary software. The device is AC operated at a power setting of 120V/60MzJO.8A. The Processor is housed in a steel-polycarbonate case measuring 390X105X460mm.
The XL-4450 Light Source: The Fujinon endoscope employs fiber bundles to transmit light from the light source and subsequently to the body cavity. The Light Source employs a 300W Xenon lamp with a 75W emergency back-up Halogen lamp. Brightness control is performed by the user. The device is AC operated at a power setting of 120V/60MzJ3.3A. The Processor is housed in a steel-polycarbonate case measuring 390X155X485mm.
The DK-440E Keyboard: The Keyboard is used to enter pertinent procedural information (patient, physician, date, etc) for display on the video monitor and digital/analog storage systems. The Keyboard is also used to control operational features of the VP-4440HD Processor. The Keyboard resembles a standard computer keyboard in size and shape.
Here's an analysis of the provided text regarding the acceptance criteria and study for the Fujinon EPX-4440HD Digital Video Processor:
NOTE: The provided 510(k) summary focuses primarily on the safety and performance aspects of a modified device compared to a predicate device, rather than defining and proving specific acceptance criteria in the way one might for a novel AI/software-as-a-medical-device (SaMD) product. The "acceptance criteria" discussed are more about meeting established regulatory and engineering standards, and the "study" is functional testing with physician evaluators for general performance and safety.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Reported Device Performance (as per document) |
---|---|
Software Validation | All testing criteria were met. |
Electrical Safety | All testing criteria were met (in accordance with IEC 60601 requirements). |
Performance Testing (Functional) | "Functional testing consisting of visual comparison tests were conducted using physician evaluators. In all cases they passed the required testing regimens." |
Predicate Equivalence | The sponsor believes the modified device maintains the same safety and performance levels as the previous generation processor. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not explicitly stated. The document mentions "physician evaluators" without specifying how many cases, images, or evaluation instances were part of their assessment.
- Data Provenance: Not explicitly stated. The context suggests general functional testing, likely performed in a controlled environment by the manufacturer. It does not mention country of origin or whether clinical patient data was used. It's likely simulated or controlled test conditions given the nature of the evaluation.
- Retrospective/Prospective: Not explicitly stated, but given it's functional testing for a device modification, it would typically be a prospective evaluation of the new device's capabilities in a controlled setting.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Number of Experts: Not explicitly stated. The document refers to "physician evaluators" (plural), but the exact number is not provided.
- Qualifications of Experts: Not explicitly stated beyond "physician evaluators." No specialty (e.g., gastroenterologist who performs endoscopy) or years of experience are listed.
4. Adjudication Method for the Test Set
- Adjudication Method: Not explicitly stated. The statement "In all cases they passed the required testing regimens" suggests a unanimous agreement or a pass/fail criterion that was met by all evaluators, rather than a formal adjudication process between conflicting interpretations.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Was an MRMC study done? No, a traditional MRMC comparative effectiveness study as typically understood for AI algorithms (i.e., comparing human readers with and without AI assistance on diagnostic accuracy) was not performed. The study mentioned was a functional performance test for a hardware/software update, not an AI-assisted diagnostic tool.
- Effect size of human readers with vs. without AI assistance: Not applicable, as no such study was conducted.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- Was a standalone study done? Yes, in a sense. The "software validation" and "electrical safety" testing would represent a standalone evaluation of the device's technical functionality. The "performance testing using physician evaluators" also tested the device itself, although with human interaction. However, this is not an AI algorithm performing a diagnostic task. The spectral and structural enhancements are "achieved through proprietary software," implying an algorithm, but its performance was evaluated by physicians using the system, not as a standalone diagnosis.
7. Type of Ground Truth Used
- Type of Ground Truth: For the "visual comparison tests conducted using physician evaluators," the ground truth was essentially the expert opinion/consensus of the physician evaluators regarding the successful functionality and acceptable image quality produced by the system. This is not pathology, outcomes data, or a pre-established clinical diagnosis.
8. Sample Size for the Training Set
- Sample Size for Training Set: Not applicable. This device is a video processor and light source, not a machine learning model that requires a "training set" in the conventional AI sense. While there is proprietary software for "spectral and structural enhancements," the document does not describe the development of this software in terms of machine learning training, but rather as a feature set of the device.
9. How the Ground Truth for the Training Set Was Established
- How Ground Truth for Training Set Was Established: Not applicable, as there is no mention of a training set for a machine learning model. If the "proprietary software" involved some form of learned enhancement, the document does not provide details on how such a model would have been trained or its ground truth established.
Ask a specific question about this device
(30 days)
FLEXIBLE TRANS-NASAL VIDEO ESOPHAGOSCOPE WITH DIGITAL VIDEO PROCESSOR AND DISPOSABLE ENDOSHEATH SYSTEMS
The flexible trans-nasal video esophagoscope with digital video processor and disposable sheath system is intended for use in endoscopic access and examination of the larynx, esophagus and gastro-esophageal junction. The system can also be used to aid in intubation.
The digital video processor is intended for use with the VSI flexible video scope.
The VSI flexible endoscope is a flexible endoscope with connections to a video processor and display monitor. The EndoSheath® Systems are sterile, single-use protective sheath systems, with or without a working channel, that are intended to cover the entire insertion tube of the videoscope. The digital video processors are used with the flexible videoscope for image visualization and capture.
The provided text describes a Special 510(k) Premarket Notification for a modified medical device. This type of submission focuses on demonstrating substantial equivalence to a predicate device through certification of compliance to design control requirements and a description of risk analysis procedures, rather than extensive new clinical or performance studies.
Therefore, the supplied text does not contain detailed information regarding:
- A table of specific acceptance criteria and reported device performance metrics.
- Sample sizes for test sets, data provenance, number of experts, adjudication methods, MRMC studies, or standalone performance data.
- Specific types of ground truth used or sample sizes for training sets (as this is a modification of an existing device, not a de novo AI device).
Summary of available information:
1. Acceptance Criteria and Reported Device Performance:
The submission states that "V & V activities, including scope/sheath/processor system functional and performance testing, and software validation was addressed through comprehensive Design Validation and Verification planning." It concludes that "Based on the indications for use, technological characteristics, performance testing and comparison to predicate devices, the VSI flexible video TNE endoscope with digital video processors and disposable EndoSheath® Systems have been shown to be safe and effective for their intended use."
However, specific quantitative acceptance criteria (e.g., in terms of sensitivity, specificity, accuracy, or other performance metrics) and the numerical results against these criteria are not provided in the document. The general acceptance criteria would be that the modified device performs as intended and is as safe and effective as the predicate device.
2. Sample size used for the test set and the data provenance:
Not explicitly stated. The submission focuses on design control compliance and functional/performance testing of the device system, rather than a clinical study evaluating diagnostic accuracy on a specific patient test set.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable or not provided. This type of information is typically relevant for studies validating diagnostic algorithms, which is not the primary focus of this Special 510(k) for a hardware modification.
4. Adjudication method for the test set:
Not applicable or not provided.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This device is a modified endoscope system, not an AI-powered diagnostic algorithm designed to assist human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This is a hardware device (endoscope, video processor, and sheaths), not a standalone algorithm.
7. The type of ground truth used:
Not explicitly stated, but for a physical device, "ground truth" would likely relate to objective measurements of device function (e.g., image quality, illumination, navigational capabilities, integrity of sheaths, software functionality) rather than clinical diagnostic outcomes requiring expert consensus or pathology.
8. The sample size for the training set:
Not applicable or not provided, as this is a device modification submission for hardware, not an AI algorithm.
9. How the ground truth for the training set was established:
Not applicable or not provided.
Ask a specific question about this device
Page 1 of 1