Search Results
Found 3 results
510(k) Data Aggregation
(199 days)
The Diagnostic X-ray System is intended to be used and operated by: adequately trained, qualified and authorized health care professionals who have full understanding of the safety information and emergency procedures as well as the capabilities and functions of the device. The device is used for radiological guidance and visualization during diagnostic, interventional and surgical procedures on all patients, except neonates (birth to one month), within the limits of the device. The device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures.
The Diagnostic X-ray System is a mobile (within an imaging facility) general-purpose diagnostic fluoroscopic X-ray system that uses a C-arm and digital techniques for image capture, display and manipulation and is designed to be used in a variety of general-purpose applications requiring real-time fluoroscopic imaging capabilities.
The Diagnostic X-ray System is consists of X-ray source assembly (combined type), collimator, flat-panel detector, image processing workstation, C-arm and mobile rack, Medical Image Workstation Software.
It appears that the provided FDA 510(k) clearance letter and associated summary pertain to a Diagnostic X-ray System (Trade Name: Diagnostic X-ray System, Model: PLX119C) manufactured by Nanjing Perlove Medical Equipment Co., Ltd.
Crucially, the document explicitly states: "No clinical study is included in this submission."
Therefore, I cannot provide details about acceptance criteria derived from a clinical study, as no such study was presented for this device's 510(k) clearance.
However, I can extract the information provided regarding the non-clinical performance and testing. It's important to understand that for a general-purpose diagnostic X-ray system, substantial equivalence is often demonstrated through comparisons of technical specifications and non-clinical performance to a legally marketed predicate device, rather than a full-scale clinical trial proving "improved" diagnostic accuracy in a specific clinical context.
Here's a breakdown of the available information based on your request, with a clear note about the absence of a clinical study for demonstrating performance against acceptance criteria in a clinical setting:
1. A table of acceptance criteria and the reported device performance
Since no clinical study was performed to establish clinical performance acceptance criteria and then demonstrate the device meets them, I can only present the reported non-clinical performance and compliance with relevant standards. The "acceptance criteria" here are implied by meeting recognized standards and demonstrating comparable technical characteristics to the predicate device.
| Category | Acceptance Criteria / Standard Compliance | Reported Device Performance |
|---|---|---|
| Electrical Safety | Compliance with IEC 60601-1-2: 2014+AMD1:2020 / EN 60601-1-2: 2015+A1: 2021 | The system complies with IEC 60601-1-2: 2014+AMD1:2020 / EN 60601-1-2: 2015+A1: 2021. |
| EMC Testing | Compliance with IEC 60601-1-2: 2014+AMD1:2020 / EN 60601-1-2: 2015+A1: 2021 | The system complies with IEC 60601-1-2: 2014+AMD1:2020 / EN 60601-1-2: 2015+A1: 2021. |
| X-ray Equipment | Compliance with IEC 60601-2-54: 2009+AMD2:2018 / EN 60601-2-54: 2009+A2:2019 (For X-ray equipment for radiography and radioscopy) | The system complies with IEC 60601-2-54: 2009+AMD2:2018 / EN 60601-2-54: 2009+A2:2019. |
| Radiation Protection | Compliance with IEC 60601-1-3:2008+A1:2013+A2:2021 / EN 60601-1-3:2008+A1:2013+A2:2021 (For radiation protection in diagnostic X-ray equipment) | The system complies with IEC 60601-1-3:2008+A1:2013+A2:2021 / EN 60601-1-3:2008+A1:2013+A2:2021. |
| Interventional X-ray | Compliance with IEC 60601-2-43:2010+A1:2017+A2:2019 / EN 60601-2-43:2010+A1:2018+A2:2020 (For X-ray equipment for interventional procedures) | The system complies with IEC 60601-2-43:2010+A1:2017+A2:2019 / EN 60601-2-43:2010+A1:2018+A2:2020. |
| Detector Performance | Bench testing of imaging metrics (MTF, DQE comparable to predicate). Expected to meet industry standards for diagnostic image quality. | MTF @ 1.0 LP/mm: Typical 59% (Predicate: Typical 64%)DQE @ 0 LP/min: Typical 77% (Predicate: Typical 77%)Note: While MTF is slightly lower than predicate, it's implied to be within acceptable diagnostic limits for the intended use and often compensated by other factors in real-world imaging. |
| Software V&V | Compliance with FDA's "Content of Premarket Submissions for Software Contained in Medical Devices" guidance; basic level documentation sufficient as failure not expected to lead to death or serious injury. | Software verification and validation testing were conducted and documentation was provided as recommended. The embedded and workstation software for this device required basic level documentation, as a failure of software function(s) would not present a hazardous situation with a probable risk of death or serious injury. |
| Cybersecurity | Compliance with FDA guidance "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" | Cybersecurity documentation was provided according to FDA guidance. |
| Usability | Compliance with IEC 60601-1-6 and IEC 62366-1 | Usability validation was performed and documentation was provided, complying with IEC 60601-1-6 and IEC 62366-1. |
| Biocompatibility | Compliance with ISO 10993-1:2018 "Biological Evaluation of Medical Devices − Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. | Biocompatibility evaluation for the proposed device was conducted in accordance with ISO 10993-1:2018. |
| Clinical Image Quality | Implied acceptance criteria: acceptable imaging performance for diagnostic/interventional guidance, comparability to predicate. | Clinical images of multiple body parts were taken using radiography and fluoroscopy, with motion, to demonstrate acceptable imaging performance of the Diagnostic X-ray System. These images were reviewed and assessed by a qualified radiologist. (Specificity of "acceptable" is not defined, but the assessment implies it met expectations for its intended use.) |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
-
Clinical Images: "Clinical images of multiple body parts were taken using radiography and fluoroscopy, with motion".
- Sample Size: Not specified (described as "multiple body parts").
- Data Provenance: Not specified, but generally, for 510(k) submissions from foreign manufacturers, testing would occur at a site capable of generating such data, potentially in the country of origin (China, in this case, for the manufacturer). The images would be prospective as they were "taken" for the purpose of demonstrating performance.
-
Bench Testing (Detector Imaging Metrics):
- Sample Size: Not specified.
- Data Provenance: Not specified, but typically conducted in a controlled laboratory environment by the manufacturer or a contracted testing facility.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Clinical Images: "reviewed and assessed by a qualified radiologist."
- Number of Experts: One ("a qualified radiologist").
- Qualifications: "qualified radiologist" – further specific experience or board certification is not detailed in the provided text.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Clinical Images: The document states the images were "reviewed and assessed by a qualified radiologist." This implies a single-reader assessment, with no mention of an adjudication process (e.g., if multiple readers disagreed). Therefore, the adjudication method was none.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No MRMC study was done. The device is a diagnostic X-ray system, not an AI-powered diagnostic aide. The submission explicitly states: "No clinical study is included in this submission." Therefore, there is no information about human reader improvement with or without AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- Not applicable. This device is a diagnostic X-ray system and does not appear to contain an AI algorithm for standalone diagnostic performance. The primary focus is on the image acquisition, processing, and display capabilities for human interpretation.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- For the "clinical images," the "ground truth" or standard of reference for "acceptable imaging performance" was the assessment by a qualified radiologist. This is a form of expert assessment/consensus (though by a single expert in this case for the review of performance). There is no mention of pathology or outcomes data being used as ground truth for performance evaluation of the imager itself.
8. The sample size for the training set
- Not applicable / Not specified. This device is hardware with associated software for image acquisition and processing. There is no mention of machine learning or deep learning algorithms requiring a distinct "training set" in the context of AI. The software verification and validation are standard procedures for medical device software, not AI training.
9. How the ground truth for the training set was established
- Not applicable. As there is no mention of a traditional "training set" for an AI algorithm, the concept of establishing ground truth for it does not apply here. The software development follows general software engineering principles and V&V, not AI model training.
Ask a specific question about this device
(48 days)
The PLX5200A is intended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, extremities, and other body parts. Applications can be performed with patient sitting, standing or lying in the prone or supine positions. It is not intended for mammography.
The proposed device is a mobile x-ray system, it uses digital techniques for image capture, display and manipulation and the mobile design allows it to operate on mains or battery power and to be driven or pushed by an operator to various locations within a building or facility. It is commonly used for bedside imaging. This system consists of a mobile base unit, combined X-ray tube assembly, collimator, frame, X-ray flat panel detector, and image processing system.
This device is a Diagnostic X-ray System, which is a hardware device. The provided text indicates that no clinical study was included in this submission. Therefore, there is no acceptance criteria or study data for device performance as requested for AI/software devices.
However, the submission does mention various performance data related to the hardware itself, which are not based on clinical efficacy or diagnostic accuracy. These include:
- Electrical safety and electromagnetic compatibility (EMC): The system complies with IEC 60601-1:2005+A1:2012 / ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, IEC 60601-1-3:2008+A1:2013, IEC 60601-2-54:2009, IEC 62471 standard for safety, and IEC 60601-1-2: 2014 standard for EMC. Radio frequency (RF) wireless coexistence of equipment testing was performed according to IEEE ANSI C63.27-2017.
- Software Verification and Validation: The Embedded software and the workstation software were considered a "Moderate" level of concern and were verified and validated according to FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." They comply with FDA recognized standard IEC 62304.
- Cybersecurity: Documentation was provided according to "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices."
- DICOM: DICOM declaration was provided according to FDA recognized standard NEMA PS 3.1 - 3.20 (2016).
- Usability Testing: Usability validation was performed and complies with FDA recognized standard IEC 60601-1-6 Edition 3.1 and EC 62366-1 Edition 1.0.
- Risk Management: Activities were performed and documented according to FDA recognized standard ISO 14971 Second edition 2007-03-01.
- Biocompatibility Testing: Conducted in accordance with ISO 10993-1, including Cytotoxicity, Skin Sensitization, and Skin Irritation for the x-ray flat panel detector.
Since this is a hardware device clearance, and not an AI/software as a medical device (SaMD) or an AI-powered diagnostic device, the typical acceptance criteria and study design elements requested (like sample size for test/training sets, expert ground truth, MRMC studies, standalone performance) are not applicable or provided in this 510(k) summary.
Ask a specific question about this device
(52 days)
The XR5 diagnostic X-ray system is intended for use on adult and pediatric patients for taking diagnostics radiographic exposure of all body parts and operated by a qualified/trained doctor or technician. The XR5 diagnostic X-ray system is designed to be used with conventional film/screen, CR cassettes or digital detectors. NOT intended for Mammography use.
The XR5 diagnostic X-ray system is used for image capture by using X-rays on a patient's body. The XR5 is a conventional X-ray machine, with an intuitive operation console to provide the user an easy way to manage optimal conditions for quality images. The high frequency generator is capable of delivering the exposition dose appropriate for general X-ray and diagnostic images of a patient. This system is made to use other detector options including: conventional films, CR, or Digital Flat Panel Detectors. Please note that the quality of the image in any detector depends on the manufacture of the receptor device. The XR5 diagnostic X-ray system doesn't provide an AEC feature.
The provided text describes the OSKO, INC. XR5 Diagnostic X-ray System, which is a conventional X-ray machine. However, the document (a 510(k) summary) focuses on demonstrating substantial equivalence to a predicate device (Multix Fusion by Siemens Medical Solutions USA, Inc.) rather than outlining specific acceptance criteria or a dedicated study proving the device meets those specific acceptance criteria in terms of clinical performance metrics.
Instead, the document emphasizes:
- Technical equivalence: Comparing specifications like generator type, image acquisition methods, and the use of FDA-cleared digital detectors.
- Compliance with safety and electrical standards: Referencing IEC 60601-1, IEC 60601-2-54, and EN60601-1-2.
- Performance of individual components: The digital detectors are evaluated based on DQE, MTF, and line resolution, with these parameters serving as indicators of their performance.
Therefore, many of the requested items (e.g., specific acceptance criteria for diagnostic accuracy, sample sizes for test/training sets, expert qualifications, MRMC studies, standalone performance) are not detailed in this type of regulatory submission, as it's not a clinical performance study report for diagnostic accuracy of the overall system.
Here's an attempt to answer the questions based only on the provided text, with many fields explicitly stated as "Not Applicable" or "Not Provided" due to the nature of the document:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly define specific "acceptance criteria" for diagnostic performance of the overall XR5 system as a diagnostic device (e.g., sensitivity, specificity, accuracy for a particular condition). Instead, it relies on demonstrating substantial equivalence to a predicate device by comparing technical specifications and performance of key components (digital detectors). The "reported device performance" primarily refers to the technical specifications and the performance metrics of the digital detectors used with the system.
| Acceptance Criteria Category (Implied by document) | Specific Criteria (from document) | Reported Device Performance (XR5 / Digital Detectors) |
|---|---|---|
| System Functionality | General X-ray and diagnostic imaging of all body parts for adult and pediatric patients. | The XR5 diagnostic X-ray system is intended for use on adult and pediatric patients for taking diagnostics radiographic exposure of all body parts. Operated by a qualified/trained doctor or technician. Designed to be used with conventional film/screen, CR cassettes, or digital detectors. |
| Safety Standards Compliance | Compliance with IEC 60601-1; 2005 + A1 (2012) and IEC 60601-2-54; Edition 1.0.2009. Applicable EMC requirements (EN60601-1-2:2007 / AC : 2010). | "The XR5 diagnostic X-ray system has been tested in accordance with Safety standard of IEC 60601-1; 2005 + A1 (2012) and IEC 60601-2-54; Edition 1.0.2009. The test reports have shown good performance, substantially equivalent to the predicate device. The XR5 has also met applicable Electro Magnetic Compatibility (EMC) requirements; EN60601-1-2:2007 / AC : 2010 (IEC60601-1-2:2007)." |
| Digital Detector Performance (Component Level) | Evaluation according to IEC 62220-1:2003 Standard (for DQE, MTF, line resolution). | MTF @ 3.5 lp/mm: 1717SCC (K122173): 13%, 1417WCC (K141566): 12.1%, FLAATZ600 (K132842): 50.02%. DQE @ 0 spatial frequency: 1717SCC: 54%, 1417WCC: 76.7%, FLAATZ600: 50%. Line resolution: Similar between 3.0~3.5 lp/mm for all listed detectors. |
| Software Validation | Firmware validation testing to verify risk analysis and individual performance results within predetermined acceptance criteria. | "The complete system configuration including the firmware function has been assessed and tested by the manufacturer and passed all in-house testing criteria. The firmware validation test was designed to evaluate for the X-ray system including X-ray exposure control and exposure switch performed by XR5. ... individual performance results were within the predetermined acceptance criteria." |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: Not provided. The document states that "Bench performance parameters for the digital detectors are sufficient to demonstrate substantial equivalence to a predicate device" and that "Clinical images are not necessary to establish substantial equivalence based on the previously FDA cleared detectors." This implies that a formal clinical test set with a specific sample size for diagnostic performance was either not conducted or not deemed necessary for this 510(k) submission.
- Data Provenance: Not provided for any clinical data. Non-clinical performance data for the detectors references the IEC 62220-1:2003 Standard.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable/Not provided. No clinical test set to establish ground truth for diagnostic accuracy is described.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable/Not provided. No clinical test set for diagnostic accuracy is described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable/Not provided. This device is a conventional X-ray system, not an AI-powered diagnostic tool, and no MRMC study is mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable/Not provided. This device is a conventional X-ray system, not an algorithm, and no standalone performance study in this context is mentioned.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
Not applicable/Not provided for diagnostic accuracy of the overall system. The document does not describe the establishment of clinical ground truth for diagnostic accuracy for the XR5 system itself. For the digital detectors, their performance metrics (DQE, MTF, line resolution) are physics-based measurements rather than clinical ground truth on patient outcomes.
8. The sample size for the training set
Not applicable/Not provided. This device is a hardware X-ray system, not an AI algorithm requiring a training set. The firmware is "programmed with ANSI C language," suggesting traditional software development and testing rather than machine learning training.
9. How the ground truth for the training set was established
Not applicable/Not provided. As above, this device is a hardware X-ray system; no training set for an AI algorithm is mentioned.
Ask a specific question about this device
Page 1 of 1