510(k) Data Aggregation
(25 days)
The Dental Pain Eraser DPE-C300 is intended to provide temporary dental anesthesia (pain relief) and is indicated for use before, during, or after dental and orthodontic procedures, and other sources of oral pain. It is intended for use by dental/orthodontic professionals for the alleviation of oral pain.
The Dental Pain Eraser DPE-C300 is a non-invasive electrical stimulation device that is used to numb the nerve pain associated with orthodontic procedures and to dull the pain of canker sores often developed inside the mouths of patients wearing braces or retainers. This method of treatment is referred to as Transcutaneous Electrical Nerve Stimulation (TENS). The Dental Pain Eraser DPE-C300 has a shape and size similar to a common toothbrush, except that in place of the bristles are two metal electrodes. In operation, the device is turned on for 30 seconds to 10 minutes, the two electrodes are rubbed gently up and down the gums above the patient's aching tooth or on the intact skin around the canker sore. Current flowing between the electrodes through the gum anesthetizes the nearby nerves thus alleviating pain. The Dental Pain Eraser is intended for the alleviation of oral pain for pediatric (10 years through 22 years of age) and adult patients. The modified Dental Pain Eraser DPE-C300 is a multi-use device to be used in a dental/orthodontic facility.
This document describes a 510(k) premarket notification for the "Dental Pain Eraser DPE-C300," which is a non-invasive electrical stimulation device for temporary dental anesthesia. The primary purpose of this submission is to demonstrate that a modified version of the device, intended for multi-use, is substantially equivalent to a predicate single-use device (K182947). Therefore, the study focuses on validating the reprocessing (cleaning and disinfection) and durability for multi-use, rather than clinical efficacy studies of the pain relief itself.
Here's an analysis based on your request:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state acceptance criteria in a quantifiable manner for the overall "performance" of the device in terms of pain relief. Instead, it focuses on demonstrating substantial equivalence to a predicate device, particularly regarding its new multi-use characteristic. The "reported device performance" in this context refers to the successful outcomes of the nonclinical tests designed to support the multi-use claim.
Acceptance Criteria Category | Specific Criteria/Focus | Reported Device Performance |
---|---|---|
Cleaning and Disinfection Validation | Adherence to standards for reprocessing medical devices. | "The results of the collective testing demonstrate that differences between the modified Dental Pain Eraser DPE-C300 as a multi-use device and the predicate as a single-use device do not raise different questions of safety or effectiveness." |
Residual Cytotoxicity Testing | Ensuring no harmful residues remain after disinfection. | Implied to be met as part of "Disinfecting Validation/Residual Cytotoxicity Testing" and the overall conclusion of no new safety/effectiveness questions. |
Cycle Testing | Durability and consistent performance over multiple uses. | "meet the established specifications necessary for consistent performance during its intended use." (Specific quantitative cycles not provided in summary). |
Tip Assembly Pull Testing | Mechanical integrity of the tip assembly. | "meet the established specifications necessary for consistent performance during its intended use." (Specific quantitative force not provided in summary). |
Biocompatibility | Compliance with relevant ISO standards and FDA guidance. | "Biocompatibility testing was performed per ISO 10993-1 and FDA Guidance Document... issued June 16, 2016." (No specific results provided in summary, but implied compliance). |
Electrical Safety & EMC | Compliance with IEC 60601 series and ANSI/AAMI NS4:2013 | "Compliant with applicable requirements of: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 62304, ANSI/AAMI NS4:2013." (General statement of compliance without specific test results in summary). |
Risk Management & Usability | Compliance with ISO 14971 and IEC 62366-1 | "Compliant with applicable requirements of: IEC 62366-1:2015, Medical devices. Part 1: Application of usability engineering to medical devices ISO 14971:2012, Medical Devices. Application of Risk Management to Medical Devices." (General statement of compliance without specific details). |
Output Specifications (Voltage, Current, etc.) | Identical to predicate device. | All output specifications (patient leakage current, output modes, waveform, shape, voltage, current, pulse duration, frequency, charge, power density, burst mode, on/off time) for the proposed device are listed as identical to the predicate device. |
General Physical Characteristics | Identical to predicate device (except multi-use capability). | Indications for Use, Intended User/Patient Population, Use Environment, Product Code, Classification, Manufacturer, Power Source, Method of Line Current Isolation, Indicator Display, Therapy Range, Weight, Unit Body Dimensions, Materials and Construction, Ingress Protection Rating are all stated as "N/A (same)" as the predicate device. |
2. Sample Size Used for the Test Set and Data Provenance
The document explicitly states: "Clinical testing was not required to establish substantial equivalence."
Therefore, there is no "test set" in the context of clinical performance data. The "tests" performed were nonclinical (bench testing) for cleaning, disinfection, cycle testing, and pull testing. The sample sizes for these specific bench tests are not provided in the summary. Data provenance is implied to be from Synapse Dental's internal testing as part of their 510(k) submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable, as no clinical test set was used to establish ground truth for clinical efficacy. The evaluation was based on nonclinical bench testing and comparison to a predicate device.
4. Adjudication Method for the Test Set
Not applicable, as no clinical test set requiring expert adjudication was used.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
Not applicable. This device is a direct-use electrical stimulation device, not an AI-powered diagnostic tool requiring human reader interpretation. No MRMC study was conducted.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This question is not applicable, as the device is not an algorithm or AI system. It is a physical medical device.
7. The Type of Ground Truth Used
For the nonclinical tests (cleaning, disinfection, cycle, pull), the "ground truth" was established by compliance with regulatory standards and pre-established specifications for those specific engineering and microbiological tests. For example, for cleaning and disinfection, it would be the absence of detectable bioburden or remaining residues, meeting the criteria outlined in AAMI TIR standards. For cycle and pull testing, it would be the structural integrity and continued function after a specified number of cycles or under a specified load.
For the substantial equivalence claim, the overarching "ground truth" is that the modified multi-use device performs equivalently to the predicate single-use device without raising new safety or effectiveness concerns, justified by the nonclinical testing results.
8. The Sample Size for the Training Set
Not applicable, as this is not an AI/machine learning device that requires a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as no training set was used.
Ask a specific question about this device
(131 days)
The Dental Pain Eraser DPE-T300 is intended to provide temporary dental anesthesia (pain relief). It is intended for prescriptive use only for the alleviation of oral pain for use by pediatric (10 years of age) and adult patients for treatment at home. Pediatric patients 10 through 12 years of age must be supervised by an adult.
The Dental Pain Eraser DPE-T300 is a non-invasive electrical stimulation device that is used to numb the nerve pain associated with orthodontic procedures and to dull the pain of canker sores often developed inside the mouths of patients wearing braces or retainers. This method of treatment is referred to as Transcutaneous Electrical Nerve Stimulation (TENS). The Dental Pain Eraser DPE-T300 has a shape and size similar to a common toothbrush, except that in place of the bristles are two metal electrodes. In operation, the device is turned on for 30 seconds to 10 minutes, the two electrodes are rubbed gently up and down the gums above the patient's aching tooth or on the intact skin around the canker sore. Current flowing between the electrodes through the gum anesthetizes the nearby nerves thus alleviating pain.
The provided text describes the Synapse Dental Pain Eraser DPE-T300, a device intended for temporary dental anesthesia (pain relief) for at-home use. The document focuses on demonstrating its substantial equivalence to a predicate device (Dental Pain Eraser DPE-C300) through nonclinical and usability testing, rather than an effectiveness study proving specific acceptance criteria in terms of pain relief metrics.
Therefore, the acceptance criteria are primarily related to safety, essential performance, and usability for at-home use, ensuring it does not raise new issues of safety or effectiveness compared to the predicate.
Here's an analysis based on the provided information:
1. Table of Acceptance Criteria and Reported Device Performance
Given that this is a 510(k) submission focused on substantial equivalence to a predicate device, the "acceptance criteria" are implied by the standards and testing performed to demonstrate that the new device (DPE-T300) is as safe and effective as the predicate (DPE-C300), particularly for a change in use environment (from professional facility to home use).
Acceptance Criteria (Implied) | Reported Device Performance (DPE-T300) |
---|---|
Safety and Essential Performance (Non-Clinical) | |
Compliance with IEC 60601-1-11 (Home Healthcare Environment) | Compliant. Synapse performed safety testing in accordance with IEC 60601-1-11, Edition 2.0, 2015-01, demonstrating basic safety and essential performance for lay or trained healthcare personnel in the home healthcare environment. |
Ingress Protection (IP) Rating appropriate for home use | IP24. Tested in accordance with IEC 60529, Edition 2.2, 2013-08. This rating is considered appropriate for home use and does not raise different safety/effectiveness questions compared to the predicate's IP21 rating. |
Electrical Safety (Patient Leakage Current, Output Specs) | Compliant. Matches predicate (DC ~0 µA AC |
Ask a specific question about this device
(154 days)
The Dental Pain Eraser DPE-C300 is intended to provide temporary dental anesthesia (pain relief) and is indicated for use before, during, or after dental and orthodontic procedures, and other sources of oral pain. It is intended for use by dental/orthodontic professionals for the alleviation of oral pain.
The Dental Pain Eraser is a non-invasive electrical stimulation device that is used to numb the nerve pain associated with orthodontic procedures and to dull the pain of canker sores often developed inside the mouths of patients wearing braces or retainers. This method of treatment is referred to as Transcutaneous Electrical Nerve Stimulation (TENS). The Dental Pain Eraser has a shape and size similar to a common toothbrush, except that in place of the bristles are two metal electrodes. In operation, the device is turned on and for 30 seconds to 10 minutes the two electrodes are rubbed gently up and down the gums above the patient's aching tooth or on the intact skin around the canker sore. Current flowing between the electrodes through the gum anesthetizes the nearby nerves thus alleviating pain.
This document describes the regulatory submission for the "Dental Pain Eraser DPE-C300" (K182947), which is a non-invasive electrical stimulation device intended to provide temporary dental pain relief. The submission aims to demonstrate substantial equivalence to a predicate device, the "CEDETA Mk2" (K915717).
Based on the provided document, here's a breakdown of the acceptance criteria and the supporting study information:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not present a formal table of pre-defined acceptance criteria with specific numerical targets. Instead, it demonstrates substantial equivalence by comparing the proposed device to a predicate device across various characteristics and stating that comprehensive testing confirmed its performance and safety are comparable. The "Acceptance Criteria" here are implicitly meeting the specifications and safety/effectiveness profile comparable to the predicate device.
However, we can infer some "performance" points from the clinical study summaries.
Characteristic | Acceptance Criteria (Implicit) | Reported Device Performance |
---|---|---|
Pain Relief (General Dental/Orthodontic Pain) | Device performs as intended in relieving periodontal, mucosal, and dentinal pain related to orthodontic treatment. | CSP-TENS-002: Demonstrated that the device performs as intended at relieving periodontal, mucosal, and dentinal pain related to orthodontic treatment of adolescents and adults. |
Pain Relief (Orthodontic Separator Pain) | Device effectively alleviates tooth pain after orthodontic separator placement, especially during the period of greatest pain. | CSP-TENS-003: Confirmed that the device relieves periodontal pain following separator placement during the first 3 days cumulatively when the greatest pain is normally experienced. |
Pain Relief (Post-Orthodontic Adjustment Pain) | Device alleviates post-orthodontic appliance adjustment tooth pain. | CSP-TENS-001: Suggested that the device alleviates post-orthodontic appliance adjustment tooth pain. |
Safety | No adverse device effects reported. | All three clinical studies (CSP-TENS-002, CSP-TENS-003, CSP-TENS-001): "No adverse device effects were reported." |
Biocompatibility | Materials are biocompatible and meet relevant ISO standards. | Table 3: All tests for Cytotoxicity, Sensitization (Sodium Chloride & Sesame oil Extract), and Irritation (Sodium Chloride & Sesame oil Extract) Passed per ISO 10993-5 and ISO 10993-10 standards. |
Electrica Safety & EMC | Compliant with applicable standards. | Nonclinical Testing Summary: "Electrical Safety and Electromagnetic Compatibility Testing" was performed and passed. Compliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10. |
Water Ingress | Device meets specified ingress protection rating. | Nonclinical Testing Summary: "Water Ingress Testing" was performed and passed. Ingress Protection Rating IP21. |
Overall Performance | Meets established specifications and exhibits required mechanical and functional characteristics. | Conclusions: "The collective results confirm that the Dental Pain Eraser meets its specifications and exhibits the required mechanical and functional characteristics for its intended use..." |
2. Sample Sizes Used for the Test Set and Data Provenance
The document does not explicitly state the exact sample sizes for the clinical studies (CSP-TENS-001, CSP-TENS-002, CSP-TENS-003). It mentions "patients" in general terms.
- CSP-TENS-002: Patients selected during the leveling and aligning phase of orthodontic treatment with moderate to severe crowding experiencing immediate orthodontic pain.
- CSP-TENS-003: Patients randomly assigned to one of three treatment groups (Experimental, Sham, Control).
- CSP-TENS-001: Patients with periodontal pain associated with orthodontic appliance adjustment.
Data Provenance:
- The clinical studies were conducted by Synapse Dental, LLC.
- CSP-TENS-001 was a preliminary pilot study done by Doctor Cosmo Haralambidis at Cranston Orthodontics, Cranston, RI, United States.
- The document does not specify whether the studies were retrospective or prospective, but clinical trials are typically prospective. The mention of patient selection, randomization (for CSP-TENS-003), and pain assessment questionnaires strongly suggests prospective studies.
- The country of origin for the clinical data appears to be USA (Cranston, RI mentioned for CSP-TENS-001).
3. Number of Experts Used to Establish Ground Truth and Qualifications
The document does not discuss "ground truth" in the context of expert review for the clinical studies, as the device is for pain relief, which is a subjective patient experience. The "ground truth" for efficacy would be the reported pain levels by the patients themselves. Pain assessment questionnaires were used for CSP-TENS-003.
For the non-clinical testing, the "ground truth" is established by the specifications and standards (e.g., IEC, ISO, ANSI/AAMI) the device was tested against. The expertise lies in the certified laboratories and engineers performing these tests.
4. Adjudication Method for the Test Set
Not applicable in the context of this device and the clinical studies described. Pain relief is self-reported, not externally adjudicated by multiple experts viewing objective data.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
An MRMC study is typically performed for diagnostic imaging devices where human readers interpret images. This is not applicable to the "Dental Pain Eraser," which is a therapeutic electrical stimulation device for pain relief. The studies conducted compare the device's effect to a control/sham, not the improvement of human readers' performance with AI assistance.
6. Standalone Performance
The clinical studies describe the performance of the device itself (with a human operator, the dental/orthodontic professional, applying it), not an algorithm in isolation. The "performance" in this context is the alleviation of pain in patients when the device is used as intended. The studies assess the direct effect of the device on patient pain levels.
7. Type of Ground Truth Used
The primary "ground truth" for efficacy in the clinical studies is patient-reported pain levels validated via pain assessment questionnaires (explicitly mentioned for CSP-TENS-003). Safety "ground truth" is based on the absence of reported adverse device effects. For non-clinical tests, the ground truth is adherence to engineering specifications and international standards.
8. Sample Size for the Training Set
This product is a medical device for therapeutic use (pain relief), not an AI/ML model that requires a "training set" in the machine learning sense. Therefore, the concept of a training set sample size is not applicable here. The device's design is based on established principles of Transcutaneous Electrical Nerve Stimulation (TENS) and engineering.
9. How the Ground Truth for the Training Set Was Established
As this is not an AI/ML device, there isn't a "training set" or "ground truth" for it in that context. The device's design and operating parameters (e.g., output voltage, current, waveform) are based on the intended physiological effects of TENS and are validated through the non-clinical and clinical performance data.
Ask a specific question about this device