Search Results
Found 1 results
510(k) Data Aggregation
(171 days)
DePuy Synthes MAXFRAME Multi-Axial Correction System
The DePuy Synthes MAXFRAME™ Multi-Axial Correction System is indicated for the following treatments in adults, and in both children (3-12) and adolescents (12-21) in which the growth plates have fused or will not be crossed with hardware: fracture fixation (open and closed) pseudoarthrosis of long bones limb lengthening (epiphyseal or metaphyseal distraction) joint arthrodesis infected fractures or nonunions correction of bony or soft tissue deformities correction of segmental defects.
The DePuy Synthes MAXFRAME™ Multi-Axial Correction System, with adult and pediatric indications, is an external ring fixation system that is combined with web-based software in treatment of soft tissue and bone deformities. In the MAXFRAME system, implanted transfixion wires, Schanz screws and pins are attached to the rings and plates surrounding the deformity using bolts, nuts and connecting plates. Upper and lower rings are then connected to one another using six telescoping struts, creating a "Stewart Platform" type device. Adjusting the strut length allows for correction of length, rotation, and angular deformity at the same time. The modular nature of a ring fixation frame allows multiple frame options. A ring fixation frame is assembled individually by the surgeon to address the different characteristics of each case. The materials of construction for this system include aluminum, elgiloy, stainless steel and various polymeric materials. This system has been validated as MR Conditional.
The provided text describes the DePuy Synthes MAXFRAME™ Multi-Axial Correction System, a medical device, and its FDA 510(k) summary. However, it does not contain information about acceptance criteria or a study proving that an AI device meets acceptance criteria. The device itself is a physical external ring fixation system combined with web-based software for treatment planning, not an AI diagnostic or predictive tool in the context of typical AI performance studies.
Therefore, I cannot fulfill the request to provide a table of acceptance criteria and reported device performance related to AI, nor can I provide information on sample sizes, ground truth establishment, expert qualifications, adjudication methods, MRMC studies, or standalone performance for an AI component.
The document discusses performance in terms of:
- Construct testing: ASTM 1541-02 (2011)
- Component testing: Static & Dynamic Compression testing, Static Torsion, Static Cantilever, Static and Dynamic Fatigue, Wire Slip and Ring to Wire Interface Testing
- System Validation
- Software Verification: In accordance with Special Controls & EN IEC 62304:2006-05 Medical device Software life cycle processes
- MR Conditional Technical Data: In accordance with FDA Guidance and ASTM F2503-13
- Reprocessing of SUD in accordance with Special Controls
The conclusion states that "The objective evidence presented demonstrates that the mechanical performance of the proposed device is comparable to that of the predicate device(s) and supports substantial equivalence for safety and effectiveness."
No clinical evaluation was deemed necessary for the determination of substantial equivalence for this device.
Ask a specific question about this device
Page 1 of 1