    Intended Use

    The DYNAMIC Hollow Fiber Dialyzer DC-Series are indicated for use whenever a patient is in acute or chronic renal failure and hemodialysis is prescribed by a physician. Therefore, use of this device should be only on the direction of a physician who has evaluated all of the aspects of the patient's illness.

    Device Description

    The membrane used in this device which is substantially equivalent to the membranes utilized in the Cobe CentrySystem 160E and Baxter CF25 Dialyzers, which have been previously approved under a 510(k) Notification (K8G4831).

    Blood enters a blood inlet port where it is distributed to hollow fibers. Each hollow fiber has an inner diameter of 200 microns and a wall thickness of 8 microns. The fibers used in this device are substantially equivalent in design to the previously approved Cobe CentrySystem 160E Dialyzers. The wall thickness of the hollow fibers in Cobe CentrySystem 160)E Dialyzers, Baxter CF25 Dialyzers and the proposed device is 8 microns. The inner diameller of hollow fibers in both Baxter CF25 Dialyzers and the proposed device is 200 microns. The patient's blood traverses the inside of the hollow fibers and exits the device via a blood exit port.

    AI/ML Overview

    The provided text describes the 510(k) submission for the DYNAMIC Hollow Fiber Dialyzer DC-Series. The study conducted is an in vitro performance comparison to predicate devices, not a clinical study involving human patients or detailed statistical analysis for AI model validation. As such, many of the requested categories are not applicable to this type of device submission.

    Here's the information based on the provided text, with clarifications where data is not available or relevant:


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the DYNAMIC Hollow Fiber Dialyzer DC-Series appear to be substantial equivalence to existing predicate devices (Cobe CentrySystem 160E, Baxter CF25, and Terumo Clirans T175 dialyzers) in terms of in vitro performance. The table below presents the reported performance of the DYNAMIC DC-160 and DC-190 models alongside the predicate devices. The implicit acceptance criterion is that the DYNAMIC devices' performance values are comparable to or better than those of the predicate devices.

    MetricDYNAMIC DC-160 (Reported Performance)DYNAMIC DC-190 (Reported Performance)TERUMO CLIRANS T1.75 (Predicate Performance)Baxter CF25 (Predicate Performance)COBE CentrySystem 160E (Predicate Performance)Acceptance Criteria (Implicit)
    BUN Clearance (ml/min)185190192191-Comparable to predicate devices (e.g., 191-192 ml/min for Terumo/Baxter)
    Creatinine Clearance (ml/min)167179177170130Comparable to predicate devices (e.g., 130-177 ml/min)
    Phosphate Clearance (ml/min)155155151159-Comparable to predicate devices (e.g., 151-159 ml/min)
    Vitamin B12 Clearance (ml/min)6773796235Comparable to predicate devices (e.g., 35-79 ml/min)
    In Vitro Ultrafiltration Coefficient (ml/hr/mmHg)7.58.58.36.54.3Comparable to predicate devices (e.g., 4.3-8.3 ml/hr/mmHg)
    Effective Surface Area (m²)1.61.91.751.60.9Comparable to predicate devices (e.g., 0.9-1.75 m²)
    Number of Fibers10,30012,00011,80012,0006,240Comparable to predicate devices (e.g., 6,240-12,000)
    Inner Diameter (µm)200200200200-Comparable to predicate devices (e.g., 200 µm)
    Wall Thickness (µm)88988Comparable to predicate devices (e.g., 8-9 µm)
    Sterilization MethodETOETOETOETOETOSame sterilization method as predicate devices (ETO)
    Priming Volume (ml)9110412311242Comparable to predicate devices (e.g., 42-123 ml)
    Max. TMP (mmHg)500500500500-Comparable to predicate devices (e.g., 500 mmHg)

    The conclusion states: "The result indicates that the proposed device is substantially equivalent to Terumo Clirans T175 and Baxter CF25 for in vitro performance." This indicates that the reported values were determined to be within an acceptable range of the predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify a distinct "test set" sample size in terms of number of dialyzers tested. The data presented are reported performance metrics, likely averages or representative values from multiple measurements.
    • Data Provenance: The study was an "in vitro performance" study conducted by Dynamic Technology Corporation, located in Hsinchu, Taiwan, R.O.C. The data are therefore laboratory-generated, not from patient studies. It is retrospective in the sense that it compares new device data to established data from predicate devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This is not applicable as the study involved in vitro performance testing of a physical device, not an AI model or clinical assessment requiring expert human interpretation for "ground truth." The "ground truth" here is the measured physical performance parameters of the device as per standard laboratory protocols for dialyzer testing.

    4. Adjudication Method for the Test Set

    Not applicable. There was no need for adjudication as the study measured objective physical parameters in an in vitro setting.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI device, and no MRMC study was performed.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI device or algorithm. The performance described is of the physical device itself.

    7. The Type of Ground Truth Used

    The "ground truth" for this study is based on in vitro measurements and established engineering/medical device performance standards for clearance rates, ultrafiltration, and physical characteristics of dialyzers. This is an objective measurement of device function in a laboratory setting. The comparison is made against the reported performance of predicate devices which would have been established through similar in vitro testing.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI device trained on a dataset.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This is not an AI device, so there is no training set or associated ground truth establishment process in the context of machine learning.