Search Results
Found 1 results
510(k) Data Aggregation
(35 days)
DUS 2GP Reagent Strips, DUS 5 Reagent Strips, DUS 10 Reagent Strips
This device is intended for the in vitro measurement of the following in urine: Leukocyte, Nitrite, Urobilinogen, Protein, pH, Blood, Specific gravity, Ketone, Bilirubin, Glucose. These strips are intended for prescription, in vitro diagnostic use only and they are visually read.
DUS 2GP reagent strips provide qualitative and semiquantitative measurements for protein, and glucose in urine specimens. Test results may provide information regarding the status of carbohydrate metabolism and kidney function.
DUS 5 reagent strips provide qualitative and semiquantitative measurements for leukocytes, nitrite, blood , protein, and glucose in urine specimens. These measurements are used to aid in the diagnosis of metabolic disorders, kidney function anomalies and urinary tract infections.
DUS 10 reagent strips provide qualitative and semiquantitative measurements for specific gravity, pH, leukocytes, nitrite, protein, glucose, ketone, urobilinogen, bilirubin and blood in urine speciments are used to aid in the diagnosis of metabolic disorders, kidney function anomalies, urinary tract infections and liver function.
The DUS Series are urine test strips with different reagent pads for the determination of specific gravity, pH, leukocytes, nitrite, protein, glucose, ketone, urobilinogen, bilirubin and blood affixed onto plastic strips. of which leukocyte, nitrite, urobilinogen, protein, pH, blood, specific gravity, ketone, bilirubin and glucose reagent pads are affixed onto the plastic strips. The reagent pads react with analytes in the urine giving a visible color. Results are confirmed by comparison of the test strip with the color chart on the container. For each color result for each analyte, a semiquantitative value is available on the box label (e.g. bilirubin results include negative, 1, 2, and 4 mg/dL) and the associated qualitative result (e.g. bilirubin results include negative. +. ++, +++).
Here's an analysis of the provided document, outlining the acceptance criteria and study details for the DUS Reagent Strips:
Acceptance Criteria and Device Performance
The provided document does not explicitly state pre-defined acceptance criteria for the "Exact Agreement" or "Agreement within +/- one block (%)" in the method comparison study. However, the study results, which consistently show high percentages (generally in the high 90s and 100%) for both metrics across all analytes, implicitly represent the device meeting an expected high level of agreement with the predicate device.
For the linearity/assay reportable range, the acceptance criteria are implied by the reported "% Exact match." A high percentage (generally 97.7% to 100%) indicates acceptance.
For the detection limit study, the acceptance criteria are stated as "Sensitivity was defined as the cutoff for which ≥95% of the contrived pooled measurements were trace or the first positive result."
The interference study implicitly accepts the device if interference is either not observed at clinically relevant concentrations or if observed interferences are adequately addressed in the labeling.
Here's a table summarizing the reported device performance, where the values themselves act as the demonstration that the implied or stated acceptance criteria (high agreement, 95% detection, or non-interference/labeling for interference) were met.
Table of Reported Device Performance
Analyte | Test Type | Reported Device Performance (Exact agreement / % Exact match / Positive Agreement / Notes) |
---|---|---|
Precision (Within-run, N=90) | ||
Urobilinogen | Level 1 (Normal) | Exact: 100%, +/-1 block: 100% |
Level 2 (4mg/dL) | Exact: 100%, +/-1 block: 100% | |
Glucose | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (1000mg/dL) | Exact: 100%, +/-1 block: 100% | |
Bilirubin | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (4mg/dL) | Exact: 100%, +/-1 block: 100% | |
Ketones | Level 1 (Positive) | Exact: 100%, +/-1 block: 100% |
Level 2 (Positive) | Exact: 100%, +/-1 block: 100% | |
SG | Level 1 (1.020) | Exact: 100%, +/-1 block: 100% |
Level 2 (1.020) | Exact: 98.9%, +/-1 block: 100% | |
Blood | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (200 RBC/uL) | Exact: 100%, +/-1 block: 100% | |
pH | Level 1 (6) | Exact: 100%, +/-1 block: 100% |
Level 2 (7) | Exact: 100%, +/-1 block: 100% | |
Protein | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (100 mg/dL) | Exact: 100%, +/-1 block: 100% | |
Nitrite | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (Pos) | Exact: 100%, +/-1 block: 100% | |
Leukocytes | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (70 WBC/uL) | Exact: 100%, +/-1 block: 100% | |
Precision (Within-day, N=90) | ||
Urobilinogen | Level 1 (Normal) | Exact: 100%, +/-1 block: 100% |
Level 2 (4mg/dL) | Exact: 100%, +/-1 block: 100% | |
Glucose | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (1000mg/dL) | Exact: 100%, +/-1 block: 100% | |
Bilirubin | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (4mg/dL) | Exact: 100%, +/-1 block: 100% | |
Ketones | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (40mg/dL) | Exact: 100%, +/-1 block: 100% | |
SG | Level 1 (1.020) | Exact: 100%, +/-1 block: 100% |
Level 2 (1.020) | Exact: 100%, +/-1 block: 100% | |
Blood | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (200 RBC/uL) | Exact: 100%, +/-1 block: 100% | |
pH | Level 1 (6) | Exact: 100%, +/-1 block: 100% |
Level 2 (7) | Exact: 98.9%, +/-1 block: 100% | |
Protein | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (100 mg/dL) | Exact: 100%, +/-1 block: 100% | |
Nitrite | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (Pos) | Exact: 100%, +/-1 block: 100% | |
Leukocytes | Level 1 (Negative) | Exact: 100%, +/-1 block: 100% |
Level 2 (70 WBC/uL) | Exact: 100%, +/-1 block: 100% | |
Linearity / Assay Reportable Range (N=90 per concentration) | ||
Urobilinogen | Various Conc. | 97.7% - 100% Exact match |
Glucose | Various Conc. | 97.7% - 100% Exact match |
Bilirubin | Various Conc. | 96.6% - 100% Exact match |
Ketones | Various Conc. | 97.7% - 100% Exact match |
Specific Gravity | Various Conc. | 94.4% - 100% Exact match |
Nitrite | Various Conc. | 100% Exact match |
Blood | Various Conc. | 98.8% - 100% Exact match |
pH | Various Conc. | 97.7% - 100% Exact match |
Protein | Various Conc. | 96.6% - 100% Exact match |
Leukocytes | Various Conc. | 97.7% - 100% Exact match |
Detection Limit (N=90 per concentration) | ||
Urobilinogen | 2 mg/dL | 100% Positive Agreement |
Glucose | 100 mg/dL | 100% Positive Agreement |
Bilirubin | 1.0 mg/dL | 100% Positive Agreement |
Ketones | 5 mg/dL | 100% Positive Agreement |
Blood | 10 RBC/μL | 100% Positive Agreement |
Protein | 15 mg/dL | 100% Positive Agreement |
Nitrite | 0.05 mg/dL | 100% Positive Agreement |
Leukocytes | 15 WBC/μL | 100% Positive Agreement |
Analytical Specificity | ||
(Various Analytes) | (Various Interferents) | No interference observed for a long list of compounds at specified concentrations. Observed interferences were noted and addressed in labeling limitations. |
Method Comparison Study (Total N=867 samples; comparisons to Multistix 10SG) | ||
Urobilinogen | All levels | Exact: 92-100%, Within 1 block: 100% |
Glucose | All levels | Exact: 92-100%, Within 1 block: 100% |
Bilirubin | All levels | Exact: 92-100%, Within 1 block: 100% |
Ketones | All levels | Exact: 91-100%, Within 1 block: 100% |
Blood | All levels | Exact: 92-100%, Within 1 block: 100% |
Protein | All levels | Exact: 91-100%, Within 1 block: 100% |
Nitrite | All levels | Exact: 97-100%, Within 1 block: 100% |
Leukocytes | All levels | Exact: 93-100%, Within 1 block: 100% |
pH | All levels | Exact: 91-100%, Within 1 block: 100% |
Specific Gravity | All levels | Exact: 91-100%, Within 1 block: 100% |
Study Details
-
Sample size used for the test set and the data provenance:
- Method Comparison Study: Total of 867 samples.
- Provenance: Fresh urine samples obtained at three medical facilities. The country of origin is not explicitly stated but implied to be South Korea, given the submitter's address. The data is prospective as samples were "processed within 4 hours."
- Precision/Reproducibility: Two levels of commercially available urine-based control solutions.
- Sample Size: 90 replicates for within-run (10 tests from 3 lots at 3 sites) and 90 replicates for within-day (1 test a day from 3 lots, at 3 sites for 10 days) for each level.
- Provenance: Commercially available control solutions.
- Linearity/Assay Reportable Range:
- Sample Size: 90 replicates per concentration level (10 replicates with each of 3 lots of test strips).
- Provenance: Samples created by spiking known concentrations of standard materials or by serial dilution of a high concentration urine sample with negative urine.
- Detection Limit:
- Sample Size: 90 replicates for each concentration (each sample concentration analyzed 30 times using 3 reagent strip lots).
- Provenance: Negative urine spiked with standard materials.
- Analytical Specificity:
- Sample Size: 3 replicates using 3 lots of DUS 10 test strips for each concentration level of interfering substance.
- Provenance: Urine sample pools prepared at 3 analyte concentrations (negative, low, high positive) spiked with potential interfering substances.
- Method Comparison Study: Total of 867 samples.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Precision, Linearity, Detection Limit, Analytical Specificity: Ground truth was established by the known concentrations of control solutions or spiked samples. Testing was performed by 3 medical technicians as reported in the linearity study and precision study, but their specific qualifications (e.g., years of experience) are not stated beyond being "medical technicians."
- Method Comparison Study: The ground truth for the method comparison study was established by comparing the DUS 10 test strips results against a predicate device (Multistix 10SG). The testing was performed by three medical technicians at each of the clinical sites. Similar to above, their specific qualifications are not detailed beyond "medical technicians."
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- The document does not describe an explicit adjudication method (like 2+1 or 3+1) involving multiple human readers or experts resolving discrepancies for the test set in the traditional sense of image-based AI studies.
- The method comparison study compares the new device's readings to the predicate device's readings, and the percentage agreement is calculated. The process for resolving discrepancies between the new device and the predicate device is not detailed, nor is there a mention of an expert panel reviewing cases.
- For other studies (precision, linearity, detection limit), the ground truth is analytically determined by control concentrations.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a multi-reader, multi-case (MRMC) comparative effectiveness study as typically understood for AI-assisted diagnostic tools (i.e., human readers with and without AI assistance) was not explicitly performed or described in this document.
- The studies focus on the performance characteristics of the device itself (the reagent strips) and its comparison to a predicate device, which is also a reagent strip for visual reading. The device's use is "visually read" by operators, but the study design is not one of AI assistance to human readers.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- This device is a reagent strip that is visually read. It is not an AI algorithm. Therefore, the concept of "standalone (algorithm only without human-in-the-loop performance)" does not directly apply in the context of an AI device. The performance data presented (precision, linearity, detection limits, analytical specificity, method comparison) are essentially standalone performance characteristics of the physical reagent strip, with human visual interpretation being the intended mode of operation for reading the results. The method comparison specifically assesses this human-read performance against a predicate device.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Method Comparison Study: The ground truth for comparative analysis was the results obtained from the predicate device (Multistix 10SG).
- Precision, Linearity, Detection Limit, Analytical Specificity: The ground truth was analytically determined concentrations derived from commercial control solutions or spiked urine samples with known analyte levels.
-
The sample size for the training set:
- This document is for a traditional in-vitro diagnostic (IVD) reagent strip that is visually read, not an AI/ML device. Therefore, there is no "training set" in the context of machine learning model development. The assays are based on chemical reactions, and the performance is inherent to the chemical formulation and strip manufacturing.
-
How the ground truth for the training set was established:
- As there is no AI/ML component or "training set" in the context of an algorithm, this question is not applicable. The chemical reactions on the strips are designed to react to specific analytes, and their performance is validated through the studies described.
Ask a specific question about this device
Page 1 of 1