Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K123264
    Date Cleared
    2013-01-16

    (90 days)

    Product Code
    Regulation Number
    870.5100
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    DK-PTCA BALLOON CATHETER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Intuit Medical, LLC, DK-PTCA Balloon Catheter is indicated for balloon dilatation of the stenotic portion of coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion.

    Device Description

    The DK-PTCA Balloon Catheter is a torqueable, rapid-exchange, percutaneous transluminal coronary angioplasty catheter. The device is compatible with commonly used accessories including standard 0.014" coronary guide wires and 6F guide catheters (MIN inner diameter of 0.074"). Catheter working length is approximately 142cm.

    The distal end of the catheter has a semi-compliant balloon that expands to known diameters and lengths at specific pressures. The balloon has two radiopaque markers to assist with positioning. The proximal end of the device is a common PTCA catheter design consisting of a hypo-tube connected to a plastic hub and strain relief and which assists with torque transmission. The hub is used to inflate the balloon and the luer connector is compatible with standard inflation devices. A second lumen within the catheter, intended for guidewire use, extends from the rapid exchange port to the distal tip. The DK-PTCA Balloon Catheter is supplied sterile and intended for single use.

    AI/ML Overview

    This document describes the DK-PTCA Balloon Catheter, a device intended for balloon dilatation of coronary artery or bypass graft stenosis. It's a 510(k) submission, meaning the manufacturer is demonstrating that its device is "substantially equivalent" to an already legally marketed predicate device.

    Hence, the acceptance criteria and study information provided here are not for an AI/ML algorithm or a new novel device proving its standalone accuracy against a ground truth. Instead, these are performance criteria for a physical medical device (a balloon catheter) to demonstrate its safety and effectiveness through engineering and biological testing, showing it performs comparably to existing, approved devices.

    Given this context, I will adapt the requested information structure to fit the provided document's content.


    Device: DK-PTCA Balloon Catheter
    510(k) Number: K123264

    1. A table of acceptance criteria and the reported device performance

    The document lists performance tests conducted, implying these tests are designed to meet established acceptance criteria for PTCA catheters. However, the specific numerical acceptance criteria for each test (e.g., maximum allowable balloon burst pressure, minimum bond strength) and the exact quantitative reported performance for each test are not explicitly detailed in this summary. The summary states that "No new questions of safety or effectiveness were identified during device testing," which indicates the device met its internal acceptance criteria.

    Acceptance Criteria (Implied)Reported Device Performance (Implied)
    Dimensional Verification (e.g., balloon diameters, lengths)Met specifications; details not provided.
    Balloon Preparation, Deployment & Retraction (e.g., smooth operation)Met specifications; details not provided.
    Flexibility & Kink Resistance (e.g., catheter maneuverability)Met specifications; details not provided.
    Balloon Rated Burst Pressure (RBP)Met specifications; details not provided.
    Balloon Fatigue (e.g., durability under pulsatile flow)Met specifications; details not provided.
    Balloon Compliance (e.g., expansion characteristics)Met specifications; details not provided.
    Balloon Inflation and Deflation TimesMet specifications; details not provided.
    Catheter Bond Strength (e.g., integrity of joints)Met specifications; details not provided.
    Tip Pull Test (e.g., secureness of distal tip)Met specifications; details not provided.
    Torque Strength (e.g., ability to transmit rotational force)Met specifications; details not provided.
    Radiopacity (e.g., visibility under fluoroscopy)Met specifications; details not provided.
    Biocompatibility (Cytotoxicity, Sensitization, Irritation, Systemic Toxicity, Hemocompatibility, Genotoxicity, Pyrogenicity)Passed all tests according to ISO 10993-1 and FDA Bluebook Memorandum (G-95).

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample sizes used for each performance test. The data provenance is internal testing conducted by Intuit Medical, LLC, likely on manufactured samples of the DK-PTCA Balloon Catheter. The data would be prospective in the sense that the tests were designed and executed specifically for this 510(k) submission. There is no information on country of origin of the data beyond the manufacturer being in the USA.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable in the typical AI/ML context. For a physical medical device like a balloon catheter, "ground truth" is established by adherence to engineering standards, validated test methods, and regulatory guidelines (e.g., ISO, FDA guidance). The "experts" involved would be engineers, material scientists, and biomedical experts responsible for designing, conducting, and interpreting these physical performance and biocompatibility tests. Their qualifications are not specified but are implied by their roles in a medical device company and adherence to regulatory standards.

    4. Adjudication method for the test set

    This question is not applicable. Adjudication methods are typically used for expert consensus in clinical data or image interpretation. For physical device testing, results are typically objective measurements against pre-defined specifications. Any discrepancies would involve re-testing, calibration, or review of test methodology, not "adjudication" in the clinical sense.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. An MRMC study is relevant for diagnostic AI/ML devices involving human readers. The DK-PTCA Balloon Catheter is a therapeutic physical device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. This is a physical medical device, not an algorithm. The "standalone performance" refers to the device's physical and biological performance as measured by the tests listed.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For this physical device, the "ground truth" is based on:

    • Engineering Specifications and Standards: (e.g., material properties, dimensional tolerances, force requirements).
    • Biocompatibility Standards: (e.g., ISO 10993-1, FDA Bluebook Memorandum G-95) which dictate acceptable levels of biological response.
    • Performance Requirements: Established by predicate device characteristics and regulatory guidance for PTCA catheters.

    8. The sample size for the training set

    This question is not applicable. There is no "training set" in the context of a physical medical device. The device's design and manufacturing processes are refined through engineering development and iterative testing, but not in the "training set" sense of machine learning.

    9. How the ground truth for the training set was established

    This question is not applicable for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1