Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K063756
    Manufacturer
    Date Cleared
    2007-03-19

    (90 days)

    Product Code
    Regulation Number
    862.1215
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CTNI method is an in vitro diagnostic test for the quantitative measurement of cardiac troponin I in human serum and plasma on the Dimension Vista™ System. Measurements of cardiac troponin I are used to aid in the diagnosis of acute myocardial infarction (AMI) and in the risk stratification of patients with acute coronary syndromes with respect to their relative risk of mortality.

    Device Description

    The CTNI method is a one-step sandwich chemiluminescent immunoassay based on LOCI™ technology. LOCI™ reagents include two synthetic bead reagents and a biotinylated anti-cardiac troponin I monoclonal antibody fragment. The first bead reagent (Sensibeads) is coated with streptavidin and contains photosensitive dye. The second bead reagent (Chemibeads) is coated with a second anti-cardiac troponin I monoclonal antibody and contains chemiluminescent dye. Sample is incubated with Chemibeads and biotinylated antibody to form a particle/cardiac troponin I/biotinylated antibody sandwich. Sensibeads then are added and bind to the biotin to form bead-aggregated immunocomplexes. Illumination of the complex by light at 680 nm generates singlet oxygen from Sensibeads, which diffuses into the Chemibeads and triggers a chemiluminescent reaction. The resulting chemiluminescent signal is measured at 612 nm and is a direct function of the cardiac troponin I concentration in the sample.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the Dimension Vista™ CTNI Flex® reagent cartridge, structured according to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission doesn't explicitly define 'acceptance criteria' in a dedicated section with pass/fail thresholds. Instead, it presents performance characteristics and demonstrates substantial equivalence to a predicate device. The implied acceptance is that the new device's performance is comparable to or improved upon the predicate.

    FeaturePredicate Device Performance (Dimension® CTNI)New Device Performance (Dimension Vista™ CTNI)Implied Acceptance Criteria (relative to predicate)Met Criteria?
    Intended UseQuantitative determination of cardiac troponin-I in human serum and heparinized plasma for AMI diagnosis and risk stratification in ACS.Identical.Identical intended use.Yes
    Assay TypePhotometric immunoassayChemiluminescent immunoassayNew technology is acceptable if other performance is comparable or better.Yes
    Reportable Range0.04 to 40 ng/mL0.015 to 40 ng/mLEqual or wider (especially lower) range.Yes (improved)
    Analytical Sensitivity0.04 ng/mL0.015 ng/mLEqual or better (lower) sensitivity.Yes (improved)
    Functional SensitivityNot specified0.04 ng/mLFunctional sensitivity in line with clinical needs (not directly comparable to predicate's unspecified value, but a specified value is a positive).Yes
    Analytical Specificity (Skeletal muscle troponin-I)0.04 ng/mL (at 1000 ng/mL)0.14 ng/mL (at 1000 ng/mL)Comparable specificity.Yes
    Analytical Specificity (Cardiac troponin-T)0.34 ng/mL (at 1000 ng/mL)0.05 ng/mL (at 1000 ng/mL)Comparable or better specificity.Yes (improved)
    Analytical Specificity (Cardiac troponin-C)0 ng/mL (at 1000 ng/mL)0 ng/mL (at 1000 ng/mL)Identical specificity.Yes
    Hook EffectNo high dose effect up to 1800 ng/mLNo high dose effect up to 1000 ng/mLNo hook effect within a relevant clinical range. (Slightly narrower range than predicate but still clinically acceptable).Yes
    Calibration IntervalUpdated for each lot, 5 levels, every 60 days.Updated for each lot, 6 levels, every 30 days.Updated for each lot with appropriate frequency and levels. (More frequent calibration and more levels could be seen as an improvement in control).Yes
    Sample Volume50 µL20 µLEqual or smaller sample volume.Yes (improved)
    Method Comparison (Slope)N/A (Predicate vs. itself)0.99Close to 1.0 (indicating good agreement).Yes
    Method Comparison (Intercept)N/A-0.10Close to 0 (indicating good agreement).Yes
    Method Comparison (Correlation Coefficient)N/A0.993High (indicating strong linear correlation).Yes
    Serum vs. Plasma (Slope)N/A1.02Close to 1.0 (indicating good agreement).Yes
    Serum vs. Plasma (Intercept)N/A0.031Close to 0 (indicating good agreement).Yes
    Serum vs. Plasma (Correlation Coefficient)N/A0.999High (indicating strong linear correlation).Yes
    Li Heparin vs. Na Heparin (Slope)N/A0.99Close to 1.0 (indicating good agreement).Yes
    Li Heparin vs. Na Heparin (Intercept)N/A-0.05Close to 0 (indicating good agreement).Yes
    Li Heparin vs. Na Heparin (Correlation Coefficient)N/A0.998High (indicating strong linear correlation).Yes
    Reproducibility (Repeatability %CV)Not explicitly compared with predicate as a criterion.Low %CVs (3.8%, 1.25%, 1.05%, 1.0%, 1.7%)Low %CVs demonstrate good precision. (Acceptance is usually based on pre-defined internal company limits or industry standards for similar assays).Yes
    Reproducibility (Within Lab %CV)Not explicitly compared with predicate as a criterion.Low %CVs (7.36%, 3.18%, 2.22%, 3.76%, 1.05%)Low %CVs demonstrate good precision.Yes

    2. Sample Sizes and Data Provenance

    • Method Comparison (Test Set for predicate comparison):
      • Sample Size: 91 split serum patient samples.
      • Data Provenance: Not explicitly stated, but implied to be human patient samples. The country of origin is not mentioned. It is retrospective since existing patient samples were used to compare two methods.
    • Serum versus Plasma Results:
      • Sample Size: 63 matched serum and lithium heparin plasma samples.
      • Data Provenance: Not explicitly stated (country of origin), implied to be human patient samples. Retrospective.
    • Lithium Heparin versus Sodium Heparin Results:
      • Sample Size: 50 matched lithium and sodium heparin plasma samples.
      • Data Provenance: Not explicitly stated (country of origin), implied to be human patient samples. Retrospective.
    • Reproducibility (Test Set for precision):
      • Sample Size: 4 serum pools and 1 Biorad Cardiac Quality Control, Level 2. The number of runs/replicates per sample is specified: "a single test from two independent cups was analyzed twice per day." The total number of individual measurements is not directly provided but can be inferred from the GLSI document referenced.
      • Data Provenance: Not explicitly stated (country of origin). Implied to be in-house laboratory testing. Not directly patient samples for this part, but control materials.

    3. Number of Experts and Qualifications for Ground Truth

    • Not applicable. This submission is for an in vitro diagnostic (IVD) test measuring a biomarker (cardiac troponin I). The "ground truth" for method comparison and precision studies is the measurement itself, typically compared against a reference method (the predicate device in this case) or internal controls with known values. There are no human expert adjudicators involved in establishing ground truth for IVD performance characteristics like analytical sensitivity, range, precision, or method correlation.

    4. Adjudication Method for the Test Set

    • Not applicable. As described above, there's no human adjudication for establishing ground truth in these types of analytical performance studies for IVD devices. The comparison is statistical between the two methods' results.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No. This is an IVD device, specifically a lab assay. MRMC studies are typically used for imaging devices or other diagnostic systems where human readers interpret output and their performance (with/without AI assistance) is evaluated. This submission focuses on the analytical performance of the assay itself, not human interpretation of results.

    6. Standalone (Algorithm Only) Performance Study

    • Yes, implicitly. The entire submission describes the "standalone" performance of the Dimension Vista™ CTNI Flex® reagent cartridge system. This device is an automated immunoassay, meaning the algorithm/system produces a quantitative result without direct human intervention in the measurement process itself. The data presented reflects the performance of the algorithm and instrumentation together. It's a "device only" performance inherent in IVD submissions.

    7. Type of Ground Truth Used

    • For Method Comparison Studies: The ground truth for the comparison was established by the predicate device (Dade Behring Dimension® CTNI method). The study aims to show good agreement between the new device's results and the predicate's results when testing the same patient samples.
    • For Reproducibility Studies: The ground truth for reproducibility is the mean concentration of the control materials/patient pools, established through repeated measurements. The goal is to demonstrate consistency and precision around this mean.
    • For Specificity Studies: The ground truth is the known concentration of interfering substances (skeletal muscle troponin-I, cardiac troponin-T, cardiac troponin-C) in spiked samples, to see if they cross-react with the assay.

    8. Sample Size for the Training Set

    • Training Set Sample Size: Not specified in the provided text. IVD submissions typically focus on validation studies (test sets) rather than detailing the internal development/training an assay might undergo. For an immunoassay, "training" might refer to assay development, optimization of reagents, and initial performance characterization, rather than a distinct "training set" in the machine learning sense.

    9. How Ground Truth for the Training Set Was Established

    • Not explicitly stated/applicable in the same way as AI/ML. For an IVD immunoassay, the "ground truth" during development involves using reference materials, characterized samples, and known concentrations of analytes and interferents to optimize the assay's chemical and enzymatic reactions, antibody binding, and signal detection to accurately measure the target analyte (cardiac troponin I). This iterative process involves internal laboratory testing and characterization rather than an external "ground truth establishment" phase for a discrete "training set."
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1