Search Results
Found 7 results
510(k) Data Aggregation
(41 days)
ELIAN Digital Diagnostic X-ray System
The ELIAN diagnostic X-ray system is intended for use on adult and pediatric patients radiographic exposure of all body parts and operated by a qualified/trained doctor or technician. NOT intended for Mammography use.
The ELIAN is a digital radiographic image acquisition device. It is a fully integrated image capture and routing system under human operator control. This system may be usable by a technician in a typical radiology environment. The ELIAN system includes a Detector Panel, Soft Ware, Case, Grid, Power Box, Switch Box, Interconnecting Cables, U-arm Mechanical and Generator. The Detector Panel is an indirect conversion device in the form of a rectangular plate in which the input x-ray photons are absorbed in an a-Si layer with CSI. The Power Box functions as a buffer between the Detector Panel and Operating PC while also supplying power to the Detector Panel. The Flugel SW transfers signals between the Detector and X-ray Generator and also indicates the status of the panel using console. Finally, the Flugel SW contains for image data capture and correction of defects on the image data.
The device in question is the ELIAN Digital Diagnostic X-ray System.
1. Acceptance Criteria and Reported Device Performance
The provided document describes a 510(k) submission for substantial equivalence, focusing on comparative performance against predicate devices rather than specific acceptance criteria for standalone performance. However, some performance metrics are reported.
Metric | Predicate: CXDI-50G (K031447) | New Device: ELIAN (1717SCC) |
---|---|---|
MTF (1lp/mm) | 0.50 | 0.49 |
MTF (2lp/mm) | 0.22 | 0.234 |
MTF (3lp/mm) | 0.11 | 0.11 |
MTF (3.5lp/mm) | N/A | 0.083 |
DQE(0) | 0.1 | 0.22 |
DQE(1) | 0.25 | 0.35 |
DQE(2) | 0.17 | 0.275 |
DQE(3) | 0.09 | 0.14 |
Pixel Size | 160 x 160 microns | 127 x 127 microns |
Dynamic Range | 12 bits | 14 bits |
Spatial Resolution | 3.1 lp/mm | 3.5 lp/mm |
The document states: "The overall test results conclude that ELIAN has outperformed the predicate device Sedecal X-Plus LP Plus Digital (K090238) in terms of modulation transfer function and detective quantum efficiency."
2. Sample Size Used for the Test Set and Data Provenance
The document mentions "Clinical images were provided" but does not specify the sample size for a test set in terms of number of cases or images.
The data provenance is not explicitly stated in terms of country of origin or whether it was retrospective or prospective. It implies a clinical evaluation, but details are absent.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document mentions a "clinical study" that compared diagnostic equivalency but does not specify the number of experts used to establish ground truth or their qualifications. It states, "The result of tests demonstrated that ELIAN produces diagnostic images of equivalent quality as the predicate device..." This implies expert assessment, but no details are provided.
4. Adjudication Method for the Test Set
The document does not describe any adjudication method used for the test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
An MRMC comparative effectiveness study was not specifically mentioned or detailed. The document alludes to a clinical study to "investigate the diagnostic equivalency of detector panels," where ELIAN's images were compared to those from a predicate device across eight body parts. However, it does not describe a setup to measure the improvement of human readers with AI assistance. The ELIAN system is an X-ray imaging device, not an AI-powered diagnostic tool for interpretation assistance in MRMC studies.
6. Standalone Performance Study
The primary performance studies mentioned are non-clinical, focusing on physical parameters like MTF and DQE, in accordance with IEC 62220-1:2003 Standard. The document states: "The overall test results conclude that ELIAN has outperformed the predicate device Sedecal X-Plus LP Plus Digital (K090238) in terms of modulation transfer function and detective quantum efficiency." This indicates a standalone algorithm-only (device-only, in this case) performance assessment against a standard and a predicate, particularly for image quality metrics.
7. Type of Ground Truth Used
For the non-clinical testing (MTF, DQE), the "ground truth" is established through physical measurements and standards, specifically IEC 62220-1:2003.
For the clinical study, the "ground truth" used for comparison appears to be the diagnostic images produced by the predicate device, with the goal of showing "equivalent quality." This suggests a comparative diagnostic performance rather than an absolute ground truth (e.g., pathology, outcomes data).
8. Sample Size for the Training Set
The document does not mention a training set sample size. The ELIAN system is described as an image acquisition device, not a machine learning or AI algorithm that typically requires a training set for model development.
9. How the Ground Truth for the Training Set Was Established
Since a training set for an AI algorithm is not mentioned, the method for establishing its ground truth is not applicable/provided.
Ask a specific question about this device
(106 days)
Digital Diagnostic X-ray System (GC80)
The GC80 Series Digital X-ray imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.
The GC80 digital X-ray imaging system consists of High voltage generator (HVG), Ceiling Suspension, X-ray tube, Collimator, Detector, AEC, DAP, CIB(Control Interface Box), Wall Stand, Patient Table, Collimator, Detector, Remote controller, Grid, Foot switch, Barcode scanner and Auto-stitching stand. This system is used to capture images by transmitting X-ray to a patient's body. The X-ray passing through a patient's body is sent to the detector and then converted into electrical signals. These signals go through the process of amplification and digital data conversion in the signal process device being sent to the S-Station (Operation Software) and saved in DICOM file, a standard for medical imaging. The captured images are sent to the Picture Archiving & Communication System (PACS) server, and can be used for reading images.
The provided document is a 510(k) Premarket Notification for the SAMSUNG ELECTRONICS Co., Ltd. Digital Diagnostic X-ray System (GC80). It primarily focuses on demonstrating substantial equivalence to a predicate device (XGEO GC80) and does not detail specific acceptance criteria and a study dedicated to proving the device meets those criteria in the typical sense of a novel AI/CAD device.
Instead, the submission for the GC80 X-ray system focuses on confirming that its performance, despite the addition of a new detector, remains consistent with established medical device standards and is substantially equivalent to a previously cleared device. Therefore, the "acceptance criteria" here are effectively the performance standards met by the predicate device and relevant international/national standards, and the "study" is a comparative assessment.
Here's a breakdown of the requested information based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present a table of acceptance criteria for specific diagnostic performance metrics (e.g., sensitivity, specificity for disease detection) that would be common for AI/CAD devices.
Instead, it focuses on the equivalence of technical specifications and image quality. The implicit acceptance criteria are that the device performs as well as the predicate device and meets relevant international standards.
Acceptance Criteria Category (Implicit) | Reported Device Performance (GC80) |
---|---|
Image Quality (Non-clinical) | MTF (Modulation Transfer Function) and DQE (Detective Quantum Efficiency) curves and measurements do not differ from the predicate device (XGEO GC80) when tested by IEC 62220-1. This indicates equivalent image quality at the physical detector level. |
Image Quality (Clinical) | Radiologists evaluated the images of GC80 as substantially equivalent, and superior in some images, to the predicate device. |
Safety and Performance Standards | Electrical, mechanical, environmental safety and performance testing according to ES 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, ISO14971, 21CFR1020.30 and 21CFR1020.31 were performed. EMC testing was conducted in accordance with IEC 60601-1-2:2007. Wireless function was tested and verified followed by guidance, Radio frequency Wireless Technology in Medical Devices. All test results were satisfying the standards. This indicates compliance with recognized safety and performance standards for X-ray systems. |
Technological Characteristics | The GC80, despite adding new detectors (S4343-W, S3025-W), maintains substantially equivalent technological characteristics to the predicate XGEO GC80 (K140334) across various components including High Voltage Generator, Ceiling Suspension, Wall Stand, Patient Table, and Collimator, as detailed in the comparative tables on pages 5-7. Key identical specs include max power (80kW), kV/mA/exposure time ranges, AEC, APR, mechanical movement ranges, brake locking, control types, etc. The differences in detector area and number of pixels are stated to not impact safety or performance because pixel pitch and type are identical. |
Intended Use | Identical intended use as the predicate device: "intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications." |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Sample Size for Clinical Data: The document states, "Clinical images were provided," but it does not specify the sample size (number of images or patients) used for this clinical evaluation.
- Data Provenance: Not explicitly stated in terms of country of origin or whether it was retrospective/prospective. Given "Clinical images were provided," it implies a retrospective collection of existing images for evaluation.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
- The document states, "In clinical data, the radiologists evaluate the image of GC80 is substantially equivalent, and superior in some images to the predicate device."
- Number of experts: Not specified (plural "radiologists" suggests more than one).
- Qualifications of experts: Not specified beyond "radiologists."
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- The document does not specify an adjudication method. It only mentions that "radiologists evaluate the image."
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, an MRMC comparative effectiveness study was not done. This submission is for an X-ray system, not an AI/CAD device. The "clinical data" involved radiologists evaluating the images themselves, not evaluating AI output or comparing their performance with and without AI. Therefore, there's no discussion of human reader improvement with AI assistance.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Not applicable. This device is an X-ray imaging system, not an AI algorithm. Its performance is inherent in its ability to generate radiographic images, not in its interpretation.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- For the clinical evaluation, the "ground truth" for the image quality assessment was the expert evaluation of radiologists. The goal was to confirm that the images generated by the GC80 were of comparable, or superior, quality for diagnostic purposes as those from the predicate device.
8. The sample size for the training set
- Not applicable / Not specified. As this is an X-ray imaging system and not an AI/CAD device, there is no "training set" in the context of machine learning. The system's performance is based on its hardware and software engineering meeting physical and technical standards.
9. How the ground truth for the training set was established
- Not applicable. See point 8.
Ask a specific question about this device
(261 days)
DIGITAL DIAGNOSTIC X-RAY SYSTEM MODEL XGEO GU60A
The XGEO GU60A Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.
The XGEO GU60A digital X-ray imaging system is to be used to take and store image for diagnosis of patients. It consists of the High voltage generator (HVG), U-arm positioner, Detector, X-ray tube, Collimator and etc.
The provided text describes a 510(k) premarket notification for the "XGEO GU60A Digital Diagnostic X-ray System." It outlines the device's intended use and compares it to a predicate device, but it does not contain acceptance criteria or a study proving the device meets those criteria in the format requested.
The document primarily focuses on demonstrating substantial equivalence to a predicate device (Revolution XR/d of General Electric Company) for regulatory purposes. It mentions various tests conducted for safety, EMC, and performance according to established standards (e.g., IEC 60601 series, 21 CFR 1020.30, 21 CFR 1020.31, and FDA guidance for 510(k)s for Solid State X-ray Imaging Devices). It states, "All test results were satisfied," which implies the device passed, but it does not specify what those acceptance criteria were or present the detailed results of those tests.
Therefore, most of the requested information cannot be extracted from this document, specifically:
1. A table of acceptance criteria and the reported device performance: Not provided. The document states "All test results were satisfied" for safety, EMC, and performance tests, but does not list specific criteria or performance metrics.
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective): Not provided. The document mentions "Non-clinical testing and Clinical Testing were conducted," but gives no details about sample size, provenance, or study design.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience): Not provided.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not provided.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not explicitly stated. The device is a "Digital Diagnostic X-ray System" for generating images, not an AI-powered diagnostic aid that assists human readers. Therefore, an MRMC study comparing human readers with and without AI assistance is unlikely for this type of device as described.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. The device is a hardware system for image acquisition, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not provided, as detailed clinical study results are absent.
8. The sample size for the training set: Not provided. This device is an X-ray imaging system, not an AI model that typically requires a training set in the same context.
9. How the ground truth for the training set was established: Not provided. (See point 8).
Summary of what is present:
- Device Name: XGEO GU60A Digital Diagnostic X-ray System (K123106)
- Intended Use: Generating radiographic images of human anatomy by a qualified/trained doctor or technician. Not for mammographic applications.
- Predicate Device: Revolution XR/d of General Electric Company.
- Studies Conducted (mentioned, not detailed): Electrical, mechanical, environmental safety and performance testing according to IEC 60601 series, 21CFR1020.30, 21CFR1020.31; EMC testing according to IEC 60601-1-2(2001); Non-clinical testing and Clinical Testing in accordance with FDA guidance for 510(k)s for Solid State X-ray Imaging Devices.
- Conclusion: The device is safe and effective and substantially equivalent to the predicate device, with all test results reported as "satisfied."
Ask a specific question about this device
(126 days)
DIGITAL DIAGNOSTIC X-RAY SYSTEM
The DIAMOND-5A,6A,8A Digital Diagnostic X-ray System is indicated for use in generating radiographic images of human anatomy. This device is not intended for mammographic applications.
The digital X-ray imaging system consists of a high voltage (HV) generator, a tube support unit, an X-ray beam limiting device, a wall stand unit, a bucky table unit, a detector, operating software, and a tube, operates on a high-frequency inverter method, and is primarily used in a hospital for diagnosis of diseases in skeletal. respiratory and urinary systems.
The provided 510(k) summary (K102408) for the DRGEM Corporation's DIAMOND-5A, 6A, 8A Digital Diagnostic X-ray System does not contain information related to specific acceptance criteria for device performance or a study demonstrating the device meets such criteria in terms of diagnostic accuracy or clinical effectiveness.
Instead, the submission focuses on demonstrating substantial equivalence to a predicate device (Sedecal X-Plus LP Plus Digital, K090238) by showing compliance with electrical, mechanical, environmental safety, and EMC standards, which are considered sufficient for this type of device in the 510(k) pathway.
Therefore, many of the requested details about acceptance criteria, clinical studies, ground truth, and expert involvement are not present in the provided document.
Here's a breakdown of what can be extracted based on the input:
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria | Reported Device Performance |
---|---|
Safety and Performance Standards Compliance: |
- EN/IEC 60601-1 (General medical electrical equipment safety)
- EN/IEC 60601-1-1 (Safety requirements for medical electrical systems)
- EN/IEC 60601-1-3 (General requirements for radiation protection in diagnostic X-ray equipment)
- EN/IEC 60601-2-7 (Safety of high-voltage generators for diagnostic X-ray equipment)
- EN/IEC 60601-2-28 (Safety for X-ray source assemblies and X-ray tube assemblies for medical diagnosis)
- EN/IEC 60601-2-32 (Safety of associated equipment for diagnostic X-ray equipment)
- EN/IEC 60601-1-2(2001) (Electromagnetic compatibility - Requirements and tests) | All test results were satisfactory for compliance with the listed electrical, mechanical, environmental safety, and EMC standards. (This is a general statement and specific performance metrics against these standards are not provided in the summary.) |
2. Sample size used for the test set and the data provenance
- Not Applicable / Not Provided. The submission details compliance with engineering and safety standards, not a clinical performance study with a test set of images or patients.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not Applicable / Not Provided. No clinical performance study requiring ground truth establishment is mentioned.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- Not Applicable / Not Provided. No clinical performance study requiring adjudication is mentioned.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This is a traditional X-ray imaging system, not an AI-assisted diagnostic device. A MRMC study was not conducted or mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- No. This is a traditional X-ray imaging system, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Not Applicable / Not Provided. No clinical performance study requiring ground truth is mentioned.
8. The sample size for the training set
- Not Applicable / Not Provided. This is not an AI/machine learning device that would typically have a "training set."
9. How the ground truth for the training set was established
- Not Applicable / Not Provided. This is not an AI/machine learning device.
Summary of the Study (as described in the 510(k) summary):
The "study" described in this 510(k) summary is primarily a technical verification and validation (V&V) against recognized electrical, mechanical, environmental, and EMC safety and performance standards (EN/IEC 60601 series). The manufacturer performed these tests to demonstrate that the DIAMOND-5A,6A,8A system operates safely and effectively as an X-ray imaging system, and is substantially equivalent to the predicate device. The results of these tests were deemed "satisfactory," indicating compliance with the standards, which serves as the basis for the safety and effectiveness conclusion. There is no mention of clinical trials or studies related to diagnostic accuracy or clinical outcomes.
Ask a specific question about this device
(35 days)
Z-MOTION DIGITAL DIAGNOSTIC X-RAY SYSTEM
Ask a specific question about this device
(25 days)
SEDECAL X-PLUS LP PLUS DIGITAL DIAGNOSTIC X-RAY SYSTEMS
These Digital Radiographic Systems are intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.
These digital diagnostic x-ray systems consist of a tubehead/collimator assembly mounted on a U-Arm, along with a generator, generator control, and an x-ray table. Power ratings for the available generators are in the rage of 50 kw to 80 kW. Exposure voltage range varies from 40 - 125 KV or 40 - 150 kV, with current of 300 - 100 mA. Exposure time is 1 ms - 10 s. This device represents a new combination of already cleared Sedecal and Canon products.
The provided 510(k) summary for the Sedecal X-Plus LP Plus Digital Diagnostic X-Ray System (K090238) does not contain the information requested regarding acceptance criteria and a specific study proving the device meets those criteria, as you would typically expect for performance data of AI/CADe devices.
This submission is for a traditional X-ray system, and the approval is based on substantial equivalence to predicate devices, rather than a performance study with defined acceptance criteria for diagnostic accuracy.
However, I can extract and infer some information based on the document's content:
1. Table of Acceptance Criteria and Reported Device Performance:
- Acceptance Criteria: For this type of device, the acceptance criteria are generally related to demonstrating substantial equivalence to existing legally marketed devices, primarily focusing on safety and effectiveness through comparison of technical characteristics and adherence to relevant performance standards.
- Reported Device Performance: The document states, "The results of bench and test laboratory indicates that the new device is as safe and effective as the predicate devices."
Acceptance Criteria Category | Reported Device Performance |
---|---|
Intended Use | Met: Same intended use as predicate devices: diagnostic radiographic exposures of skull, spinal column, chest, abdomen, extremities, and other body parts for adult and pediatric subjects, in various patient positions. |
Configuration | Met: Similar configuration (U-Arm mount only) to predicate K062335. Differs from K042876 (U-Arm or Ceiling Suspension) but is considered a minor technological difference not impacting safety/effectiveness. |
Positioning Controls | Met: "Enhanced, same as X Plus LP Plus" (predicate K062335), an improvement over K042876's "Standard" but still within the scope of safe and effective operation for an X-ray system. |
Performance Standard | Met: Adheres to 21 CFR 1020.30 (Performance standards for ionizing radiation emitting products - radiographic equipment). This is a statutory requirement for all such devices. |
Generator | Met: High frequency made by Sedecal, same as both predicate devices. |
Collimator | Met: Manual (Ralco) or Automatic (Huestis). Offers an automatic option, which is an enhancement over the manual-only collimators of the predicate devices but is a common and accepted component in X-ray systems. |
Image Acquisition | Met: Utilizes CANON panels (multiple models: 50C, 50G, 40C & 40G). This is a key technological difference from predicate K062335 (Film) and an update from K042876 (Digital CANON 50G). The use of cleared Canon digital panels (referenced by K numbers K060433, K031633, K031447, K023750) demonstrates their established safety and effectiveness. |
Electrical Safety | Met: Adheres to IEC-60601 UL listed. This is a standard safety requirement for medical electrical equipment. |
Overall Conclusion | "as safe and effective as the predicate device, have few technological differences, and has no new indications for use, thus rendering them substantially equivalent to the predicate devices." |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- This information is not provided in the document. The approval is based on "bench and test laboratory" results and comparison of specifications, not a clinical test set with patient data for diagnostic accuracy assessment.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not applicable/provided. No clinical reader study or ground truth establishment relevant to diagnostic accuracy is described.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- This information is not applicable/provided.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done. This device is a digital X-ray system, not an AI or CADe device designed to assist human readers or provide diagnostic interpretations.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- No, a standalone performance study was not done. This is an imaging acquisition device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Not applicable. The "ground truth" for this submission revolves around engineering performance verified by "bench and external laboratory testing to applicable standards" (e.g., 21 CFR 1020.30, IEC-60601) to demonstrate the device functions as intended and is safe. It's not about diagnostic accuracy.
8. The sample size for the training set:
- Not applicable. This isn't an AI/machine learning device that requires a training set.
9. How the ground truth for the training set was established:
- Not applicable.
Ask a specific question about this device
(35 days)
DIGITAL DIAGNOSTIC X-RAY SYSTEM, MODELS MAK 1500, MAK 2000 AND MAK 3000
The MAK 1500, MAK 2000, MAK 3000 Digital Diagnostic X-Ray System (STATIONARY). (Digital Diagnostic X-Ray Systems) are intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.
The MAK 1500 employs a ceiling mounted tubehead/collimator combination made by Suinsa, the NOVA . The system has one digital panel inside the universal bucky in the chest stand. It employs the Suinsa NBS-2100 Universal Bucky. The MAK 2000 uses a conventional vertical column mount and adds a Suinsa NET-4100 table. The MAK 3000, like the MAK 1500, employs a ceiling mount and universal bucky in the chest stand and adds the Suinsa NET-4100 table with a digital panel mounted inside. The digital pancl is the same one we supplied in K080582, manufactured by DRTcch Corporation (K080064) and the software is also unchanged.
The provided text is a 510(k) summary for a Digital Diagnostic X-Ray System (MAK 1500, MAK 2000, MAK 3000) and an FDA clearance letter. It describes the device, its intended use, and its equivalence to predicate devices. However, this document does not contain information about specific acceptance criteria or a study proving that the device meets such criteria in terms of performance metrics.
The 510(k) summary primarily focuses on:
- Identification of the Device: Proprietary-Trade Name, Classification Name, Product Codes, Common/Usual Name.
- Equivalent legally marketed devices: Stating it's a modified device based on K080582.
- Indications for Use: What the device is intended to be used for (diagnostic radiographic exposures of various body parts in adults and pediatrics).
- Description of the Devices: Details about the components (tubehead, bucky, digital panel, software).
- Safety and Effectiveness, comparison to predicate device: A general statement that "The results of bench and standards testing indicates that the new device is as safe and effective as the predicate devices."
The FDA letter confirms the substantial equivalence determination but does not delve into specific performance data or acceptance criteria.
Therefore, based solely on the provided text, I cannot complete the requested tables and sections related to acceptance criteria, detailed study information, sample sizes, expert qualifications, or ground truth establishment. The document does not contain this level of detail regarding performance studies.
Here's what I can extract and state based on the provided text, noting the missing information:
Description of Device and General Regulatory Context:
- Device Name: Model MAK 1500, MAK 2000, MAK 3000 Digital Diagnostic X-Ray System (Stationary).
- Manufacturer: MedicaTech USA
- 510(k) Number: K082798
- Indications for Use: Intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.
- Predicate Device: K080582 (The current device represents a variation of the tubehead mounting method described in this 510(k)).
Missing Information from the Provided Text:
1. Table of Acceptance Criteria and Reported Device Performance:
- Not provided in the document. The document states "The results of bench and standards testing indicates that the new device is as safe and effective as the predicate devices," but no specific performance metrics, acceptance criteria, or comparative data are listed.
2. Sample size used for the test set and the data provenance:
- Not provided in the document. No information on a specific "test set" for performance evaluation (e.g., number of images, patient demographics, country, retrospective/prospective nature) is available.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not provided in the document. There is no mention of expert reviews or ground truth establishment for a test set.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not provided in the document. No adjudication method is mentioned as there is no described test set evaluation.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable/Not provided. This is a diagnostic imaging system (X-ray machine), not an AI-assisted diagnostic software. Therefore, an MRMC study comparing human readers with/without AI assistance would not be relevant to this specific device's clearance. The document does not describe any AI component.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable/Not provided. As stated above, this is an X-ray system, not an algorithm, so a standalone algorithm performance evaluation is not relevant or described.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Not provided in the document. No ground truth type is mentioned as no specific performance study is detailed.
8. The sample size for the training set:
- Not applicable/Not provided. As this is a hardware X-ray system and not an AI/machine learning algorithm, there is no "training set" in the conventional sense.
9. How the ground truth for the training set was established:
- Not applicable/Not provided. (See point 8).
Conclusion:
The provided 510(k) summary and FDA letter pertain to the substantial equivalence of a general-purpose digital X-ray system. The documentation focuses on hardware specifications, intended use, and comparison to a predicate device based on bench and standards testing for safety and effectiveness. It does not provide the detailed performance study information, acceptance criteria, sample sizes, expert qualifications, or ground truth establishment methods typically found for software algorithms or devices requiring specific diagnostic accuracy claims. For an X-ray system, "safety and effectiveness" often refers to image quality (resolution, contrast, dose), electrical safety, mechanical safety, and compliance with performance standards, rather than diagnostic accuracy metrics of a distinct algorithm.
Ask a specific question about this device
Page 1 of 1