Search Results
Found 1 results
510(k) Data Aggregation
(59 days)
DENU Light Body(Regular, Fast) / DENU Heavy Body(Regular, Fast)
- Crown and bridge impression
- Inlay and onlay impression
- Denture impression
- Model impression
DENU Putty Set(Regular, Fast)
- Crown and bridge impression
- Inlay and onlay impression
The DENU family of silicone impression materials consists of three different viscosities (light body, heavy body, putty) for various application systems. Products are further classified into Regular type and Fast type, depending on curing time, but there is no difference on viscosity, indications for use, or scope of use. Each device is consisted of 1:1 base and catalyst component, packaged in 560mL jars for DENU Putty set and 50mL syringes for other devices.
This is a 510(k) premarket notification for a dental impression material, DENU Light Body (Regular, Fast), DENU Heavy Body (Regular, Fast), and DENU Putty Set (Regular, Fast). The submission aims to demonstrate substantial equivalence to legally marketed predicate devices.
Here's an analysis of the acceptance criteria and supporting study information:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are primarily derived from the international standard ISO 4823:2021 (Dentistry - Elastomeric impression materials) and the device's performance is compared to these standards and the predicate devices.
| Performance Characteristic | Acceptance Criteria (from ISO 4823:2021 or Predicate) | DENU Light Body (Regular, Fast) Reported Performance | DENU Heavy Body (Regular, Fast) Reported Performance | DENU Putty Set (Regular, Fast) Reported Performance | Performance Conclusion |
|---|---|---|---|---|---|
| Working Time | Specified by manufacturer or within predicate's range (e.g., 1'30" - 2'30" for light body) | Light Body, Regular: 1'30" | Heavy Body, Regular: 1'30" | Putty Set Regular: 1'30" | Equivalent / Similar (conforms to standard) |
| Light Body, Fast: 1'30" | Heavy Body, Fast: 1'30" | Putty Set Fast: 1'30" | Equivalent / Similar (conforms to standard) | ||
| Setting Reaction Time | Specified by manufacturer or within predicate's range (e.g., 2'30" - 4'00" for light body) | Light Body, Regular: 4'10" | Heavy Body, Regular: 4'10" | Not directly reported (comparison to predicate shows equivalence) | Equivalent |
| Light Body, Fast: 2'40" | Heavy Body, Fast: 2'40" | Equivalent | |||
| Consistency / Flow | Light Body: >= 36 mm (ISO 4823:2021 7.2) | Light Body, Regular: 39.18 mm | Equivalent | ||
| Light Body, Fast: 36.16 mm | Equivalent | ||||
| Heavy Body: < 35 mm (ISO 4823:2021 7.2) | Heavy Body, Regular: 34.10 mm | Equivalent | |||
| Heavy Body, Fast: 32.54 mm | Equivalent | ||||
| Putty Set: < 35 mm (ISO 4823:2021 7.2) | Putty Set Regular: 26.72 mm | Equivalent | |||
| Putty Set Fast: 27.94 mm | Equivalent | ||||
| Compatibility with Gypsum | 50 µm to 75 µm should be reproduced without interruption (ISO 4823:2021 7.6) | Light Body, Regular: Reproduced for 0.05 mm | Heavy Body, Regular: Reproduced for 0.05 mm | Putty Set Regular: Reproduced for 75 µm | Equivalent |
| Light Body, Fast: Reproduced for 0.05 mm | Heavy Body, Fast: Reproduced for 0.05 mm | Putty Set Fast: Reproduced for 75 µm | Equivalent | ||
| Strain in Compression | 0.8% - 20% (ISO 4823:2021 7.8) | Light Body, Regular: 2.52% | Heavy Body, Regular: 1.54% | Putty Set Regular: 1.74% | Similar (conforms to standard) |
| Light Body, Fast: 2.58% | Heavy Body, Fast: 1.10% | Putty Set Fast: 1.60% | Similar (conforms to standard) | ||
| Linear Dimensional Change / Accuracy | <= 1.5% (ISO 4823:2021 7.5) | Light Body, Regular: 0.09% | Heavy Body, Regular: 0.20% | Putty Set Regular: 0.21% | Equivalent / Similar (conforms to standard) |
| Light Body, Fast: 0.13% | Heavy Body, Fast: 0.27% | Putty Set Fast: 0.17% | Equivalent / Similar (conforms to standard) | ||
| Mixing Time | <= 60 sec (ISO 4823:2021 7.1) | Not directly reported for L/H body | Not directly reported for L/H body | Putty Set Regular: 26.2" | Similar (conforms to standard) |
| Putty Set Fast: 21.4" | Similar (conforms to standard) | ||||
| Appearance and Component colors | Not explicitly quantified, but implied to be similar to predicate. | Met preset test criteria. | Met preset test criteria. | Met preset test criteria. | Met criteria |
| Detail Reproduction | Not explicitly quantified, but implied to be similar to predicate and meet ISO 4823. | Met preset test criteria. | Met preset test criteria. | Met preset test criteria. | Met criteria |
| Elastic recovery | Not explicitly quantified, but implied to be similar to predicate. | Met preset test criteria. | Met preset test criteria. | Met preset test criteria. | Met criteria |
| Cytotoxicity | Conforms to ISO 7405 and ISO 10993-1. | Met preset test criteria. | Met preset test criteria. | Met preset test criteria. | Met criteria |
| Skin Sensitization | Conforms to ISO 10993-10. | Met preset test criteria. | Met preset test criteria. | Met preset test criteria. | Met criteria |
| Oral Mucosa Irritation | Conforms to ISO 10993-10. | Met preset test criteria. | Met preset test criteria. | Met preset test criteria. | Met criteria |
| Acute Systemic Toxicity | Conforms to ISO 10993-11. | Met preset test criteria. | Met preset test criteria. | Met preset test criteria. | Met criteria |
2. Sample size used for the test set and the data provenance
The document does not explicitly state the sample sizes for each specific physical property test (e.g., how many samples were tested for working time, consistency, etc.). The testing was based on ISO 4823:2021 and other biocompatibility standards (ISO 7405, ISO 10993-10, ISO 10993-11). These standards typically specify the number of samples required for each test. Given the context of a 510(k) submission, it is assumed that the protocol for each standard was followed, including the specified sample sizes (though not explicitly listed in this summary).
The data provenance is retrospective, as the tests have already been performed and the results are being presented. The manufacturer, HDI, Inc., is based in South Korea, implying the tests were likely conducted there or by affiliated labs.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This submission describes physical and biological bench testing of an impression material against established international standards. There is no mention of human expert involvement or a "ground truth" established by experts in the context of diagnostic interpretation for this type of device. The "ground truth" for these tests is the objective measurement of physical and chemical properties as defined by the ISO standards.
4. Adjudication method for the test set
Not applicable. As described in point 3, this is bench testing of physical and biological properties. There is no expert adjudication process in this context.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a medical device submission for a dental impression material, not an AI-powered diagnostic device. No MRMC study was performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a dental impression material, not an algorithm or AI.
7. The type of ground truth used
The ground truth used for this device's evaluation is primarily objective physical and chemical measurements as defined by:
- International Standards: Specifically, ISO 4823:2021 (Dentistry - Elastomeric impression materials) and ISO 10993 series (Biological evaluation of medical devices).
- Predicate Device Performance: The performance characteristics of the proposed device are directly compared to those of the predicate devices (Vonflex STM and Vonflex S Putty).
8. The sample size for the training set
Not applicable. There is no AI or machine learning component to this device, so there is no training set in the conventional sense. The "training" for the device's performance could be considered the iterative formulation and testing by the manufacturer to meet the required specifications, but this is not a data-driven training set like in AI.
9. How the ground truth for the training set was established
Not applicable. As there is no training set for an AI/ML algorithm, this question does not apply.
Ask a specific question about this device
Page 1 of 1