Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K151373
    Date Cleared
    2015-08-07

    (78 days)

    Product Code
    Regulation Number
    876.1075
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Corvocet Coaxial Introducer is intended for use as a guiding needle in obtaining core biopsy samples from soft tissues such as liver, kidney, prostate, spleen, breast, lung, lymph nodes and various soft tissue tumors.

    Device Description

    The Merit Corvocet Coaxial Introducer is used as a guiding needle for a biopsy device obtaining core biopsy samples from soft tissues. It is a three-part disposable device consisting of the following: a stainless steel outer cannula with a female Luer lock hub and an echo-enhanced tip to aid with visibility under ultrasound; a stainless steel inner trocar attached to a unique trocar hub design that allows it to be released by squeezing both sides of the hub; and a flexible silicone slip ring-style depth stop that is color-coded for gauge identification and can be manually adjusted to maintain the depth of the coaxial introducer in a procedure. An optional stainless steel blunt stylet with hub can be included with 17G and 19G coaxial introducers. The blunt stylet may be used to manipulate through soft tissue and around vasculature or other organs to minimize the risk of unintentional damage to these areas. The Merit Corvocet Coaxial Introducer is supplied sterile and is single use only.

    AI/ML Overview

    This document describes the performance testing and safety evaluations for the Merit Corvocet Coaxial Introducer (K151373) to demonstrate its substantial equivalence to a predicate device.

    Acceptance Criteria and Device Performance:

    The document lists various performance and biocompatibility tests conducted. For each test category, the report states that the device "met the standards' established acceptance criteria applicable to the safety and efficacy of the device." Specific numerical acceptance criteria and reported device performance values are not provided in this document. Instead, it indicates conformity to recognized standards and guidance.

    Acceptance Criteria CategoryReported Device Performance
    Performance Testing - Bench
    DimensionsMet established acceptance criteria per ISO standards
    Tensile of Joints Needle-to-HubMet established acceptance criteria per ISO standards
    Tensile of Joints Trocar-to-HubMet established acceptance criteria per ISO standards
    Tensile of Joints Blunt Stylet-to-HubMet established acceptance criteria per ISO standards
    Luer - GaugingMet established acceptance criteria per ISO standards
    Luer Liquid LeakageMet established acceptance criteria per ISO standards
    Luer Air LeakageMet established acceptance criteria per ISO standards
    Luer Separation ForceMet established acceptance criteria per ISO standards
    Luer Unscrewing TorqueMet established acceptance criteria per ISO standards
    Luer Ease of AssemblyMet established acceptance criteria per ISO standards
    Luer Resistance to OverridingMet established acceptance criteria per ISO standards
    Luer Stress CrackingMet established acceptance criteria per ISO standards
    Depth StopMet established acceptance criteria per ISO standards
    Introducer VisibilityMet established acceptance criteria
    Simulated UseMet established acceptance criteria
    Biocompatibility Testing
    CytotoxicityMet established acceptance criteria per ISO 10993
    SensitizationMet established acceptance criteria per ISO 10993
    IrritationMet established acceptance criteria per ISO 10993
    Acute Systemic ToxicityMet established acceptance criteria per ISO 10993
    PyrogenicityMet established acceptance criteria per USP 37, NF 32
    Chemical CharacterizationMet established acceptance criteria per ISO 10993

    Details of the Study:

    The provided document describes a series of performance and biocompatibility tests rather than a single "study" in the context of clinical trials involving human subjects or AI algorithms.

    1. Sample size for the test set and data provenance:

      • The document does not specify the sample sizes (e.g., number of needles/introducers tested) for each individual bench test or biocompatibility test.
      • The data provenance is from bench testing and laboratory biocompatibility testing of the device components and final product. There is no indication of human subject data or data from specific countries of origin for these tests beyond the device manufacturer's location (South Jordan, Utah, USA).
    2. Number of experts used to establish the ground truth for the test set and their qualifications:

      • This is not applicable as the tests described are primarily engineering performance tests and laboratory-based biocompatibility evaluations against predefined standards. There is no "ground truth" established by experts in the context of diagnosing conditions or interpreting images.
    3. Adjudication method for the test set:

      • Not applicable for the types of tests described. Results are evaluated against objective criteria outlined in the referenced standards.
    4. Multi-Reader Multi-Case (MRMC) comparative effectiveness study:

      • No MRMC study was done. This device is an introducer needle for biopsy, not an imaging or diagnostic AI device that would involve human readers.
    5. Standalone (i.e., algorithm only without human-in-the-loop performance) study:

      • No standalone algorithm performance study was done. This device is a physical medical instrument, not an AI algorithm.
    6. Type of ground truth used:

      • The "ground truth" for the performance tests is derived from recognized international and national standards and guidance documents (e.g., ISO standards, ASTM standards, FDA guidance, USP). These standards define the acceptable physical properties, functionality, and biological safety limits for such medical devices.
    7. Sample size for the training set:

      • Not applicable. This is not a machine learning device; therefore, there is no training set in the AI sense. The design and manufacturing processes are likely informed by prior experience and engineering principles.
    8. How the ground truth for the training set was established:

      • Not applicable, as there is no training set mentioned in the context of AI. The device's design and manufacturing are based on established engineering principles and compliance with the referenced standards.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1