510(k) Data Aggregation

    K Number
    K193287
    Device Name
    CorInsights MRI
    Date Cleared
    2020-11-20

    (359 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CorInsights MRI Medical Image Processing Software is intended for automatic labeling, visualization and volumetric quantification of segmentable brain structures from MRI images. Volumetric measurements are compared to reference percentile data. CorInsights MRI is for adults age 45 to 95.

    Device Description

    CorInsights MRI is a fully automated MR medical image processing software intended for automatic labeling, visualization and volumetric quantification of identifiable brain structures from DICOM formatted magnetic resonance images. The resulting output consists of a pdf report for review, which can be used in research and clinical use, and a DICOM image showing the anatomical structure boundaries identified by the software. The proposed device provides morphometric measurements based on T1 MRI series. The output of this software only device includes morphometric reports that provide comparison of measured volumes to age and gender-matched reference data and an image volume that has been annotated with color overlays representing each segmented region. The architecture has a proprietary automated internal process that includes artifact correction, atlas-based segmentation, volume calculation, and report generation. Quality control measures include automated quality control including image header checks to verify that the scan acquisition protocol and provided data adhere to system requirements, an image morphometry check, a tissue contrast check, and value range checks.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly state "acceptance criteria" as a separate, pre-defined column with specific numeric thresholds. Instead, it describes accuracy and reproducibility goals in terms of achieving certain statistical measures (ICC, DICE coefficient, mean absolute percentage difference) against ground truth or in test-retest scenarios.

    I will infer the "acceptance criteria" based on the reported "Performance" values, as these are the results presented to demonstrate the device's capability against known ground truths or expected variations.

    CategorySpecific MeasureAcceptance Criteria (Implied from Reported Performance)Reported Device Performance
    AccuracyHippocampal Volume
    IntraClass Correlation (ICC)ICC >= 0.95 (Highly correlated with ground truth)0.95
    DICE Coefficient (Left)DICE >= 83% (Good overlap with ground truth)83% (SD 2.5%)
    DICE Coefficient (Right)DICE >= 83% (Good overlap with ground truth)83% (SD 2.7%)
    Mean Absolute % Difference (Left)MAD = 0.99 (Highly correlated with ground truth)0.99
    DICE CoefficientDICE >= 95% (Excellent overlap with ground truth)95% (SD 1.6%)
    Mean Absolute % DifferenceMAD = 0.89 (Strong correlation with ground truth)0.89
    DICE CoefficientDICE >= 95% (Excellent overlap with ground truth)95% (SD 1.1%)
    Mean Absolute % DifferenceMAD = 0.98 (Very strong correlation with ground truth)0.98 (left and right)
    DICE Coefficient (Left)DICE >= 88% (Good overlap with ground truth)88% (SD 5.2%)
    DICE Coefficient (Right)DICE >= 87% (Good overlap with ground truth)87% (SD 5.6%)
    Mean Absolute % Difference (Left)MAD = 0.97 (Very high test-retest consistency)0.97
    DICE CoefficientDICE >= 89% (Good test-retest overlap)89% (SD 4.0%)
    Mean Absolute % Difference Range (Across all volumes)MAD (range) 0.7-5.8%, Average
    Ask a Question

    Ask a specific question about this device