Search Results
Found 7 results
510(k) Data Aggregation
(66 days)
For storage of soft (hydrophilic), hard and rigid gas permeable (RGP) contact lenses during chemical disinfection. Use for storage during chemical disinfection only. Do not use during heat disinfection.
The Contact Lens Case is a lens case product to be used by the contact lens wearer or practitioner for storing soft (hydrophilic), hard and rigid gas permeable contact lenses during chemical disinfection. The Case is to be designed during chemical disinfection only. No designed for heat disinfecting system.
The Contact Lens Case includes A-1 and B-1 two models. Both have adopted the same structure design, consisting of two parts: holder and cover. The holder is based with adjoining dual wells for the containment of fluid, and two covers are designed for screwing. At the same time, the bottom of each well is marked with L (left) and R (right) letters, and the covers are also labeled with L (left) and R (right) letters.
The Contact Lens Case is simple with a well volume of 4.2mL $\pm$0.15. This volume capacity can provide sufficient space to insert and retrieve the contact lenses. It also provides sufficient space to introduce the chemical cleaning solution and adequately over the lenses for cleaning.
The provided document describes a 510(k) premarket notification for a Contact Lens Case, Model: A-1, B-1. The primary focus of this submission is to demonstrate substantial equivalence to legally marketed predicate devices, rather than establishing novel safety and effectiveness through extensive clinical trials for a new medical device. Therefore, the information regarding acceptance criteria and studies will be different from what might be found for a device requiring more comprehensive performance validation (e.g., AI algorithms or complex medical devices).
Here's an analysis based on the provided text:
Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device are largely based on demonstrating equivalence in product design, material biocompatibility, and functional performance (leakage) to existing, approved predicate devices.
Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criterion | Requirement | Reported Device Performance |
|---|---|---|
| I. Biocompatibility | Must comply with ISO 10993-1, specifically tests for: | All tests were passed (see details below). |
| Cytotoxicity | ISO 10993-5:2009/(R)2014, Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity | Passed |
| Skin Sensitization | ISO 10993-10:2010, Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization | Passed |
| Ocular Irritation | ISO 10993-10:2010, Biological Evaluation of Medical Devices - Part 10: Tests for Ocular Irritation | Passed |
| Systemic Toxicity | ISO 10993-11:2006/(R)2010, Biological Evaluation of Medical Devices - Part 11: Tests for Systemic Toxicity | Passed |
| II. Leakage Testing | No leakage when filled to 2/3 with liquid, turned upside down for 15 minutes, repeated 3 times. | None of the tested contact lens cases showed any leakage and all passed the leakage tests successfully. (120 combinations of different tops and bottoms were tested). |
| III. Effectiveness (Volume) | The capacity must be sufficient for the contact lens to be fully immersed under use conditions. (Predicate devices had volumes of 4.2mL ± 0.15 and 4.8mL, and "over 3.0mL"). Subject device has a well volume of 4.2mL ± 0.15. | The capacity is sufficient for contact lens to be fully immersed under use conditions. (Stated in the comparison table and device description, implicitly meeting the criterion by matching/exceeding predicate volume). |
| IV. Material Equivalence | The subject device materials should be similar to or have demonstrated equivalent safety profile to predicate devices. (A-1: Polypropylene (holder), High-density Polyethylene (HDPE) (cover); B-1: Acrylonitrile-Butadiene-Styrene copolymer (ABS)) | A-1 Model is made of similar materials as Predicate Device 1 (Polypropylene holder, Polyethylene lid). B-1 Model is made of the same material as Predicate Device 2 (Acrylonitrile-Butadiene-Styrene copolymer (ABS)). Relevant biocompatibility tests were conducted and met all requirements, demonstrating comparability. |
| V. Design & Intended Use | Similar design (holder with two wells, screw covers, L/R markings) and identical intended use (storage of soft, hard, RGP contact lenses during chemical disinfection, not heat disinfection) to predicate devices. | The subject device has the same classification, intended use, and similar product design as the predicate devices. Minor differences in material and appearance were addressed by safety testing. The subject device has "the same performance effectiveness as the predicate devices." (Stated in comparison tables). |
Study Details
Given the nature of a 510(k) for a contact lens case, the "study" is primarily a set of non-clinical tests to demonstrate safety and performance equivalence.
1. Sample size used for the test set and the data provenance:
- Biocompatibility Testing: The specific sample sizes for each individual biocompatibility test (cytotoxicity, sensitization, ocular irritation, systemic toxicity) are not explicitly stated in the provided text. These tests are typically performed on extracts from a representative number of device components or material samples under controlled laboratory conditions, not on a "test set" in the sense of clinical cases.
- Leakage Testing:
- Sample Size: 120 combinations of different tops and bottoms.
- Data Provenance: Not explicitly stated but implied to be conducted by the manufacturer or a contracted testing lab as part of the device's validation.
2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not applicable or not provided for this type of device and submission. "Ground truth" established by experts is typically relevant for diagnostic devices or AI algorithms where human interpretation is being evaluated or augmented. For a contact lens case, the "ground truth" for performance is defined by adherence to physical and biological testing standards.
3. Adjudication method for the test set:
- This information is not applicable or not provided. Adjudication methods (e.g., 2+1, 3+1) are for resolving discrepancies in expert interpretations of data, usually in diagnostic imaging or clinical trials. For a non-clinical device like a contact lens case, the test results are usually measured objectively against predefined pass/fail criteria.
4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices involving human readers. The contact lens case is a physical medical device, not an AI diagnostic tool.
5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- No, a standalone algorithm performance study was not done. This device does not involve an algorithm.
6. The type of ground truth used:
- The "ground truth" for this device's performance is established by objective measurements against recognized international standards and specifications:
- Biocompatibility: Adherence to ISO 10993 standards for cytotoxicity, sensitization, ocular irritation, and systemic toxicity.
- Leakage: Absence of visible liquid escape under specified test conditions.
- Volume/Effectiveness: Physical measurement of fluid capacity compared to the functional requirement (sufficient immersion) and predicate devices.
7. The sample size for the training set:
- Not applicable / Not provided. This device is not an AI algorithm that requires a "training set." The testing performed (biocompatibility, leakage) is for validation against established standards.
8. How the ground truth for the training set was established:
- Not applicable / Not provided. As there is no "training set" for an AI algorithm, this question is irrelevant to this device submission.
Ask a specific question about this device
(49 days)
The Contact Lens Case is a device intended for storage of soft (hydrophilic), rigid gas permeable, and/or hard contact lenses. Used for storage during chemical disinfection only. Not for use during heat disinfection.
The Contact lens case is medical device for the storage of soft (hydrophilic), rigid gas permeable, and/or hard contact lenses. The applicant device of Contact Lens Case consists of two parts: case body and case lids. The case body based with adjoining dual wells for the containment of fluid, and the two lids are designed for screwing. All the four variant models of this device have a capacity of 5.3 ml to allow contact lenses fully immersed into the well, and the well from these models accommodates all lenses currently being sold in the market. In addition, the contact lens case are made of polypropylene (96%) and Polyethylene (4%), which allow the same design principle with the same intended use. Speaking of the labeling, the inner lids of the case are marked with L (left) or R (right), meanwhile the bottom of each well is also labeled with L (left) or R (right) to distinguish the left and right lenses. In addition, at the bottom of the case body, two sides of which have labeled REPLACE MONTHLY respectively to remind user to replace the contact lens case at least once every month.
The provided document is a 510(k) premarket notification for a contact lens case, not an AI/ML medical device. Therefore, it does not contain the information requested regarding acceptance criteria, performance studies (like MRMC or standalone), ground truth establishment, or sample sizes related to AI/ML model development and validation.
The document focuses on demonstrating substantial equivalence to a predicate device by comparing:
- Intended Use: Both the proposed and predicate devices are for storing soft, rigid gas permeable, and/or hard contact lenses for chemical disinfection only, not heat disinfection.
- Design: Both have two adjoining wells with screw-down lids.
- Materials: While different (Polypropylene/Polyethylene vs. ABS), the proposed device's materials were tested for biocompatibility (cytotoxicity, irritation, systemic toxicity) and passed.
- Performance: Non-clinical tests were conducted for leakage and color fastness.
Since the request is specifically about AI/ML device validation, and this document is for a physical contact lens case, I cannot extract the requested information.
Ask a specific question about this device
(68 days)
The CONTACT LENS CASE ROUND GLASSES is intended for the storage of soft (hydrophilic), rigid gas permeable, and/or hard contact lenses. Used for storage during chemical disinfection only. Not for use during heat disinfection.
The Contact Lens Case Round Glasses is leak-proof and made from durable plastic with secure, snap lids. Note: can be used with hard and soft contact lenses; for chemical disinfection only; rinse case thoroughly before initial use; replace contact case every 2 months. The secure snap lids are connected to the body by stainless steel hinges and securely close by use of the silicon seal within the plastic Each compartment has a capacity of 4,4ml and diameter of 26,5mm and an overall dimension of 76mm x 29mm x 17mm The Case body is in White Color and the Lids in Black Color with transparent windows. Contact lenses can be fully immersed in each compartment, and each compartment fit all standard lenses currently being sold in the market. The side of each compartment is marked with L (left) or R (right)
The document describes a contact lens case, not an AI device, so many of the requested criteria related to AI performance, such as MRMC studies, ground truth establishment, or training set size, are not applicable.
However, based on the provided text, I can extract the relevant information regarding the acceptance criteria and the study performed for this medical device.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criterion | Reported Device Performance |
|---|---|
| Anti-Leakage | None of the tested lens cases showed any leakage. All of the Leakage Test Passed Successfully. |
| Cytotoxicity | No cytotoxicity |
| Eye Irritation | No Eye Irritation |
| Systemic Toxicity | No systemic Toxicity |
2. Sample Size for Test Set and Data Provenance
- Anti-Leakage Test: 48 products (units of Contact Lens Case Round Glasses) were used. The provenance is internal testing performed by the manufacturer, FiftyTwoWavs Ltd, presumably in Hong Kong S.A.R. (based on their address). This appears to be a prospective test as it was performed on units of their product.
- Biocompatibility Tests (Cytotoxicity, Eye Irritation, Systemic Toxicity): The document states "All results show that the material is not cytotoxic, causes no eye irritation and has no systematic cytotoxic." However, it does not explicitly state the number of samples or the specific methodology/sample size used for these biocompatibility tests (ISO-10993-5, ISO 10993-10, ISO-10993-11). The results were submitted in the application.
3. Number of Experts and Qualifications
Not applicable for tests described (leakage, biocompatibility). These are laboratory/engineering tests, not expert-adjudicated clinical assessments.
4. Adjudication Method for Test Set
Not applicable. The tests performed are objective physical/chemical assessments (e.g., visual inspection for leakage, lab assays for biocompatibility).
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
Not applicable. This device is a contact lens case, not an AI-assisted diagnostic or treatment device. No MRMC study was performed or is relevant. The submission explicitly states: "No Clinical Test Were Performed."
6. Standalone (Algorithm Only) Performance Study
Not applicable. This device is a physical product (contact lens case), not an algorithm or software.
7. Type of Ground Truth Used
- Anti-Leakage: Empirical observation (visual inspection for liquid leakage).
- Biocompatibility: Established ISO standards (ISO-10993-5, ISO 10993-10, ISO-10993-11) for evaluating material safety. The "ground truth" is defined by the passing criteria of these standardized tests.
8. Sample Size for Training Set
Not applicable. This device is a physical product and does not involve AI/machine learning, therefore no training set is relevant.
9. How Ground Truth for Training Set Was Established
Not applicable, as there is no training set for this device.
Ask a specific question about this device
(57 days)
For storage of soft (hydrophilic), hard and rigid gas permeable contact lenses during chemical disinfection. Use for storage during chemical disinfection only. DO NOT USE WITH HEAT.
The Contact Lens Case is a simple device made with a polypropylene base and a polyethylene lid to store soft (hydrophilic), hard, and rigid gas permeable contact lenses during chemical disinfections. This device is manufactured only for chemical disinfections with approved chemical cleaning agents to clean contact lenses.
The physical style is simple with a base volume of 4.8 milliliter and dimensions of 6.25 mm wide and 9.25 mm high. This provides sufficient space to insert and retrieve the contact lenses. It also provides sufficient space to introduce the chemical cleaning solution and adequately cover the lenses for cleaning.
The provided text is related to a 510(k) premarket notification for a Contact Lens Case. It outlines the device description, indications for use, and a summary of testing conducted to demonstrate substantial equivalence to a predicate device. However, it does not contain the specific acceptance criteria and detailed study that proves the device meets those criteria, as typically seen for AI-based medical devices.
The information provided describes a physical medical device (contact lens case), not an AI or algorithm-based device. Therefore, the questions related to AI performance metrics (like sample sizes for test/training sets, data provenance, expert ground truth, adjudication, MRMC studies, standalone performance, and effect size improvements) are not applicable to this document. The provided text details testing for biocompatibility and leakage, which are relevant for a physical contact lens case.
Here's an attempt to answer the questions based only on the provided text, recognizing the mismatch with typical AI/algorithmic device data:
1. Table of Acceptance Criteria and Reported Device Performance
| Test/Acceptance Criteria | Reported Device Performance |
|---|---|
| Biocompatibility: | |
| ISO 10993-5: In Vitro Cytotoxicity | Test article extract did not show potential toxicity to L-929 cells with polar and non-polar extracts. |
| ISO 10993-10: Ocular Irritation | Test article extracts showed no significant evidence of causing ocular irritation (response of ocular on testing side does not exceed that on the control side; grade 0). |
| ISO 10993-10: Skin Sensitization | Test article extract showed no significant evidence of causing skin sensitization in the guinea pig (positive rate of all test group animals was 0%). |
| ISO 10993-11: Systemic Toxicity | Test article extracts showed no significant evidence of causing acute systemic toxicity in the mice (no animals found with abnormal clinical symptoms, all had normal weight change). |
| Leakage Testing: | None of the tested contact lens cases showed any leakage and all passed the leakage tests successfully. |
2. Sample size used for the test set and the data provenance
- Biocompatibility Testing: The text mentions "L929 mouse fibroblast cells" for cytotoxicity, "guinea pig" for skin sensitization, and "mouse" for systemic toxicity. The specific number of cells or animals used as a "sample size" is not provided.
- Leakage Testing: 240 combinations of different tops and bottoms were tested.
- Data Provenance: Not explicitly stated, but the tests were conducted according to ISO standards, implying laboratory testing. The manufacturing company is in China, so it's likely the testing was conducted there or by a certified lab. The data is prospective in the sense that the manufacturer conducted these tests specifically for this 510(k) submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- This question is not applicable. The "ground truth" for these tests (cytotoxicity, irritation, systemic toxicity, leakage) is objective biological responses or physical observation, not expert interpretation of medical images or data.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- This question is not applicable. Adjudication methods are typically used for subjective assessments where multiple experts might disagree (e.g., classifying medical images). The tests performed are objective laboratory and physical tests.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This device is a physical contact lens case, not an AI or algorithmic device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This device is a physical contact lens case, not an AI or algorithmic device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The "ground truth" for the biocompatibility tests involved observable biological responses in cells and animals (e.g., cell viability, irritation scores, systemic effects). For leakage testing, the ground truth was the observable absence or presence of liquid leakage.
8. The sample size for the training set
- Not applicable. This device is a physical contact lens case, not an AI or algorithmic device that requires a training set.
9. How the ground truth for the training set was established
- Not applicable. This device is a physical contact lens case, not an AI or algorithmic device that requires a training set.
Ask a specific question about this device
(155 days)
Contact Lens Case is a lens care product to be used by the contact lens wearer or practitioner for storing soft (hydrophilic) contact lenses while not being worn. Not designed for heat disinfecting system. Use only with chemical disinfection.
Contact Lens Case is a lens care product to be used by the contact lens wearer or practitioner for storing soft(hydrophilic) contact lenses while not being worn. This device is not sterile and not for heat-disinfection. Use only with chemical disinfection.
There are two main series includes the variant model DC-2001, DC-3001 and DC-6001, the other includes DC-7001 and DC-8001. All these five variant models are made of Polypropylene, which follow the same design principle with the same intended use.
All the five variant models of this device have a capacity of over 1.5 ml in each case well. And the inner height of the all wells exceeds 8.5 mm. With regard to the Center Thickness of the normal hydrophilic and hydrophobic contact lens will not outnumber 8.5 mm, the capacity is sufficient for contact lens to be fully immersed under use condition.
This document is a 510(k) premarket notification for a Contact Lens Case. It does not describe a study that uses a machine learning algorithm or AI. Therefore, I cannot provide the requested information about acceptance criteria for an AI device.
The document primarily focuses on establishing substantial equivalence to a predicate device (K071081) for a physical product (contact lens case). The "tests" mentioned are non-clinical leakage testing and biological evaluation for safety.
Here's a breakdown of why the requested information cannot be extracted:
- No AI/Machine Learning Device: The device in question is a "Contact Lens Case," a physical product for storing contact lenses. There is no mention of any AI or machine learning component.
- No Acceptance Criteria for AI Performance: Since it's a physical product, the acceptance criteria are related to its physical properties, safety, and functionality (e.g., lack of leakage, biocompatibility, sufficient capacity for lenses), not performance metrics like sensitivity, specificity, or AUC for an AI algorithm.
- No Study for AI Performance: The document explicitly states, "The subject of this premarket submission, Contact lens case did not require clinical studies to support substantial equivalence." The "Summary of Non-Clinical Leakage Testing" is a performance test for the physical case, not an AI study.
- No Ground Truth, Training Set, Experts, etc.: These concepts are relevant to AI/machine learning studies, not to the approval process for a simple medical device like a contact lens case.
Therefore, I cannot populate the table or answer the specific questions related to AI studies.
Ask a specific question about this device
(195 days)
For storage of soft (hydrophilic), hard, and rigid gas permeable contact lenses during chemical disinfections. Use for storage during chemical disinfections only. Do not use heat disinfection.
Our contact lens case will consist of a lens base with dual adjoining wells for the containment of fluid. The case covers are two screw top caps. The contact lens case has a capacity of over 3.0 ml, therefore, any contact lens can be submerged into the chambers.
The provided text is a 510(k) premarket notification summary for a contact lens case. It describes the device, its intended use, and compares it to a predicate device to demonstrate substantial equivalence. However, it does not contain a study that proves the device meets explicit acceptance criteria in the way an AI/ML device submission would.
The document is a regulatory submission for a medical device (contact lens case), not an AI/ML diagnostic or prognostic device. Therefore, the detailed information requested regarding acceptance criteria, performance metrics, sample sizes for test/training sets, expert ground truth, adjudication methods, or MRMC studies for AI/ML performance is not applicable and not present in this document.
For traditional medical devices like a contact lens case, the "acceptance criteria" are generally tied to demonstrating substantial equivalence to a legally marketed predicate device, ensuring the new device has the same intended use, technological characteristics, and raises no new issues of safety or effectiveness. The "study" in this context is the comparison against the predicate device, along with adherence to general controls and relevant regulations.
Here's how the provided information relates to your request, noting the differences for this type of device:
1. A table of acceptance criteria and the reported device performance
-
Acceptance Criteria (Implicit for a 510(k) submission): That the device has the same intended use, similar technological characteristics, and raises no new issues of safety or effectiveness compared to the predicate device.
-
Reported Device Performance: The document states that "The applicant contact lens case is similar to the predicate device in - . Intended use - . Materials - . Design" and "The applicant contact lens case introduces no new questions concerning safety and efficacy." This is the core "performance" claim for substantial equivalence. No quantitative performance metrics (like sensitivity, specificity, AUC) are reported because they are not relevant for a contact lens case's function in the way they would be for a diagnostic tool.
Acceptance Criteria (Implied for 510(k) Equivalence) Reported Device Performance Same Intended Use as Predicate Yes (Intended use is identical; see comparison table) Similar Technological Characteristics as Predicate Yes (Materials and design are similar; see comparison table) No New Questions of Safety or Efficacy Yes (Explicitly stated in the conclusion)
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not Applicable. This is not an AI/ML device that uses a test set of data for performance evaluation. The "test" is the comparison against the predicate device based on its design, materials, and intended use as described in the submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. This is not an AI/ML device requiring expert consensus for ground truth. "Ground truth" for a contact lens case would relate to its physical properties and intended function, which are evaluated through standard engineering and manufacturing processes, and regulatory review based on established standards and the predicate device.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. No adjudication method for a test set is mentioned or required for this type of device submission.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This is a physical non-AI medical device; MRMC studies are irrelevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is a physical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Regulatory Substantial Equivalence: The "ground truth" here is the established safety and effectiveness profile of the predicate device (Arlington Contact Lens Service, Inc. Contact Lens Case, K120969), as determined by its prior FDA clearance. The new device demonstrates "truth" by showing it aligns with this established profile. Its own physical properties and performance (e.g., fluid containment capacity) are verified through internal company testing and manufacturing controls, but not typically detailed in this type of summary.
8. The sample size for the training set
- Not Applicable. This is not an AI/ML device that uses a training set.
9. How the ground truth for the training set was established
- Not Applicable. This is not an AI/ML device that uses a training set.
Ask a specific question about this device
(14 days)
The Applicant contact lens case is a lens care product to be used by the contact lens wearer or practitioner for storing soft, rigid gas permeable or hard contact lenses while not being worn. Not designed for heat disinfecting system. Use only with chemical disinfection.
The applicant device of Contact Lens Case is a device intended to be used by the contact lens wearer or practitioner for storing contact lenses while not being worn. The applicant device is not sterile and not for heat-disinfection.
Applicant Device Variants: SL-295, SL-298, SL-326, SL-338, SL-316, SL-511, SL-800, SL-863, SL-884, SL-886, SL-869, SL-876, SL-880, SL-899
All the variant models as mention above follow the same design principle with the same intended use. The only difference is the dimension and appearance of each model, which do no effect on the usage and intended use and just for commercial purpose.
The applicant device of Contact Lens Case consists of 2 parts: case body and case cover. The case body is based with adjoining dual wells for the containment of fluid, and the cap is designed for screwing. All the variant models of the applicant device have a capacity of over 1.5 ml in each case well. And the inner height of the all well exceeds 8.5 mm. With regard to the Center Thickness of the normal hydrophilic and hydrophobic contact lens will not outnumber 8.5 mm, the capacity is sufficient for contact lens to be fully immersed under use condition.
The applicant device of Contact Lens Case made of 2 types of material of Polypropylene (PP) and Acrylonitrile - butadiene - styrene copolymer (ABS) for each variant model.
The provided 510(k) summary is for a Contact Lens Case, which is a medical device. The summary focuses on demonstrating substantial equivalence to existing predicate devices, rather than establishing efficacy through a traditional clinical study with acceptance criteria for device performance in terms of diagnostic accuracy or impact on human readers.
Therefore, many of the typical bullet points for AI/diagnostic device studies (like sample size for test sets, number of experts, adjudication methods, MRMC studies, standalone performance, and ground truth for training) are not applicable to this type of device and submission.
Here's an analysis based on the information provided, tailored to a non-AI/diagnostic device and its substantial equivalence determination:
1. Table of Acceptance Criteria and the Reported Device Performance
For this type of device, "acceptance criteria" are primarily related to safety (biocompatibility) and functional equivalence, rather than a quantifiable performance metric like sensitivity or specificity. The "reported device performance" is framed in terms of meeting these safety standards and having comparable functionality to the predicate.
| Acceptance Criteria (Safety/Functionality) | Reported Device Performance |
|---|---|
| Biocompatibility of Materials: | |
| - In Vitro Cytotoxicity | - No Cytotoxicity (for Polypropylene, ABS, and PP with Carbazole violet) |
| - Delayed-type Hypersensitivity | - No delayed dermal contact sensitization (for Polypropylene, ABS, and PP with Carbazole violet) |
| - Eye Irritation / Intracutaneous Reactivity (for Systemic Toxicity test) | - No intracutaneous reactivity (for Polypropylene, ABS, and PP with Carbazole violet) |
| - Systemic Toxicity | - No systemic toxicity (for Polypropylene, ABS, and PP with Carbazole violet) |
| - Compliance with ISO 10993 "Biological Evaluation of Medical Devices" | - All conducted Biological Evaluation Tests are in compliance with ISO 10993. The compatibility of skin-contact component material meets biocompatibility requirements. |
| Functional Equivalence: | |
| - Capacity of each case well | - Over 1.5 ml (stated to be sufficient for full immersion of contact lenses) |
| - Inner height of well | - Exceeds 8.5 mm (stated to be sufficient for full immersion given typical contact lens thickness) |
| - Intended Use | - The same as the predicate device: storing soft, rigid gas permeable, or hard contact lenses with chemical disinfection (not heat disinfection). |
| - Design | - Two adjoining wells, screwed with screw top (similar to the predicate's screwed with integral hinged cap, deemed similar for intended use). |
| - Disinfection Type | - Chemical Disinfection, Not Heat-Disinfection (same as predicate). |
| - Indications | - Storage and Disinfection of Soft, Rigid Gas Permeable or Hard Contact Lenses (same as predicate). |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: Specific sample sizes for the individual biocompatibility tests (e.g., number of test specimens for cytotoxicity, number of animals for sensitization/irritation) are not detailed in this summary. The summary states that representative samples (SL-884 for PP, SL-800 for ABS, and SL-295 for PP with colorant) were used for testing their respective material types.
- Data Provenance: The origin of the raw test data is not explicitly stated beyond being "test results." Given the manufacturer's location in China, it is plausible the tests were conducted there or at labs compliant with ISO standards. This is a prospective assessment of newly manufactured materials/devices.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not Applicable. This is a safety and functional equivalence submission, not a diagnostic accuracy study relying on expert interpretation of images or patient data. The "ground truth" here is objective laboratory measurements for biocompatibility (e.g., cell viability, dermal reactions) and physical dimensions.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not Applicable. There is no expert adjudication process for the types of tests conducted (biocompatibility, physical measurements). The results of these objective tests are generally interpreted against established criteria or controls.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not Applicable. This is not an AI or diagnostic imaging device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not Applicable. This is not an algorithm or AI device. The tests performed are standalone evaluations of the physical device's characteristics and material safety.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" consists of objective laboratory measurements and internationally recognized standards. Specifically:
- Biocompatibility: Results of tests (cytotoxicity, delayed-type hypersensitivity, eye irritation, systemic toxicity) are compared against established biological responses and compliance with ISO 10993 standards.
- Physical Characteristics: Direct physical measurements (e.g., volume capacity, inner height) of the device.
8. The sample size for the training set
Not Applicable. This device uses standard materials and manufacturing processes; there is no "training set" in the context of machine learning or AI.
9. How the ground truth for the training set was established
Not Applicable. As there is no training set for an AI model, this question does not apply. The safety and effectiveness are established through direct testing against recognized standards and comparison to a legally marketed predicate device.
Ask a specific question about this device
Page 1 of 1