Search Results
Found 1 results
510(k) Data Aggregation
(165 days)
Comarch e-Care Platform
The Comarch e-Care Platform is intended to connect with physiological measurement devices (weight scales, blood pressure meters, pulse oximeters, peak flow meters, thermometers, glucometers) intended to use at home and send the measurement results to central server. Comarch e-Care Platform serves as Software as a Medical Device and can be used only with FDA cleared measurement devices.
Comarch e-Care Platform displays the collected measurements on the Web application and securely stores them in a database server, where the caregiver can view them, leave comments and contact patient if necessary. Caregivers are able to set thresholds individually for each patient. Measurement results sent to e-Care Platform from connected devices are analyzed and if result is beyond the threshold, caregiver gets the notification.
The Comarch e-Care Platform is not interpretive, nor is it intended for diagnosis or as a substitute for medical care, and it is not intended to provide real time data. It is made available to patients when time-critical care is not required.
The Comarch e-Care Platform is contraindicated for patients requiring direct - medical supervision or emergency intervention. It is intended for patients who are willing and capable of managing its use. Clinical judgment and experience by a caregiver are required to check and interpret the information delivered.
Comarch e-Care Platform is a software intended for use in remote patient monitoring outside of traditional healthcare settings (e.g. at home). Components of Comarch e-Care Platform are: Comarch SMA application, Comarch e-Care application, application server, database server.
Comarch SMA is a software application intended to use by patients. It is designed to collect, display and transmit vital sign measurements captured from commercially available home monitoring devices.
The following vital signs are collected: temperature, glucose, noninvasive blood pressure, pulse oximetry, weight and spirometry.
The Comarch e-Care Platform is a software intended for use in remote patient monitoring outside of traditional healthcare settings (e.g. at home). It connects with physiological measurement devices (weight scales, blood pressure meters, pulse oximeters, peak flow meters, thermometers, spirometers, glucometers) and sends the measurement results to a central server. The platform displays collected measurements on a web application, securely stores them, and allows caregivers to view and analyze them, leave comments, and contact patients. Caregivers can set individual thresholds for each patient, and the platform notifies the caregiver if a measurement result is beyond the set threshold. The device is not interpretive, not intended for diagnosis or as a substitute for medical care, and does not provide real-time data. It is for patients who are able and willing to manage its use, and clinical judgment by a caregiver is required to interpret the information.
Here's an analysis of the acceptance criteria and supporting study information:
1. Table of Acceptance Criteria and Reported Device Performance:
The provided document does not explicitly present a table of quantitative acceptance criteria for device performance. Instead, the substantial equivalence determination for the Comarch e-Care Platform is based on similarities to predicate devices in terms of:
Feature | Acceptance Criteria (Implied) | Reported Device Performance (Comarch e-Care Platform) |
---|---|---|
Indications for Use | Collect physiological measurements from home-use devices, transmit results to a central server, display measurements on a web application, securely store data, allow caregivers to view/analyze results, comment, and contact patients, enable setting of individual thresholds, and notify caregivers if thresholds are exceeded. Not interpretive, not for diagnosis, not real-time, for patients managing its use, and requiring caregiver judgment for interpretation. | Intended to connect with physiological measurement devices (glucose meters, weight scales, blood pressure meters, pulse oximeters, peak flow meters, thermometers, spirometers) for home use and send results to a central server. Displays collected measurements on a web application and securely stores them. Caregiver can view, analyze, comment, and contact patient. Caregivers can set individual thresholds, and receive notifications if results are beyond thresholds. Not interpretive, not for diagnosis/substitute for medical care, not real-time. For patients willing and capable of managing its use; clinical judgment by caregiver required. |
Intended Use | Telemedicine system | Telemedicine system |
Intended Users | Home users and healthcare providers | Home users and healthcare providers |
Patient Population | Adult users | Adult users |
Site of Use | Healthcare related environment or home | Healthcare related environment or home |
Rx/OTC | Prescription Use (Rx) | Prescription Use (Rx) |
Data Collection Software | Software application for collecting, displaying, and transmitting vital sign measurements from home monitoring devices. | Comarch e-Care software application; Comarch SMA software application. Intended to collect, display, and transmit vital sign measurements captured from commercially available home monitoring devices. |
Data Collection Software Functionality | Transmit data from measuring devices and store data in a central database. | Transmit data from measuring devices and store data in a central database. |
Communication Method of Hub with Central Server | Internet access method (e.g., Ethernet, Wi-Fi, 3G/4G network) | Any Internet access method with Ethernet or Wi-Fi 802.11 b/g/n interface or 3G/4G network. |
Types of Measuring Devices Interfaced | Glucose meters, weight scales, blood pressure meters, pulse oximeters, peak flow meters, thermometers, spirometers (FDA-cleared devices for home use). | Glucose meters, weight scales, blood pressure meters, pulse oximeters, peak flow meters, thermometers, spirometers (from a list of FDA cleared devices for home use). |
Implementation Method of Collecting Data from Measuring Devices | Short range radio system using Bluetooth (v2.0 and v4.0). | Short range radio system using Bluetooth (v2.0 and v4.0). |
Measuring Device Software | Unchanged | Unchanged |
Measuring Devices Communication Frequency | Bluetooth 2.4 GHz | Bluetooth 2.4 GHz |
Power Source | Wall power plug (120 VAC/50-60) | Wall power plug (120 VAC/50-60) |
Display | On devices and hub, and monitors connected to central server | On devices and hub, and monitors connected to central server |
Communications with Patients | Visual reading and feedback on display, phone call/email messages from caregiver. | Visual reading and feedback on hub's screen. Phone call and email messages from caregiver. |
Use of Thresholds/Algorithms | Thresholds set by healthcare professionals in server software, and sent to the hub for analysis, with results sent back to server software. | Thresholds are set by Healthcare professionals in server software and sent to the hub. Hub performs the analysis and sends it back with results to server software. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document states that "no clinical tests were conducted." Therefore, there is no specific test set or data provenance mentioned for clinical performance evaluation. The evaluation was based on non-clinical verification and validation.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Since no clinical tests were conducted and no specific test set-based ground truth was established, this information is not applicable and not provided in the document.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Since no clinical tests were conducted and no test set ground truth was established by experts, this information is not applicable and not provided.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
The document explicitly states "no clinical tests were conducted." Therefore, no MRMC comparative effectiveness study was done, and no effect size regarding human reader improvement with/without AI assistance is provided.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
The substantial equivalence is based on the device's functional similarities to predicate devices in remote patient monitoring, data transmission, storage, and caregiver notification, rather than a standalone performance evaluation of a diagnostic algorithm. The device itself is described as "not interpretive" and "not intended for diagnosis." The non-clinical verification and validation focused on the software's functionality and adherence to design specifications. Therefore, a standalone performance study in the context of diagnostic accuracy was not performed for this device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
As no clinical tests were performed, there was no "ground truth" derived from patient data, expert consensus, pathology, or outcomes data for the purpose of validating diagnostic or interpretive accuracy. The "ground truth" for the non-clinical verification and validation would have been the design specifications and requirements of the software.
8. The sample size for the training set
The document states "no clinical tests were conducted." This implies that the device did not undergo a process involving a "training set" for machine learning model development in the context of clinical performance. The platform's functionality is about data handling and alerts based on pre-set thresholds, not on learning from a dataset to perform interpretations. Therefore, this information is not applicable and not provided.
9. How the ground truth for the training set was established
As no training set was used for clinical performance evaluation, this information is not applicable and not provided.
In summary, the Comarch e-Care Platform's acceptance was based on a demonstration of substantial equivalence to existing predicate devices through comprehensive non-clinical verification and validation testing, ensuring the software's functionality, adherence to design specifications, and addressing identified risks. No clinical studies or performance evaluations requiring test sets, ground truth establishment, or expert adjudication were conducted, as the device is not interpretive and does not provide diagnostic insights.