Search Results
Found 2 results
510(k) Data Aggregation
(31 days)
CastleLoc Pectus Bar System
The CastleLoc Pectus Bar System is indicated for the treatment of Pectus Excavatum and other anterior chest deformities. It is intended to be used in pediatric (children and adolescents) and adult populations.
The Castleloc Pectus Bar System is a thoracic support product that repairs the thoracic wall, using minimally invasive surgical techniques to elevate the ribs and sternum to correct a type of chest wall deformity called pectus excavatum. Recommended implantation time is 2~3 years but may vary based on surgeon preference and patient.
This system includes various sizes of straight and curved Castleloc Pectus Bars. The appropriate bar is selected based on suitability for the patient body size. The Castleloc Pectus Bar, and Castleloc Pectus Stabilizer, Nut and Castleloc Pectus Claw Fixator for fixing the Castleloc Pectus Bar, are made of 316L (ASTM F138).
This FDA 510(k) clearance letter is for a medical device, specifically a Pectus Bar System, not an AI/ML powered device. As such, many of the typical acceptance criteria and study elements associated with AI/ML (like sample size for test/training sets, expert ground truth, adjudication methods, MRMC studies, or standalone algorithm performance) are not applicable to this submission.
The acceptance criteria and supporting study described here are focused on demonstrating the mechanical performance and safety/effectiveness equivalence of the new device to existing legally marketed predicate devices, as required for traditional medical devices.
Here's an analysis based on the provided text, highlighting the differences from AI/ML criteria where relevant:
CastleLoc Pectus Bar System Acceptance Criteria and Study Summary
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criterion (Type of Test) | Device Performance (CastleLoc Pectus Bar System - Subject Device) | How Performance Meets Criterion |
---|---|---|
Static 4-Point Bending (per ASTM F382-17) | Values were demonstrated to be similar or higher than the reference product (Park's Pectus System). | Meets or exceeds mechanical strength requirements established by the predicate device and standard. |
Dynamic (fatigue) 4-Point Bending (per ASTM F382-17) | Values were demonstrated to be similar or higher than the reference product (Park's Pectus System). | Meets or exceeds mechanical fatigue life requirements established by the predicate device and standard. |
Vertical Tensile Tests (per ASTM F382-17) | Values were demonstrated to be similar or higher than the reference product (Park's Pectus System). | Meets or exceeds tensile strength requirements established by the predicate device and standard. |
Overall Mechanical Performance | Demonstrated to have substantially equivalent mechanical performance compared to predicate devices. | Supports conclusion that the device is as safe and effective as the predicate devices. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify the exact number of physical samples (e.g., number of bars tested) for the mechanical bench testing. However, it implicitly refers to "test values" for the CastleLoc Pectus Bar System.
- Data Provenance: Not applicable in the context of mechanical bench testing; the "data" is generated from physical tests performed on the manufactured device components themselves rather than patient data.
3. Number of Experts Used to Establish Ground Truth and Qualifications
- Not applicable. This relates to mechanical bench testing of a physical device, not an AI/ML algorithm requiring expert interpretation of medical images or data. Ground truth for mechanical testing is established by physical measurement against engineering standards.
4. Adjudication Method for the Test Set
- Not applicable. This relates to mechanical bench testing, where performance is measured objectively against an ASTM standard and compared to predicate device performance, not debated or adjudicated by experts.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No. An MRMC study is relevant for AI/ML devices that impact human reader performance. This submission concerns a physical surgical implant.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Not applicable. This is a physical medical device, not an algorithm. The "standalone performance" here refers to the device's mechanical integrity under various loads, which was tested.
7. The Type of Ground Truth Used
- The ground truth for this evaluation is based on established engineering standards (ASTM F382-17) for metallic bone fixation appliances and the performance characteristics of legally marketed predicate devices (K243357 CastleLoc Pectus Bar System and K191057 Park's Pectus System). The acceptance criterion is that the subject device's mechanical test results must be "similar or higher" than the reference product.
8. The Sample Size for the Training Set
- Not applicable. This is a physical device submission, not an AI/ML submission that uses a training set of data.
9. How the Ground Truth for the Training Set Was Established
- Not applicable. There is no training set for a physical medical device.
Ask a specific question about this device
(121 days)
CastleLoc Pectus Bar System
The CastleLoc Pectus Bar System is indicated for the treatment of Pectus Excavatum and other anterior chest deformities. It is intended to be used in pediatric (children and adolescents) and adult populations.
The CastleLoc Pectus Bar System is a thoracic support product that repairs the thoracic wall, using minimally invasive surgical techniques to elevate the ribs and sternum to correct a type of chest wall deformity called pectus excavatum. Recommended implantation time is 2~3 years but may vary based on surgeon preference and patient.
This system includes various sizes of straight and curved CastleLoc Pectus Bars. The apropriate bar is selected based on suitability for the patient body size. The CastleLoc Pectus Bar, and Castleloc Pectus Stabilizer, Nut and Castleloc Pectus Claw Fixator for fixing the Castleloc Pectus Bar are made of Ti-6Al-4V (ASTM F136).
The provided text describes a 510(k) premarket notification for the "CastleLoc Pectus Bar System." This document primarily focuses on demonstrating substantial equivalence to legally marketed predicate devices, rather than an AI/ML medical device's performance evaluation against specific acceptance criteria.
Therefore, the information required to answer your prompt, which pertains to acceptance criteria, device performance, sample sizes, ground truth establishment, expert qualifications, and specific study types (like MRMC or standalone performance for AI), is not present in the provided text.
The document discusses:
- Trade/Device Name: CastleLoc Pectus Bar System
- Regulation Number/Name: 21 CFR 888.3030, Single/Multiple Component Metallic Bone Fixation Appliances And Accessories
- Regulatory Class: Class II
- Indications for Use: Treatment of Pectus Excavatum and other anterior chest deformities in pediatric and adult populations.
- Predicate Devices: Park's Pectus System (K191057) (primary), Pectus Blu Support Bar System (K241709), KLS Martin Pure Pectus System (K221938).
- Technological Comparison: States that the subject device is substantially equivalent to predicate devices in materials, design, indications for use, intended use, and operational principles.
- Non-Clinical Tests: Mentions static and dynamic 4-point bending test and vertical test according to ASTM F382-17 standards.
- Acceptance Criteria (for mechanical tests only): The mechanical performance test results were "shown to be similar or higher than the predicate device, Park's Pectus System (K191057), and meets the specified acceptance criteria." However, the specific numerical values of these acceptance criteria and the reported device performance are not detailed in the text.
- Conclusion: The tests verified mechanical performance and supported substantial equivalence in mechanical performance compared to the predicate device.
In summary, this document is a regulatory approval letter for a medical device (a pectus bar system), not a study report for an AI/ML medical device. It does not contain the detailed performance metrics, ground truth information, or study methodologies that your prompt requests for AI devices.
Ask a specific question about this device
Page 1 of 1