Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K110061
    Date Cleared
    2011-09-21

    (254 days)

    Product Code
    Regulation Number
    862.1660
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    CVM, UNCONJUGATED ESTRIOL (UE3) CALIBRATION VERIFICATION MATERIAL

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For in vitro diagnostic use as a control for calibration verification of the IMMULITE® Unconjugated Estriol (uE3) assays on the IMMULITE 1000, 2000 systems.
    The calibration verification material is assayed control with four levels. The analyte levels for IMMULITE 1000 are 0.00, 0.16, 2.70 and 11.2 ng/ml, and 0.00, 0.19, 2.90 and 12.00 for IMULITE 2000 systems. The matrix is estriol in processed horse serum.

    Device Description

    One set of four vials, 2 mL each, containing low, intermediate and high levels of unconjugated estriol in processed horse serum, with preservative, and an unconjugated estriol free sample. The calibration verification material (CVM) is supplied in liquid form, ready to use. Store unopened materials refrigerated at 2-8°C until expiration date. CVM's are for single use only, immediately after opening, and discard after use.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the IMMULITE® Unconjugated Estriol (uE3) Calibration Verification Material, based on the provided 510(K) summary:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The primary acceptance criteria for this device revolve around the precision and accuracy of the assigned values for the Calibration Verification Materials (CVMs) across different platforms.

    ParameterAcceptance Criteria (IMMULITE/IMMULITE® 1000)Reported Device Performance (IMMULITE/IMMULITE® 1000)Acceptance Criteria (IMMULITE® 2000)Reported Device Performance (IMMULITE® 2000)
    CVM 1 (0 ng uE3/mL)Range: 0 - 0.07 ng uE3/mLCVM 1 = 0.0 ng/mL (Target)Range: 0 - 0.07 ng uE3/mLCVM 1 = 0.0 ng/mL (Target)
    Precision: NAConsistent with 0 targetPrecision: NAConsistent with 0 target
    CVM 2 (Target u: 0.16 ng uE3/mL)Range: 0.07 - 0.25 ng uE3/mLWithin range (based on value assignment)Range: 0.09 - 0.29 ng uE3/mLWithin range (based on value assignment)
    Precision: