Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K963065
    Date Cleared
    1997-05-29

    (295 days)

    Product Code
    Regulation Number
    892.1560
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    CTS-285 WITH EZU-PL21 AND EZU-PC3B TRANSDUCERS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Diagnostic ultrasound imaging of the human body as follows: Fetal, Abdominal, Cardiac Adult (B mode, M mode)

    Device Description

    Model CTS-285 is a compact-type linear/convex electronic scanning ultrasound system with a built-in digital scan converter (DSC). The unit allows adult heart, abdominal organ and fetal tomographic images to be observable on a video monitor. The main unit is portable and is separable from other equipment to be carried for its use at another place as well as being usable in combination with a full keyboard, 9-inch video monitor and a special photographic unit.

    AI/ML Overview

    This 510(k) premarket notification describes the CTS-285 Diagnostic Ultrasound System. The primary method for proving substantial equivalence to the predicate device, Hitachi Medical Corporation EUB-310 (K862867), is through a comparison of technological characteristics. There are no direct performance studies with acceptance criteria in the document that would be typical for AI/ML devices. Therefore, I will create a table to compare the reported specifications of the new device (SIUI CTS-285) against its predicate (Hitachi EUB-310A) to infer "acceptance criteria" based on the predicate's established performance, and then note the new device's "performance" based on its characteristics.

    1. Table of Acceptance Criteria (Inferred from Predicate) and Reported Device Performance

    Feature/ParameterInferred Acceptance Criteria (Based on Hitachi EUB-310A)Reported Device Performance (SIUI CTS-285)Device Meets Criteria?
    Scanning ModesElectronic linear scanning (compatible 80 elements linear probe), Electronic convex sector (compatible convex probe of 80 elements)Electronic linear scanning (compatible 64 elements linear probe), Electronic convex sector (compatible convex probe of 64 elements)Partially (fewer elements)
    Display ModesB mode, B/B mode, M mode, B/M mode simultaneouslyB mode, B/B mode, M mode, B/M mode simultaneouslyYes
    Measurements (B-mode)Distance, area, circumferenceDistance, area, circumferenceYes
    Measurements (M-mode)Time interval, velocity, depth, heart rateTime interval, velocity, depth, heart rateYes
    CalculationsArea, circumference, volume, heart rate, pregnant weeks, heart functionArea, circumference, volume, heart rate, pregnant weeks, heart functionYes
    Focusing Mode4-steps dynamic focusing with variable aperture and lens focusing4-steps dynamic focusing with variable aperture and lens focusingYes
    Scanning Width (3.5MHz linear)104mm102mmYes (within tolerance)
    Scanning Width (Convex sector)60°60°Yes
    Transmitting VoltagePulse height 130VPulse height 130VYes
    Transmitting Pulse Width (3.5MHz)140 μs160 μsYes (within acceptable range for diagnostic ultrasound)
    Detecting Depth (3.5MHz)Maximum 210mmMaximum 210mmYes
    Zoom (3.5MHz probe)x1.0, x1.2, x1.5, x2.0 selectable, depth shiftx1.0, x1.2, x1.5, x2.0 selectable, depth shiftYes
    Frame RateMaximum 40 frame/secondMaximum 40 frame/secondYes
    Grey Scale1616Yes
    Memory512x512x4 bit512x512x4 bitYes
    Video OutputPAL or NTSC system TV signal (confirmed in order)PAL or NTSC system TV signal (confirmed in order)Yes
    Power Supply100-234V, ±10%, 50/60Hz, about 250W100-234V, ±10%, 50/60Hz, about 140WYes (lower power consumption is generally an advantage)
    Monitor5.5" and 8" black and white monitor5.5" black and white monitorSlightly lower (no 8" option)
    Volume and WeightApprox. 405(W)x710(L)x1310(H)mm, approx. 40kgApprox. 280(W)x235(H)x415(L)mm, approx. 13kgYes (more compact and portable)
    Electrical Safety StandardConform to requirement of I class B type apparatus of IEC 601-1; Isolate resistor testing: 1000V L-L, L-G>10MΩ; Leakage current: U-G10MΩ; Leakage current: U-G
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1