Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    Why did this record match?
    Device Name :

    CT Syringe for Stellant Injectors (with & without Spikes, Prime Tubes, & Fill Tubes), CT Syringe Dual

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For use with the Medrad® Stellant® CT Injection System equipped with the appropriate Coeur Fascia and Ram Tip for the injection of contrast media or saline to the vascular system for diagnostic purposes.

    Device Description

    The 200mL Syringe for Stellant Injectors are syringes / kits intended for use with the Medrad® Stellant® CT Injection System equipped with the appropriate Coeur Fascia and Ram Tip for the injection of contrast media or saline to the vascular system for diagnostic purposes. Each configuration is listed below in the "Models"section and provided sterile. To further define the device, the kits may include (depending on package configuration) extension lines, fill tubes, prime tubes, and spikes. The extension lines are used to connect the svringe to a patient's IV. The prime tube is used for catching fluid while purging air from the extension line. The fill tubes and spikes are used to connect the syringe to contrast media or saline containers for filling. The Fascia (installed on the surface of the injector case and through which the ram/ram tip moves) and the RamTip (connected to the injector ram and to which the syringe is connected), used in adapting the Medrad* Stellant* Injector for use with the Coeur 200mL Syringe for Stellant Injectors are not sterile, and after installation onto the injector, remain installed as long as the Coeur 200mL Syringe for Stellant Injectors is used with the injector. (Cleaning instructions for the Fascia and Ram Tip are provided in the IFU.)

    Principle of operation- The syringe plunger is engaged by the equipment ram and moved forward and back for filling and injection purposes.

    AI/ML Overview

    The provided text is a 510(k) summary for the Coeur 200mL Syringe for Stellant Injectors. It describes the device, its intended use, and the testing performed to demonstrate substantial equivalence to predicate devices. However, this document does not contain information about an AI/ML powered device, a multi-reader multi-case (MRMC) comparative effectiveness study, or details about ground truth establishment as would be relevant for such a device.

    Here's an analysis based on the information provided, focusing on what can be extracted and noting what cannot, given the nature of the device (a syringe, not an AI system):

    1. Table of Acceptance Criteria and Reported Device Performance

    The document doesn't present acceptance criteria and device performance in a formal table format as requested, nor does it define acceptance criteria in numerical targets. Instead, it lists the types of nonclinical tests performed and states that the product conforms to various standards or functions as intended. Since this is a mechanical device, performance is typically assessed against engineering specifications and international standards rather than AI-specific metrics like AUC, sensitivity, or specificity.

    Test TypeStandard/RequirementReported Performance
    Visual VerificationConformance to visual requirementsInspection of the product for conformance to visual requirements
    Dimensional EvaluationConformance to Coeur drawing specificationsInspection of the product to verify dimensional acceptance
    ISO 594-2:1998Liquid leakage, air leakage, separation force, ease of assembly, resistance to overriding, and stress crackingEvaluation to applicable requirements of the ISO 594 standard
    ISO 7886-2Cleanliness, limits for acidity and alkalinity, limits for extractable metals, lubricant, tolerance on graduated capacity, graduated scale, piston plunger assembly, and nozzleEvaluation to applicable requirements of the ISO 7886-2 standard
    Injector Fit EvaluationFit with Medrad® Stellant® InjectorEvaluation of syringe, fascia and ram tip for fit with the Medrad® Stellant® Injector
    Volume EvaluationAccuracy of volumeEvaluation of volume accuracy
    Functional Testing (Syringe/Fascia/Ram Tip)Function with Medrad® Stellant® InjectorFunctional testing was conducted to evaluate function with the Medrad® Stellant® Injector
    Pressure Testing (Syringe)400 psi for 10 secondsSyringe held at 400psi for 10 seconds
    Pressure Testing (Extension Line)400 psi for 2.5 minutesExtension Line held at 400psi for 2.5 minutes
    Pull Testing (Extension Lines)Correct assembly and bonding processConducted to verify correct assembly and bonding process
    Functional Testing (Spike)Ability to connect, allow fluid flowEvaluated ability to connect to syringe, bottle port, bag, and allow fluid flow
    Age VerificationMaintain performance for 1 yearLeveraged 1-year expiration from similar marketed devices and included aging study
    BiocompatibilitySuitability for intended purposesAll test results indicate suitability of materials per ISO 10993-1 guidance

    2. Sample size used for the test set and the data provenance

    The document does not specify exact sample sizes for each nonclinical test (e.g., how many syringes were pressure tested). It refers to "inspection of the product," "evaluation of product," and "functional testing" without providing the number of units tested. This is standard for a 510(k) for a mechanical device, where testing often involves a statistically determined sample or a representative sample.

    Data provenance is not applicable here as the "performance data" refers to in-house laboratory testing of the physical syringe components, not clinical data from patients. There's no country of origin for "data" in the sense of patient records, nor is it retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is a mechanical syringe, not an AI/ML device relying on expert-annotated ground truth. "Ground truth" in this context would likely refer to engineering specifications and measurements performed by qualified technicians.

    4. Adjudication method for the test set

    Not applicable. There is no adjudicated "ground truth" for this type of mechanical device testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices, especially those using AI, to compare human performance with and without AI assistance. This document is for a medical syringe.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or AI device.

    7. The type of ground truth used

    Not applicable in the AI/ML context. For a syringe, "ground truth" corresponds to established engineering specifications, international standards (ISO 594-2, ISO 7886-2), and the physical properties of the materials. Performance is verified against these objective, predetermined values.

    8. The sample size for the training set

    Not applicable. There is no training set for a mechanical syringe.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set or associated ground truth for a mechanical syringe.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1