    Device Name :

    CP620G LCD monitor, CP620 LCD monitor

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CP620G/CP620 LCD Monitor is intended to be used in displaying digital images for review and analysis by trained medical practitioners. It does not support the display of mammography images for diagnosis.

    Device Description

    CP620G/CP620 LCD Monitor is the display system with the high resolution (30002000), high luminance (500 cd/m2), and 4398 billion colors, built-in DICOM standard LUT. Monitor is portable and light, the 750g weight make it is easy to carry. With these this display can automatic adjustment according to different requirements in order to achieve the best results.(There is no difference between CP620G and CP620 except for labeling as they are marketed in different areas. CP620G is global sales, CP620 is domestic sales .It does not affect their safety or effectiveness in any terms. ) The product is consisted of the following components: - 13.5" Color-TFT LCD Panel - DMS3400AR0/6 layers/214150mm/1.0mm/REV:1.0 - CP620G LCD Monitor software - Power Adapter - Data Cable. In accordance with the May 11, 2005 Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, the software level of concern for the CP620G/CP620 LCD Monitor was determined to be Moderate on account of a failure or latent flaw could indirectly result in minor injury to the patient or operator through incorrect or delayed information or through the action of a care provider. the software doesn't include any functions of image manipulation. The LCD Monitor is designed, tested, and will be manufactured in accordance with both mandatory and voluntary standards: 1. IEC 60601-1:2012, EN 60601-1:2013, ANSI/AAMI ES60601-1:2005+A1:2012+C1:2009+A2:2010,CAN/CSA C22.2 NO.60601-1:14, Medical equipment medical electrical equipment - Part 1: General requirements for basic safety and essential performance. 2. IEC 60601-1-2 Edition 4:2014, EN 60601-1-2:2015, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests.

    AI/ML Overview

    The information provided describes the Nanjing Jusha Display Technology Co., Ltd. CP620G LCD Monitor / CP620 LCD Monitor. Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly state "acceptance criteria" in the traditional sense of measurable thresholds for diagnostic accuracy (e.g., sensitivity, specificity for an AI algorithm). Instead, the performance data presented focuses on the technical specifications of the display and compliance with established electrical safety and electromagnetic compatibility (EMC) standards. The "predicate device" is used as the benchmark for substantial equivalence.

    AttributeAcceptance Criteria (Predicate Device K200472)Reported Device Performance (CP620G / CP620 LCD Monitor - K230643)
    Screen Technology13.5" Color TFT LCD Panel13.5" Color TFT LCD Panel
    Viewing Angle (H, V)Horizontal 170°, Vertical 170°Horizontal 170°, Vertical 170°
    Resolution3000x20003000x2000
    Display Area285.3(V)x190.2(H)mm285.3(V)x190.2(H)mm
    Contrast Ratio1800:11800:1
    Scanning frequency (H; V)94.94~123.84kHz; 60Hz94.94~123.84kHz; 60Hz
    Recommended Luminance300cd/m2500cd/m2
    Pixel Pitch0.0951x0.0951mm0.0951x0.0951mm
    BacklightLEDLED
    Display Colors14-bit, 4398 billion colors14-bit, 4398 billion colors
    Luminance CalibrationGeneral calibration sensorGeneral calibration sensor
    Video Signal InputType-C x1, Mini DP x1Type-C x2, HDMI x1
    Display ControllerOff the shelfOff the shelf
    Power RequirementAC 100240V 5060HzAC 100240V 5060Hz
    Power Consumption/Save Mode15W / N/A20W / NA
    Applicable StandardsIEC 60601-1:2012, EN 60601-1:2013, ANSI/AAMI ES60601-1:2005+A1:2012+C1:2009+A2:2010, CAN/CSA C22.2 NO.60601-1:14; IEC 60601-1-2 Edition 4:2014, EN 60601-1-2:2015IEC 60601-1:2012, EN 60601-1:2013, ANSI/AAMI ES60601-1:2005+A1:2012+C1:2009+A2:2010, CAN/CSA C22.2 NO.60601-1:14; IEC 60601-1-2 Edition 4:2014, EN 60601-1-2:2015

    Discussion of Differences:

    • Recommended Luminance: The proposed device has a higher recommended luminance (500cd/m2) than the predicate (300cd/m2), which is considered an improvement ("CP620G is better than CP610").
    • Video Signal Input: The proposed device has different input ports (Type-C x2, HDMI x1 vs. Type-C x1, Mini DP x1). This difference is stated to "only shows that they have different input, has nothing to do with the display function."
    • Power Consumption/Save Mode: The proposed device has a higher power consumption (20W vs. 15W). This is attributed to "components used in the LCD Monitor" and is stated to have "nothing to do with the display function."
    • Dimensions w/o stand: Slight differences are noted (312.3mmx220mmx11.5mm vs. 323mmx225x11.5mm), explained as "Different housing design due to the different glass size."
    • New Features: The proposed device includes new features: Bidirectional power supply function, dual-screen display function, gravity sensing function, adaptive environment brightness function, and screen touch function. These are all described as making the product "more convenient" and "has nothing to do with the display function" (except for dual-screen display, which allows two independent 3MP windows side-by-side).

    2. Sample size used for the test set and the data provenance

    The document describes bench testing for the CP620/CP620G. There is no mention of a "test set" in the context of clinical images or patient data. The testing pertains to the physical and electrical performance of the display monitor itself. Therefore, concepts like sample size and data provenance (country, retrospective/prospective) are not applicable in the way they would be for an AI diagnostic algorithm. The tests conducted include:

    • Measurement of the angular dependency of luminance response in horizontal, vertical and diagonal directions
    • Measurement of the luminance non-uniformity characteristics of the display screen
    • Measurement of the chromaticity non-uniformity characteristics of the display screen
    • Measurement of small-spot contrast ratio
    • Measurement of temporal response
    • Performance data on luminance stability

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable. The device is an LCD monitor for displaying medical images, not an AI algorithm that provides diagnostic interpretations requiring expert ground truth setting for a dataset. The "ground truth" for this device relates to its physical performance characteristics meeting technical specifications, which are verified through various physical and electrical measurements.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable for the reasons stated above.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. This device is an LCD monitor, not an AI diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable. This device is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance acceptance is based on:

    • Technical Specifications: Measured physical and electrical properties of the display (luminance, contrast, resolution, viewing angle, color, etc.) matching or exceeding the predicate device's specifications.
    • Compliance with Standards: Adherence to recognized medical device safety (IEC 60601-1) and electromagnetic compatibility (IEC 60601-1-2) standards.

    8. The sample size for the training set

    This information is not applicable. The device is a monitor, not an AI algorithm that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable for the reasons stated above.