Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K994244
    Date Cleared
    2000-02-18

    (64 days)

    Product Code
    Regulation Number
    878.4040
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Convertors® Surgical Gowns are devices intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from the transfer of microorganisms, body fluids and particulate material.

    Device Description

    The gowns are comprised of a single base layer of degradable spunlaced nonwoven fabric in gown configurations of unreinforced, fabric reinforced and poly-reinforced. The fabric reinforced gown contains an additional layer of spunlaced nonwoven fabric in the chest and sleeves area. The poly-reinforced gown contains an additional layer of polyolefin film in the chest and sleeves of the gown.

    AI/ML Overview

    This document describes the "Convertors® Surgical Gowns" manufactured by Allegiance Healthcare Corporation. The information provided focuses on the regulatory submission (510(k)) for this medical device, which is a Class II device (per 21 CFR § 878.4040) used as surgical apparel.

    Here's an analysis of the provided information regarding acceptance criteria and the study that proves the device meets those criteria:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not present a formal table of specific, quantifiable acceptance criteria with corresponding performance data. Instead, it describes general categories of testing and statements of acceptability, primarily focused on biocompatibility and general performance as compared to a predicate device.

    Acceptance Criteria (Implied/General)Reported Device Performance
    Biocompatibility (Safety)All materials used in the fabrication of Convertors® Surgical Gowns were evaluated through biological qualification safety tests as outlined in ISO 10993 Part-1 "Biological Evaluation of Medical Devices." The biocompatibility tests performed were cytotoxicity, sensitization, and primary skin irritation. These materials were found to be acceptable for the intended use.
    Performance Attributes (Protective Barrier)The performance attributes are similar to the predicate device (Isolyser Industries Enviroguard Surgeons Gowns). This implies meeting the functional requirements for surgeons' gowns, such as protection against the transfer of microorganisms, body fluids, and particulate material. The term "similar" suggests that the Convertors® gowns perform comparably in these aspects to the predicate device. The document does not provide specific quantitative metrics (e.g., AAMI barrier levels, tear strength, linting) for either the acceptance criteria or reported performance.
    Intended Use EquivalenceThe intended use is the same as the predicate device (Isolyser Enviroguard gowns). This is explicitly stated as a point of substantial equivalence. The intended use is "to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from the transfer of microorganisms, body fluids and particulate material."

    2. Sample Size Used for the Test Set and the Data Provenance:

    The document does not specify the sample size used for the test set(s) for either biocompatibility or performance testing. It also does not explicitly state the data provenance (e.g., country of origin, retrospective or prospective). Given the nature of a 510(k) submission for a non-active medical device like a surgical gown, testing would typically be conducted under controlled laboratory conditions, and the data would likely be prospective from these dedicated tests.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    This information is not provided in the document. For device testing (e.g., biocompatibility, protective barrier), the "ground truth" is typically established by recognized standards and validated test methods, not necessarily by experts reviewing individual test results in a consensus manner. The "experts" in this context would be the technicians and scientists performing the tests and interpreting the results according to the test standards.

    4. Adjudication Method for the Test Set:

    This information is not provided and is generally not applicable to the type of device testing described (biocompatibility, material performance). Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or image-based diagnostic device evaluations where subjective interpretation plays a significant role.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:

    An MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices, particularly those involving interpretation by human readers (e.g., radiologists, pathologists) where AI might assist in improving performance. Surgical gowns do not fit this category.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    No standalone (algorithm-only) performance was done. This is not applicable to a surgical gown, which is a physical barrier device, not an algorithm-driven diagnostic or treatment device.

    7. The Type of Ground Truth Used:

    The ground truth for the device's performance is established by:

    • Biocompatibility Standards: Specifically, ISO 10993 Part-1 "Biological Evaluation of Medical Devices" for cytotoxicity, sensitization, and primary skin irritation tests. The "ground truth" here is the pass/fail criteria defined within these international standards.
    • Industry Recognized Test Methods: For other performance attributes (e.g., barrier effectiveness, strength), the document mentions testing in accordance with "industry recognized test methods." The "ground truth" would be the established performance thresholds or benchmarks within those specific test methods.
    • Predicate Device Equivalence: A significant part of the "ground truth" for substantial equivalence is the demonstration that the new device's intended use and performance are "similar" to a legally marketed predicate device (Isolyser Industries Enviroguard Surgeons Gowns).

    8. The Sample Size for the Training Set:

    This information is not provided. This concept is primarily relevant for machine learning algorithms, which are not applicable to the testing of a surgical gown.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not provided, as the concept of a "training set" and establishing its ground truth is not applicable to the evaluation of a surgical gown.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1