510(k) Data Aggregation

    K Number
    K130100
    Date Cleared
    2013-04-18

    (93 days)

    Product Code
    Regulation Number
    876.5820
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    COMINED STANDARD SOLUTION

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Conductivity/pH Calibrator Solutions are a secondary standard solution used for the calibration of conductivity/TDS cells together with conductivity/TDS and pH measurement instruments. The conductivity /TDS cells and instruments may be indicated for calibrating the conductivity and pH measurement function of hemodialysis machines and water purification equipment for hemodialysis, or for verifying proper function of hemodialysis machine or water purification equipment measurement functions. These solutions are used remotely from the hemodialysis machine or water purification equipment, and does not come into contact with the patient.

    Device Description

    The device consists of salts (Sodium Chloride, Sodium Phosphate Dibasic, and Potassium Phosphate Monobasic) dissolved in de-ionized water. The proportion of total salt determines the solution's conductivity. The proportions of Sodium Phosphate Dibasic and Potassium Phosphate Monobasic determine the solution's pH. The solution is packaged into sealed polyethylene bottles. The performance specifications of the device are a pH value of pH 7.0 ± 0.03 pH units and a conductivity value of 14.00 mS/cm ± 0.03 mS/cm. The solution should maintain these values for 18 months stored in a sealed bottle. The solution should also maintain these values for 30 days within an opened but recapped bottle. There will be two bottle sizes of this solution, a 32-oz bottle and a 16-oz bottle.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Device: Conductivity/pH Calibrator Solution

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Performance Specifications)Reported Device Performance
    pH value of pH 7.0 ± 0.03 pH unitsNew Device (2013): 7.0 pH ± 0.03 pH @ 25°C.
    Study Results: The "PQ-RM017 pH and Conductivity Solution Manufacturing Process" validated production to specification. The "PQ-RM019 pH and Conductivity Solution Storage Stability" demonstrated pH values remain within specification for 18 months, with extrapolation from 13-41 weeks. The "PQ-RM018 pH and Conductivity Solution Open Bottle Stability" demonstrated pH values remain within specification for 30 days after opening.
    Conductivity value of 14.00 mS/cm ± 0.03 mS/cmNew Device (2013): 14.0 millisiemen ±0.03 millisiemen @ 25°C.
    Study Results: The "PQ-RM017 pH and Conductivity Solution Manufacturing Process" validated production to specification. The "PQ-RM019 pH and Conductivity Solution Storage Stability" demonstrated conductivity values remain within specification for 18 months, with extrapolation from 13-41 weeks. The "PQ-RM018 pH and Conductivity Solution Open Bottle Stability" demonstrated conductivity values remain within specification for 30 days after opening.
    Shelf life: 18 months (sealed bottle)New Device (2013): 18 months.
    Study Results: The "PQ-RM019 pH and Conductivity Solution Storage Stability" demonstrated that values remain within specification for 18 months. Results were extrapolated from data taken between 13 and 41 weeks.
    Shelf life: 30 days (opened but recapped bottle)New Device (2013): 30 days.
    Study Results: The "PQ-RM018 pH and Conductivity Solution Open Bottle Stability" demonstrated the device is within specification after 30 days of the bottle being opened and recapped several times.
    Independence of measurement instrumentStudy Results: The "PQ-RM020 pH and Conductivity Solution Instrument Independence" demonstrated that the new device's conductivity measurements are independent of the type of measurement instrument, similar to the predicate.

    2. Sample Sizes Used for the Test Set and Data Provenance

    The document does not explicitly state specific sample sizes (number of bottles, measurements, etc.) for each validation test (PQ-RM017, PQ-RM018, PQ-RM019, PQ-RM020). It refers to "production lots" for the initial manufacturing process validation.

    The data provenance is retrospective, as the studies were performed on "production lots" and involved testing for stability over time. The country of origin of the data is Mesa Laboratories, Inc., located in Lakewood, CO, USA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable to this type of device. The ground truth for calibrator solutions is established through comparison to N.I.S.T. (National Institute of Standards and Technology) SRM's (Standard Reference Materials), which are highly precise and certified standards. It does not involve human expert interpretation or consensus.

    4. Adjudication Method for the Test Set

    This information is not applicable. As discussed above, the validation involves objective measurement against NIST standards, not human interpretation that would require adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging or interpretation tasks where human readers' performance is evaluated and compared with and without AI assistance. This device is a calibrator solution, not an interpretive medical device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    This concept is not directly applicable in the traditional sense of an "algorithm only" performance for an AI/ML device. However, the performance of the calibrator solution itself, independent of the user, is assessed in the various validation studies (e.g., manufacturing process, storage stability, open bottle stability, instrument independence). The device's performance characteristics (pH, conductivity) are measured directly using instrumentation.

    7. The Type of Ground Truth Used

    The ground truth used for these studies is N.I.S.T. SRM's (Standard Reference Materials). These are primary calibration standards produced in accordance with NIST instructions, providing highly accurate and traceable reference values.

    8. The Sample Size for the Training Set

    This information is not applicable. This device is a chemical calibrator solution, not an AI/ML algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable, as there is no "training set" for this device.