Search Results
Found 1 results
510(k) Data Aggregation
(251 days)
COLORMATE III
The Colormate III™ is to be used as an aid to the physician in monitoring the status of newborn babies for the development of hyperbilirubinemia. Following the physician's examination within the first hours after birth, newborn babies are initially measured and periodically monitored by the Colormate III™ for incremental changes in the yellow content of the skin color as compared to the baseline Colormate III™ measurements. Babies with Colormate III™ test results indicative of hyperbilirubinemia are to be re-evaluated by the attending physician for appropriate patient management.
Colormate III™ (Colormate III or CM III) is a computer assisted noninvasive transcutaneous bilirubinometer which, through colorimetric technology, illuminates the skin of newborn babies and measures the yellow content of the skin color. The incremental changes of these color readings measured over time are compared to the newborn's baseline readings. These data are then automatically processed to provide a numerical index of estimated bilirubin count in mg/dl that has been shown to correlate with total serum bilirubin within a clinically useful range. The Colormate III, like other colorimeters, generates a set of tristimulus color coordinates for each measurement. It then performs a series of algorithms to produce final measurement results.
The provided document describes the Colormate III™, a transcutaneous bilirubinometer, and a clinical study conducted to demonstrate its performance. Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The document doesn't explicitly state "acceptance criteria" as a set of predefined thresholds that the device must meet. Instead, it presents performance metrics against visual assessment as a comparator, focusing on correlation, bias, and sensitivity. We can infer the performance goals from the reported results.
Metric (Inferred Acceptance Criteria) | Reported Device Performance (Colormate III™) | Comparator (Visual Assessment) |
---|---|---|
Correlation with Serum Bilirubin Concentration (r) | r = 0.90 (p 12.0 mg/dl** | 89% (statistically significantly better than visual assessment, p 12.0 mg/dl): |
* Colormate III: 89%
* Clinicians' visual assessment: 61%
* The Colormate III shows **significantly higher accuracy and sensitivity** than visual assessment.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Yes, a standalone performance assessment was done. The "Colormate III" performance metrics (correlation r=0.90, bias, sensitivity=89%) represent the algorithm's performance when measuring skin color and processing it to provide a numerical index of estimated bilirubin. The device's output is then used by a physician, but the reported performance metrics focus on the device's ability to correlate with blood serum levels.
7. Type of Ground Truth Used
- The primary ground truth used for evaluating the Colormate III and visual assessment was total serum bilirubin concentration (BRC), measured via blood tests.
- The document also describes how "a cutoff of 12.0 mg/dl of bilirubin" was used as an "acceptable level of clinical concern" for calculating sensitivity.
8. Sample Size for the Training Set
- The document states: "Prior to undertaking a pivotal clinical trial to evaluate the Colormate III's performance, two phases of studies were conducted to validate the clinical approach and test criteria and determine calibration algorithms for subsequent validation."
- This indicates that a training/calibration phase was performed before the pivotal clinical trial, but the sample size for these training/calibration studies is not explicitly provided.
9. How the Ground Truth for the Training Set Was Established
- The document implies that the ground truth for the training/calibration phase was also established using serum bilirubin blood tests in conjunction with Colormate III measurements. It states these studies were used to "determine calibration algorithms," which would necessitate comparing device readings to a reference standard (serum bilirubin) to build or refine its internal models. However, detailed methodology for training ground truth establishment is not provided.
Ask a specific question about this device
Page 1 of 1