Search Results
Found 2 results
510(k) Data Aggregation
(78 days)
The CODMAN Certas Programmable Valve is an implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus.
The CODMAN Certas Therapy Management System allows the noninvasive reading or adjustment of the valve setting.
The Certas Programmable Valve System is designed for shunting cerebrospinal fluid (CSF) for the treatment of hydrocephalus. The valve can be set to a choice of eight opening pressure settings for constant intraventricular pressure and drainage of CSF. The Therapy Management System allows the user to non-invasively adiust the opening pressure setting before and after implant.
The Certas Programmable Valve is available with and without the legally marketed SIPHONGUARD CSF Control Device and the Codman BACTISEAL Catheters as well as accessories needed to facilitate placement and use of the valve.
The Certas Programmable Valve is an implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus.
The Certas Therapy Management System allows the noninvasive reading or adjustment of the valve setting.
The provided text describes a 510(k) submission for the CODMAN CERTAS Programmable Valve System and CODMAN CERTAS Therapy Management System, focusing on demonstrating substantial equivalence to a predicate device after extending its shelf life from 1 year to 5 years.
Here's an analysis based on your requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are not explicitly stated in numerical thresholds, but rather implied by "no change" and "maintained" functionality. The performance data demonstrates that these criteria were met.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Packaging integrity maintained for 5 years after accelerated aging | "The packaging stability study shows that there was no change in post-sterile seal strength after accelerated aging." |
Product functionality maintained for 5 years after accelerated aging | "The product stability testing evaluated Adjustment and Indication capability, Pressure-Flow, as well as Leak and Reflux testing per EN ISO 7197:2009 after 5 years of accelerated aging. The test results demonstrate that the Certas Valve is substantially equivalent to the predicate device." |
No changes to intended use, materials, or principles of operation | "No changes are being made to the intended use, performance characteristics, materials, and principles of operation of the valve or its accessories." |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample size for the stability testing (both packaging and product). It refers to "packaging stability study" and "product stability testing" without providing specific numbers of units tested.
- Sample Size: Not explicitly stated.
- Data Provenance: The studies were conducted by the manufacturer, Medos International Sarl, as part of their regulatory submission. The context suggests these were prospective tests designed specifically to demonstrate the extended shelf life. The country of origin of the data is not specified beyond the company's location (Switzerland/US branch).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This type of information is not applicable (N/A) to this device submission. This submission concerns a physical medical device (a valve system) and its mechanical/functional performance after aging, not an AI/software device that requires expert-established ground truth for diagnostic or interpretative tasks. The "ground truth" here is the physical measurement of performance parameters (e.g., seal strength, pressure-flow, leak, reflux) against established engineering standards (EN ISO 7197:2009).
4. Adjudication Method for the Test Set
This is N/A for the same reasons as point 3. Adjudication methods like 2+1 or 3+1 are typically used for expert reviews in image analysis or clinical assessments where there might be inter-observer variability. For objective physical device testing, the results are measured against defined performance specifications.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is N/A. This submission is for a physical medical device (hydrocephalus shunt valve) for which MRMC comparative effectiveness studies with or without AI assistance are not relevant. This type of study is typically performed for AI-powered diagnostic tools or image analysis software.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
This is N/A. The device is a physical valve system and a programming system, not an algorithm, so a standalone algorithm performance study is not relevant.
7. The Type of Ground Truth Used
The ground truth used for this testing is based on:
- Objective physical measurements: Assessing parameters like seal strength, adjustment/indication capability, pressure-flow characteristics, and leak/reflux.
- Engineering Standards: Conformance to recognized international standards like EN ISO 7197:2009 for hydrocephalus shunts. The "ground truth" is that the device must continue to meet these functional and safety specifications after accelerated aging.
8. The Sample Size for the Training Set
This is N/A. This device is not an AI/machine learning product, so there is no "training set" in the context of data-driven model development. The product development likely involved engineering design and testing, but not a data-based training set.
9. How the Ground Truth for the Training Set was Established
This is N/A for the same reasons as point 8.
Ask a specific question about this device
(82 days)
The CODMAN Certas Programmable Valve is an implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus. The CODMAN Certas Therapy Management System allows the noninvasive reading or adjustment of the valve setting.
The Certas Programmable Valve System is designed for shunting cerebrospinal fluid (CSF) for the treatment of hydrocephalus. The valve can be set to a choice of eight opening pressure settings for constant intraventricular pressure and drainage of CSF. The Therapy Management System allows the user to non-invasively adjust the opening pressure setting before and after implant. The Certas Programmable Valve will be available with and without the legally marketed SIPHONGUARD CSF Control Device and the Codman BACTISEAL Catheters. The Certas Programmable Valve includes the same legally marketed accessories that are available with the Hakim Programmable Valve. The Certas Programmable Valve is an implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus. The Certas Therapy Management System allows the noninvasive reading or adjustment of the valve setting. The SIPHONGUARD device is designed to reduce the potential hazards of excessive lowering of intraventricular pressure (with respect to atmospheric pressure) when a patient is in an upright position. The BACTISEAL catheters are intended for use in the treatment of hydrocephalus as a component of a shunt system when draining or shunting of cerebrospinal fluid is indicated.
Here's an analysis of the provided text, focusing on the acceptance criteria and study information for the Codman Certas Programmable Valve System:
1. Table of Acceptance Criteria and Reported Device Performance
The provided text describes pre-clinical testing for the Codman Certas Programmable Valve System. While it doesn't explicitly state quantitative "acceptance criteria" values in a table, it lists the types of tests performed and the general outcome.
Acceptance Criterion (Test Performed) | Reported Device Performance |
---|---|
Leak Testing | Performs as intended, safe and effective. |
Pressure-Flow Testing | Performs as intended, safe and effective. |
Overpressure Testing | Performs as intended, safe and effective. |
Bursting Pressure Testing | Performs as intended, safe and effective. |
Dynamic Break Strength Testing | Performs as intended, safe and effective. |
Long-Term Stability Testing | Performs as intended, safe and effective. |
Puncture Testing | Performs as intended, safe and effective. |
MRI Translational Attraction (ASTM F2052) | MR Conditional in a 3 Tesla MRI system. |
MRI Torque (ASTM F2213) | MR Conditional in a 3 Tesla MRI system. |
MRI Artifact (ASTM F2119) | MR Conditional in a 3 Tesla MRI system. |
MRI Heating (ASTM F2182) | MR Conditional in a 3 Tesla MRI system. |
Biocompatibility (ISO 10993-1, FDA G95-1) | All materials meet requirements. |
Sterilization (ISO 17665 Part 1 & 2, 11737) | Sterility assurance level of 10^-6^ achieved. |
Overall Substantial Equivalence | The Certas Valve and Certas Therapy Management System are substantially equivalent to predicate devices. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not provide specific sample sizes for the pre-clinical tests or details on the provenance of data (e.g., country of origin, retrospective/prospective). The testing appears to be primarily bench or laboratory-based, as is common for pre-clinical performance evaluations of medical devices.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
The document does not mention the use of experts to establish a "ground truth" for the test set. The tests described are primarily objective, quantitative engineering and materials tests (e.g., pressure-flow, strength, MRI compatibility, biocompatibility, sterility) rather than clinical evaluations requiring expert interpretation of outcomes.
4. Adjudication Method for the Test Set
Not applicable. As noted above, the tests are objective pre-clinical evaluations, not interpretive clinical assessments that would require an adjudication method.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. The document describes pre-clinical (bench) testing to demonstrate substantial equivalence based on performance characteristics, materials, and principles of operation. It does not mention any clinical studies, MRMC studies, or human reader performance evaluations.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Not applicable. The device is a physical medical implant and associated management system, not an AI algorithm. The performance evaluation is for the physical device itself.
7. Type of Ground Truth Used
The "ground truth" for the pre-clinical tests would be the established scientific and engineering standards and methodologies defined by the cited ISO and ASTM standards. For example, for sterility, the ground truth is a sterility assurance level of 10^-6^. For MRI compatibility, the ground truth is the criteria defined by ASTM F2503 for "MR Conditional."
8. Sample Size for the Training Set
Not applicable. The device is a physical medical device, not an AI/ML algorithm that requires a training set. The "training" of the device itself refers to its design and manufacturing processes based on engineering principles and regulatory standards.
9. How the Ground Truth for the Training Set Was Established
Not applicable. As above, this is a physical medical device, not an AI/ML algorithm. The "ground truth" for its development would be based on established engineering principles, material science, and prior knowledge from predicate devices and relevant standards.
Ask a specific question about this device
Page 1 of 1